Disposable Circular Staple, Disposable Hemorrhoidal Stapler, Disposable Linear Stapler and Reloads, Disposable Linear Cutter Stapler and Reloads, Disposable Endoscopic Linear Cutter Stapler and Reloads
K201639 · Beijing Biosis Healing Biolog Ical Technology Co., Ltd. · GDW · May 27, 2021 · General, Plastic Surgery
Device Facts
Record ID
K201639
Device Name
Disposable Circular Staple, Disposable Hemorrhoidal Stapler, Disposable Linear Stapler and Reloads, Disposable Linear Cutter Stapler and Reloads, Disposable Endoscopic Linear Cutter Stapler and Reloads
The Disposable Circular Stapler has application throughout the alimentary tract to create end-to-end, side-to-side anastomoses in both open and laparoscopic surgeries. The Disposable Hemorrhoidal Circular Stapler has application throughout the anal to perform surgical treatment of hemorrhoid disease. The Disposable Linear Stapler and Reloads can be applied in abdominal, thoracic and pediatric surgical procedures for transection or resection of tissue. The Disposable Linear Cutter Stapler and Reloads can be applied in abdominal, gynecological, pediatric and thoracic surgical procedures for resection, transection and creation of anastomosis. The Disposable Endoscopic Linear Cutter Stapler and Reloads has applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transaction and creation of anastomosis.
Device Story
Manual surgical staplers; used in open and laparoscopic procedures; alimentary tract, abdominal, thoracic, gynecological, and pediatric applications. Devices place titanium staples (double or triple staggered rows) to compress and join tissue; integrated circular or linear knives simultaneously resect excess tissue or divide tissue between staple rows. Operated by surgeons; manual activation via handles/firing mechanisms. Safety features include insurance/locking mechanisms to prevent misfiring. Benefits include efficient tissue closure, resection, and anastomosis creation. Devices are single-use, sterile, and disposable.
Clinical Evidence
No clinical data. Bench testing only. Testing included ISO 10993 biocompatibility (cytotoxicity, irritation, sensitization, systemic toxicity), ASTM F88/F1929 packaging/seal integrity, and ex-vivo porcine tissue performance testing (pressure, staple formation, force-to-fire).
Technological Characteristics
Manual surgical staplers; unalloyed titanium staples; stainless steel and various polymers (ABS, PC, PPA, POM, PA-757, PE, AL6063). Sterilization: Ethylene Oxide. Endotoxin limit: 20EU. Configurations: Circular, Linear, Linear Cutter, and Endoscopic Linear Cutter. Staple heights: 2.5mm to 4.8mm. Connectivity: None (mechanical).
Indications for Use
Indicated for patients undergoing abdominal, thoracic, gynecological, or pediatric surgical procedures requiring tissue transection, resection, or anastomosis, and for the surgical treatment of hemorrhoid disease.
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
Panther Endo Linear Cutter Staplers with Single Use Loading Unit (K142577)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
May 27, 2021
Beijing Biosis Healing Biological Technology Co., Ltd. % Diana Hong General Manager Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120 China
Re: K201639
Trade/Device Name: Disposable Circular Stapler, Disposable Hemorrhoidal Stapler, Disposable Linear Stapler and Reloads, Disposable Linear Cutter Stapler and Reloads, Disposable Endoscopic Linear Cutter Stapler and Reloads
Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable Staple Regulatory Class: Class II Product Code: GDW, GAG Dated: June 2, 2020 Received: June 16,2020
Dear Diana Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies.combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation -emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
for Cindy Chowdhury, Ph.D., MBA Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K210639
### Device Name
Disposable Circular Stapler; Disposable Hemorrhoidal Stapler and Reloads; Disposable Linear Cutter Stapler and Reloads; Disposable Endoscopic Linear Cutter Stapler and Reloads
### Indications for Use (Describe)
The Disposable Circular Stapler has application throughout the alimentary tract to create end-to-end, sideto-side anastomoses in both open and laparoscopic surgeries.
