K201525 · Apple, Inc. · QDA · Oct 8, 2020 · Cardiovascular
Device Facts
Record ID
K201525
Device Name
ECG App
Applicant
Apple, Inc.
Product Code
QDA · Cardiovascular
Decision Date
Oct 8, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2345
Device Class
Class 2
Attributes
Software as a Medical Device
AI Performance
Output
Algorithm
Acceptance
Observed
Dev DS
Dev Readers
Test DS
Test Readers
Atrial Fibrillation Detection
—
—
98.5% sensitivity
—
—
Pivotal, prospective, multi-center clinical trial of 546 subjects
1 (cardiologist)
Sinus Rhythm Detection
—
—
99.3% specificity
—
—
Pivotal, prospective, multi-center clinical trial of 546 subjects
1 (cardiologist)
Indications for Use
The ECG app is a software-only mobile medical application intended for use with the Apple Watch to create, record, store, transfer, and display a single channel electrocardiogram (ECG) similar to a Lead I ECG. The ECG app determines the presence of atrial fibrillation (AFib), sinus rhythm, and high heart rate (no detected AF with heart rate 100-150 bpm) on a classifiable waveform. The ECG app is not recommended for users with other known arrhythmias. The ECG app is intended for over-the-counter (OTC) use. The ECG data displayed by the ECG app is intended for informational use only. The user is not intended to interpret or take clinical action based on the device output without consultation of a qualified healthcare professional. The ECG waveform is meant to supplement rhythm classification for the purposes of discriminating AFib from sinus rhythm and is not intended to replace traditional methods of diagnosis or treatment. The ECG app is not intended for use by people under 22 years old.
Device Story
Software-only mobile application; operates on Apple Watch and iPhone. User initiates 30-second recording by wearing Watch on one wrist and touching digital crown with opposite hand; completes electrical circuit. Device performs signal processing, feature extraction, and rhythm classification. Outputs include ECG waveform, average heart rate, and rhythm classification (AFib, sinus rhythm, high heart rate, inconclusive, or poor recording). Used in home/OTC setting by patients. Data displayed on Watch/iPhone; intended to supplement rhythm classification. Does not replace traditional diagnostic methods; requires physician consultation for clinical action.
Clinical Evidence
Prospective, multi-center clinical trial (n=546; 305 AFib, 241 sinus rhythm; mean age 58). Compared app-generated ECG to cardiologist-interpreted 12-lead ECG. Results: 98.5% sensitivity for AFib (HR 50-150 bpm) and 99.3% specificity for sinus rhythm (HR 50-150 bpm). Waveform morphology assessment showed 100% pass rating for PQRST and 97.2% for R-wave amplitude compared to reference.
Technological Characteristics
Software-only mobile application. Uses integrated electrical sensors on Apple Watch for signal acquisition. Connectivity via iOS/WatchOS platforms. Complies with IEC-60950 (thermal), EN 301-489-17, and FCC Part 15 (RF/EMC). Software level of concern: moderate.
Indications for Use
Indicated for adults 22+ years old to record, store, and display single-channel ECG (Lead I) via Apple Watch. Determines presence of AFib, sinus rhythm, or high heart rate (100-150 bpm). Not for users with known arrhythmias other than AFib. OTC use; informational only; not for diagnosis or clinical action without professional consultation.
Regulatory Classification
Identification
An electrocardiograph software device for over-the-counter use creates, analyzes, and displays electrocardiograph data and can provide information for identifying cardiac arrhythmias. This device is not intended to provide a diagnosis.
Special Controls
In combination with the general controls of the FD&C Act, the electrocardiograph software for over-the-counter use is subject to the following special controls:
*Classification.* Class II (special controls). The special controls for this device are:(1) Clinical performance testing under anticipated conditions of use must demonstrate the following:
(i) The ability to obtain an electrocardiograph of sufficient quality for display and analysis; and
(ii) The performance characteristics of the detection algorithm as reported by sensitivity and either specificity or positive predictive value.
(2) Software verification, validation, and hazard analysis must be performed. Documentation must include a characterization of the technical specifications of the software, including the detection algorithm and its inputs and outputs.
(3) Non-clinical performance testing must validate detection algorithm performance using a previously adjudicated data set.
