The MouthLab Vital Signs Monitoring System is a hand-held, wireless device intended to record, transfer, store and display single lead electrocardiography (ECG), heart rate variability (HRV), functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), respiration rate (RR), respiration pattern (RP), oral temperature (TEMP) and basic lung function measurements: peak expiratory flow (PEF) and forced expiratory volume in one second (FEV ). This system is for spot checking and does not have continuous monitoring capability or any alarm features. The device comes in contact with the user for approximately 60 seconds at each use and the captured medical parameters shall be displayed on a mobile application. It is intended for use by adults in the home environment and in healthcare facilities. This system makes no specific diagnosis. Respiration rate is not intended for adults with underlying or suspected medical conditions. The device is for single user use. It is intended for use with users who are well perfused and during no motion condition. Users with implanted pacemakers and/or implanted cardioverter-defibrillators (ICDs) are not recommended to use the device.
Device Story
Hand-held, wireless, battery-powered device for non-invasive, intermittent spot-check of vital signs. Inputs: ECG (dry electrodes), SpO2 (integrated sensor), temperature (thermistor), and respiration/lung function (acoustic air turbulence via microphones). Device contacts user for ~60 seconds; data transmitted via cellular network to Aidar Cloud for processing/storage; results displayed on Aidar Care mobile application. Used by adults in home or healthcare facilities. Provides measurements for clinician review; no diagnostic interpretation or alarms. Benefits include convenient, multi-parameter physiological data collection for patient health tracking.
Clinical Evidence
Clinical agreement study for SpO2 conducted per ISO 80601-2-61:2017 in controlled desaturation study vs. co-oximeter arterial blood gas. Respiration rate validated in 20 healthy adults vs. clinician-scored capnography (EtCO2). Bench testing validated ECG/HR (IEC 60601-2-27), pulse rate (ISO 80601-2-61), temperature (ISO 80601-2-56), and lung function (ATS 2005).
Technological Characteristics
Handheld, battery-powered (2000mAh Li-polymer). Sensors: dry conductive ECG electrodes, SpO2 sensor, thermistor, microphones for acoustic respiration/lung function. Connectivity: cellular to cloud. Software: Moderate Level of Concern. Biocompatibility: ISO 10993-1. Standards: IEC 60601-1, 60601-1-2, 60601-1-6, 60601-1-11.
Indications for Use
Indicated for adult patients in home or healthcare settings for spot-check monitoring of ECG, HRV, SpO2, PR, RR, RP, oral temperature, PEF, and FEV1. Contraindicated for users with implanted pacemakers or ICDs. Not intended for patients with underlying or suspected medical conditions regarding respiration rate. Requires well-perfused, no-motion conditions.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. Underneath the square are the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 6, 2021
MultiSensor Diagnostics (dba Aidar Health) % Allison Komiyama Principal Consultant AcKnowledge Regulatory Strategies, LLC 2251 San Diego Ave, Suite B-257 San Diego, California 92110
Re: K201635
Trade/Device Name: MouthLab Vital Signs Monitoring System Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (Including Cardiotachometer And Rate Alarm) Regulatory Class: Class II Product Code: MWI, DPS, FLL, BZQ, BZH, DPS Dated: December 23, 2020 Received: December 28, 2020
Dear Allison Komiyama:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Stephen Browning Assistant Director Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K201635
Device Name MouthLab Vital Signs Monitoring System
### Indications for Use (Describe)
The MouthLab Vital Signs Monitoring System is a hand-held, wireless device intended to record, transfer, store and display single lead electrocardiography (ECG), heart rate variability (HRV), functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), respiration rate (RR), respiration pattern (RP), oral temperature (TEMP) and basic lung function measurements: peak expiratory flow (PEF) and forced expiratory volume in one second (FEV ).
This system is for spot checking and does not have continuous monitoring capability or any alarm features. The device comes in contact with the user for approximately 60 seconds at each use and the captured medical parameters shall be displayed on a mobile application.
