K201730 · Foggiare Technologies, LLC · IRP · Aug 22, 2022 · Physical Medicine
Device Facts
Record ID
K201730
Device Name
Powered Inflatable Tube Massager - SLIM UP DRAIN
Applicant
Foggiare Technologies, LLC
Product Code
IRP · Physical Medicine
Decision Date
Aug 22, 2022
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The IAF-SYSTEM is intended for the temporary relief of minor muscle aches and pains and for the temporary increase in circulation to the treated areas in people who are in good health. IAF-SYSTEM simulates kneading of tissues by using an inflatable garment.
Device Story
IAF-SYSTEM is a powered inflatable tube massager for use in clinics, hospitals, athlete training, or home environments. The system comprises an air pump, air pressure sensor, microprocessor, and inflatable sleeves (leg, hip, or arm attachments) connected via hoses. Each sleeve contains four compression chambers. The device operates by sequentially inflating and deflating these chambers to simulate kneading and stroking of tissues, moving from limb end to body center. Users select therapy times (10, 20, 25, or 30 minutes) and pressure levels (0-119 mmHg). The microprocessor controls the timing and pressure based on sensor feedback to ensure safe operation. A user-accessible button allows for immediate pause or stop of the therapy session. The device provides non-invasive, mechanical compression to assist with muscle recovery and circulation, offering a similar therapeutic effect to established compression therapy systems.
Clinical Evidence
No clinical data. Safety and effectiveness were demonstrated through bench testing, including electrical safety (IEC 60601-1), EMC testing, and functional performance testing of the air pump, pressure sensors, and inflation cycles.
Indicated for healthy individuals seeking temporary relief of minor muscle aches and pains and temporary increase in circulation to treated areas via simulated tissue kneading and stroking.
Regulatory Classification
Identification
A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol is a stylized representation of a human figure, while the FDA name is written in blue and includes the words "U.S. Food & Drug Administration".
August 22, 2022
Foggiare Technologies LLC % Jay Mansour Principal Mansour Consulting LLC 845 Aronson Lake Court Roswell, Georgia 30075
Re: K201730
Trade/Device Name: IAF-SYSTEM Regulation Number: 21 CFR 890.5650 Regulation Name: Powered Inflatable Tube Massager Regulatory Class: Class II Product Code: IRP Dated: July 6, 2022 Received: July 7, 2022
Dear Jay Mansour:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Tushar Bansal, PhD Acting Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K201730
Device Name IAF-SYSTEM
Indications for Use (Describe)
IAF-SYSTEM is intended for the temporary relief of minor muscle aches and pains and for the temporary increase in circulation to the treated areas in people who are in good health. IAF-SYSTEM simulates kneading of tissues by using an inflatable garment.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------|--|
|-------------------------------------------------|--|
Prescription Use (Part 21 CFR 801 Subpart D)
X Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) #K201730
# 510k Summary:
As required by 21 CFR 807.92 (c)
1 - Date Summary Prepared: August 13, 2022
- 2 Owner/Submitter/Sponsor/Applicant information:
#### OWNER/SUBMITTER/SPONSOR/APPLICANT
Contact name: Edouard Sezionale- Foggiare Technologies LLC Title: owner/CEO Telephone: (561) 923-9161 Email address: thierry@foggiare.com Address: 426 E Palmetto Park Rd. Boca Raton, FL 33432 USA
### APPLICATION CORRESPONDENT
Contact name: Jay Mansour, MSQA, BE, RAC- Mansour Consulting LLC Title: regulatory consultant- Principal Telephone: (678) 908-8180 Email address: jay@mansourconsulting.com Address: 845 Aronson Lake Court Roswell, GA 30075 USA
#### 3 - Device Information:
Common/usual/classification name:
massager, powered inflatable tube
Device name:
| | Device Model Name | Model Number |
|---|-------------------|-------------------|
| 1 | IAF-SYSTEM | (NO MODEL NUMBER) |
| FDA 3 Letter Code | IRP |
|-------------------------------|-------------------|
| FDA regulation number: 21 CFR | 890.5650 |
| Regulation medical specialty | Physical Medicine |
| Review panel | Physical Medicine |
| Class | 2 |
{4}------------------------------------------------
#### 4 - Substantial equivalency is claimed against the following predicate device(s):
| 510k<br>Number | Trade or Proprietary or Model Name | Manufacturer | Primary Predicate |
|----------------|-----------------------------------------|-------------------------------------------|-------------------|
| k182668 | Rapid Reboot Compression Therapy System | Rapid Reboot<br>Recovery<br>Products, LLC | Yes |
#### 5 - Description of the device
Subject device consists of an air pump, air pressure sensor, and sleeves working together as one unit. The air pump is connected to the dedicated sleeves via a series of hoses; each sleeve has four (4) compression chambers. The compression massage direction is from limb end to body center. By inflating the air chambers sequentially and then deflating as one cycle, the pressure can be adjusted to avoid any discomfort to the patient. The sleeve works under the action of sensor and microprocessor. Software controls the timing and pressure reflected by the sensor, cycling airflow into and out of the sleeves to compress body
#### 6 - Intended Use + indications for use
| Intended Use: | The IAF-SYSTEM is intended for the temporary relief of minor muscle aches<br>and pains and for the temporary increase in circulation to the treated areas in |
|----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|
| indications for use: | people who are in good health. The IAF-SYSTEM simulates kneading and<br>stroking of tissues by using an inflatable garment. |
#### 7 - Basis for determination of substantial equivalency:
(a) Indications for use:
The indication for use is identical to the predicate's.
