The 'ANCORIS' PET Scanner is designed to acquire data, process and display images by appropriated trained medical professionals via measuring the distribution of injected positron emitting radiopharmaceuticals in humans for the purpose of determining various metabolic and physiologic functions within the human body.
Device Story
ANCORIS is a stand-alone PET scanner for metabolic and physiologic imaging. It acquires data via a three-detector-ring gantry using Ge68 rod sources for transmission scans. The system integrates a patient handling system (PHS) supporting weights up to 227kg (500lbs) and an integrated workstation for image reconstruction, processing, and display. Operated by trained medical professionals in a clinical setting, the device includes a touch interface for acquisition control and system status monitoring. It features an integrated trigger unit for ECG and respiratory gating. Output images allow clinicians to assess metabolic/physiologic functions, aiding diagnostic decision-making. The device combines technology from the ECAT ACCEL PET scanner and ECAT Scintron PET workstation.
Clinical Evidence
No clinical data provided. Evidence consists of bench testing, performance verification, and validation. Testing included verification of the PHS for higher patient loads, safety testing per IEC 60601-1, and comparison of transmission technology (rod source holder, septa, parking positions, and attenuation correction) against the ECAT ACCEL predicate.
Technological Characteristics
Three-detector-ring PET system; 16.2cm field of view; Ge68 rod source transmission technology; PHS supports up to 227kg; integrated workstation; touch-screen user interface; ECG and respiratory gating trigger unit; conforms to IEC 60601-1 safety standards.
Indications for Use
Indicated for use by trained medical professionals to acquire, process, and display PET images by measuring the distribution of injected positron-emitting radiopharmaceuticals in humans to determine metabolic and physiologic functions. No specific age or disease state contraindications provided.
Regulatory Classification
Identification
An emission computed tomography system is a device intended to detect the location and distribution of gamma ray- and positron-emitting radionuclides in the body and produce cross-sectional images through computer reconstruction of the data. This generic type of device may include signal analysis and display equipment, patient and equipment supports, radionuclide anatomical markers, component parts, and accessories.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the letters "FDA" followed by "U.S. FOOD & DRUG ADMINISTRATION" in a smaller font size.
MiEGmbH % Mr. Thomas Kuehl CEO Hauptstrasse 112 Seth, Schleswig-Holstein 23845 GERMANY
August 27, 2020
Re: K201807
Trade/Device Name: ANCORIS Regulation Number: 21 CFR 892.1200 Regulation Name: Emission computed tomography system Regulatory Class: Class II Product Code: KPS Dated: June 25, 2020 Received: July 1, 2020
Dear Mr. Kuehl:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801 and Part 809); medical device reporting of medical device-related adverse events) (21 CFR
{1}------------------------------------------------
803) for devices or post-marketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K201807
Device Name
ANCORIS
Indications for Use (Describe)
The 'ANCORIS' PET Scanner is designed to acquire data, process and display images by appropriated trained medical professionals via measuring the distribution of injected positron emitting radiopharmaceuticals in humans for the purpose of determining various metabolic and physiologic functions within the human body.
| Type of Use (Select one or both, as applicable) |
|---------------------------------------------------------------------------------------|
| <div> <span> </span> <span>Prescription Use (Part 21 CFR 801 Subpart D)</span> </div> |
| <div> <span> </span> <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for MiE, which is displayed in white text on a blue rectangle. The letters "MiE" are stylized with the "M" and "i" connected at the top, and the "E" is slightly offset to the right. The blue rectangle provides a solid background for the logo, making it stand out.
