This instrument is intended to be used with a video system center, light source, documentation equipment, monitor. hand instruments, electrosurgical unit, and other ancillary equipment for endoscopic surgery within the thoracic and abdominal cavities including female reproductive organs.
Device Story
The ENDOEYE FLEX LTF-S190-5 is a deflectable videoscope used in thoracic and abdominal endoscopic surgery. It captures visual data via a distal objective lens and light guide, transmitting images to a video system center for display on a monitor. Operated by surgeons in healthcare facilities, the device provides visualization to assist in surgical procedures. It features a deflectable tip for maneuverability and supports Narrow Band Imaging (NBI). The device is a repeat-use, non-sterile instrument requiring reprocessing between uses. It integrates with standard surgical ancillary equipment (electrosurgical units, monitors). By providing high-quality visualization, it enables surgeons to perform minimally invasive procedures, potentially reducing patient trauma compared to open surgery.
Clinical Evidence
Bench testing only. Performance data included reprocessing validation, biocompatibility (ISO 10993-5, -10, -11), electrical safety (ANSI/AAMI ES 60601-1, IEC 60601-2-18), EMC (IEC 60601-1-2), and bench performance testing (thermal safety, durability, imaging performance, photobiological safety). All acceptance criteria were met.
Technological Characteristics
Deflectable videoscope; distal end Ø 5.4 mm; 85° FOV; 0° direction of view; 18-100 mm depth of field; 1-chip color CCD; NBI support. Materials: Fluoro Rubber (bending rubber), Stainless steel (distal end/tube/pin), Epoxy Resin (glue), Glass (lens). Sterilization: Ethylene oxide, Autoclave, Hydrogen peroxide (Sterrad 100S/NX). Connectivity: Proprietary interface to Olympus video system center.
Indications for Use
Indicated for use in endoscopic surgery within the thoracic and abdominal cavities, including female reproductive organs, in conjunction with compatible video system centers, light sources, and ancillary surgical equipment.
Regulatory Classification
Identification
A gynecologic laparoscope is a device used to permit direct viewing of the organs within the peritoneum by a telescopic system introduced through the abdominal wall. It is used to perform diagnostic and surgical procedures on the female genital organs. This generic type of device may include: Trocar and cannula, instruments used through an operating channel, scope preheater, light source and cables, and component parts.
Predicate Devices
ENDOEYE FLEX 3D DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-190-10-3D (K123365)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 17, 2021
Olympus Medical Systems Corp. % Jon Gilbert Regulatory Affairs Consultant for Olympus Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610
K201832 Re:
> Trade/Device Name: ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-5 Regulation Number: 21 CFR§ 884.1720 Regulation Name: Gynecologic Laparoscope and Accessories Regulatory Class: II Product Code: HET, GCJ, NWB Dated: August 13, 2021 Received: August 16, 2021
Dear Jon Gilbert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
# Jason Roberts -S
Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K201832
Device Name ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-5
Indications for Use (Describe)
This instrument is intended to be used with a video system center, light source, documentation equipment, monitor. hand instruments, electrosurgical unit, and other ancillary equipment for endoscopic surgery within the thoracic and abdominal cavities including female reproductive organs.
Type of Use (Select one or both, as applicable)
| <span style="font-size:14px">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size:14px">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The word is positioned above a thick, horizontal, yellow line. The font is sans-serif and the letters are closely spaced together. There is a registered trademark symbol to the right of the word.
## 510(k) Summary
## I. GENERAL INFORMATION
| ■ 510(k) Submitter: | OLYMPUS MEDICAL SYSTEMS CORP. |
|---------------------|--------------------------------------------------|
| | 2951 Ishikawa-cho, Hachioji-shi, Tokyo 192-8507, |
| | Japan |
- Contact Person: Jon Gilbert c/o Olympus Corporation of the Americas 3500 Corporate Parkway PO Box 610 Center Valley, PA 18034-0610, USA Phone: 906-361-3237 Fax: 484-896-7128 Email: jgilb.raca@gmail.com
- Aizu Olympus Co., Ltd., ■ Manufacturing site: 500 Muranishi, Niidera, Monden-machi, Aizuwakamatsu-shi, Fukushima 965-8520, Japan
Date Prepared: September 14, 2021
## II. DEVICE IDENTIFICATION
| ■ Device Name | ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE<br>OLYMPUS LTF-S190-5 |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|
| ■ Common Name | VIDEOSCOPE |
| ■ Regulation Number | 21 CFR 884.1720 |
| ■ Regulation Name | Gynecologic laparoscope and accessories |
| ■ Regulatory Class | II |
| ■ Product Code | HET; Laparoscope, Gynecologic (And Accessories)<br>GCJ; Laparoscope, General & Plastic Surgery<br>NWB; Endoscope, Accessories, Narrow Band Spectrum |
- Classification Panel Obstetrics/Gynecology
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the Olympus logo in blue, with a gold line underneath. To the right of the logo, the text "Traditional 510(k) Notification" is present. Below the logo, the text "ENDOEYE FLEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-5" is displayed.
