← Product Code [NDC](/productcode/NDC) · K201619

# EndoTool IV System (K201619)

_Monarch Medical Technologies, LLC · NDC · Aug 7, 2020 · Anesthesiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K201619

## Device Facts

- **Applicant:** Monarch Medical Technologies, LLC
- **Product Code:** [NDC](/productcode/NDC.md)
- **Decision Date:** Aug 7, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 868.1890
- **Device Class:** Class 2
- **Review Panel:** Anesthesiology
- **Attributes:** Software as a Medical Device, Therapeutic, Pediatric

## Indications for Use

EndoTool IV is a glucose management software system, designed for use by healthcare professionals in all patient care settings to recommend intravenous and subcutaneous transition dosing, as well as carbohydrates. Evaluating current and cumulative glucose levels, the software adjusts and maintains the glucose level within a configurable clinician-determined turget range. The system is indicated for use in adult and pediatric patients ages 2 years and above, who weigh 12 kgs. or more. EndoTool IV logic is not a substitute for, but rather an adjunct to clinical reasoning. No medical decision should be based solely on the recommended guidance provided by this software system.

## Device Story

EndoTool IV is a glycemic management software system used by healthcare professionals in clinical settings. It takes current and cumulative blood glucose values as input; uses proprietary mathematical modeling and feedback control to calculate insulin, dextrose, or carbohydrate doses; and recommends IV and subcutaneous transition dosing. The software adjusts and maintains glucose levels within a clinician-determined target range. Calculations are repeated every 30 minutes to 2 hours based on glucose patterns. The system acts as an adjunct to clinical reasoning; it does not replace professional judgment. Healthcare providers review recommendations to inform clinical decision-making, potentially improving glycemic control and patient safety. The device is a modified version of the predicate, featuring updated user interface, patient management aspects, and additional software features.

## Clinical Evidence

No clinical data. Substantial equivalence is based on benchtop performance testing, including requirements-based module and integration testing, regression testing, and static analysis of software code.

## Technological Characteristics

Standalone software; no physical materials. Operates on standard server/PC hardware. Connectivity: online web-based and offline server-based. Software validation per IEC 62304. Uses 'Algorithm B' for glucose management calculations.

## Regulatory Identification

A predictive pulmonary-function value calculator is a device used to calculate normal pulmonary-function values based on empirical equations.

## Predicate Devices

- EndoTool IV 1.10 ([K200443](/device/K200443.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA

U.S. FOOD &amp; DRUG

ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

INSTRUMENT ONLY

## I Background Information:

A 510(k) Number

K201619

B Applicant

Monarch Medical Technologies, LLC

C Proprietary and Established Names

EndoTool IV System

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  NDC | Class II | 21 CFR 868.1890 - Predictive Pulmonary-Function Value Calculator | Clinical Chemistry (75)  |

## II Submission/Device Overview:

A Purpose for Submission:

Modification to a cleared device to update the user interface, change certain aspects of patient management and addition of software features.

B Type of Test:

Insulin or dextrose/carbohydrate intravenous and subcutaneous transition dose calculation.

## III Intended Use/Indications for Use:

A Intended Use(s):

See Indications for Use below.

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

{1}

B Indication(s) for Use:

EndoTool IV is a glucose management software system, designed for use by healthcare professionals in all patient care settings to recommend intravenous and subcutaneous transition dosing, as well as carbohydrates. Evaluating current and cumulative glucose levels, the software adjusts and maintains the glucose level within a configurable clinician-determined target range. The system is indicated for use in adult and pediatric patients ages 2 years and above, who weigh 12 kgs. or more. EndoTool IV logic is not a substitute for, but rather an adjunct to clinical reasoning. No medical decision should be based solely on the recommended guidance provided by this software system.

C Special Conditions for Use Statement(s):

Rx - For Prescription Use Only

A patient's potassium level must be assessed prior to administering insulin. Patients with low to low normal potassium levels are at risk for a life threatening drop in their circulating potassium level when insulin therapy is initiated. Consult with the patient's provider to determine if the patient's potassium level is adequate or if treatment should be initiated prior to insulin administration.

IV Device/System Characteristics:

A Device Description:

EndoTool IV is a glycemic management software device for blood glucose management which evaluates current and cumulative blood glucose values to recommend intravenous insulin/dextrose or oral carbohydrate dose, to adjust and maintain the blood glucose level within a provider-ordered target range. In addition, the application will recommend a subcutaneous basal transition insulin dose when IV insulin therapy is no longer required. It is used by healthcare professionals in patient care settings to recommend IV and subcutaneous transition dosing, as well as carbohydrate, to adjust and maintain the blood glucose level within a configurable clinician-determined target range.

