Avinent Biomimetic Coral and Ocean-Iceberg Sterilization Cassettes

K201533 · Avinent Implant System S.L.U. · KCT · Mar 1, 2021 · General Hospital

Device Facts

Record IDK201533
Device NameAvinent Biomimetic Coral and Ocean-Iceberg Sterilization Cassettes
ApplicantAvinent Implant System S.L.U.
Product CodeKCT · General Hospital
Decision DateMar 1, 2021
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 880.6850
Device ClassClass 2

Indications for Use

Avinent Sterilization Cassettes are designed to hold various dental surgical drills and tools in order to organize, steam sterilize, and transport the instruments between uses. The tray is to be enclosed in an FDA cleared steam sterilizable wrap and sterilized in an FDA cleared sterilizer for the following cycle: Prevaccum Steam - At 132° for 4 minutes with a 30 minute dry time. -The trays are intended for sterilization of non-porous loads. -The trays are recommended not to be stacked during sterilization. -The Biomimetic Coral Sterilization Cassette Guided Surgery represents the worst case validated load due to the number of components (32 instruments) and the weight of 599.6 grams. -Avinent Implant System SLU does not make any lumen claims for the Avinent Sterilization Cassettes.

Device Story

Reusable rigid sterilization container/organizing tray; holds dental surgical drills and tools. Composed of Radel R-5000 polymer resin base, lid, and internal insert tray; features medical-grade silicone grommets/cradles to secure instruments. Perforated design allows steam penetration; requires FDA-cleared sterilization wrap. Used in healthcare facilities; operated by clinical staff. Protects instruments during transport, steam sterilization, and storage. Validated for prevacuum steam sterilization (132°C, 4 min, 30 min dry). Ensures instrument organization and sterility maintenance.

Clinical Evidence

Bench testing only. Validated cleaning cycle (no visible soil, hemoglobin <2.2µg/cm², protein ≤6.4µg/cm²). Steam sterilization validation achieved 10-6 SAL. Durability testing confirmed functionality over 100 sterilization cycles. Cytotoxicity testing (ISO 10993-5) confirmed non-cytotoxic results (score 0) for instruments processed in the tray.

Technological Characteristics

Rigid, multi-piece perforated tray. Materials: Radel R-5000 polymer resin (lid, base, insert) and medical-grade silicone (grommets/cradles). Dimensions: 207.5 x 157.5 x 71.7mm. Sterilization: Prevacuum steam (132°C). Vent-to-volume ratio: 0.011 in²/in³. Reusable; non-stackable.

Indications for Use

Indicated for use in healthcare facilities to organize, steam sterilize, and transport non-porous dental surgical drills and tools. Not for use with lumen-containing devices. Not to be stacked during sterilization.

