CT-STOR is a liquid culture media intended for ex vivo cell and tissue processing applications
Device Story
CT-STOR is a serum-free liquid culture medium used for ex vivo preservation, maintenance, and growth of human or animal-derived cells and tissues. It functions by providing essential nutrients to support cellular viability during ex vivo processing. The formulation includes Dulbecco's Modified Eagle Medium (DMEM), adenosine-5'-triphosphate (ATP) encapsulated in lipid vesicles, purified soy phosphatidylcholine, 2,3-Dioleoyloxy-propyl-trimethylammoniumchloride (DOTAP), and ascorbic acid. ATP supports trophic signaling and maintains intracellular energy levels during hypoxia; lipids provide structural components for cell membranes, energy stores, and signaling molecules; ascorbic acid acts as an antioxidant to scavenge reactive oxygen species (ROS) and prevent cellular damage. The medium is used by laboratory personnel in clinical or research settings to support cell survival and development. Healthcare providers use the resulting cultured cells or tissues for downstream clinical applications. The device benefits patients by maintaining the viability and functional integrity of cells and tissues during ex vivo handling.
Clinical Evidence
Bench testing only. Performance was validated using human dermal fibroblasts (HDFa). Cells were cultured in CT-STOR versus control media (fibroblast basal medium) for 48 hours. Proliferation was assessed at 8 and 128 days of media aging. Results showed seeding comparisons of 95.09% and 173.43% respectively, exceeding the >75% acceptance limit.
Technological Characteristics
Serum-free liquid culture medium. Composition: DMEM, ATP (encapsulated in lipid vesicles), purified soy phosphatidylcholine, DOTAP, and ascorbic acid. Aseptically processed. Standalone product. No software or electronic components.
Indications for Use
Indicated for ex vivo cell and tissue processing applications using human or animal-derived cells and tissues.
Regulatory Classification
Identification
Tissue culture media for human ex vivo tissue and cell culture processing applications consist of cell and tissue culture media and components that are composed of chemically defined components (e.g., amino acids, vitamins, inorganic salts) that are essential for the ex vivo development, survival, and maintenance of tissues and cells of human origin. The solutions are indicated for use in human ex vivo tissue and cell culture processing applications.
Special Controls
*Classification.* Class II (special controls): FDA guidance document, “Class II Special Controls Guidance Document: Tissue Culture Media for Human Ex Vivo Processing Applications; Final Guidance for Industry and FDA Reviewers.”
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 20, 2021
Energy Delivery Solutions William Ehringer, Ph.D. CEO 3315 Industrial Parkway Jeffersonville, IN 47130
K201789 Re: Trade/Device Name: CT-STOR Regulation Number: 21 CFR§ 876.5885 Regulation Name: Tissue Culture Media for Human Ex Vivo Tissue and Cell Culture Processing Applications Regulatory Class: II Product Code: NDS Dated: August 19, 2021 Received: August 20, 2021
Dear William Ehringer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{1}------------------------------------------------
You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Glenn B. Bell, Ph.D. Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K201789
Device Name
CT-STOR
Indications for Use (Describe)
CT-STOR is a liquid culture media intended for ex vivo cell and tissue processing applications
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------|----------------------------------------------------------------|
| <span> ☑ Prescription Use (Part 21 CFR 801 Subpart D) </span> | <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 5. 510(k) Summary
| Owner's Name: | Energy Delivery Solutions |
|----------------------|--------------------------------------------------------------------------------------------|
| Address: | 3315 Industrial Parkway<br>Jeffersonville, IN 47130 |
| Phone Number: | (812) 920-0596 |
| Contact Person: | William Ehringer, PhD, Kristyn Smith, or Caitlin Stanton<br>Energy Delivery Solutions |
| Classification Name: | Media, Culture, ex vivo, Tissue and Cell |
| Date of Preparation: | September 11, 2021 |
| Device Name: | CT-STOR |
| Common Name: | Culture Media |
| Regulation Name: | Tissue culture media for human ex vivo tissue and cell culture processing<br>applications. |
| Regulatory Class: | II |
| Product Code: | NDS |
| Regulation Number: | 21 CFR 876.5885 |
| Panel: | Gastroenterology/Urology |
## Legally Marketed Device for Substantial Equivalence Comparison:
CT-STOR is substantially equivalent to the currently marketed device:
- AIM-V® Medium, Invitrogen Corporation; 510(k) K022086 (Invitrogen Corporation) ●
#### Device Description
CT-STOR is a tissue culture media intended for human ex vivo tissue and cell culture processing applications. AIM-V® is the predicate device and is also a tissue culture media intended for human ex vivo tissue and cell culture processing applications. CT-STOR and AIM-V® are composed of chemically defined nutrient materials in solution (with or without supplements) that are essential
Energy Delivery Solutions 510(k) Submission for CT-STOR CONFIDENTIAL 08/19/2021
{4}------------------------------------------------
for the survival and development of tissue or cells of human or other animal origin. These nutrients are provided in liquid form for use in supporting the growth or maintenance of human tissue and cells.
### Intended Use
CT-STOR is a liguid tissue culture media product intended for ex vivo tissue and cell culture processing applications.
