Tumark Vision, Tumark Professional, Tumark Q, Tumark Professional Q-Shape
K201863 · Somatex Medical Technologies GmbH · NEU · Feb 18, 2021 · General, Plastic Surgery
Device Facts
Record ID
K201863
Device Name
Tumark Vision, Tumark Professional, Tumark Q, Tumark Professional Q-Shape
Applicant
Somatex Medical Technologies GmbH
Product Code
NEU · General, Plastic Surgery
Decision Date
Feb 18, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4300
Device Class
Class 2
Indications for Use
The Tumark® Vision, the Tumark® Professional, the Tumark® Professional Q-Shape are intended to attach a marker to soft breast tissue and axillary lymph nodes, following an open or a percutaneous procedure to radiographically mark the location of the surgical site. It is not indicated to be used with magnetic resonance imaging (MRI) techniques.
Device Story
Tumark Vision, Professional, Q, and Professional Q-Shape are sterile, single-use tissue site marking systems. Each consists of a non-absorbable nickel-titanium (Nitinol) clip marker preloaded in a stainless steel introducer cannula with a plastic handle. The device is used by physicians in clinical settings to mark surgical sites in breast tissue or axillary lymph nodes. The operator uses a one-handed slide button mechanism on the handle to deploy the marker from the cannula into the target tissue. A safety catch prevents premature deployment. The cannula features a bevelled tip and depth markings. Once implanted, the marker provides a permanent radiographic reference point for the surgical site, aiding in follow-up imaging and clinical decision-making. The device is not MRI-compatible.
Clinical Evidence
No clinical trials were conducted. Evidence consists of bench testing (device function, performance, and stability) and a literature review supporting the use of markers in axillary lymph nodes. Physician statements were also provided to support the indication for use.
Technological Characteristics
Sterile, single-use, preloaded marking system. Marker: Nitinol. Cannula: Stainless steel, 18G, lengths 70-150mm. Handle: Plastic with one-handed deployment mechanism and safety catch. Biocompatibility per ISO 10993-1. Sterilization: EO. Non-MRI compatible.
Indications for Use
Indicated for patients undergoing open or percutaneous procedures requiring radiographic marking of soft breast tissue or axillary lymph nodes. Not for use with MRI.
Regulatory Classification
Identification
An implantable clip is a clip-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a symbol representing the Department of Health & Human Services. To the right of the symbol, there is a blue square with the letters "FDA" in white. Next to the blue square, the words "U.S. FOOD & DRUG ADMINISTRATION" are written in blue.
February 18, 2021
SOMATEX Medical Technologies GmbH David Vasmer Regulatory Affairs Manager Hohenzollerndamm 150/151 Berlin, Berlin 14199 Germany
Re: K201863
Trade/Device Name: Tumark Vision, Tumark Professional, Tumark Q, Tumark Professional Q-Shape Regulation Number: 21 CFR 878.4300 Regulation Name: Implantable Clip Regulatory Class: Class II Product Code: NEU Dated: July 3, 2020 Received: July 6, 2020
Dear Dr. Vasmer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Cindy Chowdhury, Ph.D., M.B.A. Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201863
Device Name
Tumark Vision, Tumark Professional, Tumark Professional Q-Shape
Indications for Use (Describe)
The Tumark® Vision, the Tumark® Professional, the Tumark® Professional Q-Shape are intended to attach a marker to soft breast tissue and axillary lymph nodes, following an open or a percutaneous procedure to radiographically mark the location of the surgical site. It is not indicated to be used with magnetic resonance imaging (MRI) techniques.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;"><b> Prescription Use (Part 21 CFR 801 Subpart D) </b></span> | <span style="text-decoration: underline;"><b> Over-The-Counter Use (21 CFR 801 Subpart C) </b></span> |
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
## DATE OF APPLICATION:
## 18.02.2021
APPLICANT:
SOMATEX Medical Technologies GmbH Hohenzollerndamm 150/151 14199 Berlin Germany Tel: + 49 (0) 30 319 82 25 00 Fax: +49 (0) 30 319 82 25 99 E-Mail: info@somatex.com
CONTACT PERSON:
David Vasmer, PhD Regulatory Affairs Manager Tel.: +49 (0) 30 319 82 25-49 E-Mail: D.Vasmer@somatex.com
{4}------------------------------------------------
#### 1 Device Name
| Trade Name: | Tumark Vision |
|-----------------------------|-----------------------------------|
| | Tumark Professional |
| | Tumark Q |
| | Tumark Professional Q-Shape |
| Common Name: | Tissue Site Marking System |
| Device Classification Name: | Marker, Radiographic, Implantable |
