K201775 · Nuvera Medical, Inc. · OBJ · Mar 5, 2021 · Cardiovascular
Device Facts
Record ID
K201775
Device Name
NuVision ICE Catheter
Applicant
Nuvera Medical, Inc.
Product Code
OBJ · Cardiovascular
Decision Date
Mar 5, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1200
Device Class
Class 2
Attributes
Pediatric
Indications for Use
The NuVision ICE Catheter is intended for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients. The catheter is intended for imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures.
Device Story
NuVision ICE Catheter is a diagnostic ultrasound imaging catheter for intracardiac/intra-luminal use. It features a 10 Fr outer shaft housing a 2D piezo-electric ultrasound transducer array capable of real-time 3D imaging. Operated by a physician in a clinical setting, the catheter is advanced, retracted, and rotated via a handle-mounted probe knob; a separate deflection knob controls an inner shaft for bidirectional tip steering. The device connects to an ultrasound cardiac imaging console via a connector cable; internal PCBs handle signal conditioning. The system uses real-time thermistor feedback to ensure the probe tip remains below 43°C. By providing structural and physiological images, the device assists clinicians in visualizing anatomy and interventional tools during percutaneous procedures, facilitating procedural guidance.
Clinical Evidence
No clinical testing included. Substantial equivalence is based on non-clinical data, including bench-top performance evaluations, animal testing, packaging validation, biological safety (ISO 10993), electrical safety (IEC 60601-1), and acoustic output testing (IEC 60601-2-37).
Technological Characteristics
10 Fr catheter with dual-shaft design (biocompatible thermopolymer over braided core). Features a multi-element 2D phased array-on-ASIC ultrasound transducer (4-10 MHz). Connectivity via mechanical snap connector to ultrasound console. Sterilized via Ethylene Oxide (ISO 11135-1). Complies with IEC 60601-1, IEC 60601-1-2, and IEC 60601-2-37. Includes integrated thermistor for temperature monitoring.
Indications for Use
Indicated for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy, physiology, and other devices in adult and pediatric patients. Intended for imaging guidance only during cardiac interventional percutaneous procedures.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 5, 2021
NuVera Medical, Inc. Robert Nardelli Vice President, RA/QA 140 Knowles Dr. Los Gatos, California 95032
Re: K201775
Trade/Device Name: NuVision ICE Catheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: OBJ Dated: March 2, 2021 Received: March 3, 2021
Dear Robert Nardelli:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Jessica Paulsen Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K201775
Device Name NuVision ICE Catheter
#### Indications for Use (Describe)
The NuVision ICE Catheter™ is intended for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediativ patients. The catheter is intended for imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures.
Type of Use (Select one or both, as applicable)
| <span style="font-size: 10px;">☑ Production Use (Part 21 CFR 601 Subpart E)</span> | ☐ Same-Site Control Use (21 CFR 601 Subpart E) |
|------------------------------------------------------------------------------------|------------------------------------------------|
|------------------------------------------------------------------------------------|------------------------------------------------|
> Prescription Use (Part 21 CFR 801 Subpart D)
__ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for NuVera Medical. The logo consists of a stylized, geometric shape in orange and yellow, resembling folded paper or ribbons. To the right of the shape is the text "NuVera" in a clean, sans-serif font, with "MEDICAL" in a smaller font size directly below it. The overall design is modern and professional.
### 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| 510(k) Number: | K201775 |
|-------------------------|----------------------------------------------------------------------------------------------------|
| Date Prepared: | June 26, 2020 |
| Applicant Information: | NuVera Medical, Inc<br>140 Knowles Drive<br>Los Gatos, CA 95032<br>Registration Number: 3016516752 |
| Contact Person: | Robert Nardelli robert@nuveramedical.com |
| Office Number: | 408-560-2500 |
| Device Information: | |
| Trade/Proprietary Name: | NuVision ICE Catheter |
| Common Name: | Intracardiac Echocardiography (ICE) Catheter |
| Classification Name(s): | Diagnostic Intravascular Catheters, 21 CFR 870.1200 (OBJ) |
Class: Class II
Panel: Division of Cardiovascular Devices
AcuNav™ Diagnostic Ultrasound Catheter (K071234) Predicate Device:
### Device Description
The NuVision ICE Catheter is a diagnostic ultrasound imaging catheter designed for intracardiac use. The catheter can be advanced, retracted and rotated within the anatomy. The catheter is comprised of a 10 Fr outer shaft capable of being rotated and deflected, which houses an ultrasound transducer. The outer shaft rotation is driven directly from a probe knob on the handle, while a deflection knob on the handle drives the deflection by controlling a deflectable inner shaft situated within the outer shaft.
