K201720 · Beijing Bywave Sensing Medical Technology Co., Ltd. · DQX · Feb 18, 2021 · Cardiovascular
Device Facts
Record ID
K201720
Device Name
LiPPS Analyzer, LiPPS Wire Pressure Guide Wire
Applicant
Beijing Bywave Sensing Medical Technology Co., Ltd.
Product Code
DQX · Cardiovascular
Decision Date
Feb 18, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1330
Device Class
Class 2
Indications for Use
LiPPS™ Intravascular Pressure Sensing System is intended to be used in an environment like cathlab and for use in blood vessels, including coronary and peripheral vessels, to measure intravascular pressure during angiography and/or interventional procedures. Pressure measurements are obtained to provide hemodynamic information, such as fractional flow reserve, for the diagnosis and treatment of blood vessel diseases.
Device Story
The LiPPS™ Intravascular Pressure Sensing System consists of the H2000 LiPPS™ Analyzer and the LiPPS™ Wire Pressure Guide Wire. The system is used in a cathlab environment by clinicians. The sterile, single-use guide wire contains an optical sensor and fiber to measure intravascular blood pressure. Signals are transmitted via an optical cable and connector to the reusable H2000 LiPPS™ Analyzer. The analyzer processes these signals to display real-time pressure curves, numerical values (systolic, diastolic, mean pressure, heart rate), and fractional flow reserve (FFR) values. The system also accepts aortic pressure input from an external transducer to calculate FFR. The output provides hemodynamic data to assist clinicians in diagnosing and treating blood vessel diseases, enabling informed clinical decision-making during interventional procedures.
Clinical Evidence
Bench testing only. Performance verified against IEC 60601-1, IEC 60601-1-2, IEC 60601-2-34, and ISO 11070 standards. Biocompatibility testing conducted per ISO 10993-1 for external communicating devices in contact with circulating blood (<24 hours), including cytotoxicity, sensitization, systemic toxicity, pyrogenicity, hemolysis, and thromboresistance studies.
Technological Characteristics
System comprises a reusable H2000 LiPPS™ Analyzer and a sterile, single-use 0.014" guide wire. Guide wire uses fiberoptic sensing (stainless steel/nitinol core, platinum-nickel alloy coil, PTFE/hydrophilic coating). Analyzer performs signal processing and display. Sterilization via ethylene oxide (ISO 11135). Connectivity includes optical cable/connector and external pressure transducer input. Software developed per IEC 62304.
Indications for Use
Indicated for use in coronary and peripheral blood vessels to measure intravascular pressure during angiography and interventional procedures to provide hemodynamic information, such as fractional flow reserve, for the diagnosis and treatment of blood vessel diseases.
Regulatory Classification
Identification
A catheter guide wire is a coiled wire that is designed to fit inside a percutaneous catheter for the purpose of directing the catheter through a blood vessel.
Special Controls
*Classification.* Class II (special controls). The device, when it is a torque device that is manually operated, non-patient contacting, and intended to manipulate non-cerebral vascular guide wires, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
PrimeWire PRESTIGE® Plus Pressure Guide Wire (K111395)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 18, 2021
Beijing Bywave Sensing Medical Technology Co., Ltd. Lizhe Zhang General Manager Room 501, Buildinctg 22, No.12, Juyuan Middle Road, Mapo Town, Shunyi District Beijing, Beijing 101399 China
Re: K201720
Trade/Device Name: LiPPSTM Intravascular Pressure Sensing System Regulation Number: 21 CFR 870.1330 Regulation Name: Catheter Guide Wire Regulatory Class: Class II Product Code: DOX, DXO Dated: January 13, 2021 Received: January 21, 2021
Dear Lizhe Zhang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
LT Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K201720
Device Name
LiPPSTM Intravascular Pressure Sensing System
Indications for Use (Describe)
LiPPS™ Intravascular Pressure Sensing System is intended to be used in an environment like cathlab and for use in blood vessels, including coronary and peripheral vessels, to measure intravascular pressure during angiography and/or interventional procedures. Pressurements are obtained to provide hemodynamic information, such as fractional flow reserve, for the diagnosis and treatment of blood vessel diseases.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------------|---------------------------------------------|
| <span style="text-decoration: underline;"><b>Prescription Use (Part 21 CFR 801 Subpart D)</b></span> | Over The Counter Use (21 CFR 801 Subpart C) |
X Prescription Use (Part 21 CFR 801 Subpart D)
he-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# Section 5: 510(K) Summary
This summary of 510(K) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1900 and 21 CFR 807.92.