#### (b) Technological characteristics:
The technological characteristics are substantially equivalent to the predicate device, meeting the same technical standards.
Substantial equivalency between our device and predicate device is tabulated below:
{5}------------------------------------------------
| Device | Subject Device<br>Primary Predicate | | Comparison |
|------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | DAM SRL | Rapid Reboot Recovery<br>Products, LLC | NA |
| 510(k)<br>Number | TO BE DETERMINED | K182668 | NA |
| Model Name | IAF-SYSTEM | Rapid Reboot Compression<br>Therapy System | NA |
| Classification | Class II Device,<br>IRP (21 CFR890.5650) | Class II Device, IRP (21 CFR890.5650) | Same |
| Indications for<br>Use (IFU) | The IAF-SYSTEM is intended for the<br>temporary relief of minor muscle aches<br>and pains and for the temporary increase<br>in circulation to the treated areas in<br>people who are in good health. The IAF-<br>SYSTEM simulates kneading and stroking<br>of tissues by using an inflatable garment. | The Rapid Reboot<br>Compression Therapy System is<br>intended for the temporary relief of<br>minor muscle aches and pains and for<br>the temporary increase in circulation<br>to the treated areas in people who<br>are in good health. The Rapid Reboot<br>Compression Therapy System<br>simulates kneading and stroking of<br>tissues by using an inflatable<br>garment. | Same |
| OTC or Rx | OTC | OTC | Same |
| Environment<br>of Use: | Clinics, hospital, athlete training, and<br>home environment. | Clinics, hospital, athlete training, and<br>home environments | Same |
| Standards | ANSI AAMI ES60601- 1:2005/ (R)2012 And<br>A1:2012<br>IEC 60601-1-2 Edition 4.0<br>2014-02 | IEC 60601-1:2014; IEC<br>60601-1-2:2014 | Same |
| | Biocompatibility not required | EN ISO 10993-5:2009 & EN ISO<br>10993-10:2010 | Different.<br>Biocompatibility<br>not required for<br>subject device, as<br>users wear<br>clothing under<br>the garments. No<br>direct contact |
| Mode of<br>Compression | Sequential"Seawave"/ Peristaltic | Sequential/Peristaltic | Same |
| Power<br>Source | 120 V, 50Hz | 120V, 60Hz | Same |
{6}------------------------------------------------
| Therapy Time | User determines therapy<br>time. Choose from 10, 20,25 or 30<br>minute session time. | User determines therapy<br>time. Choose from 10, 20, or 30<br>minute session time, with option<br>to add additional 10 minutes to<br>any therapy time. | Same |
|-------------------------------------------------------------|--------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------|
| Max Pressure<br>Min Pressure | 0-119 mmHg | 0-200 mmHg | Different.<br>Subject device<br>provides<br>substantially<br>equivalent<br>pressure, despite<br>the different<br>maximum<br>pressure |
| Number of<br>Chambers | 4 Chambers | 4 Chambers | Same |
| Compression<br>Applicator<br>Garments<br>Sleeve<br>Material | Nylon with a Polyurethane laminate | Nylon with a Polyurethane<br>laminate | Same |
| Housing<br>Materials<br>And<br>Constructions | Molded ABS enclosure | Molded ABS enclosure | Same |
| Patient contact | Non-conductive attachments | Non-conductive attachments | Same |
| Power<br>Consumption | 400 W | 30W | Different due to<br>physical<br>configuration |
{7}------------------------------------------------
| Cycle time | Leg (consisting of foot, calf, knee, upper | | | | | | 1 min 20 sec | Same and different |
|------------|-----------------------------------------------------------------------------------------|--------------------------------------------|------------|------------|------------|-----------------------------------------------------|--------------|-----------------------------------------------------|
| | leg) | | | | | | | depending on the |
| | The cycle time for all programs is | | | | | | | program. See |
| | tabulated below. | | | | | | | below. |
| | The first cycle can be different from the | | | | | | | |
| | following ones, because it starts with | | | | | | | Information below |
| | | | | | | totally empty sectors, while the following | | lists 1st cycle, |
