K201535 · Bonesupport AB · MQV · Oct 16, 2020 · Orthopedic
Device Facts
Record ID
K201535
Device Name
Cerament Bone Void Filler
Applicant
Bonesupport AB
Product Code
MQV · Orthopedic
Decision Date
Oct 16, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3045
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
CERAMENT | BONE VOID FILLER is a ceramic bone graft substitute intended for orthopedic applications as a filler for gaps and voids that are not intrinsic to the stability of the bony structure. CERAMENT | BONE VOID FILLER is indicated to be injected, or placed, into bony voids or gaps in the skeletal system, i.e. extremities, pelvis, and posterolateral spine (only during open surgery in spine). These defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 9 years old), or osseous defects created as a result of either surgery or traumatic injury to the bone. CERAMENT | BONE VOID FILLER resorbs and is replaced by bone during the healing process. CERAMENT | BONE VOID FILLER can augment hardware and support bone fragments during the surgical procedure. The cured paste acts only as a temporary support media and is not intended to provide structural support during the healing process. CERAMENT | BONE VOID FILLER can be drilled and screws can be placed through it.
Device Story
CERAMENT BONE VOID FILLER is a ceramic bone graft substitute consisting of a powder (hydroxyapatite and calcium sulfate hemihydrate) and a liquid (iohexol for radio-opacification). Components are mixed using a combined mixing injection (CMI) device to form a viscous paste. The paste is injected or manually placed into bone defects during open or percutaneous surgery. It sets in vivo to a drillable material that supports bone fragments and hardware during procedures. It is not intended for structural support during healing; it resorbs over time and is replaced by bone. The device includes tip extenders to facilitate delivery. It is used by surgeons in clinical settings.
Clinical Evidence
No clinical data provided. Substantial equivalence is supported by bench testing and engineering analyses, including biocompatibility testing and performance testing demonstrating the material's ability to support hardware and be drilled/screwed, in accordance with the FDA Class II Special Controls Guidance for Resorbable Calcium Salt Bone Void Filler devices.
Technological Characteristics
Ceramic bone graft substitute; powder (hydroxyapatite, calcium sulfate hemihydrate) and liquid (iohexol) components. Injectable/moldable paste. Sets ex vivo or in vivo. Drillable. Radio-opaque. Includes plastic tip extenders for delivery. Biocompatible; compliant with pyrogenicity requirements for orthopedic implants.
Indications for Use
Indicated for filling non-structural bony voids or gaps in the skeletal system (extremities, pelvis, posterolateral spine via open surgery) in adults and pediatric patients ≥ 9 years old. Used for defects from benign bone cysts, tumors, surgery, or trauma. Supports hardware and bone fragments during surgery; drillable.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
Norian Drillable Inject and Norian Drillable Fast Set Putty (K102722)
Submission Summary (Full Text)
{0}------------------------------------------------
October 16, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left side of the image is the Department of Health & Human Services logo. To the right of the HHS logo is the FDA logo, which consists of the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG" in blue, with the word "ADMINISTRATION" underneath.
BoneSupport AB Blerta Shuka Regulatory Affairs Specialist Scheelevagen 19. Ideon Science Park Lund. SE 223-70 Sweden
### Re: K201535
Trade/Device Name: CERAMENT BONE VOID FILLER Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable Calcium Salt Bone Void Filler Device Regulatory Class: Class II Product Code: MQV Dated: June 8, 2020 Received: June 8, 2020
Dear Blerta Shuka:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Laura C. Rose, Ph.D. Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### Indications for Use
510(k) Number (if known) K201535
Device Name CERAMENT BONE VOID FILLER
#### Indications for Use (Describe)
CERAMENT | BONE VOID FILLER is a ceramic bone graft substitute intended for orthopedic applications as a filler for gaps and voids that are not intrinsic to the stability of the bony structure.
CERAMENT | BONE VOID FILLER is indicated to be injected, or placed, into bony voids or gaps in the skeletal system, i.e. extremities, pelvis, and posterolateral spine (only during open surgery in spine). These defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 9 years old), or osseous defects created as a result of either surgery or traumatic injury to the bone.
CERAMENT | BONE VOID FILLER resorbs and is replaced by bone during the healing process.
CERAMENT | BONE VOID FILLER can augment hardware and support bone fragments during the surgical procedure. The cured paste acts only as a temporary support media and is not intended to provide structural support during the healing process. CERAMENT | BONE VOID FILLER can be drilled and screws can be placed through it.
| Type of Use (Select one or both, as applicable) | |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
CONTINUE ON A SEPARATE PAGE IF NEEDED#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Manufacturer: | BONESUPPORT AB<br>Scheelevägen 19<br>IDEON Science Park<br>SE-223 70 LUND, SWEDEN<br>+ 46 46 286 53 70 |
|----------------|--------------------------------------------------------------------------------------------------------|
| Contact: | Blerta Shuka<br>blerta.shuka@bonesupport.com<br>+ 46 70 330 27 68 (direct phone) |
| Prepared by: | MCRA, LLC<br>1050 K Street, NW, Suite 1000<br>Washington, DC 20001<br>+1.202.552.5800 |
| Date prepared: | October 9, 2020 |
## Device Information
| Trade Name: | CERAMENT BONE VOID FILLER |
|----------------------|-------------------------------------------------|
| Common Name: | Bone void filler |
| Classification Name: | Resorbable calcium salt bone void filler device |
| Regulation: | 21 CFR 888.3045 |
| Regulatory Class: | II |
| Product Code: | MQV |
### Device Description
CERAMENT| BONE VOID FILLER is a fast-setting, injectable and moldable ceramic bone graft substitute intended for filling bone voids/gaps. The material consists of a powder and a liquid component. The major constituents of the powder are hydroxyapatite and calcium sulfate hemihydrate. The liquid component (C-TRU) contains iohexol as a radio-opacification enhancer. Mixing the components with the combined mixing injection (CMI) device, results in a viscous material intended to set ex vivo or in vivo.
