GYNECARE TVT™ System, GYNECARE TVT™ with Abdominal Guides System, GYNECARE TVT™ Introducer and Catheter Guide, GYNECARE TVT EXACT™ Continence System and Trocar, GYNECARE TVT™ Obturator System and Passers and Winged Guide, GYNECARE TVT ABBREVO™ Continence System and Passers and Winged Guide
K201686 · ETHICON, Inc. · OTN · Sep 17, 2020 · General, Plastic Surgery
Device Facts
Record ID
K201686
Device Name
GYNECARE TVT™ System, GYNECARE TVT™ with Abdominal Guides System, GYNECARE TVT™ Introducer and Catheter Guide, GYNECARE TVT EXACT™ Continence System and Trocar, GYNECARE TVT™ Obturator System and Passers and Winged Guide, GYNECARE TVT ABBREVO™ Continence System and Passers and Winged Guide
Applicant
ETHICON, Inc.
Product Code
OTN · General, Plastic Surgery
Decision Date
Sep 17, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3300
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The GYNECARE TVT Device is intended to be used as a pubo-urethral sling for treatment of Stress Urinary Incontinence (SUI), for female urinary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency. The GYNECARE TVT Introducer and Rigid Catheter Guide are available separately and are intended to facilitate in the placement of the GYNECARE TVT Device. The GYNECARE TVT Device is intended to be used as a pubo-urethral sling for treatment of stress urinary incontinence (SUI), for female urinary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency. The GYNECARE TVT Introducer and Rigid Catheter Guide are available separately and GYNECARE TVT Abdominal Guides and Couplers are included in each kit. The GYNECARE TVT Introducer is a reusable device intended to aid in the placement of the GYNECARE TVT Device retropubically. The GYNECARE TVT Rigid Catheter Guide is a reusable device intended to facilitate the identification of the urethra and bladder neck during the placement of the GYNECARE TVT Device. The GYNECARE TVT Abdominal Guides and Couplers are single use devices used to facilitate placement of the GYNECARE TVT Device when placed in a top-down retropubic fashion (also known as an abdominal approach). The GYNECARE TVT EXACT Continence System is intended to be used as a pubo-urethral sling for treatment of female Stress Urinary Incontinence, resulting from urethral hypermobility and/or intrinsic sphincter deficiency. The GYNECARE TVT EXACT Continence System Trocar is a single use device intended to aid in the placement of the GYNECARE TVT EXACT Continence System retropubically. The GYNECARE TVT Rigid Catheter Guide is a reusable device intended to facilitate the identification of the urethra and bladder neck during the placement of the GYNECARE TVT EXACT Continence System. The GYNECARE TVT Obturator Device is intended to be used in women as a sub-urethral sling for the treatment of stress urinary incontinence (SUI) resulting from urethral hypermobility and/or intrinsic sphincter deficiency. The GYNECARE TVT Obturator Helical Passers and Atraumatic Winged Guide are intended to aid in the placement of the GYNECARE TVT Obturator Device. The GYNECARE TVT ABBREVO Continence System is intended for use in women as a sub-urethral sling for the treatment of SUI resulting from urethral hypermobility and/or intrinsic sphincter deficiency. The GYNECARE TVT ABBREVO Helical Passers and Atraumatic Winged Guide are intended to aid in the placement of the GYNECARE TVT ABBREVO Device.
Device Story
Surgical mesh systems for treatment of female stress urinary incontinence; consists of PROLENE polypropylene mesh tape and various stainless steel delivery instruments (introducers, helical passers, trocars, winged guides). Used by surgeons in clinical settings to place mesh sub-urethrally or retropubically to support urethra. Mesh provides mechanical support to treat urethral hypermobility/intrinsic sphincter deficiency. Delivery instruments facilitate precise placement through vaginal/abdominal incisions. Devices are either single-use (mesh/kits) or reusable (stainless steel instruments). Output is physical support of urethra; clinical decision-making based on patient anatomy and incontinence diagnosis. Benefits include restoration of continence.
Indicated for female patients with stress urinary incontinence (SUI) caused by urethral hypermobility and/or intrinsic sphincter deficiency requiring a pubo-urethral or sub-urethral sling.
