K201697 · Haolang Medical USA Corporation · FOS · May 6, 2021 · General Hospital
Device Facts
Record ID
K201697
Device Name
Umbilical Vessels Catheter
Applicant
Haolang Medical USA Corporation
Product Code
FOS · General Hospital
Decision Date
May 6, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5200
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
UMBILICAL VESSELS CATHETER is intended for short-term vascular access via umbilical vessels in neonatal patients: Via the umbilical artery: ---sampling of blood, ---measurement of intravascular arterial pressure. Via the umbilical vein: ---infusion (simultaneous infusion of several medicinal products for the dual lumen catheter).
Device Story
Umbilical Vessels Catheter provides short-term vascular access in neonates. Device consists of radiopaque polyurethane tube with heat-melted rounded tip to minimize tissue trauma; available in single and dual lumen configurations (2 Fr to 7 Fr sizes; 15-40 cm lengths). Depth markings at 1 cm intervals facilitate placement, which requires X-ray confirmation. Device is non-active, manual-use, ethylene oxide sterilized, and latex/DEHP-free. Used in clinical settings for blood sampling, arterial pressure monitoring, or fluid/medication infusion. Output is direct vascular access for clinician-managed therapy or monitoring.
Clinical Evidence
No clinical data. Substantial equivalence demonstrated via non-clinical bench testing, including biocompatibility (ISO 10993), sterilization validation (ISO 11135), and mechanical performance (ISO 10555-1, ISO 594-1/2).
Technological Characteristics
Materials: Radiopaque polyurethane, POM M370 44. Dimensions: 2-7 Fr, 15-40 cm. Sensing/Actuation: Manual. Connectivity: None. Sterilization: Ethylene Oxide. Standards: ISO 10555-1, ISO 594-1/2, ISO 10993 series, ASTM F88, F1886, F1929, F1140, D4169, D3078.
Indications for Use
Indicated for short-term vascular access via umbilical vessels in neonatal patients for blood sampling, intravascular arterial pressure measurement, or infusion of medicinal products.
Regulatory Classification
Identification
An intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings and that is inserted into the patient's vascular system for short term use (less than 30 days) to sample blood, monitor blood pressure, or administer fluids intravenously. The device may be constructed of metal, rubber, plastic, or a combination of these materials.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA acronym with the agency's name on the right. The symbol on the left is a stylized representation of human services, while the right side features the FDA acronym in a blue square, followed by the words "U.S. FOOD & DRUG" and "ADMINISTRATION" in a smaller font size.
May 6, 2021
Haolang Medical USA Corporation % Field Fu Senior Consultant Shenzhen Joyantech Consulting Co., Ltd. 1713A, 17th Floor, Bloack A, Zhongguan Times Square, Xili Town Shenzhen, Guangdong 518000 China
Re: K201697
Trade/Device Name: Umbilical Vessels Catheter Regulation Number: 21 CFR 880.5200 Regulation Name: Intravascular Catheter Regulatory Class: Class II Product Code: FOS Dated: March 29, 2021 Received: April 5, 2021
Dear Field Fu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201697
Device Name UMBILICAL VESSELS CATHETER
Indications for Use (Describe)
UMBILICAL VESSELS CATHETER is intended for short-term vascular access via umbilical vessels in neonatal patients:
Via the umbilical artery:
--- sampling of blood,
---measurement of intravascular arterial pressure.
Via the umbilical vein:
--- infusion (simultaneous infusion of several medicinal products for the dual lumen catheter).
