SmartHDM-510 System

K201765 · Mesa Laboratories, Inc. · FIZ · Sep 2, 2020 · Gastroenterology, Urology

Device Facts

Record IDK201765
Device NameSmartHDM-510 System
ApplicantMesa Laboratories, Inc.
Product CodeFIZ · Gastroenterology, Urology
Decision DateSep 2, 2020
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.5820
Device ClassClass 2

Indications for Use

The SmartHDM-510 system may be used by hemodialysis personnel to test the conductivity, temperature, pH and flow of the dialysate solution used with hemodialysis delivering systems. The SmartHDM-510 system may also be used to test the conductivity/temperature and pH of acid and sodium bicarbonate dialysate concentrates and water used in hemodialysis applications. The intended use is limited to periodic use for installation and maintenance of hemodialysis delivering systems and does not include the daily monitoring of hemodialysis delivering systems prior to treatment.

Device Story

SmartHDM-510 is a hemodialysis service system for installation and maintenance of hemodialysis delivery equipment. It utilizes external sensors (HDU-Sensors, HDM-Sensor Modules, 90XL-Sensors) to measure conductivity, temperature, pressure, pH, and flow. Data is transmitted to an Android-based display module running the SmartHDM-510 App. The system provides high-resolution graphical displays, statistical values, and user hints to assist technicians in identifying errors and verifying system conditions. Used by hemodialysis personnel in clinical settings; output allows technicians to perform precise adjustments and maintenance on dialysis machines, ensuring accurate dialysate composition and system performance, ultimately supporting safe patient treatment.

Clinical Evidence

Bench testing only. Performance verified through software testing, code reviews, system verification/validation, and electrical/electromagnetic safety compliance. Testing addressed user interface, usability, sensor communication, and measurement accuracy against specifications.

Technological Characteristics

System comprises external sensors (conductivity, temperature, pressure, pH, flow) and an Android-based display module. Connectivity includes USB and Wi-Fi (HTTPS/TLS for manuals/registration only). Cybersecurity controls include signed apps, Google Play Store updates, and no open incoming network ports. Measurement ranges: Conductivity 0-200 mS/cm, Temperature 0-100°C, Pressure -672 to +1551 mmHg, Flow 100-2000 ml/min, pH 0-14.

Indications for Use

Indicated for hemodialysis personnel to test conductivity, temperature, pH, pressure, and flow of dialysate solutions, concentrates, and water during installation and maintenance of hemodialysis delivery systems. Not for daily monitoring prior to treatment.

