K201591 · Ka Imaging, Inc. · MQB · Sep 3, 2020 · Radiology
Device Facts
Record ID
K201591
Device Name
Flat Panel Detector
Applicant
Ka Imaging, Inc.
Product Code
MQB · Radiology
Decision Date
Sep 3, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Indications for Use
The flat panel detector when used with a radiographic imaging system is intended to generate radiographic images of human anatomy wherever a conventional screen-film, digital radiography (DR) or computed radiography (CR) detector is used for general purposes. When the dual energy subtraction is enabled, it is intended to assist the physician through the visualization of anomalies by reducing the visibility of underlying or overlying anatomical structures. This device is not intended for use in mammography applications.
Device Story
Reveal 35C is a portable digital flat panel detector (FPD) for general radiography; integrates with PC workstation and X-ray source. Input: X-ray radiation captured by three stacked amorphous silicon active matrix arrays with cesium iodide scintillators. Operation: arrays read simultaneously and combined additively to produce standard digital images; optional dual-energy subtraction mode uses spectral differences between layers to compute and output three images (standard, bone-only, soft-tissue-only) from a single 120kVp exposure. Used in clinical settings by physicians/technologists. Output: digital images transmitted via wired/wireless connection to workstation for display. Benefit: enables material-specific visualization (bone vs. soft tissue) without requiring multiple exposures or motion-correction algorithms.
Clinical Evidence
Bench testing only. Device hardware and software requirements verified against specifications. Image quality validation confirmed equivalence to predicate devices. Adherence to IEC 60601-1 and IEC 60601-1-2 standards demonstrated.
Technological Characteristics
Amorphous silicon active matrix array with cesium iodide scintillator; 140-micron pixel pitch; 2500 x 3052 matrix; 16-bit dynamic range. Three-layer stacked sensor design. Battery-powered; wireless/wired connectivity. Dimensions: 38 x 46 x 1.5 cm; weight: 3.2 kg. Operates on Windows PC workstation.
Indications for Use
Indicated for general projection radiographic imaging of human anatomy. Not indicated for mammography, fluoroscopy, tomography, or angiography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
FUJIFILM DUAL ENERGY SUBTRACTION (DES) SOFTWARE OPTION (K122454)
GE DUAL ENERGY AND TISSUE EQUALIZATION SOFTWARE OPTION (K013481)
Submission Summary (Full Text)
{0}------------------------------------------------
September 3, 2020
Image /page/0/Picture/1 description: The image shows the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text. The logos are simple and professional, and they are easily recognizable.
KA Imaging Inc. % Amol Karnick QA/RA Representative 560 Parkside Dr #3 Waterloo, Ontario N2L 5Z4 CANADA
## Re: K201591
Trade/Device Name: Flat Panel Detector Regulation Number: 21 CFR 892.1680 Regulation Name: Stationary x-ray system Regulatory Class: Class II Product Code: MQB Dated: June 10, 2020 Received: June 12, 2020
Dear Amol Karnick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
K201591
Device Name Flat Panel Detector
### Indications for Use (Describe)
The flat panel detector when used with a radiographic imaging system is intended to generate radiographic images of human anatomy wherever a conventional screen-film, digital radiography (DR) or computed radiography (CR) detector is used for general purposes.
When the dual energy subtraction is enabled, it is intended to assist the physician through the visualization of anomalies by reducing the visibility of underlying or overlying anatomical structures.
This device is not intended for use in mammography applications.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows a logo with the letters 'KA' stacked on top of the word 'IMAGING'. The 'K' is blue, and the 'A' transitions from blue to yellow. The word 'IMAGING' is in a bold, sans-serif font and is also yellow. The logo is simple and modern, with a clean design.
