Light Cure Dental Adhesive can be used as a strict self-etch dental bonding agent, or in combination with the traditional total etch or enamel etch procedures: - Direct bonding under a light-cured composite or compomer fillings/resin modified glass ionomer restoratives; - Repair of dental restorations of ceramic, composites and/or metal alloys; - Indirect dental restoration applications in combination with resin cement for cementation of all sorts of dental restorations including crowns, bridges, inlays/onlays, veneers, dental posts and other restorations made from materials of metals/alloys, mental-ceramic and/or porcelain, composites, and their combinations.
Device Story
Light Cure Dental Adhesive is a single-component, ethanol-based, multifunctional methacrylate resin bonding agent. It is applied by a dental professional in a clinical setting to tooth structure or restoration surfaces. The device is light-cured to facilitate adhesion. It functions as a universal bonding agent for direct restorations, intra-oral repairs, and the cementation of indirect restorations. By creating a bond between the tooth structure and restorative materials (composites, ceramics, metals), it ensures the retention and stability of dental restorations, benefiting patients by restoring dental function and aesthetics.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including shear bond strength testing per ISO 29022:2013 and biocompatibility testing (cytotoxicity, sensitization, irritation, systemic toxicity, subchronic toxicity, and genotoxicity) per ISO 10993 standards.
Technological Characteristics
Single-component, ethanol-based, unfilled multifunctional methacrylate resin. Polymerization via visible light curing. Supports total-etch, self-etch, and selective-etch procedures. Complies with ISO 29022:2013 for shear bond strength and ISO 10993 for biocompatibility.
Indications for Use
Indicated for direct bonding of light-cured composite, compomer, or resin-modified glass ionomer restoratives; repair of ceramic, composite, or metal alloy dental restorations; and cementation of indirect restorations (crowns, bridges, inlays/onlays, veneers, posts) made of metal/alloys, metal-ceramic, all-ceramic, porcelain, or composites. Used as a self-etch, total-etch, or selective-etch bonding agent.
Regulatory Classification
Identification
A resin tooth bonding agent is a device material, such as methylmethacrylate, intended to be painted on the interior of a prepared cavity of a tooth to improve retention of a restoration, such as a filling.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 25, 2020
Maggie Zheng, Regulatory Affairs Manager Rizhao HuGe Biomaterials Company, Ltd. No.2 North Zhaoyang Road, District of Donggang Rizhao City, CHINA 276800 Shandong Province
Re: K201787
Trade/Device Name: Light Cure Dental Adhesive Regulation Number: 21 CFR 872.3200 Regulation Name: Resin Tooth Bonding Agent Regulatory Class: Class II Product Code: KLE Dated: July 17, 2020 Received: July 27, 2020
Dear Maggie Zheng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinivas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K201787
Device Name Light Cure Dental Adhesive
Indications for Use (Describe)
Light Cure Dental Adhesive can be used as a strict self-etch dental bonding agent, or in combination with the traditional total etch or enamel etch procedures:
- Direct bonding under a light-cured composite or compomer fillings/resin modified glass ionomer restoratives;
- Repair of dental restorations of ceramic, composites and/or metal alloys;
- Indirect dental restoration applications in combination with resin cement for cementation of all sorts of dental restorations including crowns, bridges, inlays/onlays, veneers, dental posts and other restorations made from materials of metals/alloys, mental-ceramic and/or porcelain, composites, and their combinations.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K201787
# 510 (k) Summary
This summary of 510(k) for the subjective device equivalence information is being submitted in accordance with the requirements of 21 C.F.R. 807.92.
- 1. Date Summary Prepared: September 25, 2020
- 2. Submitter Information:
| Owner's Name | Rizhao HuGe Biomaterials Company, Ltd. |
|----------------|-------------------------------------------------------------------------------------------------|
| Address | No.2 North Zhaoyang Road, District of Donggang, Rizhao City,<br>Shandong Province, China 276800 |
| Telephone | 0086 633 2277268 |
| Fax | 0086 633 2277298 |
| Contact Person | Mrs. Maggie Zheng |
| Contact Title | Regulatory Affairs Manager |
| E-mail | zhengxy@hugedent.com |
### 3. Device Name
Trade name: Light Cure Dental Adhesive Common name: Dental Adhesive Classification name: Resin Tooth Bonding Agent (21 CFR 872.3200) Regulatory Class: II Product Code: KLE
### 4. Predicate Device Information
| Table 1: Predicate Device Information | | | | |
|---------------------------------------|-------------------|-------------|--------------|-----------|
| Owner/Operator | Device Trade Name | 510 (k) No. | Product Code | Predicate |
| BISCO, INC. | DREAMBOND | K112118 | KLE | Primary |
This predicate device has not been subject to a design-related recall.
No reference devices were used in this submission.
#### 5. Description of Device
{4}------------------------------------------------
Light Cure Dental Adhesive is a universal self-etch visible light curable dentine bonding agent specifically designed for meeting all your bonding needs. It is not only for direct bonding under a dental composite restorative or a resin modified glass ionomer cement restorative, but also for all indirect dental restoration applications, either used as a strict self-etch dental bonding agent, or in combination with the traditional total etch or enamel etch procedures. In all of those cases, this dental bonding agent will give you exceptional results and predicting a reliable treatment outcome.