Regulatory Classification
Identification
Surgical mesh is a metallic or polymeric screen intended to be implanted to reinforce soft tissue or bone where weakness exists. Examples of surgical mesh are metallic and polymeric mesh for hernia repair, and acetabular and cement restrictor mesh used during orthopedic surgery.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains two logos. On the left is the Department of Health & Human Services logo. On the right is the FDA logo, which includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
September 17, 2020
Ethicon, Inc. Ariell Joiner Regulatory Affairs Specialist III Route 22 West Somerville, NJ 08876-0151
Re: K201686
Trade/Device Name: GYNECARE TVT™ System, GYNECARE TVT™ with Abdominal Guides System, GYNECARE TVT™ Introducer and Catheter Guide, GYNECARE TVT EXACT™ Continence System and Trocar, GYNECARE TVTTM Obturator System and Passers and Winged Guide. GYNECARE TVT ABBREVO™ Continence System and Passers and Winged Guide
Regulation Number: 21 CFR§ 878.3300 Regulation Name: Surgical Mesh Regulatory Class: II Product Code: OTN, PWJ Dated: June 19, 2020 Received: June 22, 2020
Dear Ariell Joiner:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Sharon M. Andrews Director DHT3B: Division of Reproductive, Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K201686
#### Device Name
GYNECARE TVT™ System. GYNECARE TVT™ with Abdominal Guides System. GYNECARE TVT™ Introducer and Catheter Guide, GYNECARE TVT EXACT™ Continence System and Trocar, GYNECARE TVT™ Obturator System and Passers and Winged Guide, GYNECARE TVT ABBREVO™ Continence System and Passers and Winged Guide
#### Indications for Use (Describe)
The GYNECARE TVT Device is intended to be used as a pubo-urethral sling for treatment of Stress Urinary Incontinence (SUI), for female urinary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency. The GYNECARE TVT Introducer and Rigid Catheter Guide are available separately and are intended to facilitate in the placement of the GYNECARE TVT Device.
The GYNECARE TVT Device is intended to be used as a pubo-urethral sling for treatment of stress urinary incontinence (SUI), for female urinary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency. The GYNECARE TVT Introducer and Rigid Catheter Guide are available separately and GYNECARE TVT Abdominal Guides and Couplers are included in each kit.
The GYNECARE TVT Introducer is a reusable device intended to aid in the placement of the GYNECARE TVT Device retropubically.
The GYNECARE TVT Rigid Catheter Guide is a reusable device intended to facilitate the identification of the urethra and bladder neck during the placement of the GYNECARE TVT Device.
The GYNECARE TVT Abdominal Guides and Couplers are single use devices used to facilitate placement of the GYNECARE TVT Device when placed in a top-down retropubic fashion (also known as an abdominal approach).
The GYNECARE TVT EXACT Continence System is intended to be used as a pubo-urethral sling for treatment of female Stress Urinary Incontinence, resulting from urethral hypermobility and/or intrinsic sphincter deficiency.
The GYNECARE TVT EXACT Continence System Trocar is a single use device intended to aid in the placement of the GYNECARE TVT EXACT Continence System retropubically.
The GYNECARE TVT Rigid Catheter Guide is a reusable device intended to facilitate the identification of the urethra and bladder neck during the placement of the GYNECARE TVT EXACT Continence System.
The GYNECARE TVT Obturator Device is intended to be used in women as a sub-urethral sling for the treatment of stress urinary incontinence (SUI) resulting from urethral hypermobility and/or intrinsic sphincter deficiency.
The GYNECARE TVT Obturator Helical Passers and Atraumatic Winged Guide are intended to aid in the placement of the GYNECARE TVT Obturator Device.
The GYNECARE TVT ABBREVO Continence System is intended for use in women as a sub-urethral sling for the treatment of SUI resulting from urethral hypermobility and/or intrinsic sphincter deficiency.
