← Product Code [FXX](/productcode/FXX) · K201624

# Spin Care Disposable Protective Mask (K201624)

_Spintech, LLC · FXX · Mar 18, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K201624

## Device Facts

- **Applicant:** Spintech, LLC
- **Product Code:** [FXX](/productcode/FXX.md)
- **Decision Date:** Mar 18, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4040
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery

## Indications for Use

The Spin Care Disposable Protective Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile.

## Device Story

Spin Care Disposable Protective Mask is a single-use, flat-pleated, non-sterile surgical mask. Device consists of a meltblown polypropylene filter layer sandwiched between two spunbond polypropylene layers. Features a malleable metal nose piece and elastane/nylon ear loops for secure fit. Used in clinical settings by healthcare personnel and patients to provide a physical barrier against microorganisms, body fluids, and particulates. Operates as a passive barrier; no active components. Healthcare providers wear the mask to reduce cross-contamination risk during procedures. Benefits include protection against fluid penetration and particulate exposure, supporting infection control protocols.

## Clinical Evidence

No clinical data. Bench testing only. Performance verified against ASTM F2100-19 (Level 2), ASTM F1862-17 (fluid resistance 120 mmHg), ASTM F2101-19 (BFE >99.9%), ASTM F2299-03 (PFE 99.8%), and 16 CFR 1610 (Class I flammability). Biocompatibility confirmed per ISO 10993-5 (cytotoxicity) and ISO 10993-10 (irritation and sensitization).

## Technological Characteristics

Materials: Spunbond polypropylene (outer/inner), meltblown polypropylene (filter), malleable metal wire (nose piece), elastane/nylon (ear loops). Dimensions: 175mm x 95mm. Design: Flat-pleated, non-sterile, single-use. Performance: ASTM F2100 Level 2. No energy source or software.

## Regulatory Identification

Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.

## Special Controls

*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.

## Predicate Devices

- Disposable Surgical Face Mask ([K153496](/device/K153496.md))
- Surgical Face Masks (Ear Loops and Tie-on) ([K160269](/device/K160269.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food & Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

March 18, 2021

SpinTech LLC % W. Victoria Rogers Regulatory Affairs Consultant Rogers Consulting 11110 Arranmore Cove Roanoke, Indiana 46783

Re: K201624

Trade/Device Name: Spin Care Disposable Protective Mask Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: Class II Product Code: FXX Dated: December 8, 2020 Received: December 31, 2020

Dear W. Victoria Rogers:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Ryan Ortega, PhD Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K201624

Device Name Spin Care Disposable Protective Mask

### Indications for Use (Describe)

The Spin Care Disposable Protective Mask is intended to be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intention control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile.

| Type of Use (Select one or both, as applicable)            |                                                           |
|------------------------------------------------------------|-----------------------------------------------------------|
| <span></span> Prescription Use (Part 21 CFR 801 Subpart D) | <span></span> Over-The-Counter Use (21 CFR 801 Subpart C) |

CONTINUE ON SEPARATE PAGE IF NEEDED

# |X Over-The-Counter Use (21 CFR 801 Subpart C)

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

In accordance with 21 CFR §807.92 and the Safe Medical Devices Act of 1990, the following information is provided for the Spin Care Disposable Protective Mask 510(k) premarket notification. The submission was prepared in accordance with the FDA guidance document, 'Format for Traditional and Abbreviated 510(k)s', issued on August 12, 2005.