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
-
#### Product: Umbilical Vessels Catheter
# 510(k) Summary
## 1. Contact Details
## 1.1 Applicant information
| Applicant Name | Haolang Medical USA Corporation. |
|-------------------------|--------------------------------------------------------|
| Address | 1100 Bellevue Way NE 8A-533<br>Bellevue, WA 98004, USA |
| Phone No. | 1-425-503-6325 |
| Fax No. | No. |
| Contact person | Lisa.Xu |
| Contact person's e-mail | lisa.xu@haolangmed.com |
| Date Prepared | May 30, 2020 |
## 1.2 Submission
- Correspondent
| Image: logo | Shenzhen Joyantech Consulting Co., Ltd |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------|
| | 1713A, 17th Floor, Block A, Zhongguan Times Square, Liuxian<br>Avenue, Xili Town, Nanshan District, Shenzhen, Guangdong<br>Province, China |
| Image: Chinese characters | |
| Phone No. | +86-755-86069197 |
| Contact person | Field Fu |
| Contact person's e-mail | field@cefda.com; info@cefda.com |
| Website | http://www.cefda.com |
## 2. Device information
| Trade name | Umbilical Vessels Catheter |
|---------------------|--------------------------------------------|
| Common name | Umbilical Vessels Catheter |
| Model | Single lumen catheter, dual lumen catheter |
| Classification | II |
| Classification name | Intravascular Catheter |
| Product code | FOS |
| Regulation No. | 880.5200 |
## 3. Legally Marketed Predicate Device
| Trade Name | Umbilical Vessels Catheter |
|---------------|----------------------------|
| 510(k) Number | K130725 |
| Product Code | FOS |
| Manufacturer | Covidien |
{4}------------------------------------------------
#### 4. Device Description
The UMBILICAL VESSELS CATHETER has catheters in single lumen and dual lumen. The catheters tube is made of radiopaque polyurethane which is insert molded to a polyurethane hub. The insertion tip of the catheter has been heat melted to form a smooth, rounded tip to reduce tissue trauma upon insertion.
Placement of the device is facilitated by depth markings, which are printed on the catheter at 1 cm intervals, beginning at 5 cm from the insertion tip and continuing to 40 cm from the tip. Catheter placement must be confirmed by x-ray.
The UMBILICAL VESSELS CATHETER has 17 primary configurations: Catheters in 2 Fr., 2.5 Fr., 2.8 Fr., 3 Fr., 3.5 Fr., 4 Fr., 5 Fr., 6 Fr., 6.5 Fr. and 7 Fr. for Single lumen; 2.8 Fr., 3 Fr., 3.5 Fr., 4 Fr., 5 Fr., 6 Fr. and 6.5 Fr. for dual lumen. Each device will continue to be packaged in a Tyvek pouches are packaged in a carton. The UMBILICAL VESSELS CATHETER does not contain DEHP and is Ethylene Oxide sterilized. The product and packaging are not made of natural rubber latex.
#### 5. Intended Use/Indication for Use
UMBILICAL VESSELS CATHETER is intended for short-term vascular access via umbilical vessels in neonatal patients:
Via the umbilical artery:
---sampling of blood,
---measurement of intravascular arterial pressure.
Via the umbilical vein:
---infusion (simultaneous infusion of several medicinal products for the dual lumen catheter).
{5}------------------------------------------------
| Item | Proposed Device:<br>Umbilical Vessels Catheter | Predicate Device:<br>Umbilical Vessels Catheter<br>(K130725) | Comments |
|---------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|
| Product Code | FOS | FOS | Same |