## III. PREDICATE DEVICE
| Device name | 510(k) Submitter | 510(k) No. |
|--------------------------------------------------------------------|----------------------------------|------------|
| ENDOEYE FLEX 3D<br>DEFLECTABLE VIDEOSCOPE<br>OLYMPUS LTF-190-10-3D | OLYMPUS MEDICAL<br>SYSTEMS CORP. | K123365 |
The predicate device has not been subject to a design-related recall.
## IV. DEVICE DESCRIPTION
## ■ General Description of the subject device
This instrument has been designed to be used with a video system center, light source, documentation equipment, monitor, hand instruments, electrosurgical unit, and other ancillary equipment for endoscopy and endoscopic surgery within the thoracic and abdominal cavities including female reproductive organs.
## Principle of Operation
The LTF-S190-5 is a video endoscope consisting of three parts: the control section, the insertion section, and the connector section. The device uses an objective lens and light guide lens for capturing images/videos from the distal end of the endoscope.
#### V. INDICATIONS FOR USE
## ENDOBYE ELEX DEFLECTABLE VIDEOSCOPE OLYMPUS LTF-S190-5
This instrument is intended to be used with a video system center, light source, documentation equipment, monitor, hand instruments, electrosurgical unit, and other ancillary equipment for endoscopy and endoscopic surgery within the thoracic and abdominal cavities including female reproductive organs
#### COMPARISON OF TECHNOLOGY CHARACTERISTICS WITH VI. THEPREDICATE DEIVCE
The detailed comparison chart between subject device and predicate devices are as follows.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are closely spaced together. The word appears to be a logo or brand name, and it is likely associated with a company or product.
| | Subject Device (SD) | Predicate Device (PD) |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Item | LTF-S190-5 | LTF-190-10-3D |
| Indications for Use | This instrument is intended to<br>be used with a video system<br>center, light source,<br>documentation equipment,<br>monitor, hand instruments,<br>electrosurgical unit, and other<br>ancillary equipment for<br>endoscopy andendoscopic<br>surgery within the thoracic<br>and abdominal cavities<br>including female reproductive<br>organs. | This instrument is intended to be<br>used with Olympus video system<br>center, light source, documentation<br>equipment, 3D processor, monitor,<br>hand instruments, electrosurgical<br>unit, and other ancillary equipment<br>for endoscopy and endoscopic<br>surgery. This instrument is<br>indicated for use within the thoracic<br>and abdominal cavities including<br>female reproductive organs.<br>This instrument must not be used<br>for observation or treatment of the<br>heart and must not contact the heart<br>or any area near the heart. In<br>addition, this instrument must not<br>come into contact with any device<br>or therapeutic accessory that<br>contacts the heart or any area near<br>the heart. |
| Environment of use | Healthcare facility/hospital | Healthcare facility/hospital |
| Single/ Repeat use | Repeat use | Repeat use |
| Sterile/ Non-sterile | Non-sterile | Non-sterile |
| Sterilization method | Ethylene oxide, Autoclave,<br>Hydrogen peroxide (Sterrad<br>100S/ NX) | Ethylene oxide, Hydrogen peroxide<br>(Sterrad NX) |
| Field of View | 85° | 80° |
| Direction of View | 0° (Forward viewing) | 0° (Forward viewing) |
| Depth of Field | 18-100 mm | 18-100 mm |
| Type of CCD | Color CCD | Color CCD |
| Number of CCD Chip | 1 | 2 |
| Narrow Band Imaging<br>observation | Available | Available |
| Insertion Tube<br>Diameter | Distal end size: Ø 5.4 mm<br>Insertion tube outer Diameter<br>(Maximum): Ø5.4 mm | Distal end size: Ø 10 mm<br>Insertion tube outer Diameter<br>(Maximum): Ø 10 mm |
| Working Length | 370 mm | 370 mm |
| Channel inner diameter | No instrument channel | No instrument channel |
| Angulation Range | U/D:100°/100°<br>R/L:100°/100° | U/D:100°/100°<br>R/L:100°/100° |
| Item | Subject Device (SD)<br>LTF-S190-5 | Predicate Device (PD)<br>LTF-190-10-3D |