### The assigned 510(K) Number: K201720
#### ડ. 510(K) Summary
#### 5.1. Date of Preparation: 22nd, May, 2020
#### 5.2. Sponsor
Beijing Bywave Sensing Medical Technology Co., Ltd. Room 501, Building 22, No. 12, Juyuan Middle Road, Mapo Town, Shunyi District Beijing, P. R. China Tel: (86)10-82890879 Fax: (86)10-82890879
Contact Person: Lizhe Zhang Position: General Manager Email: zhanglizhe@bws-tech.com
### 5.3. Subject Device Identification
Subject Device Name: LiPPSTM Intravascular Pressure Sensing System Common name: Intravascular Pressure Sensing System Classification Name(s): Catheter Tip Pressure Transducer/ Catheter Guide Wire Product Code: DXO, DQX Regulation Number: 21 CFR 870.2870/ 21 CFR 870.1330 Review Panel: Cardiovascular Classification: II
### 5.4. Predicate Device
510(k) Number: K142598 Device Name: OptoWire and OptoMonitor System Manufacturer: Opsens, Inc. 510(k) Number: K111395 Device Name: PrimeWire PRESTIGE® Plus Pressure Guide Wire Manufacturer: Volcano Corporation
#### 5.5. Indications for use:
LiPPS™ Intravascular Pressure Sensing System is intended to be used in an environment like cathlab and for use in blood vessels, including coronary and peripheral vessels, to measure intravascular pressure during angiography and/or interventional procedures. Pressure measurements are obtained
{4}------------------------------------------------
to provide hemodynamic information, such as fractional flow reserve, for the diagnosis and treatment of blood vessel diseases.
#### 5.6. Device Description
The subject device LiPPSTM Intravascular Pressure Sensing System contains H2000 LiPPS™ Analyzer and LiPPS™ Wire Pressure Guide Wire.
The LiPPS™ Wire Pressure Guide Wire is used together with the H2000 LiPPS™ Analyzer in order to measure intravascular pressure. The LiPPS™ Wire contains an optical sensor and an optical fiber to induce the blood pressure and transmit signal. The LiPPS™ Wire can be connected to the analyzer through the companied handle (with an optical cable and an optical connector). The LiPPS™ Analyzer is an electronic signal processing and display units which can process signals received from the LiPPSTM Wire to display intravascular blood pressure and fractional flow reserve (FFR) values, and various connection cables. LiPPS™ Wire and companied handle (with an optical cable and an optical connector) are sterile, single-use devices. The LiPPS™ Analyzer and its accessories are reusable.
{5}------------------------------------------------
K201720
| Item | Subject Device | Predicate Device 1 | Predicate Device 2 | Remark | | |
|---------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------|--------------------------|--------------------------------------|
| | LiPPSTM Intravascular<br>Pressure Sensing System | OptoWire and<br>OptoMonitor System<br>(K142598 Opsens, Inc.) | PrimeWire PRESTIGE®<br>Plus Pressure Guide Wire<br>(K111395 Volcano Corp.) | | | |
| Device<br>Common/Usual<br>Name | LiPPSTM Intravascular Pressure<br>Sensing System | Intravascular Pressure<br>Monitoring System | PrimeWire PRESTIGE® Plus<br>Pressure Guide Wire | / | | |
| Device Class | Class II | Class II | Class II | SE | | |
| Product Code/<br>Regulation Number | DXO, DQX<br>21 CFR 870.2870/ 21 CFR<br>870.1330 | DXO, DQX<br>21 CFR 870.2870/ 21 CFR<br>870.1330 | DXO, DQX<br>21 CFR 870.2870/ 21 CFR<br>870.1330 | SE | | |
| Classification<br>Name(s) | Catheter Tip Pressure Transducer<br>Catheter Guide Wire | Catheter Tip Pressure Transducer<br>Catheter Guide Wire | Catheter Tip Pressure Transducer<br>Catheter Guide Wire | SE | | |