| | | cycles can start with some air in the | | | | | | subsequent cycle |
| | | | | | | | | |
| | | cycles change over the course of the | | | | sectors; moreover, in many programs the | | and predicate cycle,<br>respectively, all in |
| | | | | | | | | |
| | | program. This is illustrated within the | | | | | | seconds, comparing |
| | | user manual.<br>minutes | | | | | | each of the 12 |
| | | | | | | | | programs with the |
| | Prog. | Treatment | Sector | cycle | cycle | subsequent | | predicate cycle |
| | n.<br>ਹੈ | time<br>25 | pause<br>ਹ | pause<br>3 | time<br>50 | cycle time<br>38 | | time: |
| | 2 | 25 | 1 | 3 | 47 | ਚ ਹ | | Program 1: 50 / 38 / 80 |
| | m | 25 | ਹੈ | 3 | ರೇ | 33 | | Program 2: 47 / 41 / 80 |
| | বাঁ | | | 4 | ਚਰੇ | | | Program 3: 46 /33 / 80 |
| | | 25 | 2 | | | ਤੌਰੇ | | Program 4: 49 / 39 / 80 |
| | ട | 25 | 2 | 4 | ਦਰੇ | eo | | Program 5: 69 / 60 / 80 |
| | б | 25 | 2 | 4 | 80 | e8 | | Program 6: 80 / 68 / 80 |
| | 7 | 25 | 1 | 3 | 97 | 35 | | Program 7: 97 / 35 / 80 |
| | 8 | 30 | 1 | 3 | 260 | 140 | | Program 8: 260/140/80 |
| | | | | | | | | Program 9: 65 / 98 / 80 |
| | ਰੇ | 25 | ਹ | 3 | દર | ಕಿ8 | | Program 10: 206/203/80 |
| | 10 | 20 | 2 | 4 | 206 | 203 | | Program 11: 34 / 28 /80<br>Program 12: 34 / 33 / 80 |
| | 11 | 30 | 2 | 3 | 34 | 28 | | |
| | 12 | 20 | 2 | 3 | 34 | 33 | | Program 6 provides the |
| | | | | | | | | same exact (1st) cycle |
| | | | | | | The cycle time is the time needed to fill air in | | time as predicate device. |
| | | the sectors with set pressure in the set | | | | | | |
| | | sequence + sectors pause. | | | | | | All other programs offer |
| | | | | | | | | substantially the same |
| | | | | | | Sector pause is the time between the end of | | cycle time, varying |
| | | | | | | fill in of air in a sector and the start of fill in | | between 34 and 260 |
| | | of the air in the sequel one. | | | | | | seconds (for 1st cycle) and |
| | | | | | | The cycle pause is the time between the end | | between 33 and 203 |
| | | of a cycle and the start of the next one. | | | | | | seconds (for subsequent |
| | | | | | | 1st cycle time is the time needed for the 1st | 3 | cycles). |
| | | complete cycle (it needs a little more | | | | | | The differences in cycle |
| | | because start with total empty sectors). | | | | | | times are within the |
| | | | | | | Subsequent cycle time is the time need for | | same order of magnitude |
| | | | | | | other cycles ( being little bit shorter because | | as predicate device, and |
| | | the sectors are not totally empty). | | | | | | do not raise new |
| | | Exception in program 9: This program | | | | | | concerns. |
| | | includes a first section where the plantar | | | | | | |
| | | | | | | | | |
| | thrust provides pressure starting from the<br>bottom-base of the thigh, then mid-thigh, | | | | | | | |
| | then knee and calf, and finally knee and | | | | | | | |
| | ankle. The effect is therefore drainage from | | | | | | | |
| | the top. After that, it starts a treatment like | | | | | | | |
| | program 1. | | | | | | | |
{8}------------------------------------------------
| Size | 49″ x 21″ x 14.5″ | 10″ x 6.5″ x 5″ | Same.<br>(Tabletop or<br>handheld portable) |
|------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------|
| Weight | 77 pounds with base | 5.8 pounds | Different due to physical<br>configuration |
| Modes (Inflation<br>sequences, all<br>preprogrammed) | 2 modes: "A" mode inflates and deflates<br>chambers from bottom up (distal to<br>proximal chambers), one at a time. "B"<br>mode also inflates chambers from<br>bottom up, but maintains pressure in<br>selected chambers as works its way to<br>top. Then all chambers release pressure<br>at same time once all chambers have<br>sequentially inflated. | 2 modes: "A" mode inflates and<br>deflates chambers from bottom up<br>(distal to proximal chambers), one<br>at a time. "B" mode also inflates<br>chambers from bottom up, but<br>maintains pressure in lower<br>chambers as works its way to top.<br>Then all chambers release pressure<br>at same time once all chambers<br>have sequentially inflated. | Same.<br>Both devices have<br>sequential mode<br>(distal to<br>proximal<br>chambers). |
