A-Silicone for Bite Registration is indicated for making occlusal registrations.
Device Story
A-Silicone for Bite Registration is a vinylpolysiloxane (addition silicone) impression material used for bite registration. It consists of two components (polydimethylsiloxane, silica, and platinum catalyst) supplied in a 1:1 double cartridge system. When mixed, the components undergo a platinum-catalyzed addition curing reaction to form an elastomeric material. The device is used by dental professionals in surgery or laboratory environments to capture the occlusal relationship of a patient's teeth. The clinician applies the material using a dispenser and mixing tips to record the bite. The resulting cured impression provides a stable, accurate reproduction of the dental structure, aiding in the fabrication of dental restorations or appliances. The device is a passive material and does not involve electronic processing or software.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing, including physical property testing (consistency, linear dimensional change, detail reproduction, elastic recovery) per ISO 4823 and biocompatibility testing (cytotoxicity, sensitization, irritation) per ISO 10993.
Technological Characteristics
Vinylpolysiloxane (addition silicone) material; two-component platinum-catalyzed system. Supplied in 1:1 double cartridge with dispenser and mixing tips. Complies with ISO 4823 (dentistry elastomeric impression materials) and ISO 10993 (biocompatibility).
Indications for Use
Indicated for making occlusal registrations in dental surgery and laboratory settings.
Regulatory Classification
Identification
Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 24, 2020
Maggie Zheng, Regulatory Affairs Manager Shandong Huge Dental Material Corporation No. 68 Shanhai Road, Donggang District Rizhao City. CHINA 276800 Shandong Province
#### Re: K201684
Trade/Device Name: A-Silicone for Bite Registration Regulation Number: 21 CFR 872.3660 Regulation Name: Impression Material Regulatory Class: Class II Product Code: ELW Dated: June 15, 2020 Received: June 22, 2020
#### Dear Maggie Zheng:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Srinvas "Nandu" Nandkumar, Ph.D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K201684
Device Name A-Silicone for Bite Registration
Indications for Use (Describe) A-Silicone for Bite Registration is indicated for making occlusal registrations.
| Type of Use ( <i>Select one or both, as applicable</i> ) | |
|----------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------|
| <div> <span> ☒ Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> | <div> <span> ☐ Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
***DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.***
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect
ete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
the
{3}------------------------------------------------
## K201684
# 510 (k) Summary
This summary of 510(k) for the subjective device equivalence information is being submitted in accordance with the requirements of 21 C.F.R. 807.92.
- 1. Date Summary Prepared: September 24, 2020
- 2. Submitter Information:
| Name | Shandong Huge Dental Material Corporation |
|----------------|----------------------------------------------------------------------------------------------|
| Address | No. 68 Shanhai Road, Donggang District, Rizhao City<br>Shandong Province, 276800, P.R. China |
| Telephone | 086-633-2277285 |
| Fax | 086-633-2277298 |
| Contact Person | Mrs. Maggie Zheng |
| Contact Title | Regulatory Affairs Manager |
| E-mail | zhengxy@hugedent.com |
#### 3. Device Name
Trade name: A-Silicone for Bite Registration
Common name: Bite Registration
Classification name: Impression material (21 CFR 872.3660)
Regulatory Class: II
Product Code: ELW
#### 4. Predicate Device Information
| Table 1: Predicate Device Information | | | | |
|---------------------------------------|-------------------|-------------|--------------|-----------|
| Owner/Operator | Device Trade Name | 510 (k) No. | Product Code | Predicate |
| Zhermack S. P. A. | Occlufast Rock | K024034 | ELW | Primary |
This predicate have not been subject to a design-related recall.
No reference devices were used in this submission.
#### 5. Description of Device
{4}------------------------------------------------
# HUGE
A-Silicone for Bite Registration is a vinylpolysiloxane (addition silicone), also known as A-Silicones, polyvinyl siloxanes (PVS), or vinyl polysiloxanes (VPS), especially made for bite registration work in the surgery and laboratory. It mainly contains polydimethylsiloxane, silica and platinum catalyst.
## 6. Indications for Use
A-Silicone for Bite Registration is indicated for making occlusal registrations.