The imaging ultrasound transducer is a 2D piezo-electric array capable of real-time 3D imaging. The imaging field emanates perpendicular to the axis of one side of the imaging tip, also referred to as a probe. The tip movement, in combination with deflection of the distal catheter, provide the user with the ability to adjust the orientation of the imaging tip to visualize the target anatomy.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image contains the logo for NuVera Medical. The logo consists of a geometric shape in orange and yellow, resembling folded paper. To the right of the shape is the company name, "NuVera," in a simple, sans-serif font. Below "NuVera" is the word "MEDICAL" in a smaller font size.
NuVision™ ICE Catheter
An electrical connector proximal to the handle allows connection to the NuVision Connector Cable which also connects to an ultrasound cardiac imaging console. Printed circuit boards within the catheter system's electrical connectors help with signal distribution and conditioning. The NuVision ICE Catheter System relies on power from the ultrasound cardiac imaging console to function.
The NuVision ICE Catheter is used with a compatible ultrasound cardiac imaging system with software cleared for market in accordance with 21 CFR 807.81(a)(3).
### Indications for Use
The NuVision ICE Catheter is intended for intracardiac and intra-luminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients. The catheter is intended for imaging guidance only, not treatment delivery, during cardiac interventional percutaneous procedures.
### Functional and Technological Comparison
The table below includes a functional and technological comparison between the NuVision ICE Catheter and the market-cleared Siemens AcuNav™ Diagnostic Ultrasound Catheter (K071234).
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image contains the logo for Nuvera Medical. The logo consists of a stylized, geometric shape in orange and yellow, resembling folded paper or a ribbon. To the right of the geometric shape, the word "NUVERA" is written in a bold, sans-serif font, with the word "MEDICAL" appearing below it in a smaller font size.
# Traditional 510(k) Notification
## Table 5-1: Catheter substantial equivalency comparison
| | Subject Device | Predicate Device | |
|-----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Component | NuVision ICE Catheter | Siemens AcuNav™ (K071234) | Comments |
| Classification | Class II | Class II | Same as predicate device |
| Regulation | 21 CFR 870.1200 | 21 CFR 870.1200 | Same as predicate device |
| Regulation Name | Diagnostic Intravascular Catheter | Diagnostic Intravascular Catheter | Same as predicate device |
| Product code | OBJ | OBJ | Same as predicate device |
| Catheter type | Intracardiac Echocardiography | Intracardiac Echocardiography | Same as predicate device |
| Clinical<br>performance data | No clinical testing is included in<br>the submission. Determination of<br>substantial equivalence performance is<br>based on an assessment of non-clinical<br>data. | No clinical testing is referenced in<br>the submission. | Same as predicate device |
| Non-clinical<br>performance data | Non-clinical data includes bench-top<br>evaluations, animal testing, packaging<br>validation, biological safety, electrical<br>safety, acoustic output testing, and animal<br>testing. | Non-clinical data includes bench-top<br>evaluations, animal testing, packaging<br>validation, biological safety, electrical<br>safety, acoustic output testing, and animal<br>testing. | Same as predicate device |
| Intended use | The NuVision ICE Catheter is intended for<br>intracardiac and intra-luminal visualization<br>of cardiac and great vessel anatomy and<br>physiology as well as visualization of other<br>devices in the heart of adult and pediatric<br>patients. The catheter is intended for<br>imaging guidance only, not treatment<br>delivery, during cardiac interventional<br>percutaneous procedures. | The catheter is intended for intra-cardiac<br>and intra-luminal visualization of cardiac<br>and great vessel anatomy and physiology<br>as well as visualization of other devices in<br>the heart of adult and pediatric patients.<br>The catheter is intended for imaging<br>guidance only, not treatment delivery,<br>during cardiac interventional percutaneous<br>procedures. | Same as predicate device |