The EndoTool IV device is a modified version of the predicate device, the EndoTool IV 1.10. Like the predicate device, the EndoTool IV system uses mathematical modeling to automatically individualize dosing by calculating the appropriate insulin infusion dose based on a patient's previous blood glucose readings and the rate of change of blood glucose concentrations. These calculations are repeated by the software every 30 minutes to 2 hours, depending on the pattern of glucose levels.

The subject device is a modified version of the predicate device and uses the same proprietary feedback control and modeling based upon a patient's glucose response to therapy. The changes to the device include modifications to the device user interface, changes to certain aspects of patient management, and addition of several software features.

The functional elements of the EndoTool IV software are the primary routines such as dosing model, therapy updates, advisories and recommendations that help achieve its intended use.

K201619 - Page 2 of 6

{2}

These functions are either unchanged or are substantially equivalent compared to the predicate device.

EndoTool IV is a standalone software device. It does not have any components or accessories. The device is designed to fit within the hospital's existing infrastructure.

## B Instrument Description Information:

1. Instrument Name:
EndoTool IV

2. Specimen Identification:
Not Applicable.

3. Specimen Sampling and Handling:
Not Applicable.

4. Calibration:
Not Applicable.

5. Quality Control:
Not Applicable.

## V Substantial Equivalence Information:

A Predicate Device Name(s):
EndoTool IV 1.10

B Predicate 510(k) Number(s):
K200443

C Comparison with Predicate(s):

|  Device & Predicate Device(s): | K201619 | K200443  |
| --- | --- | --- |
|  Device Trade Name | EndoTool IV System | EndoTool IV 1.10  |
|  General Device Characteristic Similarities |  |   |
|  Intended Use/Indications For Use | Calculate recommended doses of insulin and carbohydrate, for | Same  |

K201619 - Page 3 of 6

{3}

|  Device & Predicate Device(s): | K201619 | K200443  |
| --- | --- | --- |
|   | intravenous and/or subcutaneous delivery, to adjust and maintain the blood glucose level within a configurable clinician-determined target range. |   |
|  User group / patient range | 12 kg and 2yr or greater | Same  |
|  Dose calculation | Creates patient-specific IV drug-infusion profiles; evaluates current glucose value; evaluates cumulative glucose value; calculates next dose of insulin, glucose, or saline; calculates carbohydrate intake, calculates nutritional bolus; calculates meal intake; based on user inputs. | Same  |
|  General Device Characteristic Differences |  |   |
|  Software installation | Cloud-based or locally installed algorithm, with locally installed user interface. | Cloud-based algorithm, with locally installed user interface.  |
|  Warnings | Yes
New warnings / controls regarding low potassium added | Yes  |

VI Standards/Guidance Documents Referenced:

ANSI AAMI IEC 62304:2006/A1:2016 Medical Device Software - Software Life Cycle Processes [Including Amendment 1 (2016)]"

ISO 14971 Second Edition 2007-03-01 Medical Devices - Applications of Risk Management o Medical Devices

K201619 - Page 4 of 6

{4}

VII Performance Characteristics (if/when applicable):

A Analytical Performance:

1. Precision/Reproducibility:
Not applicable.

2. Linearity:
Not applicable.

3. Analytical Specificity/Interference:
Not applicable.

4. Accuracy (Instrument):
Not applicable.

5. Carry-Over:
Not applicable.

B Other Supportive Instrument Performance Characteristics Data:

Software Documentation

The level of concern for this software is major. Software documentation consistent with a major level of concern was provided as described in the FDA guidance for the Content of Premarket Submissions for Software Contained in Medical Devices. Documentation included complete software requirements and design specification documentation, detailed depiction of functional units and software modules, state diagrams and flowcharts, software lifecycle development plan with annotated listing of control documents generated during development process, configuration management plan, and verification and validation at unit, integration, and system levels with protocols, pass/fail criteria, and results.

Software verification and validation testing was conducted per IEC 62304 and included the following:

- Requirements-based unit, integration, and system-level tests
- Endo-Tool regression testing including:
- Requirements-based testing for risk-related requirements
- Automated algorithm test cases for the complete application
- Static analysis of the software code

K201619 - Page 5 of 6

{5}

VIII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

IX Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

K201619 - Page 6 of 6

---

**Source:** [https://fda-staging.innolitics.com/device/K201619](https://fda-staging.innolitics.com/device/K201619)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