#### Classification / Product Code 2
The subjected devices can be classified according to following device name and product code:
| Device | Regulation<br>Description | Regulation<br>Medical<br>Specialty | Review Panel | Product Code | Regulation<br>Number | Device<br>Classification |
|-----------------------------------------|---------------------------|------------------------------------|------------------------------|--------------|----------------------|--------------------------|
| Marker,<br>Radiographic,<br>Implantable | Implantable clip | General & Plastic<br>Surgery | General & Plastic<br>Surgery | NEU | 2 | 878.4300 |
#### 3 Predicate Device / Reference Device
| New Device | Predicate Device | Reference Device | 510(k) Number | 510(k) Holder |
|----------------------------------------------------------------|---------------------|----------------------------------------|---------------|--------------------------------------|
| Tumark Vision | Tumark Professional | --- | K073095 | SOMATEX Medical<br>Technologies GmbH |
| Tumark Professional<br>Tumark Q<br>Tumark Professional Q-Shape | --- | Tumark Vision | K180443 | SOMATEX Medical<br>Technologies GmbH |
| | --- | UltraCor Twirl Breast<br>Tissue Marker | K180061 | Bard Peripheral Vascular, Inc. |
#### 4 Device Description
#### 4.1 Tumark Vision
The Tumark" Vision is a sterile product for single use only and consists of a non-absorbable nickeltitanium clip marker, an introducer cannula and a plastic handle. When new and unopened, the clip marker is contained within the cannula. The cannula tip is bevelled to help insertion, has markings 1 cm apart for measuring the depth of penetration, and a textured surface behind the cannula tip. The handle is equipped with a slide button which allows one handed placement of the marker by pressing it forward. A safety catch system prevents the slide button from inadvertently moving forward and therefore prevents premature deployment of the marker has a spherical shape. The symbol of the clip marker shape is depicted on the handle.
#### 4.2 Tumark Professional
The Tumark® Professional is a sterile product for single use only and consists of a non-absorbable nickel-titanium clip marker, an introducer cannula, and a plastic handle. When new and unopened, the clip marker is contained within the cannula. The cannula tip is bevelled to help insertion, has markings 1 cm apart for measuring the depth of penetration, and a textured surface behind the cannula tip. The handle is equipped with a slide button which allows one handed placement of the marker by pressing it forward. A safety catch system prevents the slide button from inadvertently moving forward and therefore premature deployment of the marker. The clip marker has a U-shape. The symbol of the clip marker shape is depicted on the handle.
#### 4.3 Tumark Q
The Tumark® Q is a sterile product for single use only and consists of a non-absorbable nickel-titanium clip marker, an introducer cannula and a plastic handle. When new and unopened, the clip marker is contained within the
{5}------------------------------------------------
cannula. The cannula tip is bevelled to help insertion, has markings 1 cm apart for measuring the depth of penetration, and a textured surface behind the cannula tip. The clip marker can be placed using one hand by pushing the slide button on the plastic handle forward once the fixing clip has been removed. A safety catch system prevents the slide button from inadvertently moving forward and therefore prevents premature deployment of the clip marker. The clip marker has a Q-shape.
#### 4.4 Tumark Professional Q Shape
The Tumark® Professional Q Shape is a sterile product for single use only and consists of a non-absorbable nickelttanium clip marker, an introducer cannula and a plastic handle. When new and unopened, the clip marker is contained within the cannula tip is bevelled to help insertion, has markings 1 cm apart for measuring the depth of penetration, and a textured surface behind the cannula tip. The handle is equipped with a slide button which allows one handed placement of the marker by pressing it forward. A safety catch system prevents the slide button from inadvertently moving forward and therefore prevents premature deployment of the clip markers for the Tumark® Professional are Q-shaped. The symbol of the clip marker shape is depicted on the handle.
#### 5 Indication for Use
The Tumark® Vision, the Tumark® Professional, the Tumark® Professional Q-Shape are intended to attach a marker to soft breast tissue and axillary lymph nodes, following an open or a percutaneous procedure to radiographically mark the location of the surgical site. It is not indicated to be used with magnetic resonance imaging (MRI) techniques.
#### 6 Duration of Use
The clip marker is a permanent implant (> 30 days).
