K201726 · Haolang Medical USA Corporation · DQY · Jul 7, 2021 · Cardiovascular
Device Facts
Record ID
K201726
Device Name
Arterial Catheter
Applicant
Haolang Medical USA Corporation
Product Code
DQY · Cardiovascular
Decision Date
Jul 7, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Arterial Catheter permits access to the peripheral arterial circulation.
Device Story
Arterial Catheter is a lumen peripherally inserted catheter assembly for short-term (less than 7 days) access to peripheral arterial circulation and pressure monitoring. Device components include introducer needle/catheter-over-needle, spring guide wire, and catheter assembly with extension tubing, slide clamp, and flex tube end cap. Insertion utilizes Seldinger catheter-over-guide-wire technique. Available in 18-24 ga sizes with lengths 2.5-23 cm. Catheters feature polyurethane hubs with Luer lock fittings. Device is provided sterile. Used by clinicians in clinical settings to facilitate arterial access; output is the established vascular access point for monitoring or fluid/medication delivery.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including visual assessment, corrosion resistance, liquid/air leakage, flow rate, tensile testing (catheter/guidewire/needle), dimensional verification, simulated use/blood withdrawal, kink testing, torque strength, radiopacity, and needle penetration force. Biocompatibility, shelf-life validation, and sterilization (EO) validation were performed.
Technological Characteristics
Materials: SUS 304 (needle), Polyurethane (tubing/hub), ABS (needle hub), Niti/Stainless Steel (guide wire), PVC (flex tube), HDPE (guide wire slide), LDPE (protector). Standards: ISO 10555-1 (catheter performance), ISO 11070 (introducer/wire), ISO 10993-1 (biocompatibility), ISO 11607-1 (packaging), ISO 11135 (EO sterilization). Form factor: Catheter-over-guide-wire assembly with Luer lock. No software or electronic components.
Indications for Use
Indicated for patients requiring short-term access to the peripheral arterial circulation for pressure monitoring or other clinical needs.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
July 7, 2021
Haolang Medical USA Corporation Lisa Xu RA Manager 1100 Bellevue Way NE 8A-533 Bellevue, Washington 98004
Re: K201726
Trade/Device Name: Arterial Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: September 1, 2020 Received: September 18, 2020
### Dear Lisa Xu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medical-devices/ device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-andcontinuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatory-assistance/contact-usdivision-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Lydia Glaw Assistant Director DHT2C: Division of Coronary and Peripheral Intervention Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K201726
Device Name Arterial Catheter
Indications for Use (Describe)
The Arterial Catheter permits access to the peripheral arterial circulation.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) Summary
Prepared in accordance with the requirements of 21 CFR Part 807.92
## Prepared Date: June 1, 2021
## 1. Submitter's Information
The submitter of this pre-market notification is:
| Name: | Haolang Medical USA Corporation |
|-----------------|-----------------------------------------------------|
| Address: | 1100 Bellevue Way NE 8A-533, Bellevue, WA 98004 USA |
| Contact person: | Lisa Xu |
| Title: | RA Manager |
| E-mail: | lisa.xu@haolangmed.com |
| Tel: | 1-425-503-6325 |
## 2. Device Identification
| Trade/Device Name: | Arterial Catheter |
|--------------------|--------------------------------|
| Models: | 18Ga, 20Ga, 21Ga, 22Ga, 24Ga |
| Regulation Number: | 21 CFR 870.1250 |
| Common Name: | Percutaneous Arterial Catheter |
| Regulation Class: | Class II |
| Product Code: | DQY |
## 3. Predicate Device
| 510(K) number: | K171146 |
|--------------------|-------------------------------------------------|
| Device Name: | Arrow Seldinger Arterial Catheterization Device |
| Manufacturer: | Arrow International, Inc. |
| Regulation Number: | 21 CFR 870.1250 |
| Regulation Name: | Percutaneous Catheter |
| Regulation Class: | Class II |
| Product Code: | DQY |
## 4. Device Description
The Arterial Catheter is a lumen peripherally inserted catheter assembly designed to perform short term access to the peripheral arterial circulatory system for less than 7 days, it is also intended to be used for pressure monitoring. It consists of an introducer needle or catheter-over-needle assembly, spring guide wire, and a catheter assembly that incorporates an extension tubing segment with a slide clamp and a flex tube end cap. These components are used together for catheter insertion using the "Seldinger" catheter-over-guide-wire insertion technique, which divided into three types according to structure: Standard Seedinger; Modified Seedinger-Integrated; Modified Seedinger-Sleeve.