The Disposable Hemorrhoidal Circular Stapler has application throughout the anal to perform surgical treatment of hemorrhoid disease.
The Disposable Linear Stapler and Reloads can be applied in abdominal, thoracic and pediatric surgical procedures for transection or resection of tissue.
The Disposable Linear Cutter Stapler and Reloads can be applied in abdominal, gynecological, pediatric and thoracic surgical procedures for resection, transection and creation of anastomosis.
The Disposable Endoscopic Linear Cutter Stapler and Reloads has applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transaction and creation of anastomosis.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> |
| <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
This 510(k) Summary is being submitted in accordance with requirements of Title 21, CFR Section 807.92.
The assigned 510(k) Number: K201639
- 1. Date of Preparation: 05/27/2021
- 2. Sponsor Identification
Beijing Biosis Healing Biological Technology Co., Ltd. No.6 Plant West, Valley No.1 Bio-medicine Industry Park, Daxing District, 102600, Beijing, China
Establishment Registration Number: 3016668451
Contact Person: Jiali Zhou Position: RA Tel: +86-10-61252660 Fax: +86-10-61252030 Email: zhoujiali@biosishealing.com
- 3. Designated Submission Correspondent
Ms. Diana Hong (Primary Contact Person) Ms. Huifan Wang (Alternative Contact Person)
Mid-Link Consulting Co., Ltd P.O. Box 120-119, Shanghai, 200120, China
Tel: +86-21-22815850 Fax: 360-925-3199 Email: info@mid-link.net
{4}------------------------------------------------
#### Identification of Proposed Device 4.
### Trade Name:
Disposable Circular Stapler Disposable Hemorrhoidal Stapler Disposable Linear Stapler and Reloads Disposable Linear Cutter Stapler and Reloads Disposable Endoscopic Linear Cutter Stapler and Reloads Common Name: Stapler, Implantable
Regulatory Information Classification Name: Implantable Stapler; Classification: II; Product Code: GDW; Regulation Number: 21CFR 878.4750 Review Panel: General & Plastic Surgery;
### Indications for Use:
The Disposable Circular Stapler has application throughout the alimentary tract to create end-to-end, end-to-side and side-to-side anastomoses in both open and laparoscopic surgeries.
The Disposable Hemorrhoidal Circular Stapler has application throughout the anal canal to perform surgical treatment of hemorrhoid disease.
The Disposable Linear Stapler and Reloads can be applied in abdominal, thoracic and pediatric surgical procedures for transection or resection of tissue.
The Disposable Linear Cutter Stapler and Reloads can be applied in abdominal, gynecological, pediatric and thoracic surgical procedures for resection, transection and creation of anastomosis.
The Disposable Endoscopic Linear Cutter Stapler and Reloads has applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transaction and creation of anastomosis.
## Device Description:
Disposable Circular Stapler places a double staggered, circular row of titanium staples upon activation, which was achieved by squeezing the handles firmly as far as they could go. Immediately after formation of the staples, the excess tissue will be resect by the circular knife, and then a circular anastomosis is created. The stapler are available in 25mm, 28mm, 29mm and 32mm five
{5}------------------------------------------------
specifications. The staple is available in 4.5mm and 4.8mm two different heights.
Disposable Endoscopic Linear Cutter Stapler and Unit places two, triple-staggered rows of titanium staples and simultaneously divides the tissue from central line. The devices are available in 30mm, 45mm and 60mm three specifications. Reloads are available in 2.5mm, 4.0mm and 4.8mm four staple sizes to accommodate various tissue thicknesses. The device may be reloaded and fired up to 25 times in a single procedure.
Disposable Hemorrhoidal Stapler is a set of instruments that place a double staggered, circular row of titanium staples. Immediately after the formation of staples, the circular knife blade resects the excess of compressed mucosa. The stapler are available in 32mm, 33mm three specifications. The staple size is 3.8mm. It cannot be reloaded.