(4) Human factors and usability testing must demonstrate the following:
(i) The user can correctly use the device based solely on reading the device labeling; and
(ii) The user can correctly interpret the device output and understand when to seek medical care.
(5) Labeling must include:
(i) Hardware platform and operating system requirements;
(ii) Situations in which the device may not operate at an expected performance level;
(iii) A summary of the clinical performance testing conducted with the device;
(iv) A description of what the device measures and outputs to the user; and
(v) Guidance on interpretation of any results.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
October 8, 2020
Apple Inc. Luke Olson Regulatory Affairs Associate One Apple Park Way Cupertino, California 94015
Re: K201525
Trade/Device Name: ECG 2.0 App Regulation Number: 21 CFR 870.2345 Regulation Name: Electrocardiograph Software for Over-The-Counter Use Regulatory Class: Class II Product Code: QDA Dated: September 8, 2020 Received: September 9, 2020
Dear Luke Olson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
for
Jennifer Shih Kozen Assistant Director DHT2A: Division of Cardiac Electrophysiology, Diagnostics and Monitoring Devices OHT2: Office of Cardiovascular Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201525
Device Name ECG 2.0 App
#### Indications for Use (Describe)
The ECG app is a software-only mobile medical application intended for use with the Apple Watch to create, record, store, transfer, and display a single channel electrocardiogram (ECG) similar to a Lead I ECG. The ECG app determines the presence of atrial fibrillation (AFib), sinus rhythm, and high heart rate (no detected AF with heart rate 100-150 bpm) on a classifiable waveform. The ECG app is not recommended for users with other known arrhythmias.
The ECG app is intended for over-the-counter (OTC) use. The ECG data displayed by the ECG app is intended for informational use only. The user is not intended to interpret or take clinical action based on the device output without consultation of a qualified healthcare professional. The ECG waveform is meant to supplement rhythm classification for the purposes of discriminating AFib from sinus rhythm and is not intended to replace traditional methods of diagnosis or treatment.
The ECG app is not intended for use by people under 22 years old.
| Type of Use (Select one or both, as applicable) | |
|--------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| <label><input type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input checked="" type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Section 5
### 510(k) Summary
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of 21 CFR §807.92:
# 5.1 Submitter
| Applicant | Apple Inc.<br>One Apple Park Way<br>Cupertino, CA 95014 |
|-----------------------------|-------------------------------------------------------------------------------------------------------------|
| Submission<br>Correspondent | Luke Olson<br>Regulatory Affairs Associate<br>Phone: (408) 608-2001<br>Email: luke_olson@apple.com |
| Contact | David Amor<br>Sr Manager, Quality and Regulatory Affairs<br>Phone: (786) 546-1806<br>Email: damor@apple.com |
| Date Prepared | October 6, 2020 |
### 5.2 Device Names and Classifications
# Subject Device:
| Name of Device | ECG App (2.0) |
|---------------------|-----------------------------------------------------------------------|
| Classification Name | Electrocardiograph Software For Over-The-Counter Use, 21 CFR 870.2345 |
| Regulatory Class | Class II |
| Product Code | QDA |
| 510(k) Review Panel | Cardiovascular |
#### Predicate Device:
| Predicate Manufacturer | Apple Inc. |
|------------------------|------------|
| Predicate Trade Name | ECG App |
| Predicate 510(k) | DEN180044 |
### 5.3 Device Description
The ECG 2.0 app comprises a pair of mobile medical apps - one on Apple Watch and the other on the iPhone.
{4}------------------------------------------------
The ECG Watch app analyzes data collected by the integrated electrical sensors on a compatible Apple Watch to generate an ECG waveform similar to a Lead I. calculate average heart rate, and provide a rhythm classification to the user for a given 30 second session. When a user opens the ECG Watch app while wearing the Watch on one wrist, and places the finger of the opposite hand on the digital crown, they are completing the circuit across the heart which begins a recording session.
Once the recording session is complete, the ECG Watch app performs signal processing, feature extraction and rhythm classification to generate a session result.
The resulting classification and average heart rate for the session, along with educational information, will be displayed to the user within the ECG Watch app.
