The Apex Knee™ System is intended for use as a primary or revision total knee replacement. This prosthesis may be used for the following conditions, as appropriate: - Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis; - Rheumatoid arthritis; - Correction of functional deformity: - Revision procedures where other treatments or devices have failed; The porous coated femoral component may be used cemented (biological fixation). The porous coated tibial baseplate is to be used uncemented (biological fixation). All other femoral, tibial baseplate, and patellar components are indicated for cemented use only. The Apex Knee™ Modular Tibial Augments are intended to the Tibia baseplate and cemented to the prepared tibia.
Device Story
Apex Knee™ System is a semi-constrained total knee prosthesis for primary or revision arthroplasty. Components include pegged tibial baseplates (Cobalt Chrome, ASTM F75) for enhanced torsional stability and ECiMa tibial inserts/patellae made of Vitamin E-stabilized highly crosslinked polyethylene (VE-XLPE). System components are implanted by orthopedic surgeons in a clinical/OR setting. Pegged baseplates are secured via pegs; inserts attach to baseplates via dovetails and a retaining mechanism. Device provides mechanical joint replacement to restore function and alleviate pain in patients with degenerative or inflammatory joint disease. System is compatible with existing Apex Knee components.
Clinical Evidence
Bench testing only. Non-clinical evidence includes FEA simulation, peg location templating, material characterization, PS post strength testing, insert disassembly strength testing, and wear testing. Biocompatibility testing included intramuscular and subcutaneous implantation in animal models.
Indicated for patients requiring primary or revision total knee replacement due to non-inflammatory degenerative joint disease (osteoarthritis, avascular necrosis), rheumatoid arthritis, functional deformity, or failed previous treatments.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{0}------------------------------------------------
April 2, 2021
Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services seal on the left and the FDA acronym with the full name of the agency on the right. The seal features a stylized depiction of human figures, while the FDA part includes a blue square with "FDA" in white and the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.
OMNIlife science Dipti Dharia Chief Quality, Regulatory, & Clinical Officer 480 Paramount Drive Raynham, Massachusetts 02767
## Re: K201611/S001
Trade/Device Name: Apex Knee™ System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee Joint Patellofemorotibial Polymer Semi-Constrained Cemented Prosthesis Regulatory Class: Class II Product Code: JWH, MBH Dated: February 19, 2021 Received: March 4, 2021
Dear Dipti Dharia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Ting Song, Ph.D., R.A.C. Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K201611
Device Name Apex Knee™ System
Indications for Use (Describe)
The Apex Knee™ System is intended for use as a primary or revision total knee replacement. This prosthesis may be used for the following conditions, as appropriate:
- · Non-inflammatory degenerative joint disease, including osteoarthritis and avascular necrosis;
- · Rheumatoid arthritis;
- · Correction of functional deformity:
- · Revision procedures where other treatments or devices have failed;
The porous coated femoral component may be used cemented (biological fixation).
The porous coated tibial baseplate is to be used uncemented (biological fixation).
All other femoral, tibial baseplate, and patellar components are indicated for cemented use only.
The Apex Knee™ Modular Tibial Augments are intended to the Tibia baseplate and cemented to the prepared tibia.
| Type of Use (Select one or both, as applicable) | |
|------------------------------------------------------------------------------------------------|----------------------------------------------------------|
| <div><span style="font-size:100%;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D)</div> | <div>□ Over-The-Counter Use (21 CFR 801 Subpart C)</div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for OMNI. The logo consists of a circular graphic on the left and the word "OMNI" on the right. The circular graphic is blue and green and appears to be a stylized representation of a person. Below the word "OMNI" is the phrase "Engineering Orthopaedic Innovation".