It is intended for use by adults in the home environment and in healthcare facilities. This system makes no specific diagnosis. Respiration rate is not intended for adults with underlying or suspected medical conditions. The device is for single user use.
It is intended for use with users who are well perfused and during no motion condition. Users with implanted pacemakers and/or implanted cardioverter-defibrillators (ICDs) are not recommended to use the device.
| Type of Use (Select one or both, as applicable) | | |
|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|----------------------------------------------------------------------|
| <table style="border:none"><tr><td><input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) </td></tr><tr><td><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) </td></tr></table> | <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
| <input checked="true" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D) | | |
| <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) | | |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(K) SUMMARY
| Device Common Name: | Monitor, Physiological, Patient (Without Arrhythmia Detection or Alarms) | |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------|
| Device Trade Name: | MouthLab™ Vital Signs Monitoring System | |
| Applicant | Multisensor Diagnostics, LLC. (DBA Aidar Health)<br>3402 Birch Hollow Rd,<br>Pikesville, MD 21208<br>Phone: (443) 875-6456<br>Website: www.aidar.com | |
| Contact: | Sathya Elumalai<br>Founder & Chief Executive Officer<br>selumalai@aidar.com | |
| Representative/<br>Consultant: | Allison C. Komiyama, Ph.D., R.A.C.<br>AcKnowledge Regulatory Strategies, LLC<br>Telephone: +1(619) 458-9547<br>Email: akomiyama@acknowledge-rs.com<br>Website: www.AcKnowledge-RS.com | |
| Date Prepared: | February 4, 2021 | |
| Classification Regulation: | Class II | |
| Classification Name: | 870.2300: Cardiac monitor (including cardiotachometer and rate alarm) | |
| Panel: | Cardiovascular | |
| Primary Product Code: | MWI | |
| Secondary Product Codes: | DQA, DPS, FLL, BZQ, DXN, BZH | |
| Predicate Devices: | Primary: | CheckMe Pro Health Monitor (K150869)<br>Secondary: Connex Vital Signs 6000 Monitor (K171621)<br>Wing Smart FEV1 and Peak Flow Meter (K152276) |
| Reference Devices: | • Masimo Acoustic Respiration Sensor (K120984)<br>• ReadMyHeart - Model RMHI3.0 (K050620)<br>• Peak me (K180487) | |
{4}------------------------------------------------
#### INDICATIONS FOR USE: 1.
The MouthLab Vital Signs Monitoring System is a hand-held, wireless device intended to record, transfer, store and display single lead electrocardiography (ECG), heart rate (HR), heart rate variability (HRV), functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), respiration rate (RR), respiration pattern (RP), oral temperature (TEMP) and basic lung function measurements: peak expiratory flow (PEF) and forced expiratory volume in one second (FEV1).
This system is for spot checking and does not have continuous monitoring capability or any alarm features. The device comes in contact with the user for approximately 60 seconds at each use and the captured medical parameters shall be displayed on a mobile application.
It is intended for use by adults in the home environment and in healthcare facilities. This system makes no specific diagnosis. Respiration rate is not intended for adults with underlying or suspected medical conditions. The device is for single user use.
It is intended for use with users who are well perfused and during no motion condition. Users with implanted pacemakers and/or implanted cardioverter-defibrillators (ICDs) are not recommended to use the device.
#### DEVICE DESCRIPTION: 2.
The MouthLab Vital Signs Monitoring System is a non-invasive, battery-powered, handheld, wireless, personalized single-user, vital sign measuring apparatus. The device is intended to intermittently monitor the user vitals through the use of an integrated SpO2 sensor, ECG electrodes, microphones and thermistor, which measures the users' functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate (PR), oral temperature (TEMP), ECG, heart rate (HR) and heart rate variability (HRV). The users' respiration rate (RR) measurements, respiration pattern (RP) and the basic lung functions (FEV1 and PEF) measurements are obtained are obtained through the acoustic measurement of air turbulence. The vital signs data are transferred to the Aidar Cloud™ Application via a cellular network for processing and storage. This processed data is then displayed on the Aidar Care™ Mobile Application.
