← Product Code [GCJ](/productcode/GCJ) · K201526

# TIPCAM1 Rubina Video Endoscope System (K201526)

_Karl Storz Endoscopy America · GCJ · Nov 20, 2020 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K201526

## Device Facts

- **Applicant:** Karl Storz Endoscopy America
- **Product Code:** [GCJ](/productcode/GCJ.md)
- **Decision Date:** Nov 20, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The TIPCAM®1 Rubina Video Endoscope System is intended to be used together with the camera control unit during diagnostic and/or surgical procedures when endoscopic video assistance is required. For use in all endoscopy and endoscopic surgery within the peritoneal and thoracic cavity, including gynecological anatomy.

## Device Story

Integrated video endoscope system; includes camera and endoscope. Illumination light from external source travels via optical fiber bundle to distal end; objective lens receives reflected light from patient lumen. Dual CMOS sensors convert light to electrical signals; transmitted to camera control unit (Image1 S Connect II + 4U-Link) for processing and display. Used in OR by physicians for minimally invasive surgery. Features 2D auto-leveling (autorotation) to maintain fixed horizon and 2D auto-switch mode to transition from 3D to 2D when rotated outside 3D range. Provides real-time visualization of surgical field; assists surgeon in navigation and procedure performance; benefits patient through minimally invasive access.

## Clinical Evidence

No clinical data. Substantial equivalence supported by non-clinical bench testing, including electrical/mechanical/thermal safety (IEC 60601-1, 60601-2-18), EMC (IEC 60601-1-2), photobiological safety (IEC 62471), and performance verification (color reproduction, distortion, depth of field, SNR, latency, spatial resolution, FOV per ISO 8600-3). Biocompatibility and sterilization validation leveraged from predicate K150525.

## Technological Characteristics

Integrated video endoscope with dual CMOS sensors. Features 0° and 30° direction-of-view models. Connectivity via cable to camera control unit. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-2-18, IEC 62471, ISO 8600-3. Software-controlled image processing for auto-leveling and 3D-to-2D switching.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- SPIES 3D System ([K150525](/device/K150525.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

November 20, 2020

Karl Storz Endoscopy-America, Inc. Mario Trujillo Associate Regulatory Affairs Specialist 2151 E. Grand Avenue El Segundo, CA 90245

K201526 Re:

> Trade/Device Name: TIPCAM®1 Rubina Video Endoscope System Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and Accessories Regulatory Class: II Product Code: GCJ, FGB, HET Dated: October 21, 2020 Received: October 23, 2020

Dear Mario Trujillo:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{1}------------------------------------------------

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatoryinformation/postmarketing-safety-reporting-combination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Jason R. Roberts, Ph.D. Assistant Director DHT3B: Division of Reproductive. Gynecology and Urology Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K201526

Device Name TIPCAM®1 Rubina Video Endoscope System

#### Indications for Use (Describe)

The TIPCAM®1 Rubina Video Endoscope System is intended to be used together with the camera control unit during diagnostic and/or surgical procedures when endoscopic video assistance is required. For use in all endoscopy and endoscopic surgery within the peritoneal and thoracic cavity, including gynecological anatomy.

| Type of Use (Select one or both, as applicable)                                                          |                                                                                                         |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
| <span style="font-family: DejaVu Sans, sans-serif">✘</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">❍</span> Over-The-Counter Use (21 CFR 801 Subpart C) |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image is a logo for Karl Storz Endoskope. The word "STORZ" is in large, bold, blue letters. Below that, the words "KARL STORZ - ENDOSKOPE" are in smaller, blue letters.

K201526

# 510(k) Summary

This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Devices Act (SMDA) of 1990 and 21 CFR 807.92. All data included in this document is accurate and complete to the best of KARL STORZ Endoscopy-America's knowledge.