## 510(k) Summary
#### SUBMITTER l.
KA Imaging Inc. 560 Parkside Dr #3, Waterloo, ON Canada N2L 5Z4 Phone: +1 (226)-215-9897 Email: sto@kaimaging.com Contact Person: Samuel To Position: QA/RA Representative Date of Submission: June 10, 2020
#### II. DEVICE
Common name: Flat Panel Detector Model name: Reveal 35C Device Classification Name: Solid State X-Ray Imager (Flat Panel/Digital Imager) Review Panel: Radiology Product code: MQB Requlation number: 21 CFR 892.1680 Device class: Class 2
#### lll. PREDICATE DEVICE
Substantial equivalence to the following predicate devices is as follow:
| 510(k) # | Product | Company | Reference |
|----------|--------------------------------------------------------------|-----------------------------------------|----------------------------------|
| K191939 | Yushan X-Ray Flat Panel<br>Detector | InnoCare<br>Optoelectronics Corp. | Primary<br>Predicate<br>Device |
| K150766 | Carestream DRX-1 System | Carestream Health, Inc. | Reference<br>Predicate<br>Device |
| K122454 | FUJIFILM DUAL ENERGY<br>SUBTRACTION (DES)<br>SOFTWARE OPTION | FUJIFILM MEDICAL<br>SYSTEM U.S.A., INC. | Reference<br>Predicate<br>Device |
| K013481 | DUAL ENERGY AND<br>TISSUE EQUALIZATION<br>SOFTWARE OPTION | GE MEDICAL<br>SYSTEMS | Reference<br>Predicate<br>Device |
#### IV. DEVICE DESCRIPTION
The Reveal 35C Flat Panel Detector is similar to the FDA cleared Yushan X-ray Flat Panel Detector. The detectors consist of amorphous silicon flat panel image sensors with cesium iodide scintillators. The light is captured by an amorphous silicon photodetector and the resulting signal is transferred via amorphous silicon thin film transistor (TFT) switches to external readout electronics to obtain X-ray images. The Reveal 35C Flat Panel Detector is a portable digital detector that can be integrated with a PC workstation and an X-ray source to acquire digital X-
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a logo for KA Imaging. The logo consists of the letters "KA" stacked on top of each other. The "K" is blue, and the "A" is yellow. Below the letters, the word "IMAGING" is written in orange. The logo is simple and modern.
Ray images for general radiography. The detector supports wireless and wired data communication and can be used wherever a conventional screen-film, digital radiography, or computed radiography detector is used for general purposes.
The Reveal 35C Flat Panel Detector synchronize their image capture cycle with the X-Ray exposure in either of the two modes:
- 1. Wired Mode
- Wireless Mode 2.
The subject device, Reveal 35C Flat Panel Detector includes an optional Dual-Energy subtraction function. When the Dual-Energy Subtraction function is enabled, it will provide additional dual energy subtracted X-ray images. The images are intended to assist the physician through the visualization of anomalies by reducing the visibility of underlying or overlying anatomical structures.
#### V. INDICATIONS FOR USE
The flat panel detector when used with a radiographic imaging system is intended to generate radiographic images of human anatomy wherever a conventional screen-film, digital radiography (DR) or computed radiography (CR) detector is used for general purposes.
When the dual energy subtraction function is enabled, it is intended to assist the physician through the visualization of anomalies by reducing the visibility of underlying or overlying anatomical structures.
This device is not intended for use in mammography applications.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for KA Imaging. The logo consists of the letters "KA" stacked on top of the word "IMAGING". The "K" is blue, and the "A" is a gradient that transitions from blue to yellow. The word "IMAGING" is in orange.
### VI. COMPARISON OF TECHNOLOGICAL CHARACTERISTICS WITH THE PREDICATE DEVICE
KA Reveal 35C device is an X-ray flat panel detector (FPD) equivalent to its indicated predicate devices. Like the predicate devices, Reveal 35C is able to capture radiographic images in a medical diagnostic setting thanks to its amorphous silicon active matrix array with an overlying cesium iodide scintillator laver. It also contains a graphical user interface software that allows for FPD operation, image processing and image display, similar to that of the identified predicate devices. Reveal 35C is therefore equivalent to its predicate devices with respect to form-factor and operation.