The GYNECARE TVT ABBREVO Helical Passers and Atraumatic Winged Guide are intended to aid in the placement of the GYNECARE TVT ABBREVO Device.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
{3}------------------------------------------------
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) Summary
| Submitter: | Ethicon, Inc. a Johnson & Johnson company<br>P.O. Box 151<br>Route 22 West<br>Somerville, NJ 08876-0151 |
|-----------------|--------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Ariell Joiner<br>Regulatory Affairs Specialist<br>Phone: 908-218-2348<br>Fax: 908-218-2595<br>Email: ajoiner@its.jnj.com |
| Date Prepared: | September 16, 2020 |
Urogynecologic Surgical Mesh Systems and Surgical Instruments
| Device Trade Name: | GYNECARE TVTTM System<br>GYNECARE TVTTM with Abdominal Guides System<br>GYNECARE TVTTM Introducer and Catheter Guide<br>GYNECARE TVT EXACTTM Continence System and Trocar<br>GYNECARE TVTTM Obturator System and Passers and Winged Guide<br>GYNECARE TVT ABBREVOTM Continence System and Passers and<br>Winged Guide |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device Common Name: | Mesh, Surgical, Synthetic, Urogynecologic, For Stress Urinary<br>Incontinence, Retropubic Or Transobturator<br><br>Instrumentation, Surgical Mesh, Urogynecologic, Stress Urinary<br>Incontinence |
| Class: | II |
| Classification: | 21 CFR 878.3300 - Surgical mesh |
| Product Code: | OTN and PWJ |
| Panel: | General and Plastic Surgery Devices |
#### Predicate Devices:
| Device | Meshes | | Instruments | |
|---------------------------------------------------------|-----------------------------------------------|---------|--------------|---------|
| | Product Code | 510(k) | Product Code | 510(k) |
| GYNECARE TVT™ Tension-free<br>Vaginal Tape (TVT System) | OTN | K012628 | PWJ | K173019 |
| GYNECARE TVT™ Reusable<br>Introducer | Not Applicable<br>(Reusable Instruments only) | | PWJ | K173019 |
| | | | PWJ | K173162 |
{5}------------------------------------------------
| Device | Meshes | | Instruments | |
|-----------------------------------------------------------------------------------------|-----------------------------------------------|---------|--------------|---------|
| | Product Code | 510(k) | Product Code | 510(k) |
| GYNECARE TVTTM Reusable Rigid<br>Catheter Guide | Not Applicable<br>(Reusable Instruments only) | | PWJ | K173019 |
| | PWJ | K173162 | | |
| GYNECARE TVTTM with Abdominal<br>Guides Tension-Free Support for<br>Incontinence System | OTN | K012628 | PWJ | K173162 |
| GYNECARE TVT EXACTTM<br>Continence System | OTN | K132054 | PWJ | K173019 |
| GYNECARE TVTTM Obturator<br>System | OTN | K033568 | PWJ | K181151 |
| GYNECARE TVT ABBREVOTM<br>Continence System | OTN | K100936 | PWJ | K181151 |
### Device Description:
### GYNECARE TVT™ Tension-free Vaginal Tape System
GYNECARE TVT™ Device is a sterile single use device, consisting of one piece of undyed or blue (up to 0.28 weight percent phthalocyanine blue, Color Index number 74160) PROLENE™ Polypropylene Mesh (tape) approximately 1/2 inch x 18 inches (1.1 cm x 45 cm), covered by clear plastic Sheaths overlapping in the middle, and held between two stainless steel needles bonded to the mesh and sheath with plastic collars.
PROLENE Polypropylene Mesh is constructed of knitted filaments of extruded polypropylene strands identical in composition to that used in PROLENE™ Polypropylene Nonabsorbable Surgical Sutures. The mesh is approximately 0.027 inches (0.7 mm) thick. PROLENE Mesh is knitted by a process which interlinks each fiber junction.
GYNECARE TVT™ Reusable Introducer (available separately) is provided non-sterile and is reusable. The Introducer is made of stainless steel. It consists of two parts, a handle and an inserted threaded metal shaft. The Introducer is intended to facilitate the passage of the GYNECARE TVT Device from the vagina to the abdominal skin. It is connected and fixed to the Needle, via the threaded end of the shaft, prior to inserting the Needle with the tape.
{6}------------------------------------------------
GYNECARE TVT™ Rigid Catheter Guide (available separately) is a non-sterile reusable stainless-steel instrument intended to facilitate the identification of the urethra and the bladder neck during the placement of the GYNECARE TVT Device and the GYNECARE TVT EXACT Continence System. It is inserted into a Foley catheter (recommended size 18 French) positioned in the bladder via the urethra.
### GYNECARE TVT™ with Abdominal Guides Tension-free Support for Incontinence System
GYNECARE TVT™ Abdominal Guide is a sterile disposable instrument intended to facilitate the passage of the GYNECARE TVT Device. Two Abdominal Guides are included in each kit with the GYNECARE TVT Couplers.