Disposable Linear Cutter Staplers and Reload places two double staggered rows of titanium staples and simultaneously cut and divides tissue between the two double rows. The devices are available in 55mm, 60mm, 75mm, 80mm and 100mm five specifications. Reloads are available in two staple sizes to accommodate various tissue thicknesses: 3.8mm and 4.8mm. It may be reloaded and fired up to 8 times in a single procedure.
The Disposable Linear Stapler and Reload places a double staggered row of titanium staples and is available in 30mm, 45mm, 60mm four specifications for use in various applications. Two staple sizes (3.5mm and 4.8mm) are available to accommodate various tissue thicknesses. It may be reloaded and fired up to 8 times in a single procedure.
#### ર . Identification of Predicate Devices
Predicate Device 1 510(k) Number: K103470 Product Name: PANTHER Group of Surgical Staplers
Predicate Device 2 510(k) Number: K142577 Product Name: Panther Endo Linear Cutter Staplers with Single Use Loading Unit
#### 6. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
{6}------------------------------------------------
- > ISO 10993-1:2009 Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process.
- > ISO 10993-5:2009 Biological evaluation of medical devices-Part 5: Tests for in vitro cytotoxicity.
- ISO 10993-10:2010 Biological evaluation of medical devices- Part 10: Test for irritation and > delayed-type hypersensitivity
- USP42-NF37 <85> Bacterial Endotoxins Tests >
- > ASTM F 88/F88M-15 Standard test method for seal strength of flexible barrier materials;
- > ASTM F1929-15 Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration
- ▲ ISO 10993-11: 2017 Biological evaluation of medical devices - Part 11: Tests for systemic toxicity
Ex-vivo tissue test was conducted on porcine stomach and intestine tissue for both proposed device and predicate device to evaluate the device performance. The test items include Pressure Rest, Closed Staple Dimension Test, Staple Formation Test and Force Required to Fire Stapler Test. Besides tissue test, jugular vein test was conducted on a porcine model to evaluate the device performance in thin tissues. This test was conducted on both proposed device and predicate device for 2.5mm staple height. Burst pressure, closed staple height and staple formation were evaluated in jugular vein test.
- 7. Clinical Test Conclusion
No clinical study is included in this submission.
{7}------------------------------------------------
#### 8. Substantially Equivalent (SE) Comparison
| Item | Proposed Device | Predicate Device 1 |
|----------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | K103470 |
| Product Code | GDW | GDW |
| Regulation No. | 878.4750 | 878.4750 |
| Indication for Use | The Disposable Circular Stapler has<br>application throughout the alimentary<br>tract to create end-to-end, end-to-side<br>and side-to-side anastomoses in both<br>open and laparoscopic surgeries | The Panther circular staplers and<br>accessories have application<br>throughout the alimentary tract to<br>create end-to-end, end-to-side and<br>side-to-side anastomoses in both<br>open and laparoscopic surgeries |
| | Anvil | Anvil |
| | Cartridge | Cartridge |
| | Drivepipe | Drivepipe |
| | Firing Handle | Firing Handle |
| Main Configuration | Insurance | Insurance |
| | Rating Nut | Rating Nut |
| | Staple | Staple |
| | Circular Knife | Circular Knife |
| | Cartridge Cover | Cartridge Cover |
| Operate Principle | Manual | Manual |
| Cutting Mechanism | Circular Knife | Circular Knife |
| Safety Mechanism | Insurance is used for preventing from<br>mis-firing | Insurance is used for preventing<br>from mis-firing |