#### 3. BIOCOMPATIBILITY:
The MouthLab Vital Signs Monitoring System is a surface device that includes components that have prolonged contact duration (> 24h and less than 30 days) with the
{5}------------------------------------------------
user's mucosal membrane and intact skin. Testing was performed in accordance with ISO 10993-1.
#### 4. PERFORMANCE DATA:
## 4.1. Bench Testing:
The following bench testing was provided to support the substantial equivalence of the MouthLab Vital Signs Monitoring System:
- . ECG and Heart Rate Measurements: To validate the ability of the MouthLab Vital Signs Monitoring System to capture ECG waveforms and heart rate measurements, bench testing as per IEC 60601-2-27:2012 was conducted.
- Pulse Rate Measurements: To validate the ability of the MouthLab Vital Signs Monitoring System to capture pulse rate measurements, bench agreement testing as per ISO 80601-2-61:2017 was conducted.
- Temperature Measurements: To validate the ability of the MouthLab Vital Signs Monitoring System to calculate temperature measurements, a bench agreement study was conducted, as per ISO 80601-2-56:2017.
- Peak Flow Measurements: To validate the ability of the MouthLab Vital Signs ● Monitoring System to measure the basic lung functions (FEV1 and PEF), bench testing according to the ATS Standardization of Spirometry, 2005 Update was conducted.
## 4.2. Clinical Testing:
The following clinical testing was provided to support the substantial equivalence of the MouthLab Vital Signs Monitoring System:
- . SpO2 Measurements: To validate the ability of the MouthLab Vital Signs Monitoring System to make pulse oximetry measurements, a clinical agreement study as per ISO 80601-2-61:2017 was conducted. MouthLab SpO2 measurements were validated in a controlled desaturation study with adults and compared to reference co-oximeter analysis of arterial blood gas.
Respiration Rate Measurements: To validate the ability of the MouthLab Vital Signs Monitoring System to calculate respiration rate, a clinical agreement study was conducted with 20 healthy adults and MouthLab respiration rate measurements were compared against a clinician's manually scored capnography (EtCO2) waveforms.
{6}------------------------------------------------
#### 5. SOFTWARE DOCUMENTATION:
Software documentation for a Moderate Level of Concern device was provided in support of the MouthLab Vital Signs Monitoring System as recommended by FDA's Guidance, Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, dated May 11, 2005 and IEC 62304:2006 + A1:2015.
#### 6. ELECTRICAL SAFETY TESTING:
The MouthLab Vital Signs Monitoring System was found to be compliant with the following standards:
- IEC 60601-1:2005 + A1:2012 .
- IEC 60601-1-6:2010 + A1:2013 o
- IEC 60601-1-11:2015 ●
#### 7. ELECTROMAGNETIC COMPATIBILITY TESTING:
The MouthLab Vital Signs Monitoring System was found to be compliant with IEC 60601-1-2:2014 standard.
#### HUMAN FACTORS/USABILITY TESTING: 8.
Human factors/usability testing was conducted to evaluate the ability of lay people to read and understand the MouthLab Vital Signs Monitoring System instructions for use, and subsequently simulate normal use of the MouthLab Vital Signs Monitoring System with limited training. The MouthLab Vital Signs Monitoring System was also tested as per IEC 60601-1-11:2015, for suitability of the system in a home environment.
#### SUBSTANTIAL EQUIVALENCE DISCUSSION: 9.
The MouthLab Vital Signs Monitoring System is substantially equivalent to the predicate devices based on the information summarized here.