| Applicant:                        | KARL STORZ Endoscopy-America, Inc.<br>2151 E. Grand Avenue<br>El Segundo, CA 90245                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|-----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact:                          | Mario Trujillo<br>Associate Regulatory Affairs Specialist<br>Tel.: (424) 218-8481<br>Email: Mario.Trujillo@karlstorz.com                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Date of<br>Preparation:           | November 19, 2020                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Type of 510(k)<br>Submission:     | Traditional                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Device<br>Identification:         | Trade Name: TIPCAM®1 Rubina Video Endoscope System<br>Regulation Name: Endoscope and Accessories                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Regulatory Class:                 | 2                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Product Code:                     | GCJ, FGB, HET                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Regulation<br>Number:             | 21 CFR 876.1500                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |
| Predicate<br>Device(s):           | SPIES 3D System (K150525)<br>This predicate device has not been subject to a design-related recall.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Device<br>Description:            | This predicate device has not been subject to a design-related recall.<br>The TIPCAM®1 Rubina video endoscope is an integrated unit that<br>includes a camera and an endoscope. The endoscope receives the<br>illumination light from the light source by the light guide connector<br>connected to the light source device. The illumination light is<br>transferred to the distal end through the optical fiber bundle inside of<br>the endoscope and illuminates the inside of the patient body through<br>the illumination lens at the distal end. The endoscope receives the<br>reflected light from the inner lumen of a patient by the objective lens at<br>the distal end. The built-in dual CMOS sensors convert the light to an<br>electrical signal, and the signal is sent to the camera control unit<br>(Image1 S Connect II + 4U-Link) via the attached cable for processing<br>and display.                                                                                                                                                                                  |
| Intended Use:                     | TIPCAM®1 Rubina videoscopes and corresponding accessories<br>(camera control unit, light sources, monitor) are used for endoscopic<br>imaging of the surgical field during diagnostic and/or surgical<br>procedures in the abdominal and thoracic cavity.<br>TIPCAM®1 Rubina videoscopes are designed for short-term use in<br>surgically invasive procedures                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Indications For<br>Use:           | The TIPCAM®1 Rubina Video Endoscope System is intended to be<br>used together with the camera control unit during diagnostic and/or<br>surgical procedures when endoscopic video assistance is required.<br>For use in all endoscopy and endoscopic surgery within the peritoneal<br>and thoracic cavity, including gynecological and urological anatomy.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Technological<br>Characteristics: | The predicate and subject devices are both camera systems that are<br>used for observation purposes in general endoscopic surgery within<br>the thoracic and peritoneal cavity, including gynecological and<br>urological anatomy. There are some minor differences in the<br>technological characteristics. These differences are:<br>The subject device has two image viewing features: A 2D<br>auto-leveling (autorotation) mode which allows the user to<br>maintain fixed horizon while rotating the endoscope and a 2D<br>Auto-switch mode that automatically switches the image from<br>3D mode to 2D Auto-leveling mode when the scope is rotated<br>outside the 3D viewing range to prevent the user from losing<br>the perspective of the image. The subject device is also available in a 30° direction of view<br>model in addition to the 0° model. The total length of the subject endoscope is 5mm longer<br>than the predicate          These differences do not raise different questions of safety and<br>effectiveness and can be evaluated through performance testing. |
| Non-Clinical<br>Performance Data: | The TIPCAM®1 Rubina System was tested for<br>electrical/mechanical/thermal safety and EMC according to the<br>following standards:<br>IEC 60601-1 2005 + CORR. 1 (2006) + CORR. 2 (2007) + AM1<br>(2012) IEC 60601-1-2:2014 (4th Edition) IEC 60601-2-18:2009 IEC 62471:2009          The following additional bench testing for performance verification<br>and validation purposes was performed:  Photobiological Safety per IEC 62471:2006 Color Reproduction and Color Contrast Enhancement Distortion Depth of Field Detection Linearity Illumination Detection Uniformity Signal-to-Noise Ratio (SNR) & Sensitivity Latency Spatial Resolution Dynamic Range Direction of View per ISO 8600-3:2019                                                                                                                                                                                                                                                                                                                                                                                   |
|                                   | • Field of View per ISO 8600-3:2019<br>• System Horizontal Parallax<br>The subject device software was validated per the FDA guidance<br>"Guidance for the Content of Premarket Submissions for Software<br>Contained in Medical Devices" as a moderate level of concern.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|                                   | The subject device materials and manufacturing are identical to that of<br>the predicate, therefore, biocompatibility testing was leveraged from<br>K150525 to support the biocompatibility of the subject device.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                   | Cleaning and sterilization validation of the subject device was<br>leveraged from the predicate device K150525, as the subject device is<br>substantially similar in specifications to that of the predicate, and the<br>sterilization process remains identical to the predicate.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                   | The bench testing noted above confirmed that the TIPCAM®1 Rubina<br>system has met all its design specification and is as safe and effective<br>as the predicate device, SPIES 3D System.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Clinical<br>Performance Data:     | Clinical testing was not required to demonstrate the substantial<br>equivalence to the predicate device. Non-clinical bench testing was<br>sufficient to assess safety and effectiveness and to establish the<br>substantial equivalence of the modifications.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Conclusion:                       | The subject device has the same intended use as the predicate. The<br>technological differences between the subject and predicate do not<br>raise different questions of safety and effectiveness. The conclusions<br>drawn from the non-clinical performance data demonstrated that the<br>subject device is as safe as and as effective as the predicate device.<br>Therefore, the subject device is substantially equivalent to the<br>predicate.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for Karl Storz Endoskope. The logo is blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the phrase "KARL STORZ - ENDOSKOPE" in a smaller font.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for KARL STORZ. The logo is in blue and features the word "STORZ" in large, bold letters. Below the word "STORZ" is the text "KARL STORZ - ENDOSKOPE" in a smaller font.

---

**Source:** [https://fda-staging.innolitics.com/device/K201526](https://fda-staging.innolitics.com/device/K201526)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
