The Reprocessed MyoSure Tissue Removal Device (Model # 10-401/10-403) is indicated for intrauterine use by trained gynecologists to hysteroscopically resect and remove tissue such as: submucous myomas, endometrial polyps, and retained products of conception. The Reprocessed MyoSure REACH Tissue Removal Device (Model # 10-401FC/10-403FC) is indicated for intrauterine use by trained gynecologists to hysteroscopically resect and remove tissue such as: submucous myomas, endometrial polyps, and retain products of conception. The Reprocessed MyoSure LITE Tissue Removal Device (Model # 30-401LITE) is indicated for intrauterine use by trained gynecologists to hysteroscopically resect and remove tissue such as: submucous myomas, endometrial polyps, and retain products of conception. The Reprocessed MyoSure XL Tissue Removal Device (Model # 50-501XL) is indicated for intrauterine use by trained gyneologists to hysteroscopically resect and remove tissue such as: submucous myomas, endometrial polyps, and retain products of conception. The Reprocessed MyoSure XL Tissue Removal Device for Fluent (Model # 50-603XL) is indicated for intrauterine use by trained gynecologically resect and remove tissue such as: submucous myomas. endometrial polyps, and retain products of conception.
Device Story
Sterile, disposable, hand-held tissue removal device; used hysteroscopically to resect intrauterine tissue. Device connects via 6-foot flexible drive shaft to motorized control unit; foot pedal controls motor. Mechanical resection mechanism: tissue drawn through cutting window via suction; inner blade simultaneously rotates and reciprocates to cut tissue. Vacuum tubing connects to collection canister. Used in clinical settings by trained gynecologists. Enables minimally invasive removal of myomas, polyps, and retained products of conception; benefits patient by providing effective tissue resection during hysteroscopic procedures.
Clinical Evidence
Bench testing only. Evaluations included biocompatibility (ISO 10993-1), reprocessing validation (AAMI/TIR 30, ASTM D7225, AAMI/TIR 12, AAMI/ANSI ST35, ISO/TS 15883-5, ASTM F3208), sterilization (ISO 11135-1), electrical safety (IEC 60601-1), and functional performance (shaft deflection, leak decay, torque, RPM, reciprocation, and tissue cutting/removal efficacy). Results demonstrate performance equivalent to predicate devices.
Technological Characteristics
Mechanical tissue resection device; consists of hand-held probe with rotating/reciprocating inner blade, flexible drive shaft, and vacuum tubing. Materials evaluated per ISO 10993-1. Sterilized via Ethylene Oxide (ISO 11135-1). Electrical safety per IEC 60601-1. Dimensions vary by model (3mm or 4mm outer diameter; 32cm working length). Standalone device requiring external motorized control unit.
Indications for Use
Indicated for intrauterine use by trained gynecologists to hysteroscopically resect and remove submucous myomas, endometrial polyps, and retained products of conception.
Regulatory Classification
Identification
A hysteroscope is a device used to permit direct viewing of the cervical canal and the uterine cavity by a telescopic system introduced into the uterus through the cervix. It is used to perform diagnostic and surgical procedures other than sterilization. This generic type of device may include obturators and sheaths, instruments used through an operating channel, scope preheaters, light sources and cables, and component parts.
