← Product Code [OVE](/productcode/OVE) · K201646

# Shoreline ACS (Anterior Cervical System) (K201646)

_SeaSpine Orthopedics Corporation · OVE · Sep 14, 2020 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K201646

## Device Facts

- **Applicant:** SeaSpine Orthopedics Corporation
- **Product Code:** [OVE](/productcode/OVE.md)
- **Decision Date:** Sep 14, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Shoreline ACS (Anterior Cervical System) are interbody fusion devices intended for anterior cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. The Shoreline ACS implants are to be used with autograft bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/ or corticocancellous bone and implanted via an anterior approach. The device is to be used in patients who have had at least six (6) weeks of non-operative treatment. When used as a standalone system, Shoreline ACS is intended to be used as an adjunct to spinal fusion procedures at one level (C2-T1) and must be used with the Shoreline ACS bone screw fixation and locking cover. When used with supplemental fixation, such as anterior cervical plates, the Shoreline Cervical low profile (TruProfile) Interbody Spacer is intended to be used as an adjunct to spinal fusion procedures at one or two levels of the cervical spine (C2-T1).

## Device Story

Shoreline ACS is an intervertebral body fusion system for anterior cervical procedures. Implant assembly includes PEEK spacer (with or without NanoMetalene titanium coating), titanium alloy bone screws, plate, and locking cover. Device acts as disc spacer to hold bone graft and provide stability via direct purchase into vertebral endplates. Used in OR by surgeons; implanted via anterior approach. System includes instruments for placement, adjustment, and locking. Benefits include spinal fusion support in DDD patients. No software or AI components.

## Clinical Evidence

No clinical data; determination of substantial equivalence based on bench testing and engineering analysis.

## Technological Characteristics

Materials: PEEK (ASTM F2026), titanium alloy (ASTM F136), commercially pure titanium (ASTM F67). Components: box-shaped spacers with surface teeth, central canal, bone screws, plates, locking covers. Sterilization: Gamma (spacers); non-sterile (screws/plates/covers) for facility steam sterilization. No software/electronics.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Shoreline ACS (Anterior Cervical Standalone) System ([K190655](/device/K190655.md))
- Shoreline Cervical Interbody RT System ([K183083](/device/K183083.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, and then the word "ADMINISTRATION" in a smaller font size, also in blue.

September 14, 2020

SeaSpine® Orthopedics Corporation Ms. Jesse Albright Regulatory Affairs Specialist 5770 Armada Drive Carlsbad, California 92008

Re: K201646

Trade/Device Name: Shoreline ACS (Anterior Cervical System) Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral Body Fusion Device Regulatory Class: Class II Product Code: OVE, ODP Dated: June 16, 2020 Received: June 17, 2020

Dear Ms. Albright:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Brent Showalter, Ph.D. Assistant Director DHT6B: Division of Spinal Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K201646

Device Name Shoreline ACS (Anterior Cervical System)

#### Indications for Use (Describe)

The Shoreline ACS (Anterior Cervical System) are interbody fusion devices intended for anterior cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. The Shoreline ACS implants are to be used with autograft bone graft and/or allogeneic bone graft composed of cancellous, cortical, and/ or corticocancellous bone and implanted via an anterior approach. The device is to be used in patients who have had at least six (6) weeks of non-operative treatment.

When used as a standalone system, Shoreline ACS is intended to be used as an adjunct to spinal fusion procedures at one level (C2-T1) and must be used with the Shoreline ACS bone screw fixation and locking cover.

When used with supplemental fixation, such as anterior cervical plates, the Shoreline Cervical low profile (TruProfile) Interbody Spacer is intended to be used as an adjunct to spinal fusion procedures at one or two levels of the cervical spine (C2-T1).

Type of Use (Select one or both, as applicable)

| <span></span> | Prescription Use (Part 21 CFR 801 Subpart D) |
|---------------|----------------------------------------------|
| <span></span> | Over-The-Counter Use (21 CFR 801 Subpart C)  |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## K201646 - 510(k) Summary

## Contact Details

| Applicant Name:      | SeaSpine® Orthopedics Corporation                                          |  |
|----------------------|----------------------------------------------------------------------------|--|
| Address:             | 5770 Armada Drive, Carlsbad CA                                             |  |
| Phone number:        | (760) 216-5176                                                             |  |
| Fax number:          | (760) 683-6874                                                             |  |
| Primary Contact:     | Alicia McArthur, Regulatory Affairs Specialist                             |  |
| Secondary Contact:   | Jesse Albright, Regulatory Affairs Specialist                              |  |
| Date Prepared:       | August 7, 2020                                                             |  |
|                      |                                                                            |  |
| Device Name          |                                                                            |  |
| Trade Name:          | Shoreline ACS (Anterior Cervical System)                                   |  |
| Common Name:         | Intervertebral Body Fusion Device                                          |  |
| Classification Name: | Intervertebral fusion device with bone graft, cervica<br>(21 CFR 888.3080) |  |
| Class:               | II                                                                         |  |