| Indications for use | LiPPSTM Intravascular Pressure<br>Sensing System is intended to be<br>used in an environment like cathlab<br>and for use in blood vessels,<br>including coronary and peripheral<br>vessels, to measure intravascular<br>pressure during angiography and/or<br>interventional procedures. Pressure<br>measurements are obtained to<br>provide hemodynamic information,<br>such as fractional flow reserve, for<br>the diagnosis and treatment of<br>blood vessel diseases. | To measure pressure in blood<br>vessels including both coronary<br>and peripheral vessels, during<br>diagnostic angiography and/or<br>other any interventional<br>procedures. Blood pressure<br>measurements provide<br>hemodynamic information, such<br>as fractional flow reserve, for the<br>diagnosis and treatment of blood<br>vessel disease. | The Prime Wire PRESTIGE® Plus<br>Pressure Guide Wire Device is<br>indicated for use to measure<br>pressure in blood vessels including<br>both coronary and peripheral<br>vessels, during diagnostic<br>angiography and/or any<br>interventional procedures. Blood<br>pressure measurements provide<br>hemodynamic information for the<br>diagnosis and treatment of blood<br>vessel disease. | No substantial<br>difference | | |
| Intended use | to measure intravascular pressure | To measure pressure ... during | to measure pressure ... during | Though expressed | | |
| Item | Subject Device<br>LiPPSTM Intravascular<br>Pressure Sensing System | Predicate Device 1<br>OptoWire and<br>OptoMonitor System<br>(K142598 Opsens, Inc.) | Predicate Device 2<br>PrimeWire PRESTIGE®/<br>Plus Pressure Guide Wire<br>(K111395 Volcano Corp.) | Remark | | |
| | during angiography and/or<br>interventional procedures | diagnostic angiography and/or<br>other any interventional<br>procedures. | diagnostic angiography and/or any<br>interventional procedure | in different<br>wording, the<br>subject device has<br>same intended<br>purpose with<br>predicate devices.<br>SE | | |
| System Components | Sterile, disposable guidewire<br>Reusable signal processor /<br>monitor<br>Embedded software<br>Connecting cables | Sterile, disposable guidewire<br>Reusable signal processor /<br>monitor<br>Embedded software<br>Connecting cables | Sterile, disposable guidewire | SE with predicate<br>device K142598.<br>Predicate device<br>K111395 is only a<br>pressure guide<br>wire. | | |
| System Capabilities | Measurement of intravascular<br>blood pressure including FFR. | Measurement of intravascular<br>blood pressure including FFR. | Measurement of intravascular<br>blood pressure and flow including<br>FFR (when used with<br>pressure/flow system) | SE | | |
| Prescription Use | Rx Only | Rx Only | Rx Only | SE | | |
| Pressure Sensing &<br>Signal Transmission<br>Technology | Fiberoptic sensor & fiber bundle<br>embedded in guidewire | Fiberoptic sensor & fiber bundle<br>embedded in guidewire | Hard wired strain gauge<br>embedded in guidewire | No substantial<br>difference | | |
| Sterile, Single Use<br>Patient<br>Contact<br>Component? | Yes –LiPPSTM Wire Pressure<br>Guide Wire | Yes - OptoWire | Yes - PrimeWire Prestige Plus | SE | | |
| Item | Subject Device | Predicate Device 1 | Predicate Device 2 | Remark | | |
| | LiPPSTM Intravascular<br>Pressure Sensing System | OptoWire and<br>OptoMonitor System<br>(K142598 Opsens, Inc.) | PrimeWire PRESTIGE®<br>Plus Pressure Guide Wire<br>(K111395 Volcano Corp.) | | | |
| FFR Capability? | Yes | Yes | Yes | SE | | |
| FFR Viewing | Yes | Yes | N/A | SE with predicate<br>device K142598<br>Predicate device<br>K111395 is only a<br>pressure guide<br>wire. | | |
| Basis for FFR<br>Determination | Simultaneous acquisition of 2<br>pressure values: distal pressure<br>from sensor embedded in LiPPSTM<br>Wire Pressure Guide Wire; aortic<br>pressure from external pressure<br>transducer | Simultaneous acquisition of 2<br>pressure values: distal pressure<br>from sensor embedded in<br>OptoWire; aortic pressure from<br>external pressure transducer | Simultaneous acquisition of 2<br>pressure values: distal pressure<br>from sensor embedded in<br>PrimeWire; aortic pressure from<br>external pressure transducer | SE | | |