Unlike its predicate devices. Reveal 35C is composed of three (3) stacked active matrix arrays. Each array has its own substrate, TFT/photo-diode layer and scintillator layer. These arrays are simultaneously read and are then additively combined to generate a single image. This additive image is always digitally stored and transferred for viewing. This single image is equivalent to the single image obtained by any of the listed predicate devices, thus making Reveal 35C's image output equivalent to those devices.
Additionally, Reveal 35C's three-layer design allows for a Dual-Energy software option to be included with this submission. When this option is used, this software option is able to compute tissue-subtracted Dual-Energy images, thanks to the spectral differences between the different layer images. The Dual-Energy images obtained by this software option are equivalent to those of reference predicate devices (K122454 and K013481). See Table 1 below for a summary.
| SE Table 1 | | | |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Description | Proposed Device: | Primary Predicate Device: | Reference Predicate Device: |
| | KA Imaging Reveal<br>35C Flat Panel<br>Detector | InnoCare<br>Optoelectronics<br>Yushan X-Ray Flat<br>Panel Detector<br>(K191939) | Carestream DRX-1<br>System<br>(K150766) |
| Indication for Use | The flat panel<br>detector when used<br>with a radiographic<br>imaging system is<br>intended to<br>generate<br>radiographic<br>images of human<br>anatomy wherever<br>a conventional | The Wireless/Wired<br>InnoCare Yushan<br>X-Ray Flat Panel<br>Detector is intended<br>to capture for<br>display radiographic<br>images of human<br>anatomy. It is<br>intended for use in<br>general projection | The device is<br>intended to capture<br>for display<br>radiographic<br>images of human<br>anatomy including<br>both pediatric and<br>adult patients. The<br>device is intended<br>for use in general |
| | | | |
| | screen-film, digital<br>radiography (DR) or<br>computed<br>radiography (CR)<br>detector is used for<br>general purposes.<br><br>When the dual<br>energy function is<br>enabled, it is<br>intended to assist<br>the physician<br>through the<br>visualization of<br>anomalies by<br>reducing the<br>visibility of<br>underlying or<br>overlying<br>anatomical<br>structures.<br><br>This device is not<br>intended for use in<br>mammography<br>applications. | radiographic<br>applications<br>wherever<br>conventional<br>film/screen or CR<br>systems may be<br>used. The InnoCare<br>Yushan X-Ray<br>Flat Panel Detector<br>is not intended for<br>mammography,<br>fluoroscopy,<br>tomography, and<br>angiography<br>applications. | projection<br>radiographic<br>applications<br>wherever<br>conventional<br>screen-film<br>systems or CR<br>systems may be<br>used. Excluded<br>from the indications<br>for use are<br>mammography,<br>fluoroscopy, and<br>angiography<br>applications. |
| Detector Type | Flat Panel detector<br>with scintillator | Flat Panel detector<br>with scintillator | Flat Panel detector<br>with scintillator |
| Image Capture<br>Area | 35cm x 43 cm | 35cm x 43 cm | 35cm x 43 cm |
| Detector Device<br>Material | Amorphous Silicon<br>Sensor Array with<br>Cesium lodide<br>scintillator | Amorphous Silicon<br>Sensor Array with<br>Cesium lodide or<br>Gadolinium<br>Oxisulfide<br>scintillator | Amorphous Silicon<br>Sensor Array with<br>Cesium lodide or<br>Gadolinium<br>Oxisulfide<br>scintillator |
| Pixel Pitch | 140 microns | 140 microns | 139 microns |