Regulatory Classification

Identification

A hemodialysis system and accessories is a device that is used as an artificial kidney system for the treatment of patients with renal failure or toxemic conditions and that consists of an extracorporeal blood system, a conventional dialyzer, a dialysate delivery system, and accessories. Blood from a patient flows through the tubing of the extracorporeal blood system and accessories to the blood compartment of the dialyzer, then returns through further tubing of the extracorporeal blood system to the patient. The dialyzer has two compartments that are separated by a semipermeable membrane. While the blood is in the blood compartment, undesirable substances in the blood pass through the semipermeable membrane into the dialysate in the dialysate compartment. The dialysate delivery system controls and monitors the dialysate circulating through the dialysate compartment of the dialyzer.(1) The extracorporeal blood system and accessories consists of tubing, pumps, pressure monitors, air foam or bubble detectors, and alarms to keep blood moving safely from the blood access device and accessories for hemodialysis (§ 876.5540) to the blood compartment of the dialyzer and back to the patient. (2) The conventional dialyzer allows a transfer of water and solutes between the blood and the dialysate through the semipermeable membrane. The semipermeable membrane of the conventional dialyzer has a sufficiently low permeability to water that an ultrafiltration controller is not required to prevent excessive loss of water from the patient's blood. This conventional dialyzer does not include hemodialyzers with the disposable inserts (Kiil type) (§ 876.5830) or dialyzers of high permeability (§ 876.5860). (3) The dialysate delivery system consists of mechanisms that monitor and control the temperature, conductivity, flow rate, and pressure of the dialysate and circulates dialysate through the dialysate compartment of the dialyzer. The dialysate delivery system includes the dialysate concentrate for hemodialysis (liquid or powder) and alarms to indicate abnormal dialysate conditions. This dialysate delivery system does not include the sorbent regenerated dialysate delivery system for hemodialysis (§ 876.5600), the dialysate delivery system of the peritoneal dialysis system and accessories (§ 876.5630), or the controlled dialysate delivery system of the high permeability hemodialysis system § 876.5860). (4) Remote accessories to the hemodialysis system include the unpowered dialysis chair without a scale, the powered dialysis chair without a scale, the dialyzer holder set, dialysis tie gun and ties, and hemodialysis start/stop tray.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue. September 2, 2020 Mesa Laboratories, Inc. % Johannes Pfingstmann Manager for Quality Assurance & Regulatory Affairs IBP Medical GmbH Ikarusallee 15 Hannover, Lower Saxony 30179 Germany Re: K201765 Trade/Device Name: SmartHDM-510 System Regulation Number: 21 CFR§ 876.5820 Regulation Name: Hemodialysis system and accessories Regulatory Class: II Product Code: FIZ Dated: August 3, 2020 Received: August 3, 2020 Dear Johannes Pfingstmann: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {1}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Carolyn Y. Neuland, Ph.D. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use 510(k) Number (if known) K201765 Device Name SmartHDM-510 System #### Indications for Use (Describe) The SmartHDM-510 system may be used by hemodialysis personnel to test the conductivity, temperature, pH and flow of the dialysate solution used with hemodialysis delivering systems. The SmartHDM-510 system may also be used to test the conductivity/temperature and pH of acid and sodium bicarbonate dialysate concentrates and water used in hemodialysis applications. The intended use is limited to periodic use for installation and maintenance of hemodialysis delivering systems and does not include the daily monitoring of hemodialysis delivering systems prior to treatment. X Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) #### CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.* The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ # 510(k) Summary | Submitter | Mesa Laboratories, Inc. | | | |-----------------------|------------------------------------------------|-------------------------------------|--| | Address | 12100 West 6th Avenue | | | | | Lakewood, CO 80228 | | | | Internet | www.mesalabs.com | | | | Contact Person | Johannes Pfingstmann | | | | eMail | johannes.pfingstmann@ibpmt.com | | | | Phone number | +49 511 957 336 0 | | | | Date preparation | June 24, 2020 | | | | Name of the<br>device | SmartHDM-510 System | | | | Common Name | Hemodialysis Service System | | | | Classification | Classification Name: | Hemodialysis System and Accessories | | | | Product Code: | FIZ | | | | Device Class: | 2 | | | | Regulation Number: | 21 CFR 876.5540 | | | | Medical Specialty: | Gastroenterology | | | Predicate<br>Devices | K020909 HDM99 (Dialysis Meters) primary device | | | | | K020908 HDM97 (Dialysis Meters) | | | | | K050812 90XL Instrumentation System | | | ## Devices Description The SmartHDM-510 System is designed for the installation and maintenance of hemodialysis delivering systems. The system consists of different sensors and sensor-modules to measure conductivity, temperature, pressure, pH and flow and an Android operating system based display module with SmartHDM-510 App. {4}------------------------------------------------ The SmartHDM-510 system consists of: - HDU-Sensors for Conductivity/Temperature, Pressure, pH and flow ● - HDM-Sensor Modules for Conductivity/Temperature, Pressure, - 90XL-Sensors for Conductivity/Temperature, Pressure and pH - Android-based Display Module with SmartHDM-510 App. ## Indication for Use The SmartHDM-510 system may be used by hemodialysis personnel to test the conductivity, temperature, pressure, pH and flow of the dialysate solution used with hemodialysis delivering systems. The SmartHDM-510 system may also be used to test the conductivity/temperature and pH of acid and sodium bicarbonate dialysate concentrates and water used in hemodialysis applications. The intended use is limited to periodic use for installation and maintenance of hemodialysis delivering systems and does not include the daily monitoring of hemodialysis delivering systems prior to treatment. #### Comparative Analysis In comparison to the predicate device, the SmartHDM-510 System has a highresolution screen. Based on the high-resolution screen, the clear display of the measurement conditions makes the reading of the measured values more reliable. The improved graphical display of the measured values makes errors on the device to be serviced more visible. Information texts and messages make the operation safer. Listed below are the major attributes of the SmartHDM-510 System compared to the predicate devices listed: {5}------------------------------------------------ # SmartHDM-510 System 510(k) Summary | Specification | Predicate<br>Device 1 | Predicate<br>Device 2 | Predicate<br>Device 3 | New Device | |-----------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------| | | HDM99<br>(Primary) | HDM97 | 90XL | SmartHDM-510<br>System | | General | | | | | | 510(k) Number | K020909 | K020908 | K050812 | | | Design | Handheld device with<br>connectors for internal and<br>external sensors | | Handheld display unit with<br>connectors for external sensors | | | Display - General functions | | | | | | Operating System | R-THOS | None | None | Android | | Automated security patches | N/A | N/A | N/A | Yes | | Battery status reading | Yes | Yes | Yes | Yes | | Display size | 37 x 137 mm<br>1.46 x 5.39 " | 55 x 43 mm<br>2.16 x 1.69 " | 57 x 76 mm<br>2.24 x 2.99 " | HDC64 5.7"<br>HDC85 8" | | Color Display | No | No | No | Yes | | Hint Texts | No | No | No | Yes | | Display selection<br>of connected sensors | Yes | Yes | No | Yes | | Display single value | Yes | Yes | Yes | Yes | | Display single value<br>including graphic | Yes | No | No | Yes | | Display several values as list | Yes | No | Yes | Yes | | Display graphic reading as list | No | No | No | Yes | | Selection of time range for graphic<br>display | Yes | No | No | Yes | | Selection of y- range for<br>graphic display | Yes | No | No | Yes | | Statistic values | No | No | No | Yes | | User hints | No | No | No | Yes | | Display - measuring functions | HDM99<br>(Primary) | HDM97 | 90XL | SmartHDM-510<br>System | | Conductivity adjustment | Yes | Yes | Yes | Yes | | Conductivity adjustment reset | Yes | Yes | No | Yes | | Selection Conductivity/<br>Temperature Compensation<br>by vendor name | Yes | Yes | No | Yes | | Conductivity resolution<br>selection | Yes | No | No | Yes | | Pressure unit selection | Yes | Yes | Yes | Yes | | Pressure resolution selection | Yes | Yes | No | Yes | | Pressure Tare function | Yes | Yes | Yes | Yes | | Pressure change<br>measurement | No | Yes | No | Yes | | pH adjustment | Yes | Yes | Yes | Yes | | pH adjustment points | 3 | 3 | 2 | 3 | | pH temperature compensation<br>selection | Yes | Yes | No | Yes | | Flow accumulation display | Yes | Yes | Not Applicable | Yes | | Flow accumulation reset | Yes | Yes | Not Applicable | Yes | | HDU-Sensors and HDM18/19 Modules | | | | | | | HDM99 | HDM97 and<br>HDM18/19 | 90XL | HDU-Sensor | | Conductivity/Temperature<br>measuring via Dialyzer<br>Connector | Yes | Yes | Yes | Yes | | Conductivity Measuring<br>Range | 0 to 24 mS/cm | 0 to 30 mS/cm | 0 to 200 mS/cm | 0 to 200 mS/cm | | Conductivity Range<br>Accuracy | 0 to 249.9 uS/cm<br>$\pm$ 0.6 uS/cm | 0 to 199 uS/cm<br>$\pm$ 0.6 uS/cm | 0 to 1.99 mS<br>$\pm$ 0.35% of reading<br>+ 0.002 mS/cm | 0 to 199.9 uS/cm<br>$\pm$ 0.6 uS/cm | | | 250 to 2499 uS/cm<br>$\pm$ 6 uS/cm | 200 to 1999 uS/cm<br>$\pm$ 6 uS/cm | 2 to 29.99 mS/cm<br>$\pm$ 0.20% of reading<br>+ 0.002, | 200 to 1999 uS/cm<br>$\pm$ 6 uS/cm | | | 2.5 to 16.99 mS/cm<br>$\pm$ 0.06 mS/cm | 2 to 11.99 mS/cm<br>$\pm$ 0.06 mS/cm | > 30 mS/cm<br>$\pm$ 0.50% of reading | 2 to 11.99 mS/cm<br>$\pm$ 0.06 mS/cm | | | 17 to 24 mS/cm<br>$\pm$ 0.06 mS/cm | 12 to 15.99 mS/cm<br>$\pm$ 0.03 mS/cm | | 12 to 19.99 mS/cm<br>$\pm$ 0.03 mS/cm | | | | 16 to 30 mS/cm<br>$\pm$ 0.06 mS/cm | | 20 to 200 mS/cm<br>$\pm$ 0.60% of reading | | Remark to measuring<br>ranges | The important range for hemodialysis is 14 to 16 mS/cm.