K201626 · Epic Medical Pte. , Ltd. · FPA · Mar 24, 2021 · General Hospital
Device Facts
Record ID
K201626
Device Name
Safety Subcutaneous Infusion Set
Applicant
Epic Medical Pte. , Ltd.
Product Code
FPA · General Hospital
Decision Date
Mar 24, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5440
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Safety Subcutaneous Infusion Set is indicated for subcutaneous infusion of medication administered by an external infusion pump or syringe.
Device Story
Safety Subcutaneous Infusion Set facilitates subcutaneous medication delivery. Device comprises stainless-steel needle mounted to wing stabilizer at 90° angle; female luer connector at proximal end for pump/syringe attachment; open-end slide clamp on tubing for flow control during priming. User inserts needle into subcutaneous tissue; wings secure needle at site. Provided with standard medical dressing. Various needle gauges/lengths available. Device used in clinical or home settings; operated by healthcare providers or patients. Output is physical delivery of medication to subcutaneous tissue. Benefits include controlled, secure medication administration with integrated needle safety features to prevent accidental sticks.
Clinical Evidence
Bench testing only. Testing included biocompatibility (ISO 10993-1), particulate matter (USP <788>), sterilization validation, EO residuals (ISO 10993-7), shelf life (ASTM F1980-16), sharps protection (ISO 23908), tensile strength/water tightness (ISO 8536-8), and luer lock connection (ISO 80369-7).
Technological Characteristics
Stainless steel needles; ABS, LDPE, PVC, Polypropylene, Thermoplastic elastomer, and Polycarbonate components; Cyanoacrylate bonding. 90° wing stabilizer. ETO sterilization (SAL 10^-6). Compliant with ISO 23908 (sharps protection), ISO 8536-8 (tensile/tightness), ISO 80369-7 (luer connection).
Indications for Use
Indicated for subcutaneous infusion of medication via external infusion pump or syringe in patients requiring subcutaneous delivery.
Regulatory Classification
Identification
An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.
Special Controls
*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health and Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
March 24, 2021
Epic Medical Pte. Ltd. % Roshana Ahmed Sr. Consultant, Regulatory Affairs Medical Devices Quaras, LLC 25 Independence Blvd. Warren, New Jersey 07059
Re: K201626
Trade/Device Name: Safety Subcutaneous Infusion Set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular administration set Regulatory Class: Class II Product Code: FPA Dated: February 17, 2021 Received: February 19, 2021
Dear Roshana Ahmed:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201626
Device Name Safety Subcutaneous Infusion Set
Indications for Use (Describe)
The Safety Subcutaneous Infusion Set is indicated for subcutaneous infusion of medication administered by an external infusion pump or syringe.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# K201626 510(k) Summary
#### I. Submitter
Epic Medical Pte. Ltd. 105 Cecil Street #20-04 The Octagon, Singapore 069534. Phone: +65 9635 2618 / +66 81 761 5292
Contact Person: Mr. Freddie Lee, Chief Executive Officer Date Prepared: February 16, 2021
#### II. Device
| Device Proprietary Name: | Safety Subcutaneous Infusion Set |
|--------------------------|----------------------------------|
| Common or Usual Name: | Subcutaneous Administration Set |
| Classification Name: | Intravascular Administration Set |
| Regulation Number: | 21 CFR 880.5440 |
| Product Code: | FPA |
| Device Classification | II |
#### III. Predicate Device
Substantial equivalence is claimed to the following device:
- SUB-Q Subcutaneous Tissue Infusion Set, K140131, EMED Technologies Corporation ●
#### Device Description IV.
The Safety Subcutaneous Infusion Set is an intravascular administration set used to administer medication subcutaneously. The device consists of subcutaneous stainless-steel needle mounted to a wing stabilizer at 90° at the distal end of the infusion set. The subject device has closing type wings to allow the user to insert the needle into the subcutaneous tissue and also secure the needles on the infusion site. A female luer connector is provided at the proximal end to allow connection to external devices.
Each set consists of the administration set and standard medical dressing to secure the needles on the infusion site. The subject device is provided in various needle gauges and lengths. Each needle is equipped with an open end slide clamp on the tubing to facilitate stop flow when required during priming of individual needles.
{4}------------------------------------------------
#### V. Indications for Use
The Safety Subcutaneous Infusion Set is indicated for subcutaneous infusion of medication administered by an external infusion pump or syringe.
#### VI. Comparison of Technological Characteristics
The Safety Subcutaneous Infusion Set has the same intended use as the predicate device (K140131). The Safety Subcutaneous Infusion Set has the same general design construction. proximal interface, number of lumens, range of needle gauge and length, tubing length, number of needles, needle safety feature, clamp type, wing type, and sterilization method/level as the predicate device (K140131).