#### 7. Technological Characteristics
All components of the A-Silicone for Bite Registration are based upon industry well-known chemistry. A-Silicone for Bite Registration is vinylpolysiloxane based materials which has two components addition-type (platinum-catalyzed) silicone rubber. If two parts were mixed together, it will transform into elastomeric materials by the platinum-catalyzed addition curing reaction. The following table shows the technological characteristics for the subject device and indicates the following similarities and differences with the predicate device:
| Table 4: Technological Characteristics Comparison Table | | | | | | | | | | | | | | | | | | |
|---------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|--|
| Technological<br>Characteristics | Subject device<br>(A-Silicone for Bite Registration) | Primary predicate device<br>(Occlufast Rock, K024034) | | | | | | | | | | | | | | | | |
| Description<br>of Material | Vinylpolysiloxane based | Vinylpolysiloxane based | | | | | | | | | | | | | | | | |
| Mode of Action | Addition-curing bite registration | Addition-curing bite registration | | | | | | | | | | | | | | | | |
| Indications of Use | A-Silicone for Bite Registration is<br>indicated for making occlusal<br>registrations. | Zhermack OCCLUFAST ROCK a dental<br>impression material intended to be<br>placed on a preformed impression tray and<br>used to reproduce the structure of<br>a patient's teeth and gums. | | | | | | | | | | | | | | | | |
| Prescription/over-<br>the-counter use | Prescription | Prescription | | | | | | | | | | | | | | | | |
| Physical Form | Double cartridge system 1:1 | Double cartridge system 1:1 | | | | | | | | | | | | | | | | |
| Accessories | Mixing tips, Dispenser | Mixing tips, Dispenser | | | | | | | | | | | | | | | | |
| Physical Properties | HUGE ZHERMACK Remarks Trade name A Silicone For<br>Bite Registration Occlufast Rock / Consistency<br>(≤35mm) 20mm 23mm Similar Linear dimensional<br>change<br>(<1.5%) - 0.2% - 0.2% Same | | | | | | | | | | | | | | | | | |
{5}------------------------------------------------
| Table 4: Technological Characteristics Comparison Table | | |
|---------------------------------------------------------|------------------------------------------------------|-------------------------------------------------------|
| Technological<br>Characteristics | Subject device<br>(A-Silicone for Bite Registration) | Primary predicate device<br>(Occlufast Rock, K024034) |
| Detail reproduction<br>(75µm) | 75µm | Same |
| Elastic recovery<br>( $≥$ 96.5%) | 99% | Same |
The subject device and primary predicate device have minor different Indications for Use language. However, the difference in language does not change the intended use or substantial equivalence, both products are intended for bite registration work in the surgery and laboratory. Besides, other comparison items such as description of material, mode of action, physical form, accessories and physical properties, etc. are the same. And both products are supplied for prescription use.
#### 8. Summary of Non-clinical performance testing
#### Physical properties testing
A-Silicone for Bite Registration has similar physical and chemical properties as the predicate devices. A-Silicone for Bite Registration was tested and met the applicable requirements of the FDA Recognized Consensus standard: ISO 4823 Fourth edition 2015-08-01 Dentistry -Elastometric impression materials.
| Summary of Physical Properties Testing | | | |
|----------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------|-----------------|
| Items | Requirements per ISO 4823 | Requirements of our company<br>and Pre-test planning approaches<br>and test methods of our company | Test Conclusion |
| Component colours | Different components intended for<br>use in the same mixture shall be<br>supplied in contrasting colours to<br>provide a means of determining<br>when the components have been<br>thoroughly mixed. | Same as ISO 4823 | Satisfactory |
| Consistency | $\leq$ 35mm | Same as ISO 4823 | Satisfactory |
| Linear dimensional<br>change | $\leq$ 1.5% | Same as ISO 4823 | Satisfactory |
| Detail reproduction | 75 $\mu$ m | Same as ISO 4823 | Satisfactory |
| Elastic recovery | $\geq$ 96.5% | Same as ISO 4823 | Satisfactory |
{6}------------------------------------------------
## Biocompatibility testing
The subject device. A-Silicone for Bite Registration, is substantially equivalent to the predicate device that have been legally marketed for decades and with no clinical adverse events. The formulation of the subject device does not contain any non-conventional chemicals compared to the legally marketed predicate device.
Biocompatibility tests were performed fully following the ISO 10993 standards. The test items include Cytotoxicity, Sensitization and Irritation.
#### 9. Clinical Performance Data
Not applicable. Clinical performance testing has not been performed for the subject device.
#### 10. Conclusions
Based on the indications for use, technological characteristics, performance testing and other information provided in this premarket notification, the subject device has been shown to be safe and effective for its intended use and the minor differences in indications for use fall within the intended use of the predicate device affecting neither the general intended use nor substantial equivalence. Shandong Huge Dental Material Corporation concludes that the subject device is substantially equivalent to the predicate device described herein.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.