| Pressure range<br>(analyzer) | -30 to 300mmHg | -30 to 300mmHg | N/A | SE with predicate<br>device K142598<br>Predicate device<br>K111395 is only a<br>pressure guide<br>wire. | | |
| Pressure range<br>(guide wire) | -300 to 300mmHg | -300 to 300mmHg | Unknown | SE | | |
| Accuracy | $\pm$ 1mmHg plus $\pm$ 1% of reading<br>(over the range -30 to 50mmHg) or<br>$\pm$ 1mmHg plus $\pm$ 3% of reading<br>(over the range 50 to 300 mmHg) | $\pm$ 1 mmHg plus $\pm$ 1% of<br>reading (pressure range -30 to 50<br>mmHg)<br>or $\pm$ 3% of reading (pressure | Unknown | SE with predicate<br>device K142598. | | |
| Item | Subject Device<br>LiPPSTM Intravascular<br>Pressure Sensing System | Predicate Device 1<br>OptoWire and<br>OptoMonitor System<br>(K142598 Opsens, Inc.)<br>range 50 to 300 mmHg) | Predicate Device 2<br>PrimeWire PRESTIGE®<br>Plus Pressure Guide Wire<br>(K111395 Volcano Corp.) | Remark | | |
| Zero thermal effect | < 0.3 mmHg / °C | < 0.3 mmHg / °C | < 0.3 mmHg / °C | SE | | |
| Zero drift | < 1 mmHg / h | < 1 mmHg / h | Unknown | SE with predicate<br>device K142598. | | |
| User Interface | Touchscreen, Remote Control,<br>Barcode scanner | Touchscreen | N/A | No substantial<br>difference | | |
| Auto-zeroing | Yes | Yes | N/A | SE with predicate<br>device K142598. | | |
| Real Time Curves | Aortic instantaneous pressure,<br>aortic mean pressure, distal<br>instantaneous pressure, distal mean<br>pressure, Pd/Pa trend curve | Aortic instantaneous pressure,<br>aortic mean pressure, distal<br>instantaneous pressure, distal<br>mean pressure | N/A | No substantial<br>difference | | |
| Real Time<br>Numerical Values | Systolic, diastolic and mean blood<br>pressure (aortic and distal), heart<br>rate, and mean Pd/mean Pa | Mean aortic pressure, mean distal<br>pressure, mean Pd/mean Pa | N/A | No substantial<br>difference | | |
| Recording values | Instantaneous Pa, Pd and Pd/Pa;<br>mean Pa; mean Pd; mean Pd/mean<br>Pa | Instantaneous Pa, Pd and Pd/Pa;<br>mean Pa; mean Pd; mean<br>Pd/mean Pa | N/A | SE with predicate<br>device K142598. | | |
| Minimum Pd/Pa<br>Cursor<br>(Detection of FFR<br>Locus) | Yes | Yes | N/A | SE with predicate<br>device K142598. | | |
| Display Monitor | LCD | LCD | N/A | SE with predicate<br>device K142598. | | |
| Aortic Input | Low Level (5uV/V/mmHg) | High Level (100 mmHg/V) | N/A | Different | | |
| Item | Subject Device | Predicate Device 1 | Predicate Device 2 | Remark | | |
| | LiPPSTM Intravascular | OptoWire and | PrimeWire PRESTIGE® | | | |
| | Pressure Sensing System | OptoMonitor System | Plus Pressure Guide Wire | | | |
| | | (K142598 Opsens, Inc.) | (K111395 Volcano Corp.) | | | |
| Aortic Output | Low Level ( $5\mu V/V/mmHg$ ) | No | N/A | Different | | |
| Distal input | LiPPS™ Wire Pressure Guide<br>Wire (optical) | OptoWire (optical) | N/A | SE with predicate | | |
| | | | | device K142598. | | |
| Distal output | Low Level ( $5\mu V/V/mmHg$ ) | Low Level ( $5\mu V /V/mmHg$ ) | N/A | SE with predicate | | |
| | | | | device K142598. | | |
| Guide wire OD | 0.014" | 0.014" | 0.014" | SE | | |
| Guide wire length | 180 cm | 175 cm | 185 cm, 300 cm | Different | | |
| Guide wire material | Stainless Steel; Nitinol | Stainless Steel; Nitinol | Stainless Steel; SS | SE with predicate | | |
| | | | | device K142598. | | |
| Guidewire Coating | PTFE; Hydrophilic coating | Teflon; Silicone | Teflon; GlyDx Hydrophilic<br>coating | No substantial<br>difference | | |
| Guidewire Coating<br>Material | PTFE; Hydrophilic coating…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.