# CUVIS-spine (K201569)

_Curexo, Inc. · OLO · May 19, 2021 · Neurology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K201569

## Device Facts

- **Applicant:** Curexo, Inc.
- **Product Code:** [OLO](/productcode/OLO.md)
- **Decision Date:** May 19, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 882.4560
- **Device Class:** Class 2
- **Review Panel:** Neurology
- **Attributes:** Therapeutic

## Indications for Use

CUVIS-spine is intended for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of guide bush to be used by surgeons for navigating compatible surgical instruments in open or percutaneous pedicle screw placement provided that the required markers and rigid patient anatomy can be identified on O-arm or C-arm.

## Device Story

Mobile robotic system for pedicle screw placement; comprises robotic arm, main console, and optional staff console. Inputs: intraoperative 2D/3D images (O-arm/C-arm). Operation: surgeon performs surgical planning; system uses optical tracking (infrared camera) to track patient anatomy and surgical instruments; robotic arm provides force-controlled positioning (hand-guide function) for guide bush orientation. Used in OR by surgeons. Output: real-time visual display of instrument position relative to anatomy; physical guidance for instrument trajectory. Benefits: precise anatomical localization and instrument positioning during spinal surgery.

## Clinical Evidence

Bench testing included pose accuracy and repeatability validation. A cadaveric study was performed for robotic-assisted pedicle screw placement. Biocompatibility testing followed ISO 10993-5 and ISO 10993-10. Electrical safety and EMC testing followed ES 60601-1 and IEC 60601-1-2.

## Technological Characteristics

Mobile robotic system with robotic arm, main console, and staff console. Optical tracking system (infrared camera). Force-controlled movement for robotic arm positioning. Biocompatibility per ISO 10993. Electrical safety per ES 60601-1; EMC per IEC 60601-1-2. Risk management per ISO 14971; usability per IEC 60601-1-6.

## Regulatory Identification

A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.

## Predicate Devices

- Excelsius GPSTM ([K171651](/device/K171651.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

May 19, 2021

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is a blue square with the letters "FDA" in white, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Curexo, Inc. % Do Hyun Kim CEO BT Solutions, Inc. Unit 904, Eonju-ro 86-gil 5, Gangnam-gu Seoul, 06210 Republic of Korea

Re: K201569

Trade/Device Name: CUVIS-spine Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic Instrument Regulatory Class: Class II Product Code: OLO Dated: April 16, 2021 Received: April 19, 2021

Dear Do Hyun Kim:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For: Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K201569

Device Name CUVIS-spine

Indications for Use (Describe)

CUVIS-spine is intended for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of guide bush to be used by surgeons for navigating compatible surgical instruments in open or percutaneous pedicle screw placement provided that the required markers and rigid patient anatomy can be identified on O-arm or C-arm.

Type of Use (Select one or both, as applicable)

| <span style="font-family: sans-serif;"> <svg height="15" width="15"> <path d="M2 2 L13 13 M2 13 L13 2" stroke="black" stroke-width="2"></path> </svg> </span> Prescription Use (Part 21 CFR 801 Subpart D) | <input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C) |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------|

CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

| Submitter/Applicant:     | CUREXO, INC.                                                                                                                                                              |
|--------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address:                 | 577, Gangnam-Daero, Seocho-Gu<br>Seoul, 06530, Republic of Korea<br>Tel: +82-31-788-7935<br>Jungeun PARK<br>Quality Management Representative<br>Email: jepark@curexo.com |
| Applicant correspondent: | Do Hyun Kim, BT Solutions, Inc.                                                                                                                                           |
| Address:                 | Unit 904, Eonju-ro 86gil 5, Gangnam-gu<br>Seoul 06210, Korea.<br>Tel: +82-2-538-9140<br>Email: ceo@btsolutions.co.kr                                                      |
| Preparation Date:        | May 20, 2021                                                                                                                                                              |

#### Device Name and Code

| Device Trade Name:   | CUVIS-spine                |
|----------------------|----------------------------|
| Common Name:         | Pedicle Screw Guide System |
| Classification Name: | Stereotaxic instrument     |
| Product Code:        | OLO                        |
| Regulation Number:   | 21 CFR 882.4560            |
| Classification:      | II                         |
| Review Panel:        | Orthopedic                 |

## Predicate Device

CUVIS-spine is substantially equivalent to the following legally marketed predicate devices Table 1 Primary Predicate device

| Applicant           | Device Name     | 510(k) Number |
|---------------------|-----------------|---------------|
| Globus Medical Inc. | Excelsius GPSTM | K171651       |

{4}------------------------------------------------

### Device Description

The CUVIS-spine is a mobile system mainly comprising the robotic arm, the main console and the staff console as an option. The robotic arm is positioned on the floor near the side of the surgical table. The location of the main console or the staff console is appropriately determined considering the user preference and the environments.

