K201731 · Shanghai Apolo Medical Technology Co., Ltd. · PKT · Sep 10, 2020 · General, Plastic Surgery
Device Facts
Record ID
K201731
Device Name
Diode Laser Body Sculpture System
Applicant
Shanghai Apolo Medical Technology Co., Ltd.
Product Code
PKT · General, Plastic Surgery
Decision Date
Sep 10, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Diode Laser Body Sculpture Systems is intended for non-invasive lipolysis of the abdomen, flanks, back, and thighs in individuals with a Body Mass Index (BMI) of 30 or less. The device is intended to affect the appearance of visible fat bulges in the abdomen, flanks, back, and thighs.
Device Story
HS-851 Diode Laser Body Sculpture System uses 1060nm infrared laser energy for non-invasive hyperthermic lipolysis; targets adipose tissue to reduce fat cell count and contour body. System comprises main unit, control unit with software/UI, and treatment applicators. Operated by clinician in clinical setting; applicators secured to patient via belt. Device utilizes contact cooling during treatment. Output is heat-assisted disruption of adipocytes; intended to improve appearance of fat bulges. Benefits include non-invasive body contouring without surgery.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical safety, electromagnetic compatibility (IEC 60601-1, IEC 60601-1-2), laser safety (IEC 60825-1, IEC 60601-2-22), and biocompatibility (cytotoxicity, skin irritation, skin sensitization per ISO 10993-1).
Technological Characteristics
1060nm (±20nm) infrared diode laser; CW pulse width; contact cooling; power density 0.8-1.6W/cm²; peak power 35W-50W per applicator; belt-based attachment. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60825-1, IEC 60601-2-22. Biocompatibility per ISO 10993-1.
Indications for Use
Indicated for non-invasive lipolysis of abdomen, flanks, back, and thighs in individuals with BMI ≤ 30 to reduce visible fat bulges.
Regulatory Classification
Identification
A Low Level Laser System for Aesthetic Use is a device using low level laser energy for the disruption of adipocyte cells within the fat layer for the release of fat and lipids from these cells for noninvasive aesthetic use.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Guidance for Industry and Food and Drug Administration Staff; Class II Special Controls Guidance Document: Low Level Laser System for Aesthetic Use.” See § 878.1(e) for the availability of this guidance document.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is a circular seal with an abstract design. To the right of the seal, the text "FDA U.S. FOOD & DRUG ADMINISTRATION" is displayed in a bold, sans-serif font. The word "FDA" is in a blue square, and the rest of the text is in a smaller font size.
September 10, 2020
Shanghai Apolo Medical Technology Co., Ltd. Mr. Felix Li RA Supervisor Room 301-310, Building 11, No.388, Yindu Road, Xuhui District Shanghai, Shanghai 200231 China
Re: K201731
Trade/Device Name: Diode Laser Body Sculpture System Regulation Number: 21 CFR 878.5400 Regulation Name: Low Level Laser System For Aesthetic Use Regulatory Class: Class II Product Code: PKT Dated: June 18, 2020 Received: June 24, 2020
Dear Felix Li:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Purva Pandya Acting Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K201731
Device Name Diode Laser Body Sculpture Systems
#### Indications for Use (Describe)
The Diode Laser Body Sculpture Systems is intended for non-invasive lipolysis of the abdomen, flanks, back, and thighs in individuals with a Body Mass Index (BMI) of 30 or less. The device is intended to affect the appearance of visible fat bulges in the abdomen, flanks, back, and thighs.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: overline;">X</span> Prescription Use (Part 21 CFR 801 Subpart D) | Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K201731-510(k) summary
#### l Submitter
Shanghai Apolo Medical Technology Co., Ltd. Room 301-310, Building 11, No.388, Yindu Road, Xuhui District, 200231, Shanghai China Establishment Registration Number: 3007120647
Contact person: Felix Li Position: Regulatory Affairs Phone: +86-138 4919 0618 Fax: +86-21-34622840 E-mail: liqiang@apolo.com.cn
Date of preparation: Sep 9th, 2020
#### II Proposed Device
| Trade Name of Device: | Diode Laser Body Sculpture Systems |
|-----------------------|------------------------------------------|
| Common name: | Low Level Laser System for Aesthetic Use |
| Regulation Number: | 21 CFR 878.5400 |
| Regulatory Class: | Class II |
| Product code: | PKT |
| Review Panel | General & Plastic Surgery |
#### III Predicate Devices
| 510(k) Number: | K182741 |
|-----------------|------------------------------------------|
| Trade name: | SculpSure |
| Common name: | Low Level Laser System for Aesthetic Use |
| Classification: | Class II |
| Product Code: | PKT |
| Manufacturer | Cynosure |
| 510(k) Number: | K191068 |
| Trade name: | Powersculp laser lipolysis system |
| Common name: | Low Level Laser System for Aesthetic Use |
| Classification: | Class II |
| Product Code: | PKT |
| Manufacturer | Wuhan Lotuxs Technology Co., Ltd. |
{4}------------------------------------------------
### IV Device description
The HS-851 Diode Laser Body Sculpture Systems is a 1060nm Diode Hyperthermic Laser Lipolysis system and utilizes the latest technology for non-invasive body contouring. It applies a 1060nm wavelength laser to target the adipose tissue to reduce stubborn fat in areas such as flank, abdomen, back and thighs.
The purpose of the device is to breakdown the fat cells in the target area and in doing so contour the body and reduce the number of fat cells in that area.