#### 7 Technological Characteristics
The technological characteristics of Tumark "Professional, Tumark" Q and Tumark" Professional Q-Shape are the same as the technological characteristics of the predicate device.
| Company | SOMATEX Medical<br>Technologies GmbH<br>--<br>Tumark Vision,<br>Tumark Professional<br>Tumark Q<br>Tumark Professional<br>Q-Shape<br>(New Device) | SOMATEX Medical<br>Technologies GmbH<br>--<br>Tumark Professional<br>(Predicate Device) | SOMATEX Medical<br>Technologies GmbH<br>--<br>Tumark Vision<br>(Reference Device) | Bard Peripheral<br>Vascular, Inc.<br>--<br>UltraCor Twirl Breast<br>Tissue Marker<br>(Reference Device) | Result |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|
| Device Name | Tumark Vision,<br>Tumark Professional<br>Tumark Q<br>Tumark Professional<br>Q-Shape | Tumark Professional | Tumark Vision | UltraCor Twirl Breast<br>Tissue Marker | --- |
| Regulation<br>Number | 878.4300 | 878.4300 | 878.4300 | 878.4300 | Substantially<br>Equivalent |
| Class | 2 | 2 | 2 | 2 | Substantially<br>Equivalent |
| Code | NEU | NEU | NEU | NEU | Substantially<br>Equivalent |
| 510(k) number | --- | K073095 | K180443 | K180061 | --- |
| Company | SOMATEX Medical<br>Technologies GmbH<br>---<br>Tumark Vision,<br>Tumark Professional<br>Tumark Q<br>Tumark Professional<br>Q-Shape<br>(New Device) | SOMATEX Medical<br>Technologies GmbH<br>---<br>Tumark Professional<br>(Predicate Device) | SOMATEX Medical<br>Technologies GmbH<br>Tumark Vision<br>(Reference Device) | Bard Peripheral<br>Vascular, Inc.<br>UltraCor Twirl Breast<br>Tissue Marker<br>(Reference Device) | Result |
| Indication for Use | The Tumark® Vision,<br>the Tumark®<br>Professional, the<br>Tumark® Q and the<br>Tumark® Professional<br>Q-Shape are intended<br>to attach a marker to<br>soft breast tissue and<br>axillary lymph nodes,<br>following an open or a<br>percutaneous<br>procedure to<br>radiographically mark<br>the location of the<br>surgical site. It is not<br>indicated to be used<br>with magnetic<br>resonance imaging<br>(MRI) techniques. | The Tumark<br>Professional is<br>intended to attach a<br>marker to soft tissue<br>at the surgical site<br>during an open or a<br>percutaneous<br>procedure. It is<br>indicated for use to<br>radiographically and<br>radiologically mark the<br>surgical location in<br>breasts following an<br>open or percutaneous<br>procedure. It is not<br>intended to be used<br>with magnetic<br>resonance imaging<br>(MRI) techniques. | The Tumark Vision is<br>intended to attach a<br>marker to soft tissue<br>at the surgical site<br>during an open or a<br>percutaneous<br>procedure. It is<br>indicated for use to<br>radiographically and<br>radiologically mark the<br>surgical location in<br>breasts following an<br>open or percutaneous<br>procedure. It is not<br>intended to be used<br>with magnetic<br>resonance imaging<br>(MRI) techniques. | The UltraCor® Twirl®<br>Breast Tissue Marker<br>is intended for use to<br>attach to soft breast<br>tissue, including<br>axillary lymph nodes,<br>to radiographically<br>mark the location of<br>the biopsy procedure. | Substantially<br>Equivalent |
| Design | Sterile, single use,<br>preloaded tissue site<br>marking systems<br>consisting of a non-<br>absorbable nickel-<br>titanium marker, a<br>cannula and a plastic<br>handheld applier with<br>deployment<br>mechanism. | Sterile, single use,<br>preloaded tissue site<br>marking systems<br>consisting of a non-<br>absorbable nickel-<br>titanium marker, a<br>cannula and a plastic<br>handheld applier with<br>deployment<br>mechanism. | Sterile, single use,<br>preloaded tissue site<br>marking systems<br>consisting of a non-<br>absorbable nickel-<br>titanium marker, a<br>cannula and a plastic<br>handheld applier with<br>deployment<br>mechanism. | Sterile, single use,<br>preloaded tissue site<br>marking systems<br>consisting of a non-<br>absorbable nickel-<br>titanium marker, a<br>cannula and a plastic<br>handheld applier with<br>deployment<br>mechanism. | Substantially<br>Equivalent |
| Marker Design | Sphere-shaped; U-<br>shaped, Q-shaped | U-shaped | Sphere-shaped | Ring-shaped | Substantially<br>Equivalent |
| Marker Material | Nitinol | Nitinol | Nitinol | Nitinol | Substantially<br>Equivalent |
| Cannula Design | Sharp tip, markings on<br>the cannula and<br>puncture function | Sharp tip, markings on<br>the cannula and<br>puncture function | Sharp tip, markings on<br>the cannula and<br>puncture function | Sharp tip, markings on<br>the cannula and<br>puncture function | Substantially<br>Equivalent |