| Detector Element<br>Matrix | 2500 x 3052 | 2500 x 3052 | 2544 x 3056 (3543<br>GOS), 2520 x 3032<br>(3543 Csl) |
| Dynamic Range | 16 bits | 16 bits | 16 bits |
| Operating System | Windows PC | Windows PC | Windows PC |
| Operating Console | Graphical User<br>Interface Based | Graphical User<br>Interface Based | Graphical User<br>Interface Based |
| Image<br>Transmission | Image is<br>electronically | Image is<br>electronically | Image is<br>electronically |
| | transmitted in digital<br>form through tether<br>(cable) or wireless<br>connection to a<br>computer monitor<br>for display | transmitted in digital<br>form through tether<br>(cable) or wireless<br>connection to a<br>computer monitor<br>for display | transmitted in digital<br>form through tether<br>(cable) or wireless<br>connection to a<br>computer monitor<br>for display |
| Detector<br>Dimensions | 38 x 46 x 1.5 cm | 38.3 x 46 x 1.5 cm | 38 x 46 x 1.6 cm |
| Detector Weight | 3.2 kg | 2.7 Kg | 4 kg |
| Power Source | Battery powered | Battery powered | Battery powered |
| Communication | Wireless/ Wired | Wireless/ Wired | Wireless/ Wired |
{6}------------------------------------------------
Image /page/6/Picture/1 description: The image shows a logo for KA Imaging. The logo consists of the letters 'KA' stacked on top of the word 'IMAGING'. The 'K' is blue, and the 'A' transitions from blue to yellow. The word 'IMAGING' is in orange.
{7}------------------------------------------------
Image /page/7/Picture/1 description: The image shows a logo for KA Imaging. The logo features the letters "KA" stacked on top of the word "IMAGING". The "K" is blue, and the "A" is a gradient from blue to yellow. The word "IMAGING" is in yellow and is located directly below the letters "KA".
The KA Imaging Reveal 35C Flat Panel Detector also includes a dual energy subtraction software option, which can be used to optionally generate three types of radiographic images from a single X-Ray exposure. The first is a (1) standard projection radiography image, equivalent to the primary and reference predicate device (K150766). The second and third images are a (2) soft-tissue image with the bone information removed, and a (3) bone image with the soft tissue information removed. The second image and third image allow the radiologist to subtract specific materials (e.g. bone, soft tissue) and focus on the tissue-type of choice, thus increasing the visualization of the objects of interest.
The reference predicate devices Fujifilm Dual Energy Subtraction (DES) Software Option (K122454) and GE Dual Energy And Tissue Equalization Software Option (K013481) are both software options to provide Dual Energy subtracted images. For both K122454 and K013481, two spectral source images are obtained using two X-ray exposures (the first exposure at a lower kV (typically 60kV) and the second exposure at a higher kV (typically 120kV)). For Reveal 35C, the spectral source images are generated from only a single kV(typically 120kV) exposure because of the use of three active matrix arrays in the Reveal 35C design. Since only a single exposure is utilized for Reveal 35C, there is no additional software required to correct for the involuntary motion artifacts described in K122454. Please see the summary in Table 2 below:
{8}------------------------------------------------
Image /page/8/Picture/1 description: The image shows the logo for KA Imaging. The logo consists of the letters "KA" stacked on top of the word "IMAGING". The "K" is blue, and the "A" is a gradient that transitions from blue to yellow. The word "IMAGING" is in orange.