GYNECARE TVT™ Coupler is a sterile disposable polypropylene connector used to connect the GYNECARE TVT Abdominal Guide to the GYNECARE TVT Needle. Two couplers are included in each kit with Abdominal Guides.
## GYNECARE TVT EXACT™ Continence System
GYNECARE TVT EXACT™ Continence System is a sterile, single patient use procedure kit consisting of:
## A. GYNECARE TVT EXACT Continence System Trocar Sheath / Implant Assembly
The GYNECARE TVT EXACT™ Continence System Trocar Sheath / Implant Assembly consists of one piece of blue (up to 0.28 weight percent phthalocyanine blue, Color Index number 74160) PROLENE™ Polypropylene Mesh (Implant) approximately 1/2 inch x 18 inches (1.1 cm x 45 cm), covered by clear plastic Sheaths and overlapping in the middle, and held between two white Trocar Sheaths, which are bonded to the Implant and Implant Sheath. PROLENE Propylene Mesh is constructed of knitted filaments of extruded polypropylene strands identical in composition to that used in PROLENE™ Polypropylene Nonabsorbable Surgical Sutures. The Implant is approximately 0.027 inches (0.7 mm) thick. PROLENE Mesh is knitted by a process which interlinks each fiber junction.
# B. GYNECARE TVT EXACT Continence System Trocar
The GYNECARE TVT EXACT™ Continence System Trocar consists of the stainless-steel Trocar Shaft and the plastic Trocar Handle. The Trocar Shaft is designed to fit inside the white Trocar Sheaths on the GYNECARE TVT EXACT™ Continence System Implant / Trocar Sheath Assembly and is used to position the GYNECARE TVT EXACT Continence System Implant in the patient, from a vaginal incision up through the abdominal wall.
### GYNECARE TVT™ Obturator System
The GYNECARE TVT™ Obturator System is a sterile, single patient use procedure kit consisting
{7}------------------------------------------------
### A. GYNECARE TVT Obturator Device
The GYNECARE TV Obturator Device consisting of one piece of undyed or blue (up to 0.28 weight percent phthalocyanine blue, Color Index 74160) PROLENE™ polypropylene mesh (tape) approximately 1/2 inch x 18 inches (1.1 cm x 45 cm), covered by clear plastic sheath overlapping in the middle. Plastic tube receptacles are attached at each end.
PROLENE Propylene Mesh is constructed of knitted filaments of extruded polypropylene strands identical in composition to that used in PROLENE Polypropylene Non-absorbable Surgical Suture. The implant is approximately 0.027 inches (0.7 mm) thick.
## B. GYNECARE TVT™ Obturator Helical Passers and Atraumatic Winged Guide
GYNECARE TVT™ Obturator Helical Passers are two stainless steel, curved wire passers with plastic handles that are designed to deliver the GYNECARE TVT Obturator Device. Helical Passers are provided as left and right units, pre-assembled to the GYNECARE TVT Obturator Device.
GYNECARE TVT™ Obturator Atraumatic Winged Guide is a stainless steel accessory instrument, which facilitates the passage of the GYNECARE TVT Helical Passers through the dissection tract.
### GYNECARE TVT ABBREVO™ Continence System
GYNECARE TVT ABBREVO™ Continence System is a sterile, single-patient use procedure kit consisting of:
# A. GYNECARE TVT ABBREVO Implant Assembly
The GYNECARE TVT ABBREVO Implant Assembly consists of one piece of blue (up to 0.28 weight percent phthalocyanine blue, Color Index number 74160) PROLENE™ Polypropylene Mesh (Implant), approximately 1/2 inch x 4.7 inches (1.1 cm x 12 cm), covered by clear plastic Sheaths and held between two Helical Passer Sheaths (white polyethylene tube receptacles). The Helical Passer Sheaths are bonded to the mesh implant sheaths and the Positioning Lines (PROLENE™ Polypropylene Monofilament). PROLENE is constructed of knitted monofilaments of extruded polypropylene, identical in composition to that used in PROLENE™ Polypropylene Nonabsorbable Surgical Sutures. The Implant is approximately 0.027 inches (0.7 mm) thick.
# B. GYNECARE TVT ABBREVO Placement Loop
The GYNECARE TVT ABBREVO Placement Loop is a sterile, single-patient use device consisting of a loop of PROLENE Polypropylene Monofilament with an attached polypropylene button. This loop and button are pre-assembled as part of the GYNECARE TVT ABBREVO Implant Assembly at the center of the mesh to aid in symmetric placement of the mesh.