| Specification | 25, 26, 28, 29 and 32mm | 24, 26, 29, 32 and 34mm |
| Staple Height | 4.5 and 4.8mm | 4.5 and 4.8mm |
| Row number of Staple | 2 | 2 |
| Closed Staple form | B-shape | B-shape |
| Staple material | Unalloyed Titanium | Unalloyed Titanium |
| Patient-contact<br>material of stapler | Stainless Steel<br>ABS<br>AL6063 | Unknown |
| Sterilization | Ethylene Oxide | Ethylene Oxide |
| Endotoxin Limit | 20EU | 20EU |
| Labeling | Conforms with 21 CFR 801 | Conforms with 21 CFR 801 |
| Item | Proposed Device | Predicate Device 1<br>K103470 |
| Product Code | GDW | GDW |
| Regulation No. | 878.4750 | 878.4750 |
| Indication for Use | The Disposable Hemorrhoidal Circular Stapler has application throughout the anal canal to perform surgical treatment of hemorrhoid disease | The PANTHER Hemorrhoidal Circular Stapler and accessories have application throughout the anal canal to perform surgical Treatment of hemorrhoidal disease. |
| Main Configuration | Anvil<br>Cartridge Cover<br>Cartridge<br>Fixed handle<br>Firing Handle<br>Rating Nut<br>Insurance<br>Staple<br>Circular Knife | Anvil<br>Cartridge Cover<br>Cartridge<br>Fixed handle<br>Firing Handle<br>Rating Nut<br>Insurance<br>Staple<br>Circular Knife |
| Operate Principle | Manual | Manual |
| Cutting Mechanism | Circular Knife | Circular Knife |
| Safety Mechanism | Insurance is used for preventing from mis-firing | Insurance is used for preventing from mis-firing |
| Specification | 32, 33 and 34mm | 31, 32, 33 and 34mm |
| Staple Height | 3.8mm | 3.8mm |
| Row number of Staple | 2 | 2 |
| Closed Staple form | B-shape | B-shape |
| Staple material | Unalloyed Titanium | Unalloyed Titanium |
| Patient-contact material of stapler | Stainless Steel<br>PE<br>ABS<br>PC | Unknown |
| Sterilization | Ethylene Oxide | Ethylene Oxide |
| Endotoxin Limit | 20EU | 20EU |
| Labeling | Conforms with 21 CFR 801 | Conforms with 21 CFR 801 |
| Sterilization | Ethylene Oxide | Ethylene Oxide |
# Table 1 Comparison of Disposable Circular Stapler
{8}------------------------------------------------
| | | | Table 2 Comparison of Disposable Hemorrhoidal Stapler | |
|--|--|--|-------------------------------------------------------|--|
|--|--|--|-------------------------------------------------------|--|
{9}------------------------------------------------
| Item | Proposed Device | Predicate Device 1<br>K103470 |
|----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code | GDW | GDW |
| Regulation No. | 878.4750 | 878.4750 |
| Indication for Use | The Disposable Linear Stapler and<br>Reloads can be applied in<br>abdominal, thoracic and pediatric<br>surgical procedures for transection<br>or resection of tissue. | The PANTHER Linear Stapler<br>can be applied in abdominal,<br>thoracic and pediatric surgical<br>procedures for transection or<br>resection of tissue. |
| Main Configuration | Anvil<br>Reloading Unit<br>Left and right slider<br>Rest knob<br>Fixed handle<br>Firing handle | Anvil<br>Reloading Unit<br>Left and right slider<br>Rest knob<br>Fixed handle<br>Firing handle |
| Operate Principle | Manual | Manual |
| Suture length | 30 mm, 45 mm, 60 mm and 90 mm | 30mm, 45mm, 60mm, 75mm and<br>90mm |
| Staple Height | 3.5mm and 4.8mm | 2.5mm, 3.5mm and 4.8mm |
| Row number of<br>Staple | Double staggered rows | Double staggered rows |
| Closed Staple form | B-shape | B-shape |
| Staple material | Unalloyed Titanium | Unalloyed Titanium |
| Patient-contact<br>material of stapler | PC<br>Stainless Steel<br>PA-757 | Unknown |
| Sterilization | Ethylene Oxide | Ethylene Oxide |