# SE Table 2
| Description | Proposed<br>Device:<br>KA Imaging<br>Reveal 35C Flat<br>Panel Detector | Reference<br>Predicate Device:<br>Fujifilm Dual<br>Energy Subtraction<br>(DES) Software<br>Option (K122454) | Reference<br>Predicate<br>Device:<br>GE Dual Energy<br>And Tissue<br>Equalization<br>Software Option<br>(K013481) |
|--------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indication for<br>Use | The flat panel<br>detector when<br>used with a<br>radiographic<br>imaging system<br>is intended to<br>generate<br>radiographic<br>images of human<br>anatomy<br>wherever a<br>conventional<br>screen-film,<br>digital<br>radiography (DR)<br>or computed<br>radiography (CR)<br>detector is used<br>for general<br>purposes.<br>When the dual<br>energy function is<br>enabled, it is<br>intended to assist<br>the physician<br>through the<br>visualization of<br>anomalies by<br>reducing the<br>visibility of<br>underlying or<br>overlying<br>anatomical<br>structures. | Fujifilm's FDR Dual<br>Energy Subtraction<br>(DES) Option may<br>be used with<br>Fujifilm's DR X-ray<br>systems and is<br>intended to be used<br>by a<br>qualified/trained<br>doctor or<br>technologist for<br>acquiring dual<br>energy subtraction<br>images of human<br>anatomy<br>DES is intended to<br>assist the physician<br>through the<br>visualization of<br>anomalies by<br>reducing the<br>visibility of<br>underlying/overlying<br>anatomical<br>structures<br>The device is not<br>intended for<br>mammographic<br>applications | Dual Energy and<br>Tissue<br>Equalization<br>software options<br>are intended for<br>use in generating<br>digital<br>radiographic<br>images of human<br>anatomy. This<br>device is not<br>intended for<br>mammographic<br>applications. |
| | This device is not<br>intended for use<br>in mammography<br>applications. | | |
| Number of x-ray<br>exposures | 1 | 2 | 2 |
| Active imager<br>matrix arrays | 3 | 1 | 1 |
| X-ray beam<br>energies | 120kVp | 60 kVp/120kVp | 60 kVp/120kVp |
| Automatic Image<br>Subtraction | Yes | Yes | Yes |
| Involuntary<br>patient motion<br>reduction | Not required | Yes | Yes |
| Images sent to<br>destination | 3 Images<br>(Standard, Bone<br>and Soft Tissue) | 3 Images<br>(Standard, Bone<br>and Soft Tissue) | 3 Images<br>(Standard, Bone<br>and Soft Tissue) |
{9}------------------------------------------------
Image /page/9/Picture/1 description: The image shows a logo for KA Imaging. The logo features the letters "KA" in a stylized design, with the "K" in blue and the "A" transitioning from blue to yellow. Below the letters, the word "IMAGING" is written in a bold, sans-serif font, also in yellow. The overall design is clean and modern.
#### VII. PERFORMANCE DATA
Non-clinical Performance Data
The Reveal 35C Flat Panel Detector conforms to the voluntary standards including IEC 60601-1 and IEC 60601-1-2. Non-clinical bench testing has determined that the device hardware and software requirements conform to its specification.
FDA's Guidance for the Submission of 510(k)s for Solid State X-ray Imaging Devices (issued on September 1, 2016), Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices, Guidance for Content of Premarket Submissions for Management of Cybersecurity in Medical Devices, were followed to demonstrated that the performance of Reveal 35C Flat Panel Detector is substantially equivalent to the predicate devices.
Furthermore, the image quality validation confirmed that the image quality of KA Imaging Reveal 35C Flat Panel Detector is substantially equivalent to that of the predicate device.
{10}------------------------------------------------
Image /page/10/Picture/1 description: The image shows a logo for KA Imaging. The logo features the letters 'KA' in a stylized design, with the 'K' in blue and the 'A' in yellow. Below the letters, the word 'IMAGING' is written in yellow, completing the logo.
#### CONCLUSIONS VIII.
The Reveal 35C Flat Panel Detector is substantially equivalent to the predicate devices in technical characteristics, design features, operating principles, functional and performance characteristics, and for the intended uses.
The Reveal 35C Flat Panel Detector is designed to comply with applicable federal and international safety and performance standards.
Based upon the supporting data summarized above, we can conclude the subject device is substantially equivalent in safety and effectiveness as the legally marketed devices. The verification and validation activities performed on the subject device did not raise any issues related to safety and effectiveness.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.