X-Tack Endoscopic HeliX Tacking System

K201808 · Apollo Endosurgery, Inc. · PKL · Dec 15, 2020 · Gastroenterology, Urology

Device Facts

Record IDK201808
Device NameX-Tack Endoscopic HeliX Tacking System
ApplicantApollo Endosurgery, Inc.
Product CodePKL · Gastroenterology, Urology
Decision DateDec 15, 2020
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The X-Tack™ Endoscopic HeliX Tacking System is intended for approximation of soft tissue in minimally invasive gastroenterology procedures (e.g. closure and healing of ESD/EMR sites, and closing of fistula, perforation or leaks). X-Tack is not intended for hemostasis of acute bleeding ulcers.

Device Story

Sterile, single-use endoscopic device for soft tissue approximation in GI tract. Operates via endoscope (2.8 mm+ working channel); physician deploys independent helix tacks around defect; suture tension closes defect; OverStitch Suture Cinch (PEEK component) secures suture tail via press-fit. Used in clinical settings for closure of ESD/EMR sites, fistulas, perforations, or leaks. Benefits patient by enabling closure of larger defects compared to traditional clips. Implanted construct typically passes in stool within one month.

Clinical Evidence

Animal study (4 pigs, 40 defects in stomach/colon) compared X-Tack to predicate. Primary endpoints: closure rates and healing at 4 weeks. Results: 100% closure at 4 weeks for both; histological analysis confirmed wound healing consistent with predicate. Bench testing verified anchor retention, suture tensile strength, usability, packaging, shelf-life, and MR compatibility.

Technological Characteristics

Helix tack: 316L stainless steel; Cinch: VESTAKEEP i4 (PEEK); Suture: USP 3-0 Polypropylene. Delivery via endoscope (2.8 mm working channel). Sterilization: Ethylene Oxide (EO). MR Conditional (1.5T/3T).

Indications for Use

Indicated for soft tissue approximation in minimally invasive gastroenterology procedures, including closure of ESD/EMR sites, fistulas, perforations, or leaks. Not indicated for hemostasis of acute bleeding ulcers.

Regulatory Classification

Identification

A hemorrhoidal ligator is a device used to cut off the blood flow to hemorrhoidal tissue by means of a ligature or band placed around the hemorrhoid.