Regulatory Classification

Identification

A sterilization wrap (pack, sterilization wrapper, bag, or accessories, is a device intended to be used to enclose another medical device that is to be sterilized by a health care provider. It is intended to allow sterilization of the enclosed medical device and also to maintain sterility of the enclosed device until used.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: a symbol on the left and the FDA name on the right. The symbol on the left is a stylized image of a human figure, while the FDA name on the right is written in blue letters. The words "U.S. FOOD & DRUG ADMINISTRATION" are written in a clear, sans-serif font. March 1, 2021 AVINENT Implant System S.L.U. % Angela Blackwell Senior Consultant Blackwell Device Consulting P.O. Box 718 Gresham, Oregon 97030 Re: K201533 Trade/Device Name: Avinent Biomimetic Coral and Ocean-Iceberg Sterilization Cassettes Regulation Number: 21 CFR 880.6850 Regulation Name: Sterilization Wrap Regulatory Class: Class II Product Code: KCT Dated: February 01, 2021 Received: February 05, 2021 Dear Angela Blackwell: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal {1}------------------------------------------------ statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems. For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100). Sincerely, Clarence W. Murray III, Ph.D. Acting Assistant Director DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health Enclosure {2}------------------------------------------------ # Indications for Use #### 510(k) Number (if known) K201533 #### Device Name Avinent Biomimetic Coral and Ocean-Iceberg Sterilization Cassettes Indications for Use (Describe) Avinent Sterilization Cassettes are designed to hold various dental surgical drills and tools in order to organize, steam sterilize, and transport the instruments between uses. The trav is to be enclosed in an FDA cleared steam sterilizable wrap and sterilized in an FDA cleared sterilizer for the following cvcle: Prevaccum Steam - At 1329 for 4 minutes with a 30 minute dry time. -The trays are intended for sterilization of non-porous loads. -The trays are recommended not to be stacked during sterilization. -The Biomimetic Coral Sterilization Cassette Guided Surgery represents the worst case validated load due to the number of components (32 instruments) and the weight of 599.6 grams. -Avinent Implant System SLU does not make any lumen claims for the Avinent Sterilization Cassettes. - The tray will be marketed in the following variations, | Device Model Name Biomimetic Coral Sterilization Cassette | 30 | Weight of tray full (g) 593.4 | Weight of tray empty (g) 510.0 | |-----------------------------------------------------------|----|-------------------------------|--------------------------------| | Biomimetic Ocean-Iceberg Sterilization Cassette | 26 | 567.8 | 510.0 | Type of Use (Select one or both, as applicable) | Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## CONTINUE ON A SEPARATE PAGE IF NEEDED. This section applies only to requirements of the Paperwork Reduction Act of 1995. #### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number." {3}------------------------------------------------ ### Avinent Biomimetic Coral and Ocean-Iceberg Sterilization Cassettes ### AVINENT Implant System S.L.U. ### Pol. Ind. Santa Anna I - 08251 Santpedor Barcelona, Spain #### +34 902 38 38 48 ### 510k Summary – K201533 ### February 25, 2021 Submission Contact: Angela Blackwell Blackwell Device Consulting P.O. Box 718 Gresham, OR 97030-0172 (704) 450-9934 angela@blackwelldevice.com Classification Name: Sterilization Wrap Containers, Trays, Cassettes, and other accessories Common Name: Instrument Sterilization Tray Device Trade Name: Avinent Biomimetic Coral and Ocean-Iceberg Sterilization Cassettes Regulation Number: 21 CFR 880.6850 Product Code: KCT Class: Il Predicate Device: Avinent Surgical Tray K191566 Coral and Ocean Guided Surgery Sterilization Cassettes ### Device Description: The Avinent Sterilization Cassettes are a reusable rigid sterilization container or organizing tray intended for use in health care facilities for the purpose of containing reusable medical devices for sterilization. It is composed of multiple pieces, designed to be integrated into a single unit which contains and protects the interior components during sterilization. Each tray consists of three components – a base tray, a lid, and an internal individualized insert tray. All of the three components are perforated for steam penetration. The internal tray has the ability to hold individualized pieces and accessories which include dental tools, drills, and torque wrench. The tray is available in one size, 207.5 x 157.5 x 71.7mm. The tray comes in two variations, Coral and Ocean. The trays in this submission are Biomimetic Coral and Ocean-Iceberg Sterilization Cassettes. The rigid multi-piece tray holds dental device apparatus and accessories before, during and after the sterilization process. The tray set has a locking lid to contain the products. The trays are designed to fit any standard