By combining hydroxyapatite and calcium sulfate, an optimal balance is achieved between implant resorption rate and bone ingrowth rate. Calcium sulfate acts as a resorbable carrier for hydroxyapatite. Hydroxyapatite has a slow resorption rate, high osteoconductivity, promoting bone ingrowth and gives long term structural support to the newly formed bone.
{4}------------------------------------------------
The ceramic bone graft substitute is placed into the bone defect under direct visualization or under radiographic monitoring during open or percutaneous surgery. The paste may be injected into the defect, molded by hand and digitally placed into the defect, or used to prepare beads that are placed into the defect. The accompanying injection device (ID) and Tip Extenders may be used to facilitate filling the bone defect.
When fully set in vivo, CERAMENT | BONE VOID FILLER is drillable and can be used to augment hardware during the surgical procedure.
# Indications for Use
CERAMENT | BONE VOID FILLER is a ceramic bone graft substitute intended for orthopedic applications as a filler for gaps and voids that are not intrinsic to the stability of the bony structure.
CERAMENT | BONE VOID FILLER is indicated to be injected, or placed, into bony voids or gaps in the skeletal system, i.e. extremities, pelvis, and posterolateral spine (only during open surgery in spine). These defects may be the result of benign bone cysts and tumors (in adults and pediatric patients ≥ 9 years old), or osseous defects created as a result of either surgery or traumatic injury to the bone.
CERAMENT | BONE VOID FILLER resorbs and is replaced by bone during the healing process.
CERAMENT | BONE VOID FILLER can augment hardware and support bone fragments during the surgical procedure. The cured paste acts only as a temporary support media and is not intended to provide structural support during the healing process. CERAMENT | BONE VOID FILLER can be drilled and screws can be placed through it.
| 510(k)<br>number | Trade name | Manufacturer | Predicate |
|---------------------|----------------------------------------------------------------|------------------------------------|-------------------|
| K073316 | CERAMENT BONE VOID FILLER | Bonesupport AB | Primary predicate |
| K090871 | CERAMENT BONE VOID FILLER<br>(High Contrast) | Bonesupport AB | Reference device |
| K132656,<br>K113871 | PRO-DENSE Bone Graft Substitute | Wright Medical<br>Technology, Inc. | Reference device |
| K102722 | Norian Drillable Inject and Norian<br>Drillable Fast Set Putty | Synthes | Reference device |
## Predicate Devices
# Comparison of Technological Characteristics with the Predicate Device
The table below summarizes the differences between the technological characteristics of the primary predicate device (CERAMENT) BONE VOID FILLER, K073316) and the subject device.
{5}------------------------------------------------
| Feature | Any difference between the primary predicate device<br>and the new device |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Powder component | No difference |
| Liquid component | No difference |
| Mixing of paste | No difference |
| Delivery of paste to patient | Two plastic Tip Extenders have been added to the device.<br>They may be connected to the CERAMENT ID to<br>facilitate paste injection |
| Indications | Indications have been expanded to include use in:<br>• filling benign bone cysts and tumors in adults and<br>pediatric patients ≥ 9 years old,<br>• augmenting hardware and supporting bone fragments<br>during a surgical procedure, and<br>• ability to drill and place screws through the material. |
## Table 1. Comparison between the device and its primary predicate device (CERAMENT | BONE VOID FILLER, K073316)
# Performance Testing
The previously provided performance testing demonstrates the substantial equivalence of CERAMENT| BONE VOID FILLER. The prior testing was conducted in accordance with the Guidance for Industry and FDA Staff (Class II Special Controls Guidance Document: Resorbable calcium salt bone void filler device; Guidance for Industry and FDA (issued June 2, 2003)).
Additional testing has been conducted to demonstrate that CERAMENT | BONE VOID FILLER can be used to augment hardware and support bone fragments during a surgical procedure, and that the material can be drilled and screws placed through it.
## Biocompatibility
Previously conducted biocompatibility testing demonstrates that CERAMENT | BONE VOID FILLER is biocompatible.
## Conclusions
Based on the information provided in this premarket notification, CERAMENT| BONE VOID FILLER is substantially equivalent to the predicate device in intended use, materials, technological characteristics, principles of operation, and function. Testing and engineering analyses showed that CERAMENT | BONE VOID FILLER met the pre-determined acceptance criteria identified in the Design Control Activities Summary. Additionally, CERAMENT | BONE VOID FILLER is in compliance with pyrogenicity testing requirements for orthopedic implants.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.