| Patient contacting<br>material | Bending rubber: Fluoro Rubber<br>Distal end/ Distal cover/<br>Insertion tube/ Pin: Stainless steel<br>Glue: Epoxy Resin<br>Lens: Glass | Bending rubber: Fluoro Rubber<br>Distal end / Insertion tube: Stainless steel<br>Glue: Epoxy Resin<br>Lens: Glass |
| Duration and type of<br>contact | External communicating device in contact with mucosal membranes, damaged mucosal membranes and tissue/bone/dentin.<br>The contact duration is limited exposure (i.e. contact is up to 24 hours). | External communicating device in contact with mucosal membranes, damaged mucosal membranes and tissue/bone/dentin.<br>The contact duration is limited exposure (i.e. contact is up to 24 hours). |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the word "OLYMPUS" in large, bold, blue letters. The font is sans-serif and the letters are slightly blurred, giving the word a sense of depth. A small registered trademark symbol appears to the right of the letter "S".
K201832 Page 4 of 5
The indications for use of the subject device do not raise different questions of safety and effectiveness when compared to the predicate device. The subject and predicate devices have the same intended use of providing visualization during thoracic and abdominal surgery, including the female reproductive organs. The differences in technological characteristics do not raise new questions of safety and effectiveness.
## VII. PERFORMANCE DATA
The following performance data were provided in support of the substantial equivalence determination.
## 1) Reprocessing validation testing
All reprocessing methods including the Sterilizer/Sterilization process were validated in accordance with the FDA guidance "Reprocessing Medical Devices in Health Care Settings: Validation Methods and Labeling" issued on March 17, 2015. The acceptance criteria were met.
## 2) Biocompatibility testing
Biocompatibility testing for the LTF-S190-5 final product was conducted in accordance with the FDA's Guidance for Industry and Food and Drug Administration Staff, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process''.
- Cytotoxicity Study Using the Colony Assay (ISO 10993-5:2009) ।
- i Intracutaneous Study in Rabbits (ISO 10993-10:2010)
- Guinea Pig Maximization Sensitization Test (ISO 10993-10:2010) ।
- Acute Systemic Toxicity Study in Mice (ISO 10993-11:2006) i
{7}------------------------------------------------
- -Rabbit Pyrogen Study (USP <151>, USP <37>)
The device was shown to be non-cytotoxic, non-irritating, non-sensitizing, not systemically toxic, and non-pyrogenetic.
## 3) Software verification and validation testing
Software verification and validation testing for the LTF-S190-5 was conducted and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" and "Content of Premarket Submissions for Management of Cybersecurity in Medical Devices". The device met all requirements.
## 4) Electrical safety and electromagnetic compatibility (EMC)
Electrical safety and EMC testing were conducted on the LTF-S190-5. The system complies with the ANSI/AAMI ES 60601-1:2005/(R)2012 and A1:2012 and IEC 60601-2-18:2009 standards for safety and the IEC 60601-1-2:2014 standards for EMC. The device met all requirements.
## 5) Performance testing - Bench
Bench testing for the LTF-S190-5 as listed below was conducted to ensure that the subject device performs as intended and meet design specifications. Device performance was assessed the design requirements, and included process verification, design verification, and design validation. The device met all acceptance criteria.
- Thermal safety test
- Composite durability test
- Color (imaging) performance
- Photobiological Safety
## VIII. CONCLUSIONS
The indications for use and technological characteristics of the LTF-S190-5 do not raise different questions of safety and effectiveness as compared to the predicate. The performance testing demonstrate that the subject device is as safe and effective as the predicate. Therefore, and the subject device is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.