autoclave, which allows them to be effective for sterilization and be able to withstand the environment of repeated steam sterilization cycles in normal operating room. Since the trays are {4}------------------------------------------------ perforated, an FDA cleared wrap must be used for sterilization purposes to maintain the sterility of contents. The trays are reusable and the tray material allows repeated sterilization cycles. The lid, base, and insert are made of Radel R-5000. This material is a polymer resin produced by Solvay Advanced Polymers, LLC and is identical to the Radel used in predicate device cleared under K191566. Radel R 5000 CL 301 is used for the lid, Radel R-5100 BK # 937 is used for the base and Radel R-5100 GY #1037 is used for the insert tray where all the drills and tools are mounted. The small circular brackets (grommets) throughout the insert tray which are used to contain drills and the tool holder with cradle on the base tray are made of medical grade silicone material which has been manufactured to meet FDA 21 CFR 177.2600. These brackets and holders are used to secure the instruments during transport, sterilization, and storage. Although these trays are reusable they will not be serviced or repaired. The instruments to be sterilized in the proposed trays are all non-porous devices and included dental surgical drills and tools. The tools and drills are all class I exempt. #### Indications for Use: Avinent Sterilization Cassettes are designed to hold various dental surgical drills and tools in order to organize, steam sterilize, and transport the instruments between uses. The tray is to be enclosed in an FDA cleared steam sterilizable wrap and sterilized in an FDA cleared sterilizer for the following cycle: Prevaccum Steam - At 132° for 4 minutes with a 30 minute dry time. -The trays are intended for sterilization of non-porous loads. -The trays are recommended not to be stacked during sterilization. -The Biomimetic Coral Sterilization Cassette Guided Surgery represents the worst case validated load due to the number of components (32 instruments) and the weight of 599.6 grams. -Avinent Implant System SLU does not make any lument Sterilization Cassettes. -The tray will be marketed in the following variations. | DEVICE MODEL NAME | MAX no. of<br>INSTRUMENTS | WEIGHT of each tray<br>FULL (g) | WEIGHT of each tray<br>EMPTY (g) | |----------------------------------------------------|---------------------------|---------------------------------|----------------------------------| | Biomimetic Coral Sterilization<br>Cassette | 30 | 593.4 | 510.0 | | Biomimetic Ocean-Iceberg<br>Sterilization Cassette | 26 | 567.8 | 510.0 | #### Comparison of Technological Characteristics: The table below shows that the Avinent Biomimetic Coral and Ocean-Iceberg Sterilization Cassettes compared to the predicate device the Avinent Surgical Tray (Coral and Ocean Guided Surgery Variations). {5}------------------------------------------------ | Device | Subject Device: Avinent<br>Biomimetic Coral and<br>Ocean-Iceberg<br>Sterilization Cassettes<br>(K201533) | Predicate Device:<br>Avinent Surgical Tray<br>Coral and Ocean Guided<br>Surgery Sterilization<br>Cassettes (K191566) | Comparison | |---------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------| | Product Code | KCT | KCT | Same | | Indications for Use | Avinent Sterilization<br>Cassettes are designed<br>to hold various dental<br>surgical drills and tools<br>in order to organize,<br>steam sterilize, and<br>transport the<br>instruments between<br>uses. The tray is to be<br>enclosed in an FDA<br>cleared steam<br>sterilizable wrap and<br>sterilized in an FDA<br>cleared sterilizer for the<br>following cycle:<br>Prevaccum Steam - At<br>132° for 4 minutes with<br>a 30 minute dry time.<br>-The trays are intended<br>for sterilization of non-<br>porous loads.<br>-The trays are<br>recommended not to be<br>stacked during<br>sterilization.<br>-The Biomimetic Coral<br>Sterilization Cassette<br>Guided Surgery<br>represents the worst<br>case validated load due<br>to the number of<br>components (32<br>instruments) and the<br>weight of 599.6 grams.<br>-Avinent Implant System<br>SLU does not make any<br>lumen claims for the<br>Avinent Sterilization<br>Cassettes. | The Avinent Surgical<br>Tray is designed to hold<br>various dental surgical<br>drills and tools in order<br>to organize, steam<br>sterilize, and transport<br>the instruments<br>between uses. The tray<br>is to be enclosed in an<br>FDA cleared steam<br>sterilizable wrap and<br>sterilized in an FDA<br>cleared sterilizer for the<br>following cycle:<br>Prevaccum Steam - At<br>132° for 4 minutes with<br>a 30 minute dry time.<br>-The trays are intended<br>for sterilization of non-<br>porous loads.<br>-The trays are<br>recommended not to be<br>stacked during<br>sterilization.<br>-The Biomimetic Coral<br>Sterilization Cassette<br>Guided Surgery<br>represents the worst<br>case validated load due<br>to the number of<br>components (32<br>instruments) and the<br>weight of 599.6 grams.