### 6. Indications for Use
Light Cure Dental Adhesive can be used as a strict self-etch dental bonding agent, or in combination with the traditional total etch or enamel etch procedures:
- Direct bonding under a light-cured composite or compomer fillings/resin modified glass ionomer restoratives;
- Repair of dental restorations of ceramic, composites and/or metal alloys;
- Indirect dental restorations cementation applications in combination with resin cement for cementation of all sorts of dental restorations including crowns, bridges, inlays/onlays, veneers, dental posts and other restorations made from materials of metals/alloys, mental-ceramic, all-ceramic and/or porcelain, composites, and their combinations.
### 7. Summary of Physical and Chemical Properties Tests
Test standards and methods based on ISO 29022: 2013 (Dentistry - Adhesive - Notched-edge sheer bond strength test) and internal final inspection standard of Rizhao HuGe Biomaterials Company, Ltd.. And the results from testing demonstrate that Light Cure Dental Adhesive is substantially equivalent to the predicate device.
#### 8. Technological Characteristics
All components of the subject device are based upon industry well-known chemistry. The etch methods of the subject device and predicate device covered the flexibility for total-, self- and selective-etch procedures. The following table shows the significant technological characteristics for the subject device and indicates the following similarities and differences with the predicate device:
| Table 4: Technological Characteristics Comparison Table | | |
|---------------------------------------------------------|------------------------------------------------|------------------------------|
| Technological<br>Characteristics | Subject device<br>(Light Cure Dental Adhesive) | Primary predicate<br>K112118 |
{5}------------------------------------------------
| Method of<br>polymerization | Light Cured | Light Cured | |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Resin<br>composition | Unfilled, multifunctional methacrylate resin | Unfilled, multifunctional methacrylate resin | |
| Solvent | Ethanol based | Ethanol based | |
| Method of<br>application | Single component adhesive | Single component adhesive | |
| Indications of<br>Use | Light Cure Dental Adhesive can be used as a<br>strict self-etch dental bonding agent, or in<br>combination with the traditional total etch or<br>enamel etch procedures:<br>- Direct bonding under a light-cured<br>composite or compomer fillings/resin<br>modified glass ionomer restoratives;<br>- Repair of dental restorations of ceramic,<br>composites and/or metal alloys;<br>- Indirect dental restorations cementation<br>applications in combination with resin<br>cement for cementation of all sorts of dental<br>restorations including crowns, bridges,<br>inlays/onlays, veneers, dental posts and other<br>restorations made from materials of<br>metals/alloys, mental-ceramic, all-ceramic<br>and/or porcelain, composites, and their<br>combinations. | The indications of use of DREAMBOND<br>are:<br>1. all direct restorations<br>2. all indirect restorations<br>3. intra-oral repairs (i.e. repair. of any fixed<br>dental prosthesis containing zirconia,<br>alumina, metals, glass ceramics, tooth<br>structure, and composites)<br>4. desensitizing/sealing of tooth structure<br>5. protective varnish for glass ionomer<br>fillings<br>6. priming of enamel for orthodontic use | |
| Prescription/over-<br>the-counter use | Prescription | Prescription | |
| Etch Method | Total-etch, self-etch and selective-etch<br>procedures | Total-etch, self-etch and selective-etch<br>procedures | |
| Physical<br>Properties | The subject device and the predicate device have substantially equivalent physical property<br>of shear bond strength as they all conform to the specifications set by internal final<br>inspection and the test method equal to ISO 29022:2013 | | |
## 9. Summary of Biocompatibility
The new device, Light Cure Dental Adhesive, is substantially equivalent to the predicate devices that have been legally marketed for decades and with no clinical adverse events. The formulation of new device does not contain any non-conventional chemicals compared to the legally marketed predicate device.
We selected our Light Cure Dental Adhesive (6mL) as the representative model in biocompatibility tests. Because our company only has one component formula for this product. The chemical compositions, raw material suppliers, curing method and other technological characteristic in all packing specifications are identical, the difference only depends on the quantity.
{6}------------------------------------------------
The packaging bottle materials used for all packaging specifications are the same, and the packaging method is the same.
Therefore, the packaging specifications differences in the delivery system affect neither the intended use nor substantial equivalence.
Biocompatibility tests were performed fully following the ISO 10993 standards. The test items include Cytotoxicity, Sensitization, Irritation, Systemic Toxicity, Subchronic Toxicity and Genotoxicity.
## 10. Clinical Performance Data
Not applicable. Clinical performance testing has not been performed for the subject device.
### 11. Conclusions
Based on the indications for use, technological characteristics, performance testing and comparison to predicate devices, the subject device has been shown to be safe and effective for its intended use and the minor differences in indications for use fall within the intended use of the predicate devices affecting neither the general intended use nor substantial equivalence. Rizhao HuGe Biomaterials Company, Ltd. concludes that the subject device is substantially equivalent to the predicate devices described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.