The Arterial Catheter is available in 18 ga, 20 ga, 21 ga, 22 ga and 24 ga, the corresponding diameter of guide wire is 0.025", 0.018", 0.015", 0.012". Configurations with usable lengths of
Haolang Medical USA Corporation
{4}------------------------------------------------
2.5 cm to 23 cm. The catheters are attached to an injection molded polyurethane hub with Luer lock fittings for access attachment.
The Arterial Catheter is provided sterile.
### 5. Indication for use
The Arterial Catheter permits access to the peripheral arterial circulation.
### 6. Comparison to Predicate Device
Compared to the predicate devices, the subject device has same intended use, similar product design, performance effectiveness as the predicate device as summarized in the following table.
| Feature | Subject device | Predicate device | Comments |
|------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer/<br>K# | Haolang Medical USA<br>Corporation | Arrow International, Inc.<br>K171146 | -- |
| Trade name<br>and model | Arterial Catheter | Arrow Seldinger Arterial<br>Catheterization Device | -- |
| Classifications<br>name and<br>Regulation<br>Name | 21 CFR 870.1250<br>Percutaneous Catheter<br>Class II<br>DQY | 21 CFR 870.1250<br>Percutaneous Catheter<br>Class II<br>DQY | Same |
| Indications for<br>use | The Arterial Catheter<br>permits access to the<br>peripheral arterial<br>circulation. | The Arrow Seldinger<br>Arterial Catheterization<br>Devices permit access to<br>the peripheral arterial<br>circulation or to other<br>small vessels | Same |
| Catheter<br>Gauge Sizes | 18 ga, 20 ga, 21 ga, 22 ga<br>and 24 ga | 18 ga, 20 ga, 22 ga and 24<br>ga | We have one more<br>size than predicate<br>device, but we have<br>performed the<br>performance test,<br>the difference does<br>not raise the new<br>risk. |
| Catheter<br>Usable Length | 2.5 cm to 23 cm | 5cm - 23cm | According to<br>different usage<br>requirements, our<br>catheter length<br>range is larger, but<br>we have also<br>performed the<br>performance test on<br>which is smaller<br>than the range of<br>predicate device, the<br>difference does not<br>raise the new risk. |
| Introducer<br>needle size | 18Ga, 20Ga, 21Ga,<br>22Ga,23Ga, 24Ga, 26Ga | 18 ga.: 18 ga X 5cm-<br>7cm (2"-2 $\frac{3}{4}$ ")<br>20 ga : 20 ga X 4cm- | Different, The<br>coverage of the<br>Introducer needle |
| 7cm (1 ½"- 2 ¾")<br>22 ga.: 22 ga. X 4cm (1<br>1 ½") | size is more than<br>that of the predicate<br>device. We used a<br>thinner needle and<br>undergoing the<br>performance test<br>according to the ISO<br>10555-1 and ISO<br>11070. The<br>difference does not<br>affect the safety and<br>performance of the<br>device. | | |
| Spring Wire<br>Guide Size | 0.012", 0.015", 0.015",<br>0.018", 0.021", 0.025" | 18 ga.: 0.64mm diameter X<br>33.5cm-60cm (13"-23")<br>20 ga.: 0.533mm diameter X<br>35cm-50cm (13 ¾"-20")<br>22 ga.: 0.533mm<br>diameter X 35cm (13 ¾")<br>24 ga.: 0.46mm diameter<br>X 25cm (9 ¾") | Different, The size<br>coverage of spring<br>Wire Guide Size is<br>more than that of the<br>predicate device.<br>We are undergoing the<br>performance test<br>according to the ISO<br>10555-1 and ISO<br>11070. The<br>difference does not<br>affect the safety and<br>performance of the<br>device. |
| Spring Wire<br>Guide Tip | Straight | Straight and J-tip | Same |
| Spring Wire<br>Guide Depth<br>Markings | Yes | Yes, etched on straight<br>spring wire guides. Not on J-<br>tip spring wire guides | Same |
| Introducer<br>Method | Introducer needle and<br>introducer catheter-over-<br>needle | Introducer needle and<br>introducer catheter-over-<br>needle | Same |