| Principles of<br>operation | Generation of structural and physiological<br>images / representations / measurements<br>of cardiovascular anatomy using pulse-<br>echo ultrasound systems. | Generation of structural and physiological<br>images / representations / measurements<br>of cardiovascular anatomy using pulse-<br>echo ultrasound systems. | Same as predicate device |
| Outside Diameter | 10F | 10F | Same as predicate device |
| Component | Subject Device<br>NuVision ICE Catheter | Predicate Device<br>Siemens AcuNav™ (K071234) | Comments |
| Imaging energy | Ultrasound | Ultrasound | Same as predicate device |
| Transducer<br>configuration | Multi-element 2D phased array-on-ASIC<br>ultrasound transducer at distal tip. | Multi-element linear phased array<br>ultrasound transducer at distal tip. | Similar to predicate<br>device. Basic transducer<br>technology is the same.<br>The NuVision ICE<br>Catheter 2D array<br>supports 3D imaging. |
| Ultrasound imaging<br>frequency | 4-10 MHz | 4-10 MHz | Same as predicate device |
| Proximal end<br>configuration | Single connector, mechanical snap into<br>NuVision Connector Cable. | Single connector, mechanical snap into<br>SwiftLink adapter | Same as predicate device |
| Acoustic testing | As per IEC 60601-2-37 and equivalent<br>analysis to NEMA UD- 2 performance | As per IEC 60601-2-37 and equivalent<br>analysis to NEMA UD- 2 performance | Same as predicate device |
| Sterilization | Ethylene Oxide | Ethylene Oxide | Same as predicate device |
| Imaging Modes | B-Mode (2D and 3D)<br>M-Mode<br>Doppler Pulsed Wave Doppler<br>Continuous Wave Doppler<br>Color Doppler (2D and 3D)<br>Power Doppler | B-Mode (2D only)<br>M-Mode<br>Doppler Pulsed Wave Doppler<br>Continuous Wave Doppler<br>Color Doppler (2D only)<br>Power Doppler | Same as predicate device<br>Similar to predicate<br>device. The NuVision ICE<br>Catheter 2D array<br>supports 3D imaging. |
| Imaging<br>Configuration | Phased array limited to ~90-degree sector<br>emanating from the side of the catheter. | Phased array limited to ~90-degree sector<br>emanating from the side of the catheter. | Same as predicate device |
| Biocompatibility | ISO 10993, Externally Communicating<br>Device, Circulating Blood category | ISO 10993, Externally Communicating<br>Device, Circulating Blood category. | Same as predicate device |
| Insertable length | 90 cm | 90 cm | Same as predicate device |
| Catheter<br>construction | Coaxial shafts of biocompatible<br>thermopolymer over braided core | Single shaft of biocompatible<br>thermopolymer over braided core | Similar to predicate<br>device. The NuVision ICE<br>Catheter uses a dual shaft<br>design to allow<br>independent rotation of<br>the catheter tip |
| Component | Subject Device | Predicate Device | Comments |
| | NuVision ICE Catheter | Siemens AcuNav™ (K071234) | |
| Re-usability | Single use | Single use | Same as predicate device |
| Preparation | No preparation required. | No preparation required. | Same as predicate device |
| Directionality of<br>catheter tip | Integrates a tip deflection mechanism<br>using a deflection handle and pull-line to<br>provide integrated bidirectional deflection<br>of an inner shaft. Catheter tip is rotated<br>with outer shaft independent of deflection. | Integrates a tip deflection mechanism<br>using a deflection handle and pull-line to<br>provide integrated 4-way deflection of<br>catheter. | Similar to predicate<br>device. Both devices use<br>pull-lines to deflect the<br>catheter tip. NuVision ICE<br>Catheter allows<br>independent rotation of<br>the catheter tip. |
| Temperature<br>Sensor | Incorporates a thermistor within the probe<br>tip. The ultrasound system reads the<br>thermistor and adjusts power to assure the<br>probe never reaches temperature limit of<br>43° C. | Ultrasound system energy delivery was<br>calibrated to assure probe never reaches<br>temperature limit of 43° C. | Similar to predicate<br>device. NuVision ICE<br>Catheter allows use of<br>real-time probe<br>temperature feedback to<br>make energy adjustments. |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image contains the logo for NuVera Medical. The logo consists of a stylized, geometric shape in shades of orange and yellow, resembling folded or layered elements. To the right of the shape, the text "NuVera" is written in a clean, sans-serif font, with the word "MEDICAL" appearing below it in a smaller font size.