| 510(k) Number:                       | K201624                                                                                                                                                                                                                                   |                                        |                               |                                        |         |                                            |                         |
|--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------|-------------------------------|----------------------------------------|---------|--------------------------------------------|-------------------------|
| Sponsor:                             | SpinTech LLC<br>1920 Flintstone Drive<br>Statesville, North Carolina 28677 USA<br>Establishment Registration Number: 3016872494                                                                                                           |                                        |                               |                                        |         |                                            |                         |
| Contact Person:                      | Jim Jean<br>Reliability Manager<br>Telephone: (704-929-3596)                                                                                                                                                                              |                                        |                               |                                        |         |                                            |                         |
| Designated Submission Correspondent: | W. Victoria Rogers<br>Rogers Consulting<br>11110 Arranmore Cove<br>Roanoke, Indiana 46783 USA<br>574-265-8356                                                                                                                             |                                        |                               |                                        |         |                                            |                         |
| Date:                                | March 9, 2021                                                                                                                                                                                                                             |                                        |                               |                                        |         |                                            |                         |
| Subject Device:                      | Trade Name: Spin Care Disposable Protective Mask<br>Common Name: Surgical Facemask                                                                                                                                                        |                                        |                               |                                        |         |                                            |                         |
|                                      | Classification Name:<br>• FXX- Surgical Apparel (21 CFR 878.4040)                                                                                                                                                                         |                                        |                               |                                        |         |                                            |                         |
| Predicate Device(s):                 | <table><tr><td>K153496</td><td>Disposable Surgical Face Mask</td><td>Xiantao Rayxin Medical Products., Ltd.</td></tr><tr><td>K160269</td><td>Surgical Face Masks (Ear Loops and Tie-on)</td><td>SAN-M Package Co., LTD.</td></tr></table> | K153496                                | Disposable Surgical Face Mask | Xiantao Rayxin Medical Products., Ltd. | K160269 | Surgical Face Masks (Ear Loops and Tie-on) | SAN-M Package Co., LTD. |
| K153496                              | Disposable Surgical Face Mask                                                                                                                                                                                                             | Xiantao Rayxin Medical Products., Ltd. |                               |                                        |         |                                            |                         |
| K160269                              | Surgical Face Masks (Ear Loops and Tie-on)                                                                                                                                                                                                | SAN-M Package Co., LTD.                |                               |                                        |         |                                            |                         |

{4}------------------------------------------------

Purpose and Device The Spin Care Disposable Protective Mask is a single use Description: multi-layer level 2 surgical mask. It houses a meltblown polypropylene filter between an outer and inner layer of Spunbond polypropylene that covers the nose and mouth of the end user and held in place by a pliable nose piece and ear loops. Contact duration is less than 24 hours. Intended Use and The Spin Care Disposable Protective Mask is intended to Indications for Use: be worn to protect both the patient and healthcare personnel from transfer of microorganisms, body fluids and particulate material. These face masks are intended for use in infection control practices to reduce the potential exposure to blood and body fluids. This is a single use, disposable device, provided non-sterile. Summary of Technological

Characteristics:

The rationale for substantial equivalence is based on consideration of the following characteristics:

|                                                    | Spin Care Disposable<br>Protective Mask                                                                                                                                                                                                                                                                                                                                                                                               | Disposable Surgical Face<br>Mask, K153496 (Primary<br>Predicate)                                                                                                                                                                                                                                                                                                                                                                    | Surgical Face Masks (Ear<br>Loops and Tie-on),<br>K160269                                                                                                                                                                                                                                                                                                                                                         |
|----------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| ASTM 2100 Level<br>Mask                            | 2                                                                                                                                                                                                                                                                                                                                                                                                                                     | 2                                                                                                                                                                                                                                                                                                                                                                                                                                   | 2                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Intended Use/<br>Indications for<br>Use            | The Spin Care Disposable<br>Protective Mask is intended<br>to be worn to protect both<br>the patient and healthcare<br>personnel from transfer of<br>microorganisms, body fluids<br>and particulate material.<br>These face masks are<br>intended for use in infection<br>control practices to reduce<br>the potential exposure to<br>blood and body fluids. This<br>is a single use, disposable<br>device, provided non-<br>sterile. | The Disposable Surgical<br>Face Masks are intended to<br>be worn to protect both the<br>patient and healthcare<br>personnel from transfer of<br>microorganisms, body fluids<br>and particulate material.<br>These face masks are<br>intended for use in infection<br>control practices to reduce<br>the potential exposure to<br>blood and body fluids. This<br>is a single use, disposable<br>device(s), provided non-<br>sterile. | The Surgical Face Masks<br>are intended to be worn to<br>protect both the patient and<br>healthcare personnel from<br>transfer of microorganisms,<br>body fluids and particulate<br>material. These face masks<br>are intended for use in<br>infection control practices to<br>reduce the potential<br>exposure to blood and body<br>fluids. This is a single use,<br>disposable device,<br>provided non-sterile. |
| Materials                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Outer Materials                                    | Polypropylene Spunbond                                                                                                                                                                                                                                                                                                                                                                                                                | Spun-bond polypropylene                                                                                                                                                                                                                                                                                                                                                                                                             | Polypropylene                                                                                                                                                                                                                                                                                                                                                                                                     |
| Filter Media                                       | Polypropylene meltdown                                                                                                                                                                                                                                                                                                                                                                                                                | Melt blown polypropylene<br>filter                                                                                                                                                                                                                                                                                                                                                                                                  | 1. Polypropylene<br>Spunbond<br>2. Polypropylene<br>meltdown                                                                                                                                                                                                                                                                                                                                                      |
| Inner Material                                     | Polypropylene Spunbond                                                                                                                                                                                                                                                                                                                                                                                                                | Spun-bond polypropylene                                                                                                                                                                                                                                                                                                                                                                                                             | Polypropylene                                                                                                                                                                                                                                                                                                                                                                                                     |
| Nose Piece                                         | Malleable metal wire                                                                                                                                                                                                                                                                                                                                                                                                                  | Malleable Aluminum Wire                                                                                                                                                                                                                                                                                                                                                                                                             | Polyethylene coated steel wire                                                                                                                                                                                                                                                                                                                                                                                    |
| Ear Loops                                          | Elastane and nylon                                                                                                                                                                                                                                                                                                                                                                                                                    | Ear Loop or Tie On - Polyester                                                                                                                                                                                                                                                                                                                                                                                                      | Ear Loop or Tie On – Polyester Polyurethane / Polypropylene Spunbond or polyester Spunbond                                                                                                                                                                                                                                                                                                                        |
| Specifications                                     | Length: 175mm<br>Width: 95mm                                                                                                                                                                                                                                                                                                                                                                                                          | Length: 175 mm +/- 1mm<br>Width: 95 mm +/- 1mm                                                                                                                                                                                                                                                                                                                                                                                      | Length: 90 ± 3mm<br>Width: 175 ± 5mm                                                                                                                                                                                                                                                                                                                                                                              |
| Mask Style                                         | Flat-pleated                                                                                                                                                                                                                                                                                                                                                                                                                          | Flat-pleated                                                                                                                                                                                                                                                                                                                                                                                                                        | Flat-pleated                                                                                                                                                                                                                                                                                                                                                                                                      |
| Color                                              | White                                                                                                                                                                                                                                                                                                                                                                                                                                 | Blue                                                                                                                                                                                                                                                                                                                                                                                                                                | White or Blue                                                                                                                                                                                                                                                                                                                                                                                                     |
| Sterilization                                      | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                           | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                         | Non-sterile                                                                                                                                                                                                                                                                                                                                                                                                       |
| <b>Performance Testing</b><br>ASTM F2100 – Level 2 |                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Fluid resistance<br>ASTM F1862                     | 120mm Hg - Pass                                                                                                                                                                                                                                                                                                                                                                                                                       | 120mm Hg - Pass                                                                                                                                                                                                                                                                                                                                                                                                                     | 120mm Hg - Pass                                                                                                                                                                                                                                                                                                                                                                                                   |
| Particle Filtration<br>Efficiency<br>ASM F2299     | Average 99.8% - Pass                                                                                                                                                                                                                                                                                                                                                                                                                  | Average 98.46% - Pass                                                                                                                                                                                                                                                                                                                                                                                                               | Average 99.6% - Pass                                                                                                                                                                                                                                                                                                                                                                                              |
| Bacterial Filtration<br>Efficiency<br>ASTM F2101   | Average >99.9% - Pass                                                                                                                                                                                                                                                                                                                                                                                                                 | Average 98.7% - Pass                                                                                                                                                                                                                                                                                                                                                                                                                | Average > 98% - Pass                                                                                                                                                                                                                                                                                                                                                                                              |
| Flammability Class<br>16 CFR 1610                  | Class I Non-Flammable                                                                                                                                                                                                                                                                                                                                                                                                                 | Class I Non-Flammable                                                                                                                                                                                                                                                                                                                                                                                                               | Class I Non-Flammable                                                                                                                                                                                                                                                                                                                                                                                             |
| Delta-P                                            | Average 3.79mmH2O/cm²<br>(EN 14683)                                                                                                                                                                                                                                                                                                                                                                                                   | Average 4.2 mmH2O/cm²<br>(MIL-M-36954C)                                                                                                                                                                                                                                                                                                                                                                                             | Average 1.6 mmH2O/cm²<br>(MIL-M-36954C)                                                                                                                                                                                                                                                                                                                                                                           |
| <b>Biocompatibility Testing</b>                    |                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Cytotoxicity<br>ISO 10993-5                        | Comply with ISO 10993-5.<br>Under the conditions of the<br>study, the proposed device<br>extract was determined to<br>be non-cytotoxic                                                                                                                                                                                                                                                                                                | Non-cytotoxic                                                                                                                                                                                                                                                                                                                                                                                                                       | Non-cytotoxic                                                                                                                                                                                                                                                                                                                                                                                                     |
| Irritation<br>ISO 10993-10                         | Comply with ISO 10993-10.<br>Under the conditions of the<br>study, the proposed device<br>extract was determined to<br>be non-irritating                                                                                                                                                                                                                                                                                              | Non-irritating                                                                                                                                                                                                                                                                                                                                                                                                                      | Non-irritating                                                                                                                                                                                                                                                                                                                                                                                                    |
| Sensitization<br>ISO 10993-10                      | Comply with ISO 10993-10.<br>Under the conditions of the<br>study, the proposed device<br>extract was determined to<br>be non-sensitizing                                                                                                                                                                                                                                                                                             | Non-sensitizing                                                                                                                                                                                                                                                                                                                                                                                                                     | Non-sensitizing                                                                                                                                                                                                                                                                                                                                                                                                   |