| Sterilization<br>Method | EO | EO | Same |
| Shelf life | 3 years | 2 years | Different, we have<br>conducted the shelf<br>life validation based<br>on the 3 years, the<br>performance is<br>acceptable, the<br>difference does not<br>affect the safety and<br>performance of the<br>device. |
| Packaging | PET/LDPE film mated<br>with Tyvek | PET/LDPE film mated<br>with Tyvek | Same |
| | | | |
| Material | 1. Introducer Needle: SUS 304<br>2. Tubing, Catheter Hub<br>and Luer: PU<br>3. Needle Hub: ABS<br>4. Spring guide wire: Niti&<br>Stainless Steel<br>5. Flex tube: PVC<br>6. Guide wire slide: HDPE<br>7. Catheter protector tube:<br>LDPE | 1. Introducer Needle: SUS 304<br>2. Tubing, Catheter Hub and<br>Luer: Polyether urethane<br>3. Needle Hub: ABS<br>4. Spring guide wire: SUS 304<br>5. Flex tube: Not specified<br>6. Guide wire slide: Not<br>specified<br>7. Catheter protector tube:<br>Not specified | Different, The<br>introducer needle<br>and the catheter<br>body are made of<br>the same material,<br>but the production<br>process and<br>chemical<br>characterization<br>information of the<br>material is unknown,<br>so it is impossible to<br>prove that its<br>biocompatibility is<br>equivalent. We<br>conducted the<br>biocompatibility<br>evaluation on<br>subject device and<br>determined the<br>endpoint according<br>to ISO 10993-1,<br>which prove that the<br>material of subject<br>device is<br>biocompatibility, the<br>difference of<br>material does not<br>affect the safety and<br>performance of the<br>device. |
{5}------------------------------------------------
### Traditional 510(k) Submission of Arterial Catheter
{6}------------------------------------------------
## 7. Performance Data
#### Clinical test:
Clinical testing is not required.
#### Non-clinical data
Arterial Catheter comply with:
#### Performance:
According to the requirements of FDA guidance<Peripheral Percutaneous Transluminal Angioplasty (PTA) and Specialty Catheters - Premarket Notification (510(k)) Submissions> ISO 10555-1:2013 we performed the following testing:
- Visual assessment -
- -Corrosion Resistance
- Liquid Leakage Testing -
- Air Leakage Testing -
- Flow Rate Testing Report -
- Catheter Tensile Testing -
- Guidewire Tensile Testing -
Haolang Medical USA Corporation
{7}------------------------------------------------
- Dimensional verification -
- Simulated use and Simulated Blood Withdrawal -
- -Kink Testing
- -Torque Strength Testing
- Radiopacity Testing -
- Needle testing: Visual assessment; Tensile strength testing of needle and hub; Penetration force
#### Shelf Life:
Accelerated aging tests were conducted to confirm the validity of the 3 years shelf life.
#### Biocompatibility:
The compatibility of the skin-contact component material in the finished product meets the requirement of Biocompatibility. The Biological Evaluation Tests are in compliance with the standards of ISO 10993-1, "Biological Evaluation of Medical Devices".
#### Sterilization:
- -ISO 11607-1:2016 Packaging for terminally sterilized medical devices - Part 1: Requirements for materials, sterile barrier systems and packaging systems.
- -ISO 11135: 2014 Sterilization of health care products -- Ethylene oxide -- Requirements for the development, validation and routine control of a sterilization process for medical devices
#### 9. Conclusion
Based on the results of above testing, the Arterial Catheter is substantially equivalent to the predicate device Arrow Seldinger Arterial Catheterization Device (K171146).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.