### Principles of Operation and Technological Characteristics
CT-STOR is a serum-free medium developed in 2016 to support cells and tissues for ex vivo preservation. CT-STOR is substantially equivalent to AIM-V®, which is a serum-free medium developed in 1987 in collaboration with Dr. Steven Rosenberg (National Cancer Institute) to support adoptive immunotherapy (lymphokine-activated killer (LAK) cells) clinical trials that were being conducted by Dr. Rosenberg and collaborating investigators. Tissue culture media products for human ex vivo tissue and cell culture applications have been explored and tested as described in peer-reviewed literature.
The formula for CT-STOR has remained the same since it was first manufactured with only a few minor alterations to composition. CT-STOR is prepared from a formulation of Dulbeco's Modified Eagle Medium (DMEM), adenosine-5'-triphosphate (ATP), purified soy phosphatidylcholine, 2,3-Dioleoyloxy-propyl-trimethylammoniumchloride (DOTAP) and ascorbic acid (Vitamin C). Each of the components of CT-STOR play a significant role in ex vivo maintenance of cells and tissues.
The central role of ATP in cell and tissue viability and maintenance has been studied since the discovery of ATP in 1929 [Khakh B., 2009]. Slight changes in intracellular ATP can play a significant role in cell function or dysfunction, and thus cellular control of intracellular ATP is critical [Khakh B., 2009]. When cells or tissues are removed from the body, there is a disruption in cell ATP production due to hypoxia [Kim J., 2014]. Addition of ATP to cell and tissue culture media results in trophic signaling and also in helping maintain intracellular ATP levels [Burnstock, G., 2016]. CT-STOR contains micromolar quantities of ATP that are protected and encapsulated in lipid vesicles, which has been shown to increase growth factor expression and decrease inflammatory markers [Castro, A., 2019].
The role of lipids in cell maintenance is three fold. First, lipids play a structural role in cellular membranes [Whitford W., 2004]. A small cell may contain 10° phospholipids in the cell membrane, which if not supplemented in some form will lead to cell death [Alberts B., 2002]. Lipid supplementation to culture media has typically occurred by the addition of serum albumin, which acts as a carrier for lipids [Whitford W., 2004]. In serum-free media, the lipids can come from various sources, such as adding a media supplement (e.g. ExCyte, Millipore), or by adding liposomes, lipid dispersions, or micro-emulsions [Savonniere S., 1996]. Secondly, lipids act as an energy store. For example, a phosphatidylcholine with two palmitic fatty acids yields 258 ATP molecules [Bhagavan, N., 2002]. In some cells, such as cardiomyocytes, the preferred energy source is by beta-oxidation of fatty acids, specifically oleic acid [Lopaschuk, G., 2010]. Third,
{5}------------------------------------------------
lipids play a significant role in cell transport and signal transduction. Lipids are the important in endosomes, golgi vesicles, and vacuoles, as well as in several kinases.
Ascorbic acid plays a significant role in cell and tissue viability during isolation [Shaghaghi, H., 2015]. As a powerful anti-oxidant, ascorbic acid acts as a reactive oxygen species (ROS) scavenger during hypoxia [Guauquil, V., 2004]. When ROS levels in cell and tissue culture media are increased significant destruction of cell membranes, proteins. DNA and RNA does occur. Thus, addition of ascorbic acid to culture media can decrease ROS-induced cell damage.
#### Conclusion
CT-STOR and AIM-V® Medium are used for human ex vivo tissue and cell culture processing applications and have the same principles of operation, technical characteristics, efficacy (generic cellular growth and maintenance) and safety (consistent in chemical content and formulation. biocompatibility with cells, and purity). Their efficacy in supporting survival, growth, development, and maintenance of human cells or tissue culture systems has been well established in scientific publications included in this submission. Both CT-STOR and AIM-V® are manufactured in accordance with OSR requirements and are labelled as aseptically processed. Thus, CT-STOR is substantially equivalent to the legally marketed device intended for the human ex vivo tissue and cell culture processing applications.
#### Indications for Use
CT-STOR is a liquid culture media intended for ex vivo cell and tissue processing applications.
#### Summary of Non-Clinical Testing
CT-STOR has been tested on human dermal fibroblasts (HDFa) cells when demonstrating performance testing for CT-STOR stability (i.e. validation). Primary HDFa cells were purchased from American Type Culture Collection (ATCC, PCS-201-012, Primary Dermal Fibroblast; Normal, Human, Adult (HDFa), Manassas, VA) and plated at 10,000 cells per well on a 24-well plate. Cells were allowed to proliferate in complete growth media for 24 hours (ATCC, PCS-201-030, Fibroblast basal medium with added Fibroblast Growth Kit-Serum-free, PCS-201-040). After 24 hours, cells were switched to either control media (PCS-201-030, Fibroblast basal medium) or test media (CT-STOR, Energy Delivery Solutions, Jeffersonville, IN). HDFa cells were allowed to proliferate for 48 hours with the respective test media. After 48 hours cell proliferation was assessed and seeding comparisons defined (see below table).
| CT-STOR Days Aged | Seeding Comparison (%)<br>Acceptable Limit: > 75% |
|-------------------|---------------------------------------------------|
| 8 | 95.09% |
| 128 | 173.43% |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.