| Intended<br>Use/Indication for<br>use | UMBILICAL VESSELS<br>CATHETER is intended for<br>short-term vascular access<br>via umbilical vessels in<br>neonatal patients:<br>Via the umbilical artery:<br>---sampling of blood,<br>---measurement of<br>intravascular arterial<br>pressure.<br>Via the umbilical vein:<br>---infusion<br>(simultaneous infusion of<br>several medicinal products for<br>the dual lumen catheter). | The Argyle TM<br>Polyurethane<br>Umbilical Vessel<br>Catheter is<br>intended for<br>short-term vascular<br>access via<br>umbilical vessels in<br>neonatal patients. | SE, only<br>wording is<br>different (01). |
| Catheter Type | Single lumen and dual lumen | Single lumen | Different (02) |
| Catheter size | Single lumen:<br>(2 Fr, 2.5Fr, 2.8Fr, 3.0Fr,<br>3.5Fr, 4.0Fr, 5.0Fr, 6.0Fr,<br>6.5Fr, 7.0Fr) x (15cm, 20cm,<br>25cm, 28cm, 30cm, 32cm,<br>34cm, 35cm, 40cm);<br>Dual lumen:<br>(2.8Fr, 3.0Fr, 3.5Fr, 4.0Fr,<br>5.0Fr, 6.0Fr, 6.5Fr) x (15cm,<br>20cm, 25cm, 28cm, 30cm,<br>32cm, 34cm, 35cm, 40cm). | 2.5 Fr x 25 cm,<br>3.5 Fr x 25 cm,<br>and 5 Fr x 25 cm | Different (03) |
| Operation mode | Manual | Manual | Same |
| Radio-detectability | Yes | Yes | Same |
| Item | Proposed Device:<br>Umbilical Vessels Catheter | Predicate Device:<br>Umbilical Vessels<br>Catheter<br>(K130725) | Comments |
| water flow rate | The flowrate for each lumen is<br>not less than 80% of the<br>Nominal value for catheters of<br>nominal outside diameter less<br>than 1.0; and the flowrate for<br>each lumen is not less than<br>90% of the Nominal value for<br>catheters of nominal outside<br>diameter equal to 1.0 or<br>greater. | Not publicly<br>available | Comply with<br>ISO 10555-1 |
| Peak tensile force | More than 5N, 10N, or15N | Not publicly<br>available | SE (Comply<br>with ISO<br>10555-1) |
| Freedom from<br>leakage | No leakage for liquid and air | Not publicly<br>available | SE (Comply<br>with ISO<br>10555-1) |
| Material | Polyurethane<br>POM M370 44 | Not publicly<br>available | SE (Comply<br>with ISO<br>10993 series) |
# 6. Substantial Equivalence Comparison for Umbilical Vessels Catheter
{6}------------------------------------------------
Discussion on differences from the predicate device:
01:
The Umbilical Vessels catheter indications for use is more specific than the predicate. The indications for use for other currently marketed Umbilical catheters on the market are similar as they are indicated for blood sampling, pressure monitoring and administration of fluids and medications. Although the subject device and the predicate device wording is different, the intended use is the same.
## 02 & 03:
The subject device is available in single and double lumen catheter configuration, with sizes 2Fr to 7Fr and 15-40 cm lengths. The predicate device is only available in a single lumen configuration 2.5 Fr to 5 Fr 25 cm length. The Umbilical catheters currently on the market come in both single and double lumen configurations and are available in various sizes lengths between 2 Fr to 8 Fr.
Performance testing was performed per recognized standards ISO 10555-1 and ISO 594-1/ISO 594-2, demonstrating that differences in physical characteristics (e.g. single lumen or dual lumen, length, size, etc.) do not raise new or different questions of safety and effectiveness.
{7}------------------------------------------------
## 7. Non-clinical Testing
All nonclinical testing performed on new devices is to demonstrate the substantial equivalence to the predicate devices. Tests setup and execution are performed in accordance with applicable standards. Results of the tests demonstrate the compliance to the standards.