## Submitter Identification
| Holder / Headquarters: | MiE GmbH<br>Hauptstrasse 112<br>23845 Seth, Schleswig-Holstein<br>Germany |
|----------------------------------|---------------------------------------------------------------------------|
| Telephone Number:<br>Fax Number: | +49 4194 9977 0<br>+49 4194 9977 55 |
| Contact Person: | Thomas Kuehl<br>th.kuehl@miegermany.de |
| Date summary prepared: | 24th August 2020 |
| Registration Number: | 3005128583 |
| Product Identification | |
| Name: | ANCORIS |
| Common Name: | Positron Emission Tomography (PET) Scanner |
| Classification Name: | Emission Computed Tomography System<br>21 C.F.R. § 892.1200 |
Class II
Classification:
# Identification of Legally Market and Equivalent Devices
| 510(k) # | Device | Manufacturer |
|----------|-------------------------|-----------------------|
| K130269 | ECAT Scintron PET | MiE GmbH |
| K002584 | ECAT ACCEL PET Scanner | CTI PET Systems, Inc. |
| K013504 | ECAT LSO PET/CT Scanner | CTI PET Systems, Inc. |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the letters 'MiE' in white against a blue background. The letters are stylized with the 'M' and 'E' being larger and bolder than the 'i'. The overall design is simple and clean, with a focus on the acronym.
### Device Description
The proposed device 'ANCORIS' is a PET system providing volume measurements of metabolic and physiologic processes. It is the same as the legally market device, name 'ECAT ACCEL PET scanner', K002584, from CTI PET Systems (now Siemens Medical), combined with the legally market device, name 'ECAT Scintron PET', K130269 from MiE GmbH.
It is designed to close the economical gap between the demand and availability of stand-alone PET scanner.
'ANCORIS' PET Scanner is designed to perform acquisitions, image The reconstruction, processing, viewing and analysis of data. It uses the same technology and has the same performance as well as indications for use as the leqally market devices.
The system includes the gantry with transmission scan capability, the currently market integrated workstation 'ECAT Scintron PET' (K130269) and the new designed PHS. The system is available as a three detector-ring system with a 16.2cm field of view same as used in currently marketed device ECAT ACCEL PET Scanner (K002584).
#### Indications for Use
The 'ANCORIS' PET Scanner is designed to acquire data, process and display images by appropriated trained medical professionals via measuring the distribution of injected positron emitting radiopharmaceuticals in humans for the purpose of determining various metabolic and physiologic functions within the human body.
### Comparison with Predicate Devices
The 'ANCORIS' PET Scanner is similar in measuring design and function to the legally market device 'ECAT ACCEL PET scanner', K002584, as well as the PET Scanner used in the legally market device, name 'ECAT LSO PET/CT Scanner', K013504. Same as in the 'ECAT ACCEL PET scanner' rotating source are used to perform the transmission scan. The main difference of the current device 'ANCORIS' is a PHS (patient handling system) for higher patient weight.
The 'ANCORIS' PET Scanner will use the same computer hard- and software technology as used in the legally market device, name 'ECAT Scintron PET', K130269.
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo with the letters "MiE" in white against a blue background. The letters are stylized, with the "M" having a sharp, angular design and the "i" having a circular dot above it. The blue background is a solid, uniform color, providing a clean and simple backdrop for the text.
### New changes in the current device
- The patient handling system (PHS) is designed for higher patient weights up to -227kg (500lbs). The mechanical technologies are similar to the legally market device 'ECAT ACCEL PET Scanner', K002584. The PHS pallet is designed for higher patient weights and similar attenuation properties.
- -The trigger unit for ECG and respiratory gating is integrated into the rear of the ANCORIS PHS and moves together with the pallet. The operation is done via the user touch interface at the gantry.
- The graphical user touch interface displays more information of the acquisition and system status including operation of the trigger unit. The touch interface allows a modern usage of the system.
#### Performance verification and validation of the device
The performance of the ANCORIS is successful checked by verification and validation of the ANCORIS system. Furthermore, it is checked if the system meets the specifications and fulfills its intended purpose.
The higher patient load of the PHS and the general requirements for basic safety and essential performance is checked during the IEC 60601-1 tests.
The transmission technology by Ge68 rod sources is similar to transmission technology of the legally market device 'ECAT ACCEL PET Scanner', K002584. This is checked by a comparison of the rod source holder, septa, parking positions, attenuation of PHS pallet and the attenuation correction procedure.
#### Conclusion
The 'ANCORIS' PET System has similar intended use, operating principle and fundamental technologies as legally market devices. The manufacturing, design and development processes of the 'ANCORIS' PET System are conform to currently valid standards including applicable medical device safety and performance.
All test results are, in opinion of MiE GmbH, that the 'ANCORIS' PET System substantially equivalent to the predicated devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.