SensiCare PI Evolution Surgical Glove (Tested for Use with Chemotherapy Drugs)

K201829 · Medline Industries, Inc. · KGO · Sep 9, 2020 · General, Plastic Surgery

Device Facts

Record IDK201829
Device NameSensiCare PI Evolution Surgical Glove (Tested for Use with Chemotherapy Drugs)
ApplicantMedline Industries, Inc.
Product CodeKGO · General, Plastic Surgery
Decision DateSep 9, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 878.4460
Device ClassClass 1

Indications for Use

The surgeon's glove is a disposable device made of synthetic rubber latex intended to be worn by operating room personnel to protect a surgical wound from contamination. In addition, these gloves were tested for use for use with chemotherapy drugs in accordance with ASTM D6978 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs.

Device Story

SensiCare PI Evolution Surgical Glove is a disposable synthetic rubber latex glove; designed for use by operating room personnel to protect surgical wounds from contamination. Device provides a physical barrier; tested for permeation resistance against various chemotherapy drugs per ASTM D6978. Users must avoid use with Carmustine due to low breakthrough time. Output is a protective barrier; clinical benefit is reduction of contamination risk during surgical procedures and chemotherapy handling.

Clinical Evidence

Bench testing only. Permeation testing performed for various chemotherapy drugs per ASTM D6978. Breakthrough detection times reported for 22 drugs; most exceeded 240 minutes, with specific exceptions for Carmustine and Thiotepa.

Technological Characteristics

Synthetic rubber latex surgical glove. Tested for chemotherapy drug permeation per ASTM D6978. Disposable, non-powdered.

Indications for Use

Indicated for use by operating room personnel as a protective barrier for surgical wounds. Tested for use with specific chemotherapy drugs.

Regulatory Classification

Identification

A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. September 9, 2020 Medline Industries, Inc. Jennifer Mason Senior Regulatory Affairs Specialist Three Lakes Drive Northfield, Illinois 60030 Re: K201829 Trade/Device Name: SensiCare PI Evolution Surgical Glove (Tested for Use with Chemotherapy Drugs) Regulation Number: 21 CFR 880.6250 Regulation Name: Non-Powdered Patient Examination Glove Regulatory Class: Class I. reserved Product Code: LZC, KGO Dated: June 29, 2020 Received: July 2, 2020 Dear Jennifer Mason: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely. Elizabeth F. Claverie-Williams, MS Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ ## Indications for Use 510(k) Number (if known) K201829 Device Name SensiCare PI Evolution Surgical Glove (Tested for Use with Chemotherapy Drugs) Indications for Use (Describe) The surgeon's glove is a disposable device made of synthetic rubber latex intended to be worn by operating room personnel to protect a surgical wound from contamination. In addition, these gloves were tested for use for use with chemotherapy drugs in accordance with ASTM D6978 Standard Practice for Assessment of Medical Gloves to Permeation by Chemotherapy Drugs. The following chemicals have been tested with these gloves. Chemotherapy Drug Permeation (Breakthrough Detection Time) in Minutes Bleomycin (15mg/ml) >240 minutes Busulfan (6 mg/ml) >240 minutes Carboplatin (10.0 mg/ml) >240 minutes Carmustine (3.3 mg/ml) 24.3 minutes (24.3, 24.6, 25.5) Cisplatin (1.0 mg/ml) >240 minutes Cyclohosphamide (Cytoxan) (20 mg/ml) >240 minutes Dacarbazine (DTIC) (10.0 mg/ml) >240 minutes Doxorubicin Hydrochloride (2.0 mg/ml) >240 minutes Epirubicin (Ellence) (2 mg/ml) >240 minutes Etoposide (Toposar) (20.0 mg/ml) >240 minutes Fludarabine (25.0 mg/ml) >240 minutes Fluorouracil (50.0 mg/ml)>240 minutes Idarubicin (1.0 mg/ml) >240 minutes Ifosfamide (50.0 mg/ml) >240 minutes Mechlorethamine HCI (1.0 mg/ml) >240 minutes Melphalan (5 mg/ml) >240 minutes WARNING: DO NOT USE WITH CARMUSTINE. Methotrexate (25 mg/ml) >240 minutes CAUTION: Testing showed a minimum Mitomycin C (0.5 mg/ml) >240 minutes breakthrough time of 32.6 minutes for Thiotepa Mitoxantrone (2.0 mg/ml) >240 minutes Paclitaxel (Taxol) (6.0 mg/ml) >240 minutes Paraplatin (10 mg/ml) >240 minutes Rituximab (10 mg/ml)>240 minutes Thiotepa (10.0 mg/ml) 32.6 minutes (32.6, 34.7, 45.4) Vincristine Sulfate (1.0 mg/ml) >240 minutes Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) X Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. {3}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...