The proposed device consists of the main unit, control unit with dedicated software and user interface, and the treatment applicators.
#### V Indication for use
The Diode Laser Body Sculpture Systems is intended for non-invasive lipolysis of the abdomen, flanks, back, and thighs in individuals with a Body Mass Index (BMI) of 30 or less. The device is intended to affect the appearance of visible fat bulges in the abdomen, flanks, back, and thighs.
| Item | Proposed device | Primary<br>predicate<br>device<br>(K182741) | Secondary<br>predicate<br>device<br>(K191068) | Discussion |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| Product<br>Code | PKT | PKT | PKT | Identical |
| Regulation<br>No. | 21 CFR 878.5400 | 21 CFR 878.5400 | 21 CFR 878.5400 | Identical |
| Class | Class II | Class II | Class II | Identical |
| Indication<br>for use | This product is<br>intended for<br>non-invasive<br>lipolysis of the<br>abdomen, flanks,<br>back, and thighs<br>in individuals with<br>a Body Mass<br>Index (BMI) of 30<br>or less. The | The Cynosure<br>SculpSure™ is<br>intended for<br>non-invasive<br>lipolysis of the<br>abdomen,<br>flanks, back, and<br>thighs in<br>individuals with<br>a Body Mass | The Powersculp<br>laser lipolysis<br>system is<br>intended for<br>non-invasive<br>lipolysis of the<br>flank and<br>abdomen to<br>achieve<br>disruption of | Equivalent |
| | device is intended to affect the appearance of visible fat bulges in the abdomen, flanks, back, and thighs. | Index (BMI) of 30 or less. In addition, the device is intended for non-invasive lipolysis of the submental area in individuals with a BMI of 49 or less. The device is intended to affect the appearance of visible fat bulges in the abdomen, flanks, back, thighs and submental area. When using the petite mask for non-invasive lipolysis of the submental area, the device can also affect the appearance of lax tissue in the submental area. | adipocyte celles intended for non-invasive aesthetic use to achieve a desired aesthetic effect. This treatment is intended for individuals with a Body Mass Index (BMI) of 30 or less. | |
| Laser type | Diode laser | Diode laser | Diode laser | Identical |
| Wavelength | 1060nm±20 nm (infrared) | 1060nm±20 nm (infrared) | 1060nm±20 nm (infrared) | Identical |
| Lipolysis method | Heat-assisted | Heat-assisted | Heat-assisted | Identical |
| Spot size | $4 * 6 cm^2/4 * 8 cm^2$ on each of the Applicator heads (up to four | $4 * 6 cm^2$ on each of the Applicator heads (up to four | $4 * 8 cm^2$ (A single applicator of four applicators) | Equivalent |
| | applicators per<br>body treatment) | applicators per<br>body treatment)<br>14.28 cm² (for<br>standard<br>submental<br>mask)<br>10.49 cm² (for<br>petite submental<br>mask) | | |
| Pulse width | CW | CW | CW | Identical |
| Power<br>density | 0.8 ~ 1.6W/cm² | Up to 1.4 W/ cm²<br>(body)<br>Up to 3.06 W/<br>cm²<br>(submental) | Up to 0.7-1.7W/<br>cm² | Equivalent |
| Attachment<br>to patient | Belt | Belt | Belt | Identical |
| Voltage | AC110-240V,<br>50/60Hz, 20-4A | AC200-240V,<br>50/60Hz, 20A | AC100-240V,<br>50/60Hz, 15A | Equivalent |
| Peak<br>power | 35W (per<br>applicator) for 4 *<br>6 cm²<br>50W (per<br>applicator) for 4 *<br>8 cm² | 30W (per<br>applicator) | 50W (per<br>applicator) | Equivalent |
| Cooling | Contact cooling | Contact cooling | Contact cooling | Identical |
#### VI Comparison of technological characteristics with the predicate devices
{5}------------------------------------------------
{6}------------------------------------------------
## VII Non-Clinical Testing
A series of tests have been performed to verity that the proposed device met all design specification. The test result demonstrated that the proposed device complies with the following standards:
#### Electrical safety and electromagnetic compatibility
- 60601-1: 2005+corr.1:2006+Corr.2.2007+A1:2012 Medical ● IEC electrical equipment - Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2:2014 Medical electrical equipment -Part 1-2: General ●
{7}------------------------------------------------
requirements for basic safety and essential performance-Collateral Standard: Electromagnetic disturbances - Requirements and tests
- IEC 60825-1:2014 Safety of Laser products-Part 1: Equipment classification and requirements
- IEC 60601-2-22:2007(third edition)+A1:2012 for use in conjunction with IEC ● 60601-1:2005 (third edition)+A1:2012 Medical electrical equipment - Part 2-22: Particular requirements for the safety and essential performance of surgical, cosmetic, therapeutic and diagnostic laser equipment
## Biocompatibility Evaluation:
Per FDA's Biocompatibility Guidance issued on June 16, 2016 and with regard to Table A.1 Evaluation Tests for consideration in ISO, "Use of international Standard ISO 10993-1, Biological evaluation of medical - Part 1: Evaluation and testing within a risk management process," the following tests performed on the material which contacts with human for Biocompatibility:
- Cytotoxicity; ●
- Skin irritation; ●
- Skin Sensitization .; ●
## VIII Clinical Testing
It is not applicable.
fonclusion substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.