<br>-Avinent Implant System<br>SLU does not make any<br>lumen claims for the<br>Avinent Surgical Tray. | Same other than<br>the name of the<br>device. | | Material Composition | Polymer Resin Radel-<br>5000, biomedical grade<br>silicone | Polymer Resin Radel-<br>5000, biomedical grade<br>silicone | Same | | Design | Plastic tray with locking<br>lid and silicone<br>containment brackets. | Plastic tray with locking<br>lid and silicone<br>containment brackets. | Same | | Dimensions | 207.5 x 157.5 x 71.7mm | 207.5 x 157.5 x 71.7mm | Same | | Sterilization by<br>1 Gravity Steam | No | No | Same | | 2 Pre Vaccuum Steam | Yes | Yes | Same | | Air Permeance | Yes | Yes | Same | | Vent to Volume Ratio | 0.011 in²/in³ | 0.011 in²/in³ | Same | | Locking System to hold<br>lid in place | Yes | Yes | Same | | Reusable | Yes | Yes | Same | | Stackable | No | No | Same | | Material compatibility<br>with sterilization process | Yes | Yes | Same | | Sterilant Penetration<br>Studies | Yes | Yes | Same | | Biocompatibility | Provided by material<br>suppliers.<br>Cytotoxicity tests<br>conducted according to<br>ISO 10993-5 and -12. | Provided by material<br>suppliers.<br>Cytotoxicity tests<br>conducted according to<br>ISO 10993-5 and -12. | Same | | Cleaning Instructions for<br>Reusable Devices<br>Validated | Yes | Yes | Same | | Repeat Sterilization to<br>show material<br>compatibility over time | Yes | Yes | Same | | Drying Time | Yes | Yes | Same | | Method to Preserve<br>Sterility of the Tray | FDA cleared wrap | FDA cleared wrap | Same | | Reference Numbers<br>referring to tray contents | Coral 0376<br>Ocean-Iceberg 2725 | Coral G.S. 3003<br>Ocean G.S. 3005 | Different because<br>two reference<br>numbers were<br>cleared in<br>K191566 and two<br>additional ones<br>are the subject of<br>this 510k<br>(K201533) | {6}------------------------------------------------ {7}------------------------------------------------ ### Summary of Non-Clinical Testing: | Name of Testing | Purpose of Testing | Acceptance Criteria | Results | |-----------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------| | Cleaning Validation | Validate the cleaning<br>cycle in the instructions<br>for use | No visible soil.<br>Hemoglobin less than<br>$2.2µg/cm²$ . Protein<br>level $≤6.4 µg/cm²$ .<br>MEM reactivity grade<br>of 2 or less. | No visible soil. All<br>acceptance criteria<br>met.<br>$<6.4 µg/cm²$ protein<br>and $< 2.2 µg/cm²$<br>hemoglobin on device<br>after cleaning<br>PASS | | Steam Sterilization<br>Validation | Validate the half-cycle<br>of the cycle in the<br>instructions for use | No growth on half<br>cycles. No moisture on<br>full drying cycle. Pass<br>on Chemical Indicators<br>half and full cycle. | $10^{-6}$ SAL<br>Pass on chemical<br>indicators half and full<br>cycle. No moisture on<br>full drying cycle. No<br>growth on half cycle. | | 100 cycles sterilization | Show the tray will last<br>at least 100<br>sterilization cycles | Pass visual inspection<br>and test of<br>functionality. | At all cycle counts<br>passed visual<br>inspection and test of<br>functionality. | | Cytotoxicity | Sterilized instruments<br>were tested to<br>demonstrate their<br>were no residuals on<br>them which would<br>change their<br>biocompatibility. | Pass is a score of no<br>greater than 2 | Pass with score of 0<br>No evidence of lysis -<br>Non-Cytotoxic | Cleaning validation using a manual method was done following simulated use according to the cleaning method in the instructions for use. Six simulated use cycles contamination, cleaning, and sterilization were performed prior to validation. Four articles were used. The MEM article, negative article and negative device control were not simulated use articles. The positive was treated as a test article during the simulated use cycles. The test soil was defibrinated blood soil. The residuals tested were hemoglobin, protein, and minimal essential media. Steam sterilization validation (half cycle) based on the cycle listed in the indications for use was completed. Each test article was evaluated to a sterility assurance level of ≤10° using the biological indicator overkill method. Dry times were validated using full cycle parameters. A second steam sterilization validation based on this cycle was done full cycle with thermal profiling. Sterilization of the tray was done to verify the tray safely last for 100 cycles. No changes were noted in the test articles after 100 cycles. A visual inspection, photographs, and a test of functionality were done every 10 cycles. Test articles were rewrapped after every 10 cycles. Biocompatibility information was provided on all the materials used in the cassettes. Cytotoxicity (ISO 10993-5) of the instruments placed in the tray for sterilization was done after a sterilization cycle was {8}------------------------------------------------ completed to show there were no residuals from the tray on the instruments which would change their biocompatibility. Conclusion: The conclusions drawn from the non-clinical tests demonstrate the subject device is as safe, as effective, and performs as well as or better than the legally marketed predicate devices.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...