{5}------------------------------------------------

{6}------------------------------------------------

## Summary of Performance Data (Nonclinical and/or Clinical)

#### Non-Clinical Tests: .

- The product was tested in alignment with o "Guidance for Industry and FDA Staff – Surgical Masks - Premarket Notification [510(k)] Submission" Guidance Document
	- o ISO 10993-5: 2009 Biological Evaluation Of Medical Devices -- Part 5: Tests For In Vitro Cytotoxicity
	- ISO 10993-10: 2010 Biological Evaluation Of Medical Devices - Part 10: Tests For Irritation And Skin Sensitization
	- o ASTM F2100-19. Standard Specification For Performance Of Materials Used In Medical Face Masks
	- ASTM F1862-17, Standard Test Method For Resistance Of Medical Face Masks To Penetration By Synthetic Blood (Horizontal Projection Of Fixed Volume At A Known Velocity)
	- o EN 14683: European standard for face masks
	- ASTM F2101-19, Standard Test Method For Evaluating The Bacterial Filtration Efficiency (BFE) Of Medical Face Mask Materials, Using A Biological Aerosol Of Staphylococcus Aureus;
	- ASTM F2299-03, Stand test method for determining the initial efficiency of materials used in medical face masks to penetration by particulates using latex spheres:
	- o 16 CFR 1610, Standard for the Flammability of clothing textiles.
- Clinical Tests: ●
	- No clinical tests were performed. o

# Substantial Equivalence Conclusion

The conclusions drawn from the nonclinical tests demonstrate that the Spin Care Disposable Protective Mask is as safe, as effective, and performs as well as or better than the predicate devices.

---

**Source:** [https://fda-staging.innolitics.com/device/K201624](https://fda-staging.innolitics.com/device/K201624)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).