{9}------------------------------------------------
| Modes (visual<br>description) | "Sequential:" | Mode A: | Same |
|-------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | <br>With "lock" on specific area | | |
| | Mode seawave: | Mode B: | Same<br>The machine<br>permits applying<br>a set pressure in<br>a targeted way<br>to each single<br>"lock" sector to<br>determine the<br>effect and<br>effectiveness on<br>the patient.<br>One or more<br>sectors can be<br>kept under<br>pressure so that<br>during program<br>execution, these<br>do not deflate<br>during the cycle. |
| | <br>"SEA WAVE" LEG-PIECE<br>With overlapping bands, pressure which decreases towards<br>the top and programmable control of sectors,<br>the flow is physiological. | | |
4
{10}------------------------------------------------
| "Leg" Attachment | Leg (consisting of foot, calf, knee, upper leg) | Leg (consisting of foot, calf, knee, upper leg) | Same |
|--------------------------|----------------------------------------------------------|-----------------------------------------------------------------------------------------------------|---------------------------------|
| Leg Attachment<br>Photos | Image: Leg attachment photo | Image: Leg attachment photo | Same |
| Leg Attachment<br>Sizes | 51"x 26" | X-Short: 14" x 41"<br>Short: 14" x 43"<br>Medium: 14" x 45"<br>Long: 14" x 48"<br>X-Long: 14" x 52" | Different.<br>Slight variation. |
| "Hip" Attachment | Hip (consisting of upper legs, glutes, hips, lower back) | Hip (consisting of upper legs, glutes, hips, lower back) | Same |
{11}------------------------------------------------
| "Arm"<br>Attachment | Arm (consisting of entire<br>arm, shoulder, upper chest and back) | Arm (consisting of entire arm,<br>shoulder, upper chest and back) | Same |
|-------------------------------------------|--------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|-------------------------------|
| Arm Attachment<br>Photos | 66 cm<br>72 cm | | Same. |
| Arm Attachment<br>Sizes | One size: 26" x 28 "<br>72 cm<br>65 cm | Regular: 18" x 38"<br>Long: 18" x 44" | Same.<br>Slight<br>variation. |
| Safety Features | Button on display allows user to stop or<br>pause therapy session at any time. | Button on display allows user to<br>stop or pause therapy session at<br>any time. | Same |
| SW/Firmware/<br>Microprocessor<br>Control | Microprocessor | Microprocessor | Same |
| Technology | Compressor and valve system which<br>sequentially inflates cells of appliance | Compressor and valve system<br>which sequentially inflates cells of<br>appliance | Same |
(c) Non-clinical tests- brief discussion:
Safety testing and EMC testing successfully completed. In particular, the following tests were performed:
- 1- Earth circuit continuity test, leakage current, current absorption, current absorption: Summary of findings- device complies with these IEC 60601-1 safety requirements, in the lab as well as on production units (tests are conducted on on-going basis).
- 2- Main switch, air hose connector, delivered pressure, display function, programs, top card function, safety pressure switch intervention, maximum setting pressure, integrity of the inflatable parts (multiple)
{12}------------------------------------------------
Summary of findings- device complies with the acceptance criteria of these various tests, documenting that safety and effectiveness are achieved on validation units and on production units (tests are conducted on on-going basis).
(d) Clinical tests- brief discussion
Not applicable. No clinical tests were conducted.
(e) Non-clinical and clinical tests- conclusions drawn demonstrating that the device is as safe and as effective, and performs as well as the predicate device(s): Safety of subject device is confirmed via testing. Effectiveness is ensured by similar parameter of a well-understood device mechanism.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.