The subject device has the same intended use and similar technological characteristics as the predicate devices cleared in K150869, K171621, and K152276. The primary predicate, CheckMe Pro Health Monitor (K150869) is also a handheld device for monitoring physiological parameters in the home or healthcare environment. The subject device has seven primary medical device functions and each of them have been compared to predicate devices. There are no differences in the technological characteristics for SpO2, Pulse Rate, Heart Rate, ECG, Temperature, Respiration Rate, Respiration Pattern, PEF and FEV1. The subject device also measures the heart rate variability (HRV) in R-R interval, which is a mathematical analysis of the variation in the
{7}------------------------------------------------
heart rate and is not intended to produce any interpretation of those measurements or be used for any kind of diagnosis.
The similar indications for use, technological characteristics, and performance characteristics for the proposed MouthLab Vital Signs Monitoring System are assessed to be substantially equivalent to the predicate devices. A comparison of the technological characteristics between the subject device and the predicate devices is provided in the tables below.
### Table 1: Technological
| | Subject Device | Primary Predicate | Substantial<br>Equivalence to<br>Predicate Device |
|-----------------------------------------------------------|-------------------------------------------------------------------|-------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| 510(k) Number | K201635 | K150869 | NA |
| Applicant | Multisensor<br>Diagnostics, LLC | Viatom Technology Co.<br>Ltd. | NA |
| Device Name | MouthLab Vital Signs<br>Monitoring System | CheckMe Pro Health<br>Monitor | NA |
| Classification<br>Regulation | 870.2300 -<br>Physiological Patient<br>Monitor | 870.2300 -<br>Physiological Patient<br>Monitor | Same |
| Device Class | Class II | Class II | Same |
| Physiological<br>Parameters<br>Monitored | ECG, HR, HRV, SpO2,<br>PR, RR, RP,<br>TEMP, PEF, FEV1 | ECG, HR, SpO2, PR,<br>TEMP | Similar |
| Location | Home and Healthcare<br>facilities | Home and Hospital | Same |
| Rx or OTC | Rx | Rx | Same |
| Power Supply | 2000mAh rechargeable<br>lithium-polymer battery | 560mAh rechargeable<br>lithium-ion polymer<br>battery | Similar |
| User<br>Population | Adults | Adults and Pediatric | Subject device only<br>supports adult<br>users |
| Alarm | No Alarms | No Alarms | Same |
| Data<br>Collection | Minimum of 10,000<br>measurements storage | 100 measurements<br>storage | Similar |
| Memory | | | |
| | Subject Device | Primary Predicate | Substantial<br>Equivalence to<br>Predicate Device |
| Operating<br>Modes | Spot-Check | Continuous Recording<br>and Spot-Check | Subject device only<br>supports spot-<br>check |
| | Comparison of ECG and Heart Rate Measurements | | |
| Measured<br>ECG<br>Parameters | ECG Waveform, Heart<br>Rate (HR), Heart Rate<br>Variability (HRV) | ECG Waveform, Heart<br>Rate (HR) | Similar |
| ECG Rhythm<br>Classification | Not Included | Not Included | Same |
| ECG Lead<br>Type | Single Lead, 2 Contacts | External<br>ECG Cable<br>and<br>Electrodes | Similar |
| Input<br>Impedance | > 2.5MΩ | > 10MΩ | Conforms to IEC<br>60601-2-27<br>impedance<br>requirements |
| Input Dynamic<br>Range | ± 2mV | ± 3mV | Subject device<br>supports this range |
| Bandwidth | 0.5 - 50 Hz | 0.05 - 40<br>Hz | Similar |
| A/D<br>Conversion | 24 bit | 16 bit | Similar |