#### of:
{8}------------------------------------------------
# C. GYNECARE TVT ABBREVO™ Helical Passers and Atraumatic Winged Guide
GYNECARE TVT ABBREVO™ Helical Passers are two stainless steel, curved wire passers with plastic handles that are designed to deliver the GYNECARE TVT ABBREVO Implant. The Helical Passers are provided as left and right units, pre-assembled to the GYNECARE TVT ABBREVO Implant Assembly.
GYNECARE TVT ABBREVO™ Atraumatic Winged Guide is a stainless steel accessory instrument which facilitates consistent passage of the GYNECARE TVT ABBREVO Helical Passers through the dissection tract. The Winged Guide is marked with an Insertion Zone to aid the surgeon's assessment of inserted depth. The Insertion Zone indicates a distance of 3 cm to 4 cm from the tip of the Winged Guide.
### Indications for Use:
The GYNECARE TVT Device is intended to be used as a pubo-urethral sling for treatment of Stress Urinary Incontinence (SUI), for female urinary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency. The GYNECARE TVT Introducer and Rigid Catheter Guide are available separately and are intended to facilitate in the placement of the GYNECARE TVT Device.
The GYNECARE TVT Introducer is a reusable device intended to aid in the placement of the GYNECARE TVT Device retropubically.
The GYNECARE TVT Rigid Catheter Guide is a reusable device intended to facilitate the identification of the urethra and bladder neck during the placement of the GYNECARE TVT Device.
The GYNECARE TVT Device is intended to be used as a pubo-urethral sling for treatment of stress urinary incontinence (SUI), for female urinary incontinence resulting from urethral hypermobility and/or intrinsic sphincter deficiency. The GYNECARE TVT Introducer and Rigid Catheter Guide are available separately and GYNECARE TVT Abdominal Guides and Couplers are included in each kit.
The GYNECARE TVT Abdominal Guides and Couplers are single use devices used to facilitate placement of the GYNECARE TVT Device when placed in a top-down retropubic fashion (also known as an abdominal approach).
The GYNECARE TVT EXACT Continence System is intended to be used as a pubo-urethral sling for treatment of female Stress Urinary Incontinence, resulting from urethral hypermobility and/or intrinsic sphincter deficiency.
The GYNECARE TVT EXACT Continence System Trocar is a single use device intended to aid in the placement of the GYNECARE TVT EXACT Continence System retropubically.
{9}------------------------------------------------
The GYNECARE TVT Rigid Catheter Guide is a reusable device intended to facilitate the identification of the urethra and bladder neck during the placement of the GYNECARE TVT EXACT Continence System.
The GYNECARE TVT Obturator Device is intended to be used in women as a sub-urethral sling for the treatment of stress urinary incontinence (SUI) resulting from urethral hypermobility and/or intrinsic sphincter deficiency.
The GYNECARE TVT Obturator Helical Passers and Atraumatic Winged Guide are intended to aid in the placement of the GYNECARE TVT Obturator Device.
The GYNECARE TVT ABBREVO Continence System is intended for use in women as a suburethral sling for the treatment of SUI resulting from urethral hypermobility and/or intrinsic sphincter deficiency.
The GYNECARE TVT ABBREVO Helical Passers and Atraumatic Winged Guide are intended to aid in the placement of the GYNECARE TVT ABBREVO Device.
# The subject devices and their respective predicate devices have the same intended use.
### Summary of Technological Characteristics:
The subject devices and their respective predicate devices are identical and therefore have the same technological characteristics.
### Substantial Equivalence Discussion:
The subject devices differ from their respective predicate devices only in the labeling (Instructions for Use). The Instructions for Use of the subject devices have been revised to meet the new European Union (EU) Medical Device Regulation (MDR) requirements and include new sections to the Instructions or Use and clarifications throughout various existing sections of the Instructions for Use including additional warnings, adverse events, safety information, and information to be conveyed to patients. There are also minor clarifications to the Indications and Contraindications statements, and no changes to the single-use and reusable designations and patient populations. No performance data are needed to evaluate these labeling changes.
# Conclusion:
Based on the intended use, fundamental scientific technology, and technological characteristics, the subject devices are substantially equivalent to their respective predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.