# 510(k) Summary
| TABLE 1: 510(k) SUMMARY | |
|------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date Summary Prepared | 06/12/2020 |
| Manufacturer/Distributor/Sponsor | OMNIlife Science<br>480 Paramount Drive<br>Raynham, MA 02767 |
| 510(k) Contact | Dipti Dharia<br>Chief Quality, Regulatory & Clinical Officer<br>Corin Ltd.,<br>480 Paramount Drive<br>Raynham, MA 02767<br>Telephone: +1-732-754-5466<br>Email: Dipti.Dharia@coringroup.com<br><br>Secondary Contact:<br>Jack Liang<br>Senior Regulatory Affairs Specialist<br>Corin Ltd / Corin Australia<br>+61 2 9497 7400<br>Jack.Liang@coringroup.com |
| Trade Name | Apex Knee™ System |
| Common Name | prosthesis, knee, patellofemorotibial, semi-constrained,<br>cemented, polymer/metal/polymer<br><br>prosthesis, knee, patello/femorotibial, semi-constrained,<br>uncemented, porous, coated |
| Classification | 21 CFR 888.3560. Class II. Knee joint<br>patellofemorotibial polymer/metal/polymer semi-<br>constrained cemented prosthesis. JWH<br><br>21 CFR 888.3565. Class II. Knee joint<br>patellofemorotibial metal/polymer porous-coated<br>uncemented prosthesis. MBH |
| Predicate Device | OMNI TiN Coated Apex Knee™ System- K191765 |
| Purpose of Submission | This traditional 510(k) premarket notification is being<br>submitted to obtain clearance for the modifications to<br>the Apex Knee™ System to expand OMNI's product<br>offering for total knee arthroplasty. |
| Indications for Use | The Apex Knee™ System is intended for use as a<br>primary or revision total knee replacement. This<br>prosthesis may be used for the following conditions, as<br>appropriate:<br>Non-inflammatory degenerative joint disease,<br>including osteoarthritis and avascular necrosis; Rheumatoid arthritis; Correction of functional deformity; Revision procedures where other treatments or<br>devices have failed; The porous coated femoral component may be used<br>cemented or uncemented (biological fixation).<br>The Porous coated tibial baseplate is to be used<br>uncemented (biological fixation).<br>All other femoral, tibial baseplate, and patellar<br>components are indicated for cemented use only.<br>The Apex Knee™ Modular Tibia System Tibial<br>Augments are intended to be bolted to the Tibia<br>baseplate and cemented to the prepared tibia. |
| Device Description | The proposed devices are intended to be used for<br>primary and revision total knee replacement.<br><br>The Pegged Tibial Baseplates offer additional torsional<br>stability by adding pegs to the posterior end of the<br>device. The material of the proposed Pegged Tibial<br>Baseplates are Cobalt Chrome, CoCr (ASTM F75).<br><br>The ECiMa tibial inserts are machined from<br>compression molded highly crosslinked polyethylene<br>with Vitamin E (VE-XLPE) and attaches to the cobalt<br>chrome tibial baseplate via dovetails and a retaining |
| Engineering Orthopaedic Innovation | |
| Sterility and Biocompatibility | The proposed devices undergo the same validated<br>sterilization process, using ethylene oxide (EO), as the<br>predicate devices (K060192, K102578, K111184,<br>K112891) under the sterility assurance level (SAL) of<br>10-6. All ethylene oxide residuals are monitored and<br>well below standard limits.<br>In addition, OMNI has developed a plan to test<br>endotoxins on all OMNIlife science device groups<br>through Limulus amebocyte lysate (LAL) testing.<br>Products have been segregated into product families<br>based on manufacturing process and material type and<br>have generated 8 product groups that will be tested for<br>endotoxins on a quarterly basis on a yearly rotation, and<br>the subject devices are included in one of the product<br>groups. Product will not be released if the 4 EU/device<br>limit is exceeded. |
| Substantial Equivalence Summary | The proposed devices with modifications are<br>substantially equivalent to the existing Apex Knee™<br>System cleared in K191765 as the basic design,<br>interface, fundamental technology and intended use are<br>the same.<br>The use of the new Apex Knee™ System components<br>with the existing knee components do not introduce any<br>new risks of safety or efficacy.<br>Tibial Baseplate:<br>In addition to the testing performed for the Apex Knee<br>System, non-clinical testing specifically for the Pegged<br>Porous Coated Cementless Tibial Baseplates include<br>FEA simulation and peg location templating study.<br>The results of the testing show that the Apex Knee<br>Pegged Porous Coated Cementless Tibial Baseplates are |
| | safe and effective for the proposed indications and is<br>substantially equivalent to the predicate devices.<br>The justifications are described in Section 10, Device<br>Description. |
| | <b>ECiMa Tibial Inserts & Patellae</b> |
| | In addition to the testing already performed for the Apex<br>Knee System, non-clinical testing was conducted<br>specifically for the ECiMa Tibial Inserts & Patellae.<br>These tests include:<br>• material characterization,<br>• PS post strength testing,<br>• insert disassembly strength testing,<br>• wear testing. |
| | Biological testing of the ECiMa material was conducted<br>for biocompatibility. These tests include:<br>• intramuscular implantation in animal models,<br>• implant toxicity and subcutaneous implantation in<br>animal models. |
| | The results of the testing show that the Apex Knee<br>ECiMa Tibial Inserts & Patellae are safe and effective<br>for the proposed indications and is substantially<br>equivalent to the predicate devices. |
| | Based on the design, fundamental technology, material,<br>intended use and technological characteristics,<br>OMNIlife science believes the proposed Apex KneeTM<br>System devices to be substantially equivalent to legally<br>marketed predicates. |
| Conclusion Statement | The conclusions drawn from the nonclinical tests<br>demonstrate that the devices are safe, as effective, and<br>perform as well as or better than the legally marketed<br>device |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for OMNI. The logo consists of a blue and green circular graphic to the left of the word "OMNI" in large, bold, blue letters. Below the word "OMNI" is the text "Engineering Orthopaedic Innovation" in a smaller font size. The logo is clean and professional, and the colors are eye-catching.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for OMNI Engineering Orthopaedic Innovation. The logo features a stylized human figure inside of a circle on the left. The word "OMNI" is in large, bold, blue letters to the right of the figure. Below the word "OMNI" is the phrase "Engineering Orthopaedic Innovation" in a smaller font.
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows the logo for OMNI. The logo consists of a blue and green circular design on the left, followed by the word "OMNI" in large, bold, blue letters. Below the word "OMNI" is the text "Engineering Orthopaedic Innovation" in a smaller font.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.