The ECG iPhone app contains the on-boarding and educational materials that a user must review prior to taking an ECG reading. The ECG iPhone app is included in the Health App, which allows users to store, manage, and share health and fitness data, and comes pre-installed on every iPhone. The ECG 2.0 app expands the classifiable heart range, introduces new classification results, and introduces minor, non-userfacing algorithm updates. These changes will be reflected in both the Apple Watch app, and also on the corresponding iPhone app within the Health App.
# 5.4 Indications for Use
The ECG app is a software-only mobile medical application intended for use with the Apple Watch to create, record, store, transfer, and display a single channel electrocardiogram (ECG) similar to a Lead I ECG. The ECG app determines the presence of atrial fibrillation (AFib), sinus rhythm, and high heart rate (no detected AF with heart rate 100-150 bpm) on a classifiable waveform. The ECG app is not recommended for users with other known arrhythmias.
The ECG app is intended for over-the-counter (OTC) use. The ECG data displayed by the ECG app is intended for informational use only. The user is not intended to interpret or take clinical action based on the device output without consultation of a qualified healthcare professional. The ECG waveform is meant to supplement rhythm classification for the purposes of discriminating AFib from sinus rhythm and is not intended to replace traditional methods of diagnosis or treatment.
The ECG app is not intended for use by people under 22 years old.
### 5.5 Comparison with the Predicate Device
| Item | Subject Device | Predicate Device |
|---------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | ECG 2.0 App | ECG App |
| Manufacturer | Apple Inc. | Apple Inc. |
| Submission<br>Reference | K201525 | DEN180044 |
| Item | Subject Device | Predicate Device |
| | ECG 2.0 App | ECG App |
| Intended Use | An electrocardiograph software device<br>for over-the-counter use creates,<br>analyzes, and displays<br>electrocardiograph data, and can<br>provide information for identifying<br>cardiac arrhythmias. This device is not<br>intended to provide a diagnosis. | An electrocardiograph software device for<br>over-the-counter use creates, analyzes, and<br>displays electrocardiograph data, and can<br>provide information for identifying cardiac<br>arrhythmias. This device is not intended to<br>provide a diagnosis. |
| Indications for<br>Use | The ECG app is a software-only mobile<br>medical application intended for use<br>with the Apple Watch to create, record,<br>store, transfer, and display a single<br>channel electrocardiogram (ECG)<br>similar to a Lead I ECG. The ECG app<br>determines the presence of atrial<br>fibrillation (AFib), sinus rhythm, and<br>high heart rate (no detected AF with<br>heart rate 100-150 bpm) on a<br>classifiable waveform. The ECG app is<br>not recommended for users with other<br>known arrhythmias.<br>The ECG app is intended for over-the-<br>counter (OTC) use. The ECG data<br>displayed by the ECG app is intended<br>for informational use only. The user is<br>not intended to interpret or take clinical<br>action based on the device output<br>without consultation of a qualified<br>healthcare professional. The ECG<br>waveform is meant to supplement<br>rhythm classification for the purposes<br>of discriminating AFib from sinus<br>rhythm and is not intended to replace<br>traditional methods of diagnosis or<br>treatment.<br>The ECG app is not intended for use by<br>people under 22 years old. | The ECG app is a software-only mobile<br>medical application intended for use with<br>the Apple Watch to create, record, store,<br>transfer and display a single channel<br>electrocardiogram (ECG) similar to a Lead<br>I ECG. The ECG app determines the<br>presence of atrial fibrillation (AF) or sinus<br>rhythm on a classifiable waveform. The<br>ECG app is not recommended for users<br>with other known arrhythmias.<br>The ECG app is intended for over-the-<br>counter (OTC) use. The ECG data<br>displayed by the ECG app is intended for<br>informational use only. The user is not<br>intended to interpret or take clinical action<br>based on the device output without<br>consultation of a qualified healthcare<br>professional. The ECG waveform is meant<br>to supplement rhythm classification for the<br>purposes of discriminating AF from normal<br>sinus rhythm and not intended to replace<br>traditional methods of diagnosis or<br>treatment.<br>The ECG app is not intended for use by<br>people under 22 years old. |