The subject device differs from the predicate device with respect to needle protective cap design, materials of construction and packaging materials.
The table below compares the subject and predicate devices.
| | Safety Subcutaneous Infusion | SUB-Q Subcutaneous Tissue | Comparison |
|--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| | Set | Infusion Set | |
| | (K201626) | (K140131) | |
| Indication for<br>Use | The Safety Subcutaneous Infusion<br>Set is indicated for subcutaneous<br>infusion of medication<br>administered by an external<br>infusion pump or syringe. | SUB-Q Subcutaneous Infusion<br>Set is intended to provide<br>subcutaneous infusion of<br>medicine from an external<br>infusion pump or syringe. | Same as predicate |
| Proximal<br>interface | Female Luer Lock | Female Luer Lock | Same as predicate |
| Tubing length | 40 cm | 5 cm ~ 91.4 cm | Similar to predicate |
| Needle gauge | 24, 26, and 27 | 24 and 27 | Similar to predicate |
| Needle lengths | 6mm, 9mm, 12mm | 6mm, 9mm, 12mm, 14mm and<br>16mm | Same as predicate |
| Number of<br>needles | 1 - 4 | 1 - 4 (up to 8 with Y connector) | Same as predicate |
| Needle lubricant | Polydimethylsiloxane (PDMS) | Polydimethylsiloxane (PDMS) | Same as predicate |
| Wing type | Closed, Needle protecting | Standard and Closed | Same as predicate |
| Needle<br>protective cap | Sliding | Tubular sheath | Different |
{5}------------------------------------------------
| | Safety Subcutaneous Infusion<br>Set<br>(K201626) | SUB-Q Subcutaneous Tissue<br>Infusion Set<br>(K140131) | Comparison |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------|
| Needle safety<br>feature | Needle inaccessible to user before<br>engaging.<br>Sharp protection feature<br>compliant to ISO 23908:2011. | Needle inaccessible to user before<br>engaging.<br>Sharp protection feature<br>compliant to ISO 23908:2011. | Same as predicate |
| Sterilization | ETO, SAL of 10-6 | ETO, SAL of 10-6 | Same as predicate |
| Priming volume | 24-gauge needle: 0.4 ~ 1.4<br>27-gauge needle: 0.1 ~ 0.4 | 24-gauge needle: 0.16 ~ 0.8<br>27-gauge needle: 0.12 ~ 0.8 | Similar to predicate |
| Transparent<br>dressing | Off-the-shelf dressing purchased<br>from:<br>3M Healthcare | Off-the-shelf dressing purchased<br>from:<br>3M Healthcare | Same as predicate |
| Component<br>materials | Luer lock: ABS<br>Luer cap: LDPE<br>Tubing: PVC<br>Wings: Polypropylene and<br>Thermoplastic elastomer<br>(locking)<br>Protective cap: Polycarbonate<br>Needles: Stainless steel<br>Slide clamp: Polyethylene<br>Bonding agent: Cyanoacrylate | Luer lock: PVC<br>Luer cap: Polypropylene<br>Tubing: PVC<br>Wings: PVC (std) and<br>Polypropylene (locking)<br>Protective cap: Polypropylene<br>Needles: Stainless steel<br>Slide clamp: ABS<br>Bonding agent: Cyanoacrylate | Different |
## Discussion
The Safety Subcutaneous Infusion Set has the same intended use as the predicate device and both the subject and predicate device share the same design, proximal interface, needle lengths, number of needles and needle safety feature, and wing type. The subject device tubing length and needle gauges and fall within the range of tubing length and needle gauges for the predicate device.
The subject device differs from the predicate device with respect to protective cap design (sliding cap design vs. tubular sheath design), priming volume (slight difference in priming volume), and materials of construction for a subset of components. These technological differences between the subject and predicate devices do not raise different questions of safety or effectiveness and are addressed by the performance and biological safety testing identified below.
{6}------------------------------------------------
#### VII. Performance Data
The following non-clinical data were provided in support of the substantial equivalence determination:
- biocompatibility studies per ISO 10993-1:2018
- particulate matter per USP <788>
- sterilization validation
- ethylene oxide residuals per ISO 10993-7:2008 ●
- shelf life per ASTM F1980-16
- sharps protection per ISO 23908:2011
- . tensile strength and water tightness per ISO 8536-8:2015
- luer lock connection test per ISO 80369-7
## VIII. Conclusion
The information provided above supports that the Safety Subcutaneous Infusion Set is as safe and effective as the predicate device. Although there are minor technological differences between the subject and predicate devices, these difference do not raise different questions of safety and effectiveness. Therefore, it is concluded that the Safety Subcutaneous Infusion Set is substantially equivalent to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.