| Endotoxin Limit | 20EU | 20EU |
| Labeling | Conforms with 21 CFR 801 | Conforms with 21 CFR 801 |
| Item | Proposed Device | Predicate Device 1 |
| | | K103470 |
| Product Code | GDW | GDW |
| Regulation No. | 878.4750 | 878.4750 |
| Indication for Use | The Disposable Linear Cutter Stapler<br>and Reloads can be applied in<br>abdominal, gynecological, pediatric<br>and thoracic surgical procedures<br>for resection, transection and creation<br>of anastomosis | The PANTHER Linear Cutter<br>Stapler can be applied in<br>abdominal, gynecological,<br>pediatric and thoracic surgical<br>procedures for resection,<br>transection and creation of<br>anastomosis. |
| Main Configuration | Push Button<br>Rotating pin<br>Cutting knife<br>Cartridge cover<br>Cartridge | Push Button<br>Rotating pin<br>Cutting knife<br>Cartridge cover<br>Cartridge |
| Operate Principle | Manual | Manual |
| Cutting Mechanism | Linear knife | Linear knife |
| Suture length | 55mm, 60 mm, 75mm, 80 mm and<br>100 mm | 55mm, 60mm, 75mm, 80mm,<br>100mm and 110mm |
| Staple Height | 3.8mm and 4.8mm | 3.8mm and 4.8mm |
| Row<br>number<br>of<br>Staple | two double staggered mows | two double staggered mows |
| Closed Staple form | B-shape | B-shape |
| Staple material | Titanium | Titanium |
| Patient-contact<br>material of stapler | Stainless Steel<br>PC<br>ABS<br>POM<br>PPA | Unknown |
| Sterilization | Ethylene Oxide | Ethylene Oxide |
| Endotoxin Limit | 20EU | 20EU |
| Labeling | Conforms with 21 CFR 801 | Conforms with 21 CFR 801 |
| Item | Proposed Device | Predicate Device 2 |
| Product Code | GDW | GDW |
| Regulation No. | 878.4750 | 878.4750 |
| Indication for Use | Disposable Endoscopic Linear Cutter Stapler and Reloads has applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transaction and creation of anastomosis. | PANTHER Endo Linear Cutter Stapler with Single Use Loading Units has applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transaction and creation of anastomosis. They may be used for transection and resection of liver substance, hepatic vasculature and biliary structures. |
| Main Configuration | Firing rod<br>Barrel shell<br>Rotary knob<br>Insurance button<br>Left and right shell<br>Cartridge | Firing rod<br>Barrel shell<br>Rotary knob<br>Insurance button<br>Left and right shell<br>Cartridge |
| Operate Principle | Manual | Manual |
| Cutting Mechanism | Linear Knife | Linear Knife |
| Safety Mechanism | Insurance button for preventing from miss-firing | Insurance button for preventing from miss-firing |
| Suture length | 30mm, 45mm and 60mm | 30mm, 45mm and 60mm |
| Staple Height | 2.5, 3.5, 4.0 and 4.8mm | 2.0, 2.5, 3.5, 4.0 and 4.8mm |
| Row number of Staple | Two, triple-staggered rows | Two, triple-staggered rows |
| Closed Staple form | B-shape | B-shape |
| Staple material | Unalloyed Titanium | Unalloyed Titanium |
| Patient-contact material of stapler | PPA<br>Stainless Steel | Stainless steel<br>Polymeric materials<br>Adhesives<br>lubricants |
| Sterilization | Ethylene Oxide | Ethylene Oxide |
| Endotoxin Limit | 20EU | 20EU |
| Labeling | Conforms with 21 CFR 801 | Conforms with 21 CFR 801 |
Table 3 Comparison of Disposable Linear Stapler and Reloads
{10}------------------------------------------------
Table 4 Comparison of Disposable Linear Cutter Stapler and Reloads
{11}------------------------------------------------
{12}------------------------------------------------
- 9. Substantially Equivalent (SE) Conclusion
Based on the comparison and analysis above, the proposed devices are determined to be Substantially Equivalent (SE) to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.