| Cannula length<br>[mm] | 70 / 100 / 120 / 150 | 70 / 100 / 120 | 100 / 120 | 100 | Substantially<br>Equivalent |
| Gauge [G] | 18 | 18 | 18 | 17 | Substantially<br>Equivalent |
| Cannula Material | stainless steel | stainless steel | stainless steel | stainless steel | Substantially<br>Equivalent |
| Handle | One-handed<br>application with safety<br>function to prevent<br>premature<br>deployment of the<br>marker | One-handed<br>application with safety<br>function to prevent<br>premature<br>deployment of the<br>marker | One-handed<br>application with safety<br>function to prevent<br>premature<br>deployment of the<br>marker | One-handed<br>application with safety<br>function to prevent<br>premature<br>deployment of the<br>marker | Substantially<br>Equivalent |
| Handle Material | stainless steel<br>plastic material | stainless steel<br>plastic material | stainless steel<br>plastic material | plastic material | Substantially<br>Equivalent |
#### 7.1 Device Characteristics Table
{6}------------------------------------------------
#### Summary of Technological Characteristics 7.2
The proposed devices are similar in terms of design, indication for use and have similar technological characteristics as the predicate device and reference devices. It can be stated, that Tumark Vision, Tumark Q
{7}------------------------------------------------
and Tumark Professional Q-Shape are considered substantially equivalent to the predicate device/reference devices.
#### 8 Performance Data
To demonstrate that Tumark Vision, Tumark Q and Tumark Professional Q-Shape are as safe and effective as the predicate and reference devices, their technological characteristics and performance criteria were evaluated. The new devices were evaluated in vitro testing. Specific aspects included:
#### 8.1 Biocompatibility
The components and manufacturing processes of the subjected devices are similar to the predicate device Tumark Professional (K073095) and the reference device Tumark Vision (K180443). The devices are made from standard materials. EO residual values are far below the acceptable limits. In addition, cytotoxicity testing and the toxicological review confirmed the absence of toxic leachables and contaminants. Therefore, the requirements of ISO 10993-1 are met and further testing is deemed not necessary.
#### 8.2 Sterilization and Shelf Life
The devices are delivered in sterile state. All acceptance criteria are met during the sterility testing. Furthermore, the defined shelf-life could be proven based on packaging and device testing after real-time ageing. All defined acceptance criteria were met during shelf life testing.
#### 8.3 Bench Testing
Bench testing was performed to validate the device design. Specific aspects include:
| Aspect | Test Method | Results |
|-----------------------------------|-------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| Device Function | Confirm that the marker can be placed in the<br>target area | U-, Q- and Vision markers could be deployed.<br>The device performs as intended.<br>All acceptance criteria are met. |
| Device Performance | Confirm that the clip marker and cannula are<br>recognized | Clip markers and cannulas are recognized in<br>ultrasound, mammography and MR imaging.<br>All acceptance criteria are met. |
| Device Stability during transport | Drop tests | Devices were not damaged.<br>All acceptance criteria are met. |
All acceptance criteria were met during the bench testing. The devices perform as intended and are as safe and as effective as the predicate device.
#### 8.4 Clinical Analysis
A literature review was performed to clinically support the use of the markers inside axillary lymph nodes. This literature review confirms the defined indication for use. Furthermore, physician statements were obtained to support the indication for use.
#### ਰੇ Substantial Equivalence Summary / Conclusion
Based on available 510(k) information herein provided, Tumark Professional, Tumark Q and Tumark Professional Q-Shape are considered substantially equivalent to the predicate devices in terms of intended use, technology and performance specifications. There are no differences between the devices which may raise new issues concerning safety or effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.