| Sampling Rate | 1000 Hz | 500 Hz | Similar |
| Measurement<br>Time | ≤ 60 seconds | 30 seconds | Similar |
| Display | Mobile Application | 400*240 Dot-matrix<br>LCD Display | Subject device only<br>supports mobile<br>application. |
| Input | Dry Conductive<br>Electrodes | Dry Conductive<br>Electrodes and/or<br>External Auxiliary<br>Electrodes | Similar |
| Heart Rate<br>(HR) Range | 30 - 200 bpm | 30 - 250 bpm | Conforms to the<br>requirements of the<br>IEC 60601-2-27<br>heart rate<br>measurements |
| | Subject Device | Primary Predicate | Substantial<br>Equivalence to<br>Predicate Device |
| Heart Rate<br>(HR) Accuracy | $\pm$ 2 bpm or $\pm$ 2%,<br>whichever is larger | $\pm$ 2 bpm or $\pm$ 2%,<br>whichever is larger | Same |
| Comparison of Blood Oxygen Saturation (SpO2) Measurements | | | |
| Display Data<br>Mode | SpO2, Pulse Rate (PR)<br>Spot-Check | SpO2, Pulse Rate (PR)<br>Continuous Recording<br>and Spot-Check | Same<br>Subject device only<br>supports spot-<br>check |
| Sensor Types | Integrated | Integrated and External | Subject device only<br>supports integrated<br>sensors |
| SpO2 Range | 70 – 100% | 70 – 100% | Same |
| SpO2<br>Accuracy | $\pm$ 3% | Integrated: $\pm$ 3% | Same |
| SpO2<br>Resolution | 1% | 1% | Same |
| Pulse Rate<br>(PR) Range | 30 – 250 ppm | 30 – 250 ppm | Same |
| Pulse Rate<br>(PR) Accuracy | $\pm$ 2 ppm or $\pm$ 2%,<br>whichever is larger | $\pm$ 2 ppm or $\pm$ 2%,<br>whichever is larger | Same |
| Pulse Rate<br>(PR)<br>Resolution | 1 ppm | 1 ppm | Same |
### Comparison – MouthLab System vs CheckMe Pro
{8}------------------------------------------------
{9}------------------------------------------------
## Table 2: Technological Comparison – MouthLab System vs Connex Monitor
| Comparison of Respiration Rate Measurement | | | |
|--------------------------------------------|----------------------------------------------|-------------------------------|---------------------------------------------------|
| | Subject Device | Secondary<br>Predicate | Substantial<br>Equivalence to<br>Predicate Device |
| 510(k) Number | K201635 | K171621 | NA |
| Applicant | Multisensor<br>Diagnostics, LLC | Welch Allyn, Inc | NA |
| Device Name | MouthLab Vital<br>Signs Monitoring<br>System | Connex Vital Signs<br>Monitor | NA |
| Classification<br>Regulation | 870.2300 - | 870.2300 - | Same |
{10}------------------------------------------------
| | Subject Device | Secondary<br>Predicate | Substantial<br>Equivalence to<br>Predicate Device |
|---------------------------------------|--------------------------------------|-----------------------------------|------------------------------------------------------|
| | Physiological<br>Patient<br>Monitor | Physiological<br>Patient Monitor | |
| Device Class | Class II | Class II | Same |
| Location | Home and<br>Healthcare<br>Facilities | Hospital | Similar |
| Measurement<br>Techniques | Acoustic<br>Measurements | Acoustic<br>Measurements | Same |
| Display Range | 6 – 50 breaths/min | 4 – 70 breaths/min | Subject device supports<br>this range |
| Measurement<br>Range | 6 – 50 breaths/min | 4 – 70 breaths/min | Subject device supports<br>this range |
| Respiration Rate<br>(RR) Accuracy | $\pm$ 2 breaths/min | $\pm$ 1 breath/min | Subject device supports<br>this accuracy |
| Comparison of Temperature Measurement | | | |
| Measurement<br>Technique | Thermo-resistive<br>Sensor | Thermo-resistive<br>Sensor | Same |
| Measurement Site | Oral | Oral, Axillary,<br>Rectal | Subject device only<br>supports oral<br>measurements |