Product Code: OVE, ODP

## Legally Marketed Predicate Devices

| 510(k) Number               | Product Code | Trade Name                                                | Manufacturer                        |
|-----------------------------|--------------|-----------------------------------------------------------|-------------------------------------|
| PRIMARY PREDICATE Device    |              |                                                           |                                     |
| K190655                     | OVE          | Shoreline ACS (Anterior<br>Cervical Standalone)<br>System | SeaSpine Orthopedics<br>Corporation |
| Additional Predicate Device |              |                                                           |                                     |
| K183083                     | OVE, ODP     | Shoreline Cervical<br>Interbody RT System                 | SeaSpine Orthopedics<br>Corporation |

## Device Description

The Shoreline Anterior Cervical System (ACS) consists of the implant assembly composed of a single use PEEK cervical spacer (ASTM F2026) and a titanium alloy (ASTM F136) plate with titanium alloy variable angle or fixed bone screws, and a titanium alloy locking cover. Shoreline ACS is offered in a variety of footprints and heights to accommodate variations in patient anatomy and is generally box-shaped with surface teeth and a central canal for receiving autograft bone

{4}------------------------------------------------

graft material and/or allogeneic bone graft composed of cancellous, cortical, and/or corticocancellous bone. The system is implanted via an anterior approach.

The system offers spacers in low profile (TruProfile) and no profile versions. Both are available with a surface coating of commercially pure titanium (ASTM F67) referred to as NanoMetalene® (NM), bonded to PEEK. The spacers will be provided in gamma sterilized packaging; the bone screws, plate, and locking cover will be provided non-sterile for subsequent sterilization at the healthcare facility.

The instruments included with the Shoreline ACS System facilitate the placement, adjustment, and final locking of the interbody spacers, and removal if necessary. The instruments also include the trays and caddies for storage, protection, and organization prior to and during the steam sterilization process.

### Indications for Use

The Shoreline ACS (Anterior Cervical System) are interbody fusion devices intended for antenor cervical interbody fusion procedures in skeletally mature patients with degenerative disc disease (DDD) of the cervical spine. DDD is defined as discogenic pain with degeneration of the disc confirmed by history and radiographic studies. The Shoreline ACS implants are to be used with autograft bone graft and/or allogeneic bone graft composed of cancellous, cortical, and /or corticocancellous bone and implanted via an anterior approach. The device is to be used in patients who have had at least six (6) weeks of non-operative treatment.

When used as a standalone system. Shoreline ACS is intended to be used as an adjunct to spinal fusion procedures at one level (C2-T1) and must be used with the Shoreline ACS bone screw fixation and locking cover.

When used with supplemental fixation, such as anterior cervical plates, the Shoreline Cervical low profile (TruProfile) Interbody Spacer is intended to be used as an adjunct to spinal fusion procedures at one or two levels of the cervical spine (C2-T1).

#### Summary of Technological Characteristics

Shoreline ACS and predicate devices have operational principle; they act as a disc spacer and hold bone graft and include integrated fixation to maintain stability by direct purchase into the bony vertebral endplates. Shoreline ACS is substantially equivalent to the cited predicate device in areas including intended use/indications for use, technological characteristics (operating principle, design, materials, sterility, manufacturing, etc.) and performance (mechanical safety). The subject and predicate devices are based on the following similar technological elements:

- Implant Spacer Heights
- Spacer Footprints ●
- Spacer Lordotic Angles
- Screw Sizes and Lengths
- Anterior Plates

K201646 510(k) Summary Page 2 of 3

{5}------------------------------------------------

## Non-Clinical Testing

The subject implants are the same as the predicate devices in terms of materials, sizes, and intended use. The subject device does not introduce a new worst-case. Engineering analyses of the modifications to Shoreline ACS determined that no additional mechanical testing was necessary.

## Clinical Testing

Not applicable; determination of substantial equivalence is not based on an assessment of clinical performance data.

## Conclusions

The submitted data demonstrates that the subject Shoreline Anterior Cervical System (ACS) has been shown to be substantially equivalent to legally marketed predicate devices for its intended use.

---

**Source:** [https://fda-staging.innolitics.com/device/K201646](https://fda-staging.innolitics.com/device/K201646)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).