# Traditional 510(k) Notification
# NuVision™ ICE Catheter
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image contains the logo for Nuvera Medical. The logo consists of an abstract geometric shape in orange and yellow, followed by the company name "Nuvera" in a simple, sans-serif font. Below "Nuvera" is the word "Medical" in a smaller font size.
# Traditional 510(k) Notification
## NuVision™ ICE Catheter
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for Nuvera Medical. The logo consists of a geometric shape on the left and the company name on the right. The geometric shape is made up of two overlapping shapes, one orange and one yellow. The company name is written in black, with "Nuvera" on top and "Medical" below.
### Testing completed:
The NuVision ICE Catheter was developed per the Guidance for Industry and FDA Staff, Marketing Clearance of Diagnostic Ultrasound Systems and Transducers (2019). Nonclinical data includes bench-top performance evaluations, packaging validation, biological safety, electromagnetic compatibility, and acoustic output testing. The devices used for testing were manufactured in accordance with the design specifications and approved manufacturing documentation.
### Dimensional and Performance
The catheter dimensional characteristics were evaluated to verify conformance to design specifications and EN ISO 10555-1:2013: Sterile, single-use intravascular catheters - Part 1: General requirements. The physical integrity, functionality, and overall performance of the catheter were evaluated including:
- a) Visual and Dimensional Characteristics
- b) Simulated Use
- · Package opening and transfer to sterile field
- · Sterile sleeve loading
- c) Catheter Functional Performance
- · Reach
- · Deflection
- · Rotation
- · Radiopacity
- d) Electro-Acoustic Performance
- e) Durability
- · Multiple insertions
- · Rotation Cycling
- · Deflection Cycling
- · Tip Kinking and buckling
- · Leakage
- · Particulate generation
- · Tensile strength
#### Sterilization and Shelf Life
The NuVision ICE Catheter is sterilized via Ethylene Oxide gas using a cycle validated per ISO 11135-1:2014 to a Sterility Assurance Level (SAL) of 10°. Product samples were subjected to sterilization processing, distribution conditioning, and aging and then tested for compliance to the product specification. Packaging performance was evaluated to demonstrate that the packaging system preserves its integrity and that of the device through the lifecycle of the product in accordance with ISO 11607-1 and ISO 11607-2 and its applicable referenced standards.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image contains the logo for NuVera Medical. The logo consists of an abstract geometric shape in orange and yellow on the left, followed by the text "NuVera" in a simple, sans-serif font. Below "NuVera" is the word "MEDICAL" in a smaller font size.
### Biocompatibility
The NuVision ICE Catheter is made from materials that are used in commercially available medical devices, and commonly used in intravascular devices. The sterilized catheter was evaluated for biocompatibility in accordance with the EN ISO 10993 series standards.
### Electromagnetic Compatibility and Electrical Safety
Acoustic Output was evaluated in accordance with FDA Guidance. Information for Manufacturers Seeking Marketing Clearance of Diagnostic Ultrasound Systems and Transducers (June 2019) and EN 60601-2-37:2008: Medical electrical equipment — Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment (IEC 60601-2-37:2015).
Electrical and thermal safety testing was conducted in accordance with EN 60601-1:2006/ IEC 60601-1:2005/A1:2012: Medical electrical equipment — Part 1: General requirements for basic safety and essential performance and IEC 60601-2-37 Edition 2.1 2015: Medical electrical equipment - Part 2-37: Particular requirements for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment.
Electromagnetic compatibility testing was evaluated, demonstrating compliance to IEC 60601-1-2 (4th Edition).
#### Conclusion:
The NuVision ICE Catheter and the predicate device have nearly identical intended use and very similar principles of operational and technological characteristics. The subject device and predicate device are both intended for use in intravascular and/or intracardiac imaging, and the minor technological differences do not raise any new safety and effectiveness risk or concerns. After analyzing the results of bench tests, laboratory tests, electrical safety tests, and animal tests, it can be concluded that the NuVision ICE Catheter is safe and effective for the intended use, is as safe and effective as the predicate device, and is substantially equivalent to the cited predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.