The CUVIS-spine is a pedicle screw guide system which consists of Robotic Arm, Main Console, Staff Console, Guide bush, Source Calibrator, Registration Tool, Registration Tool Adapter, Robotic Arm Drape, Tool Drape, Detector Drape, Marker Ball, Patient Marker, Marker Driver, Detector Calibrator, Dilator, Serration-tip Dilator, Drill Bit, Tapper, Stylet Tapper, Screwdriver, Instrument Container.

#### Indications / Intended Use

CUVIS-spine is intended for use as an aid for precisely locating anatomical structures and for the spatial positioning and orientation of guide bush to be used by surgeons for navigating and/or guiding compatible surgical instruments in open or percutaneous pedicle screw placement provided that the required markers and rigid patient anatomy can be identified on O-arm or C-arm.

#### Technical Characteristics in Comparison to Predicate Devices

|                                                                           | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                               | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|---------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                                           | CUVIS-spine(K201569)                                                                                                                                                                                                                                                                                                                                                                                                         | Excelsius GPSTM (K171651)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Manufacturer                                                              | CUREXO, INC.                                                                                                                                                                                                                                                                                                                                                                                                                 | Globus Medical Inc.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Classification                                                            | Class II                                                                                                                                                                                                                                                                                                                                                                                                                     | Class II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Product Code                                                              | OLO                                                                                                                                                                                                                                                                                                                                                                                                                          | OLO                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Regulation No.                                                            | 882.4560                                                                                                                                                                                                                                                                                                                                                                                                                     | 882.4560                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |
|                                                                           | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                               | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                                                           | CUVIS-spine(K201569)                                                                                                                                                                                                                                                                                                                                                                                                         | Excelsius GPSTM (K171651)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| General Device<br>description                                             | The CUVIS-spine is a mobile<br>system mainly comprising the<br>robotic arm, the main console and<br>the staff console as an option. The<br>robotic arm is positioned on the<br>floor near the side of the surgical<br>table. The location of the main<br>console or the staff console is<br>appropriately determined<br>considering the user preference and<br>the environments.                                             | The EXCELSIUS GPSTM is a<br>Robotic Positioning System that<br>includes a computer controlled<br>robotic arm, hardware, and software<br>that enables real time surgical<br>navigation and robotic guidance<br>using radiological patient<br>images(preoperative CT,<br>intraoperative CT and fluoroscopy),<br>using a dynamic reference base and<br>positioning camera.                                                                                                                                                                                        |
| Indications<br>for use                                                    | CUVIS-spine is intended for use as<br>an aid for precisely locating<br>anatomical structures and for the<br>spatial positioning and orientation<br>of guide bush to be used by<br>surgeons for navigating and/or<br>guiding compatible surgical<br>instruments in open or percutaneous<br>pedicle screw placement provided<br>that the required markers and rigid<br>patient anatomy can be identified on<br>O-arm or C-arm. | The EXCELSIUS GPSTM is<br>intended for use as an aid for<br>precisely locating anatomical<br>structures and for the spatial<br>positioning and orientation of an<br>instrument holder or guide tube to<br>be used by surgeons for navigating<br>and/or guiding compatible surgical<br>instruments in open or percutaneous<br>procedures provided that the<br>required fiducial markers and rigid<br>patient anatomy can be identified on<br>CT scans or fluoroscopy. The<br>system is indicated for the<br>placement of spinal and orthopaedic<br>bone screws. |
| Technical Characteristics                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Principle of<br>operation                                                 | - Intraoperative images<br>- Patient registration<br>- Surgical planning<br>- Real-time tracking of navigated<br>instruments<br>- Guidance of instruments                                                                                                                                                                                                                                                                    | - Intraoperative/preoperative<br>images<br>- Patient registration<br>- Surgical planning<br>- Real-time tracking of navigated<br>instruments<br>- Guidance of instruments                                                                                                                                                                                                                                                                                                                                                                                      |
| Image                                                                     | O-arm, C-arm                                                                                                                                                                                                                                                                                                                                                                                                                 | O-arm, C-arm, CT                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                                                           | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                               | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                                                           | CUVIS-spine(K201569)                                                                                                                                                                                                                                                                                                                                                                                                         | Excelsius GPS TM (K171651)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Input Images                                                              | - 3D intra-operative exam<br>- 2D intra-operative exam                                                                                                                                                                                                                                                                                                                                                                       | - 3D pre-operative exam<br>- 3D intra-operative exam<br>- 2D intra-operative exam                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Tracker                                                                   | Optical Tracking System                                                                                                                                                                                                                                                                                                                                                                                                      | Optical Tracking System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Guide                                                                     | Dilator and Tapper                                                                                                                                                                                                                                                                                                                                                                                                           | Dilator and Tapper                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Target<br>Tracking                                                        | YES                                                                                                                                                                                                                                                                                                                                                                                                                          | YES                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Integrated<br>Software                                                    | - SRC (Control Software)<br>- SPN (Planner Software)                                                                                                                                                                                                                                                                                                                                                                         | Excelsius GPS Planning and<br>Navigation Application Software                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Save/load<br>Planning                                                     | YES                                                                                                                                                                                                                                                                                                                                                                                                                          | YES                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Merge images<br>functionality                                             | YES                                                                                                                                                                                                                                                                                                                                                                                                                          | YES                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Trajectory<br>planning<br>parameters                                      | - Entry point<br>- Target point<br>- Instrument length/diameter                                                                                                                                                                                                                                                                                                                                                              | - Entry point<br>- Target point<br>- Instrument length/diameter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Localization<br>means                                                     | Optical system (infrared camera)                                                                                                                                                                                                                                                                                                                                                                                             | Optical system (infrared camera)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Image-guided                                                              | YES                                                                                                                                                                                                                                                                                                                                                                                                                          | YES                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Controller                                                                | Forced-controlled movement<br>allowing robot arm positioning<br>(called hand guide function)                                                                                                                                                                                                                                                                                                                                 | Force-controlled movement<br>allowing robotic arm positioning                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Patient<br>registration<br>method                                         | - Intra-op CT: Registration tool<br>- Fluoroscopy: Source calibrator                                                                                                                                                                                                                                                                                                                                                         | - Pre-op CT: Fluoroscopic to pre-op<br>CT merge<br>- Intra-op CT: Registration fixture<br>- Fluoroscopy: Registration fixture                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Real time<br>display of<br>instrument<br>position                         | YES                                                                                                                                                                                                                                                                                                                                                                                                                          | YES                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Accessories                                                               | - Registration instruments<br>- Patient reference instruments<br>- Surgical instruments                                                                                                                                                                                                                                                                                                                                      | - Registration instruments<br>- Patient reference instruments<br>- Surgical instruments<br>- End effector                                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Performance data                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                              |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Electrical<br>Safety and<br>Electromagenti                                | - IEC60601-1<br>- IEC60601-1-2                                                                                                                                                                                                                                                                                                                                                                                               | - IEC60601-1<br>- IEC60601-1-2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
|                                                                           | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                               | Predicate Device                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
|                                                                           | CUVIS-spine(K201569)                                                                                                                                                                                                                                                                                                                                                                                                         | Excelsius GPS TM (K171651)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |
| c compatibility<br>Accuracy<br>verification on<br>anatomical<br>landmarks | Yes                                                                                                                                                                                                                                                                                                                                                                                                                          | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Biocompatibility                                                          | The biocompatibility evaluation for<br>the patient marker has been<br>conducted in accordance with EN<br>ISO 10993 standards.                                                                                                                                                                                                                                                                                                | The biocompatibility evaluation for<br>EXCELSIUS GPSTM has been<br>conducted in accordance with ISO<br>10993 standards.                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| Parts of<br>Contact with<br>Patient                                       | Patient marker                                                                                                                                                                                                                                                                                                                                                                                                               | Patient reference instruments<br>(Patient marker)                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Nature of<br>Body Contact                                                 | Implant device / bone                                                                                                                                                                                                                                                                                                                                                                                                        | Implant device / bone                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Patient Fixation                                                          | Reference is fixed to patient's bony<br>structure for tracking system                                                                                                                                                                                                                                                                                                                                                        | Reference is fixed to patient's bony<br>structure for tracking system                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |

CUVIS-spine is substantially equivalent to the following legally marketed predicate devices

{5}------------------------------------------------

{6}------------------------------------------------

{7}------------------------------------------------

#### Performance Data

Non-clinical tests: Pose accuracy and Repeatability of the CUVIS-spine were tested and validated. A cadaveric study of the proposed device had been performed for the robotic-assisted pedicle screw placement.

Biocompatibility were tested using following consensus standards:

- -Tests for in vitro cytotoxicity were tested and evaluated according to the FDArecognized consensus standard, ISO 10993-5.
- -Tests for irritation, skin sensitization and intracutaneous reactivity were tested and evaluated according to the FDA-recognized consensus standard, ISO 10993-10.

Electromagnetic compatibility and electrical safety, etc, were tested using following consensus standards:

- -Basic safety and essential performance of the CUVIS-spine is tested and evaluated according to the FDA-recognized consensus standard, ES 60601-1.
- -Effect to the device by electromagnetic disturbances were tested and evaluated according to the FDA-recognized consensus standard IEC 60601-1-2.
- Risk management was recorded by referring to ISO 14971. -

{8}------------------------------------------------

- -Usability was documented by referring to IEC 60601-1-6.
#### Substantial Equivalence

CUVIS-spine is not based on a new technology. The technological characteristics of the subject device is comparable to the predicate device for comparable indications for use. Thus, subject device CUVIS-spine is concluded to be substantially equivalent to the predicate device.

#### Conclusions

On the basis of the information provided in this Summary, CUREXO, INC. believes that CUVIS-spine is substantially equivalent to legally commercialized predicate devices for the purposes of this 510 (k) submission.

---

**Source:** [https://fda-staging.innolitics.com/device/K201569](https://fda-staging.innolitics.com/device/K201569)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).