| Mode | Direct Mode | Direct or Predictive<br>Mode | Subject device only<br>supports direct mode |
| Unit | °F or °C | °F or °C | Same |
| Measuring Range | 93.2°F – 109.4°F<br>(34°C – 43.0°C) | 80°F – 110°F<br>(26.7°C – 43.3°C) | Conforms to the ISO<br>standard requirements |
| Accuracy | $\pm$ 0.6 °F<br>( $\pm$ 0.3°C) | $\pm$ 0.2°F<br>( $\pm$ 0.1°C) | Conforms to the ISO<br>standard requirements |
## Table 3: Technological Comparison – MouthLab System vs Wing
| Comparison of Lung Function Measurements | | | |
|------------------------------------------------|-------------------------------------------------|-------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| 510(k) Number | Subject Device | Secondary<br>Predicate | Substantial<br>Equivalence |
| K201635 | K152276 | NA | |
| | Subject Device | Secondary<br>Predicate | Substantial<br>Equivalence |
| Applicant | Multisensor<br>Diagnostics, LLC | Sparo Labs | NA |
| Device Name | MouthLab Vital<br>Signs Monitoring<br>System | Wing Smart FEV1<br>and Peak Flow<br>Meter | NA |
| Classification<br>Regulation | 870.1860 – Peak<br>Flow Meter for<br>Spirometry | 870.1860 – Peak<br>Flow Meter for<br>Spirometry | Same |
| Device Class | Class II | Class II | Same |
| Physiological<br>Parameters<br>Monitored | FEV1, PEF | FEV1, PEF | Same |
| Location | Home and Hospital | Home | Similar |
| User Interface | Mobile Application | Mobile Application | Same |
| Rx or OTC | Rx | OTC | Subject device is only<br>applicable for<br>prescription use based<br>on other parameters it<br>measures |
| Use Population | Adults | Adults and pediatric<br>users above 5<br>years of age | Subject device is for<br>adult users |
| Operating Modes | Spot-Check | Spot-Check | Same |
| Measurement<br>Technique | Acoustic<br>Measurement using<br>Microphones | Acoustic<br>Measurement using<br>Microphones | Same |
| Peak Expiratory<br>Flow (PEF) Range | 2 – 15 L/s | 50 – 900 L/min | Subject device<br>supports this range |
| Peak Expiratory<br>Flow (PEF)<br>Accuracy | ± 0.33 L/s or 10%,<br>whichever is larger | ±20 L/min or 10%,<br>whichever is larger | Same |
| PEF Resolution | 0.01 L/s | 1 L/min | Same |
| Forced Expiratory<br>Volume (FEV1)<br>Range | 1.00 – 7.50 L | 0.01 – 9.99 L | Subject device<br>supports this range |
| Forced Expiratory<br>Volume (FEV1)<br>Accuracy | ± 0.1 L or 5%,<br>whichever is larger | ± 0.1 L or 5%,<br>whichever is larger | Same |
| | Subject Device | Secondary<br>Predicate | Substantial<br>Equivalence |
| FEV1 Resolution | 0.01 L | 0.01 L | Same |
{11}------------------------------------------------
{12}------------------------------------------------
#### 10. CONCLUSION:
Based on the information provided in this 510(k) premarket notification, the MouthLab Vital Signs Monitoring System is considered to be substantially equivalent (safe and effective) to its predicate devices: CheckMe Pro Health Monitor (K150869), the Welch Allyn Monitor (K171621), and Wing Smart FEV1 and Peak Flow Meter (K152276). The MouthLab Vital Signs Monitoring System has the same intended uses, similar indications, technological characteristics and principles of operation as its predicate devices. Through the performance and clinical testing conducted as mentioned herein, the minor technological differences between MouthLab Vital Signs Monitoring System and its predicate devices raise no new issues of safety and effectiveness. Hence, the MouthLab Vital Signs Monitoring System is substantially equivalent to its predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.