| Standard | Standard Title | Test Results |
|------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------|
| ISO 11135:2014 | Sterilization of health-care products - Ethylene oxide<br>-Requirements for the development, validation and<br>routine control of a sterilization process for medical<br>devices | Pass |
| ASTM F88/F88M-09 | Standard Test Method for Seal Strength of Flexible<br>Barrier Materials | Pass |
| ASTM F1886-98 | Standard Test Method for Determining Integrity of<br>Seals for Flexible Packaging by Visual Inspection | Pass |
| ASTM F1929-98<br>(Reapproved 2004) | Standard Test Method for Detecting Seal Leaks in<br>Porous Medical Packaging by Dye Penetration | Pass |
| ASTM F1140-00 | Standard Test Methods for Internal Pressurization<br>Failure Resistance of Unrestrained Packages for<br>Medical Applications | Pass |
| ASTMD4169-05 | Standard Practice for Performance Testing of<br>Shipping Containers and Systems | Pass |
| ASTM D3078-02 | Standard Test Method for Determination of Leaks in<br>Flexible Packaging by Bubble Emission | Pass |
| ASTM F1886-98 | Standard Test Method for Determining Integrity of<br>Seals for Flexible Packaging by Visual Inspection | Pass |
#### a) The tests concerned sterilization and shelf life as follow:
#### b) The tests concerned biocompatibility as follow:
| Test Item | Reference | Standard Title | Test Results |
|---------------------------------|---------------------------------|------------------------------------------------------------------------------------------------------------------------|--------------|
| Genotoxicity | ISO 10993-3:2014 | Biological evaluation of medical devices<br>—Part 3: Tests for genotoxicity, carcinogenicity and reproductive toxicity | Pass |
| Hemocompatibility,<br>Hemolytic | ISO 10993-4:2006;<br>ASTM F756, | Biological evaluation of medical devices<br>—Part 4: Selection of tests for interactions with blood; | Pass |
{8}------------------------------------------------
| Cytotoxicity | ISO<br>10993-5:2009 | Standard Practice for Assessment of<br>Hemolytic Properties of Materials. | Pass |
|-----------------------------------------------|----------------------|------------------------------------------------------------------------------------------------------|------|
| Implantation | ISO<br>10993-6:2016 | Biological evaluation of medical devices<br>—Part 6: Tests for local effects after<br>implantation | Pass |
| EO,ECH residual | ISO<br>10993-7:2008 | Biological evaluation of medical devices<br>-Part 7: Ethylene oxide sterilization<br>residuals | Pass |
| Sensitization | ISO10993-10:2<br>010 | Biological evaluation of medical devices<br>—Part 10: Tests for irritation and skin<br>sensitization | Pass |
| Irritation or<br>Intracutaneous<br>reactivity | ISO<br>10993-10:2010 | Biological evaluation of medical devices<br>-Part 10: Tests for irritation and skin<br>sensitization | Pass |
| Acute Systemic<br>toxicity | ISO<br>10993-11:2006 | Biological evaluation of medical devices<br>-Part 11: Tests for systemic toxicity | Pass |
| Subacute<br>Subchronic toxicity | ISO<br>10993-11:2006 | Biological evaluation of medical devices<br>-Part 11: Tests for systemic toxicity | Pass |
| Material-Mediated<br>Pyrogenicity | ISO<br>10993-11:2006 | Biological evaluation of medical devices<br>-Part 11: Tests for systemic toxicity | Pass |
| Limulus Amebocyte<br>Lysate | USP<85> | United States Pharmacopoeia | Pass |
## c) Software validation
Not applicable, the subject devices are non-active products.
#### d) Electric safety
Not applicable, the subject devices are non-active products.
## e) The tests concerned performance-bench as follow:
| Standard | Standard Title | Test Result |
|------------------|----------------------------------------------------------------------------------------------------------------------------------------|-------------|
| ISO 10555-1:2013 | Intravascular catheters — Sterile and single-use<br>catheters —Part 1: General requirements | Pass |
| ISO 594-1:1986 | Conical fittings with a 6% (Luer) taper for syringes,<br>needles and certain other medical equipment - Part<br>1: General requirements | Pass |
{9}------------------------------------------------
| Standard | Standard Title | Test Result |
|----------------|---------------------------------------------------------------------------------------------------------------------------------|-------------|
| ISO 594-2:1998 | Conical fittings with a 6% (Luer) taper for syringes,<br>needles and certain other medical equipment - Part<br>2: Lock fittings | Pass |
| ISO7886-1:2017 | Sterile hypodermic syringes for single use - Part 1:<br>Syringes for manual use | Pass |
#### 8. Animal study data
Substantial equivalence does not depend on animal study data.
## 9. Clinical Testing
Substantial equivalence does not depend on the clinical test data.
## 10. Conclusions
Based on device comparison information and non-clinical bench testing, the proposed device is substantially equivalent to legally marketed predicate devices (K130725).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.