Predicate Devices
MyoSure Hysteroscopic Tissue Removal System and MyoSure Tissue Removal Devices (K172566)
MyoSure Hysteroscopic Tissue Removal System and MyoSure Tissue Removal Devices (K152723)
MyoSure XL Tissue Removal Device for Fluent (K181974)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 25, 2020
Stryker Sustainability Solutions Scott English Senior Staff Regulatory Affairs Specialist 1810 W. Drake Drive Tempe, AZ 85283
#### Re: K201756
Trade/Device Name: Reprocessed MyoSure Tissue Removal Device (10-401/10-403), Reprocessed MyoSure REACH Tissue Removal Device (10-401FC/10-403FC), Reprocessed MyoSure LITE Tissue Removal Device (30-401LITE/30-403LITE), Reprocessed MyoSure XL Tissue Removal Device (50-501XL/50-503XL), Reprocessed MyoSure XL Tissue Removal Device for Fluent (50-601XL/50-603XL) Regulation Number: 21 CFR§ 884.1690 Regulation Name: Hysteroscope and Accessories Regulatory Class: II Product Code: HIH Dated: June 27, 2020 Received: June 29, 2020
Dear Scott English:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Jason R. Roberts, Ph.D. Acting Assistant Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
510(k) Number (if known) K201756
#### Device Name
Reprocessed MyoSure Tissue Removal Device (10-401/10-403). Reprocessed MyoSure REACH Tissue Removal Device (10-401FC/10-403FC), Reprocessed MyoSure LITE Tissue Removal Device (30-401LITE), Reprocessed MyoSure XL Tissue Removal Device (50-501XL), Reprocessed MyoSure XL Tissue Removal Device for Fluent (50-603XL)
#### Indications for Use (Describe)
The Reprocessed MyoSure Tissue Removal Device (Model # 10-401/10-403) is indicated for intrauterine use by trained gynecologists to hysteroscopically resect and remove tissue such as: submucous myomas, endometrial polyps, and retained products of conception.
The Reprocessed MyoSure REACH Tissue Removal Device (Model # 10-401FC/10-403FC) is indicated for intrauterine use by trained gynecologists to hysteroscopically resect and remove tissue such as: submucous myomas, endometrial polyps, and retain products of conception.
The Reprocessed MyoSure LITE Tissue Removal Device (Model # 30-401LITE) is indicated for intrauterine use by trained gynecologists to hysteroscopically resect and remove tissue such as: submucous myomas, endometrial polyps, and retain products of conception.
The Reprocessed MyoSure XL Tissue Removal Device (Model # 50-501XL/50-503XL) is indicated for intrauterine use by trained gyneologists to hysteroscopically resect and remove tissue such as: submucous myomas, endometrial polyps, and retain products of conception.
The Reprocessed MyoSure XL Tissue Removal Device for Fluent (Model # 50-601XL/50-603XL) is indicated for intrauterine use by trained gynecologically resect and remove tissue such as: submucous myomas. endometrial polyps, and retain products of conception.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> <path d="M13.6 2.4L5.6 10.4L2.4 7.2L1.6 8L5.6 12L14.4 3.2L13.6 2.4Z" fill="black"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) </div> | <div> <span> <svg fill="none" height="16" viewbox="0 0 16 16" width="16" xmlns="http://www.w3.org/2000/svg"> </svg> </span> Over-The-Counter Use (21 CFR 801 Subpart C) </div> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) SUMMARY
#### Submitter:
Stryker Sustainability Solutions 1810 W. Drake Drive Tempe, Arizona 85283
### Contact:
Scott English Senior Staff Regulatory Affairs Specialist 480-763-5333 (o) 480-763-5310 (f) scott.english@stryker.com
Date of Preparation: June 27, 2020
## Name of Device:
| Trade/Proprietary Name | Model Number |
|---------------------------------------------------------|-----------------------|
| Reprocessed MyoSure Tissue Removal Device | 10-401/10-403 |
| Reprocessed MyoSure REACH Tissue Removal Device | 10-401FC/10-403FC |
| Reprocessed MyoSure LITE Tissue Removal Device | 30-401LITE/30-403LITE |
| Reprocessed MyoSure XL Tissue Removal Device | 50-501XL/50-503XL |
| Reprocessed MyoSure XL Tissue Removal Device for Fluent | 50-601XL/50-603XL |
Common Name: Hysteroscope and Accessories
Classification Information: Hysteroscope and Accessories (21 CFR § 884.1690, HIH, Class II)
# Predicate Devices:
| Model<br>Number | 510(k)<br>Number | 510(k) Title | Original<br>Manufacturer |
|--------------------------|------------------|-----------------------------------------------------------------------------------|--------------------------|
| 10-401<br>10-403 | K172566 | MyoSure Hysteroscopic Tissue Removal System<br>and MyoSure Tissue Removal Devices | Hologic, Inc. |
| 10-401FC<br>10-403FC | K152723 | MyoSure Hysteroscopic Tissue Removal System<br>and MyoSure Tissue Removal Devices | Hologic, Inc. |
| 30-401LITE<br>30-403LITE | K172566 | MyoSure Hysteroscopic Tissue Removal System<br>and MyoSure Tissue Removal Devices | Hologic, Inc. |
| 50-501XL<br>50-503XL | K172566 | MyoSure Hysteroscopic Tissue Removal System<br>and MyoSure Tissue Removal Devices | Hologic, Inc. |
| 50-601XL<br>50-603XL | K181974 | MyoSure XL Tissue Removal Device for Fluent | Hologic, Inc. |
The predicate devices have not been subject to a design-related recall.