<br>The measuring range of predicate HDM97 and HDM18/19 are identical.<br>The measuring range of predicate 90XL- and HDU-Conductivity/Temperature<br>sensors are identical. | | | | | Temperature Measuring<br>Range | 0 to 100 °C | 0 to 100 °C | 10 to 90°C | 0 to 100 °C | | Temperature Accuracy | ± 0,1 °C | ± 0.1°C | ± 0.1°C | ± 0.1°C | | Pressure Measuring Range | -700 to +1500 mmHg | -700 to + 1900<br>mmHg | -600 to +1600<br>mmHg | -672 to +1551<br>mmHg | | Pressure Accuracy | -700 to +1500 mmHg<br>± 0.5% of reading | HDM97<br>0 to 300 mmHg<br>± 1 mmHg<br>otherwise<br>± 2 mmHg<br>HDM18/19<br>Internal<br>± 2 mmHg<br>External<br>0 to 300 mmHg<br>± 1 mmHg<br>otherwise<br>± 2 mmHg | 0 to 199 mmHg<br>±1.0 mmHg,<br>200 to 300<br>mmHg<br>± 1.5 mmHg<br>above 300<br>mmHg and<br>below 0 mmHg<br>± 0.5% of<br>reading + 1<br>mmHg | 0 to 300 mmHg<br>± 1 mmHg<br>otherwise<br>± 2 mmHg | | Flow Measuring Range | 100 to 2000 ml/min | 100 to 2000<br>ml/min | Not Applicable | 100 to 2000<br>ml/min | | Flow Measuring Accuracy | 100 to 2000 ml/min<br>± 0.5 % full scale | ± 2.0 %<br>of reading | Not Applicable | ± 2.0 %<br>of reading | | pH Measuring Range | 0...14 pH | 0...14 pH | 0...14 pH | 0...14 pH | | pH Measuring | ± 0.02 pH | ± 0.02 pH | ± 0.1 pH | ± 0.02 pH | | Communication Interface | | | | | | USB | Yes | Yes | No | Yes | | Bluetooth | No | Only HDM19 | No | No | | WIFI | No | No | No | Yes | | Remark to WIFI | SmartHDM-510 uses WIFI / Internet connections by HTTPS protocol, TCP port 443.<br>These connections are for displaying manuals, or a registration page for a user<br>or legal information which are not stored in the App itself.<br>The servers of Mesa Laboratories offer TLS1.2, TLS1.1, TLS1.0 as the encryption<br>method. WIFI is not used for measuring purposes or other SmartHDM-510 use<br>cases at all. | | | | | Cybersecurity | Not applicable | Not applicable | Not applicable | Applicable | | Remark to Cybersecurity | • SmartHDM-510 has no open incoming network ports<br>• SmartHDM-510 interacts only with HTTPS secured Mesa Laboratories websites<br>on user action (outgoing connection) such as Mesa Laboratories WEB pages and<br>Mesa Laboratories PDF's for: User manuals and legal information.<br>• SmartHDM-510 is updated by Google Play Store functionalities only<br>• SmartHDM-510-App is signed by Mesa Laboratories development and Google<br>and only spread by Google Play Store<br>• Bluetooth (LE) is only used for a special purpose and is not open to other devices<br>except these of Mesa Laboratories | | | | {6}------------------------------------------------ {7}------------------------------------------------ # SmartHDM-510 System 510(k) Summary {8}------------------------------------------------ ## Performance Testing The performance testing for the SmartHDM-510 System includes software testing and code reviews, system verification and validation testing, and testing to compliance standards for electrical and electromagnetic safety. Traceability has been documented between the system specification to verification and validation test protocols. Verification and validation test procedures also address the user interface, user manual descriptions, usability, sensor communication and general performance including measuring accuracy. The design of the device has been verified and validated both through testing and actual experience in international use. The testing and international experience provided data that demonstrated substantial equivalence to the predicate devices. ## Conclusion The characteristics of the SmartHDM-510 System are identical to those of the predicate devices in almost all categories. The device has the same intended use as its primary predicate device. Also, it has similar measuring ranges and measuring accuracy as the predicate. Different to HDM99 and the HDM97, where the display unit, all measuring electronic and pressure sensor are integrated into one housinq, and identical to the 90XL, the SmartHDM-510 System consist of separate devices for the display unit and external connected sensors with integrated measuring electronic. However, SmartHDM-510 System passed all performance tests and hence is as safe as its predicate devices. The performance of the SmartHDM-510 System is substantially equivalent to that of the predicate devices and do not raise any new questions of safety or effectiveness and performs as well or better than the predicate devices. Therefore, we conclude that it is substantially equivalent to its cleared predicate devices.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...