Special Controls

*Classification.* Class II (special controls). Except for a hemostatic metal clip intended for use in the gastrointestinal tract, the device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the seal of the Department of Health & Human Services. To the right of that is the FDA logo in blue, with the words "U.S. FOOD & DRUG" stacked on top of the word "ADMINISTRATION". December 15, 2020 Apollo Endosurgery, Inc. Natalie Allen Regulatory Affairs Specialist 1120 S. Capital of Texas Hwy Austin, Texas 78749 Re: K201808 Trade/Device Name: X-Tack Endoscopic HeliX Tacking System Regulation Number: 21 CFR 876.4400 Regulation Name: Hemorrhoidal Ligator Regulatory Class: Class II Product Code: PKL, OCW Dated: November 5, 2020 Received: November 6, 2020 Dear Natalie Allen: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part {1}------------------------------------------------ 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Shanil P. Haugen, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K201808 Device Name X-Tack Endoscopic HeliX Tacking System #### Indications for Use (Describe) The X-Tack™ Endoscopic HeliX Tacking System is intended for approximation of soft tissue in minimally invasive gastroenterology procedures (e.g. closure and healing of ESD/EMR sites, and closing of fistula, perforation or leaks). X-Tack is not intended for hemostasis of acute bleeding ulcers. | Type of Use (Select one or both, as applicable) | | |----------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------| | <span style="font-size:100%;"> Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size:100%;"> Over-The-Counter Use (21 CFR 801 Subpart C)</span> | ### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. ### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ Image /page/3/Picture/0 description: The image is a logo for Apollo Endosurgery. The logo features a stylized human figure with a blue circle for a head and a blue spiral for a body. The text "apollo" is written in blue, sans-serif font to the right of the figure. Below "apollo" is the word "endosurgery" in a smaller, blue, sans-serif font. 1120 S. Capital of TX Hwy, Bldg. 1, Ste. 300 Austin, Texas 78746 USA Phone: 512.279.5100 FAX: 512.279.5105 ## 510(k) Summary | Owner's Name & Address: | Apollo Endosurgery<br>1120 S. Capital of Texas Hwy.<br>Building 1, Suite 300<br>Austin, TX 78746 | |-------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | David M. Hooper, PhD<br>Vice President, Quality and Regulatory Affairs<br>Phone: (512) 279-5100<br>Email : david.hooper@apolloendo.com | | Date: | December 10, 2020 | | Trade Name: | X-Tack™ Endoscopic HeliX Tacking System | | Common Name: | Endoscopic Tissue Approximation Device | | Product Code: | PKL, OCW | | Classification: | Class II (21 CFR 876.4400) | | Classification Name | Hemorrhoidal Ligator | | Predicate Devices: | K151802 - Resolution™ 360 Clip | | Device Description | X-Tack™ Endoscopic HeliX Tacking System<br>The X-Tack™ Endoscopic HeliX Tacking System is a sterile,<br>single-use device that enables the user to approximate soft tissue<br>in the gastrointestinal (GI) tract using helix tacks and a 3-0 suture<br>through a 2.8 mm or larger working channel of an endoscope (e.g.<br>gastroscope or colonoscope). | | | OverStitch Suture Cinch<br>The Overstitch Suture Cinch device is comprised of thermoplastic<br>and stainless steel materials and includes an implantable PEEK<br>Cinch component designed to secure and cut the suture once<br>tissue approximation is complete. It is the final step of the X-Tack<br>procedure. The device functions by squeezing the handle and<br>deploying the PEEK components, which form a press-fit onto the<br>tail end of the suture to maintain suture position in situ. | | | Indications for Use: The X-Tack™ Endoscopic HeliX<br>Tacking System is intended for approximation of soft tissue in<br>minimally invasive gastroenterology procedures (e.g. closure and<br>healing of ESD/EMR sites, and closing of fistula, perforation or<br>leaks). X-Tack is not intended for hemostasis of acute bleeding<br>ulcers | {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Apollo Endosurgery. The logo features a stylized figure of a person with a spiral body and a circular head, all in blue. To the right of the figure is the word "apollo" in blue, with the word "endosurgery" in a smaller font size underneath it, also in blue. Page 2 of 3 1120 S. Capital of TX Hwy, Bldg. 1, Ste. 300 Austin, Texas 78746 USA Phone: 512.279.5100 FAX: 512.279.5105 K201808 #### Technological Characteristics: The X-Tack™ Endoscopic Helix Tacking System shares technological characteristics similar to the predicate device. These characteristics include: ● Equivalent intended use . Endoscopic delivery of closure device ● Manufactured from materials commonly utlilized for implant devices used in the gastrointestinal (GI) tract . Similar design concept in that a force is delivered by the implant to approximate soft tissue. ● Sterilized using an ethylene oxide (EO) process Equivalent performance in a randomized, controlled animal ● study, specifically evaluating the closure and healing of defects over the course of one month, as evaluated through direct visualization and histological analysis. Basis of Substantial Equivalence: Non-clinical testing was performed to verify anchor retention, suture tensile strength, and endoscope compatbilitity. Testing was performed to validate usability and human factors, packaging, shelf-life and the ability of the device to withstand distribution forces. The sterilization cycle and biocompatibility of the device were validated per recognized ISO standards. MR testing was performed to establish the conditions underwich the device could be safety scanned, as well as the heating and artefact that could be expected during MRI scanning. MR testing included scanning to determine induced forces, followed by laboratory tests to verify safety under maximum scan conditions. A randomized study involving 4 pigs and 40 defects (created in the stomach and colon), was done to compare the closure rates and healing between X-Tack and the predicate. That study demonstrated that the closure rates associated with X-Tack were equivalent to the predicate and that X-Tack enabled larger defects to be closed. Histological analysis of defects showed both devices resulted in healing of defects. Based on the non-clinical and animal testing, the device is as safe and effective, and performs as well or better than the legally marketed predicate device. | | X-Tack | Predicate | |------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Principle of Operation | • Device is inserted into an endoscope<br>and deployed at target site.<br>• Helix tacks are deployed<br>independently around the defect.<br>• Implanted construct closes the defect<br>through metallic anchors and suture<br>tension. | • Device is inserted into an endoscope<br>and deployed at target site.<br>• Clips are deployed independently<br>around the defect.<br>• Implanted construct consists of<br>metallic arms and spring tension. | ### Summary Table of Equivalence {5}------------------------------------------------ Image /page/5/Picture/0 description: The image contains the logo for Apollo Endosurgery. The logo features a stylized human figure with a circular head and a flowing, curved body. To the right of the figure, the word "apollo" is written in a bold, sans-serif font, with "endosurgery" written in a smaller font directly below it. The logo is rendered in a light blue color. 1120 S. Capital of TX Hwy, Bldg. 1, Ste. 300 Austin, Texas 78746 Austin, Texas 78748 SA SA SA SA SA SA SA SA SA SA SA SA SA SA SA SA SA SA SA SA SA SA SA SA SA SA SA S 12.279 | Endoscope<br>compatability | Scopes with a 2.8mm working channel. | Scopes with a 2.8mm working channel. | |----------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Working length | Gastric: 155 cm<br>Colon: 235 cm | Gastric: 160 cm<br>Colonic: 235 cm | | Respositionable prior<br>to deployment | Yes | Yes | | Physician control<br>over placement and<br>final locking | Yes | Yes | | Expected implant<br>duration. | Approximately 1 month or less, then<br>implant is passed in stool. | Approximately 1 month or less, then<br>implant is passed in stool. | | Clip opening width | Not applicable. Device uses<br>independent tacks, suture and cinch<br>and is not limited by a clip opening<br>dimension. | 11 mm | | Closure efficacy in<br>animal study | 100% closure at 4 weeks.<br>Histologic evaluations of closure were<br>consistent with wound healing. | 100% closure at 4 weeks.<br>Histologic evaluations of closure were<br>consistent with wound healing. | | Sterilization Method | EO | EO | | Usage | Single-use | Single-use | | Implanted Materials | Helix Tack: 316L Stainless Steel<br>Cinch: VESTAKEEP i4 (PEEK)<br>Suture: Polyproylene<br>USP 3-0 Polypropylene<br>Copper Phthalocyannine Blue (Below<br>0.5WT%) in accordance with 21 CFR<br>74, 3045 | Capsule: 304 Stainless Steel<br>Arms: 17-7 PH SS<br>Tension breaker: Luran 968R<br>(polystyrene)<br>Yoke: F75 Cobalt Chrome | | Biocompatibility | Tested per ISO 19993. | Tested per ISO 19993. | | MR Compatibility | MR Conditional with 1.5 and 3 T MR<br>scanners with spatial field gradient<br>of 2500 Gauss/cm (extrapolated or<br>less) and SAR of 2.0 W/kg for 15<br>minutes of continuous scanning | MR Conditional with 1.5 and 3T MR<br>scanners with spatial field gradient<br>of 2500 Gauss/cm (extrapolated or<br>less) and SAR of 2.0 W/kg for 15<br>minutes of continuous scanning |
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...