### Device Description:
The Reprocessed MyoSure Tissue Removal Devices are sterile, disposable, hand-held tissue
{4}------------------------------------------------
removal devices that are used to hysteroscopically remove intrauterine tissue. The devices are connected via a 6-foot (1.8-meter) flexible drive shaft to a motorized control unit. A foot pedal allows the user to control the cutting action of the tissue removal device by turning the motor in the control unit on and off. The proximal end of the 10-foot (3-meter) vacuum tubing is connected to a collection canister. The devices use mechanical resection by drawing the tissue through the cutting window under suction, while the inner blade simultaneously rotates and reciprocates to cut the tissue.
| Device Specifications | | | | | |
|-------------------------------|-------------------|-------------------|------------------|-----------------|----------------|
| Model | Outer<br>Diameter | Working<br>Length | Window<br>Length | Window<br>Depth | Window<br>Size |
| Reprocessed MyoSure | 3 mm | 32 cm | 14.0 mm | 1.8 mm | 54 mm3 |
| Reprocessed MyoSure REACH | 3 mm | 32 cm | 14.0 mm | 1.8 mm | 54 mm3 |
| Reprocessed MyoSure LITE | 3 mm | 32 cm | 10.2 mm | 1.5 mm | 31 mm3 |
| Reprocessed MyoSure XL | 4 mm | 32 cm | 14.0 mm | 2.4 mm | 98 mm3 |
| Reprocessed MyoSure XL-Fluent | 4 mm | 32 cm | 14.0 mm | 2.4 mm | 98 mm3 |
# Indication for Use:
The Reprocessed MyoSure Tissue Removal Device (Model # 10-401/10-403) is indicated for intrauterine use by trained qynecologists to hysteroscopically resect and remove tissue such as: submucous myomas, endometrial polyps, and retained products of conception.
The Reprocessed MyoSure REACH Tissue Removal Device (Model # 10-401FC/10-403FC) is indicated for intrauterine use by trained gynecologists to hysteroscopically resect and remove tissue such as: submucous myomas, endometrial polyps, and retain products of conception.
The Reprocessed MyoSure LITE Tissue Removal Device (Model # 30-401LITE/30-403LITE) is indicated for intrauterine use by trained qynecologists to hysteroscopically resect and remove tissue such as: submucous myomas, endometrial polyps, and retain products of conception.
The Reprocessed MyoSure XL Tissue Removal Device (Model # 50-501XL/50-503XL) is indicated for intrauterine use by trained gynecologists to hysteroscopically resect and remove tissue such as: submucous myomas, endometrial polyps, and retain products of conception.
The Reprocessed MyoSure XL Tissue Removal Device for Fluent (Model # 50-601XL/50-603XL) is indicated for intrauterine use by trained qynecologists to hysteroscopically resect and remove tissue such as: submucous myomas, endometrial polyps, and retain products of conception.
The indications for use are the same as the predicate devices.
# Comparison of Technological Characteristics:
{5}------------------------------------------------
The principles of operation and primary functional specifications of the Reprocessed MyoSure Tissue Removal Devices are identical to those of the predicate MyoSure Tissue Removal Devices. There are no changes to the claims, intended use, clinical applications, patient population, performance specifications, or method of operation. The only differences between the Reprocessed MyoSure Tissue Removal Devices and the predicate devices MyoSure Tissue Removal Devices is that devices are reprocessed, and some device components are replaced with equivalent components during the reprocessing operation.
The differences in technological characteristics do not raise different questions of safety and effectiveness.
# Performance Data:
Bench and laboratory testing were conducted to demonstrate performance (safety and effectiveness) of the Reprocessed MyoSure Tissue Removal Device. This included the following tests:
# Biocompatibility
The biocompatibility evaluation for the Reprocessed MyoSure Tissue Removal Devices was conducted in accordance with FDA Guidance Document "Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process"" dated June 16, 2016, and International Standard ISO 10993-1, "Biological Evaluation of Medical Devices – Part 1: Evaluation and Testing Within a Risk Manaqement Process," as recognized by FDA. The battery of testing included the following tests:
- Cytotoxicity ●
- Sensitization
- Irritation
- Systemic Toxicity
- Materials Mediated Pyrogenicity ●
- Chemical Characterization .
The Reprocessed MyoSure Tissue Removal Devices are considered surface contacting, applied to breached surfaces for a duration of less than 24 hours.
# Validation of Reprocessing
The reprocessing validation was derived from the recommendations of AAMI/TIR 30:2011, "A Compendium of Processes, Materials. Test Methods, and Acceptance Criteria for Cleaning Reusable Medical Devices"; ASTM D7225-13:2013, "Standard Guide for Blood Cleaning Efficiency of Detergents and Washer-Disinfectors: AAMI/TIR 12:2010: Designing, Testing, and Labeling Reusable Medical Devices for Reprocessing in Health Care Facilities: A Guide for Device Manufacturers"; AAMI/ANSI ST35:2003, "Safe handling and biological decontamination of medical devices in health care facilities and in non-clinical settings"; ISO/TS 15883-5, "Washer-disinfectors – Part 5: Test soils and methods for demonstrating cleaning efficacy", and ASTM F3208-18, "Standard Guide for Selecting Test Soils for Validation of Cleaning Methods for Reusable Medical Devices."
# Sterilization and Shelf Life
{6}------------------------------------------------
Sterility of the Reprocessed MyoSure Tissue Removal Devices was assessed in accordance with ANSI/AAMI/ISO 11135-1:2014, "Sterilization of Health Care Products - Ethylene oxide -Part 1: Requirements for Development, Validation, and Routine Control of a Sterilization Process for Medical Devices."
Shelf Life verification testing included assessment of package integrity and device functional performance following aging.
# Electrical Safety
Electrical safety testing was conducted on the Reprocessed MyoSure Tissue Removal Devices. The devices comply with the IEC 60601-1 standard for safety.
# Functional Performance
- Shaft deflection ●
- Leak decay ●
- Drive cable plug insertion
- Shaft insertion force ●
- Torque ●
- Linearity ●
- . Revolutions Per Minute
- Reciprocation ●
- Wear particulate ●
- Bend
- Tissue cutting and removal performance
- Reliability testing
- Device functional attribute tests ●
- . Equipment compatibility
The bench testing involved evaluation of the devices' performance and ability to cut and resect tissue. The results of the evaluations demonstrate that the Reprocessed MyoSure Tissue Removal Devices effectively cut and resect tissue in accordance with the intended use.
# Conclusion:
The results of bench testing and laboratory evaluations demonstrate that the Reprocessed MyoSure Tissue Removal Devices are at least as safe and effective and perform as well as the identified legally marketed predicate devices as described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.