| Principle of<br>Operation | The ECG 2.0 app acquires platform<br>sensor data from Apple Watch. After<br>acquisition, the ECG 2.0 app<br>algorithms process and classify the<br>signal and display the classification to<br>the user. | The ECG app acquires platform sensor<br>data from Apple Watch. After acquisition,<br>the ECG app algorithms process and<br>classify the signal and display the<br>classification to the user. |
| Item | Subject Device | Predicate Device |
| | ECG 2.0 App | ECG App |
| ECG Session<br>Classification<br>Results | - Low Heart Rate (< 50 bpm) | - Low Heart Rate (< 50 bpm) |
| | - Sinus Rhythm (50-99 bpm) | - Sinus Rhythm (50-100 bpm) |
| | - High Heart Rate (No AFib) (100-150 bpm) | - Atrial Fibrillation (50-120 bpm) |
| | - Atrial Fibrillation (50-99 bpm) | - Inconclusive |
| | - Atrial Fibrillation High Heart Rate (100-150 bpm) | - Inconclusive - Poor Recording |
| | - Inconclusive | - High Heart Rate (> 120 bpm) |
| | - Poor Recording | |
| | - High Heart Rate (> 150 bpm) | |
| | | |
| Clinical<br>Application | The ECG 2.0 app is intended to<br>supplement rhythm classification for<br>the purposes of discriminating Afib<br>from normal rhythms. The device is not<br>intended to replace traditional methods<br>or diagnosis. | The ECG app is intended to supplement<br>rhythm classification for the purposes of<br>discriminating Afib from normal rhythms.<br>The device is not intended to replace<br>traditional methods or diagnosis. |
| Compatibility<br>with Intended<br>Platforms | iOS version 14.0<br>WatchOS version 7.0<br>Apple Watch Series 4, Apple Watch<br>Series 5<br>iPhone 6s - iPhone 11 models | iOS 12.1.1 - 14.0.1<br>WatchOS 5.1.2 - 7.0.1<br>Apple Watch Series 4, Apple Watch Series<br>5, Apple Watch Series 6<br>iPhone 5s - iPhone 11 models |
{5}------------------------------------------------
{6}------------------------------------------------
# Performance Data
Apple conducted all necessary clinical and non-clinical performance testing on the ECG 2.0 app to support a determination of substantial equivalence to the predicate device.
# Non-Clinical Testing Summary
Apple conducted the necessary non-clinical testing on the ECG 2.0 app with passing results supporting a determination of substantial equivalence. Non-clinical testing conducted included the following:
- Software Verification Testing
Software verification testing was conducted and documentation was provided as recommended by FDA's Guidance for Industry and Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices." The software for this device was considered as a "moderate" level of concern.
- Human Factors Validation
{7}------------------------------------------------
Apple conducted a human factors validation study to verify the ECG 2.0 app is safe and effective for the intended users, uses, and use environments.
- ECG Database Testing per EC57 In compliance with special controls established under 21 CFR 870.2345, Apple conducted database testing using a previously adjudicated dataset.
- Platform compliance with the following standards:
- Thermal safety requirements under IEC-60950
- Applicable RF and EMC requirements under EN 301-489-17 v 3.2.0 and FCC Part 15
# Clinical Testing Summary
The ECG app's ability to accurately classify an ECG recording into AFib and sinus rhythm was extensively tested in a pivotal, prospective, multi-center clinical trial of approximately 546 subjects -305 were enrolled in the Atrial Fibrillation cohort, 241 were enrolled in the normal sinus rhythm cohort. The mean age of enrolled subjects was 58. Rhythm classification of a 12-lead ECG by a cardiologist was compared to the rhythm classification of a simultaneously collected ECG from the ECG 2.0 app. The ECG app demonstrated 98.5% sensitivity in classifying AFib (HR 50-150 bpm) and 99.3% specificity in classifying sinus rhythm (HR 50-150 bpm) in classifiable recordings. Subgroup analysis indicated sensitivity ranged from 98.3% - 100% across all age groups, and specificity ranged from 99.0% - 100.0%. Specificity and sensitivity estimates were slightly higher for females (99.6% and 99.2%, respectively) than for males (99.1% and 98.3%, respectively). Specificity and sensitivity estimates for subjects identifying as White were 99.1% and 98.5%, respectively, and were 100.0% each for subjects identifying as Asian, Black or African American, and Other.
In addition, The morphology of the waveform was also tested in this clinical trial by visual assessment of the PQRST wave and R wave amplitude in comparison to a reference. The ECG 2.0 app produced a visually acceptable PQRST waveform with a pass rating of 100% compared to the reference waveform. Additionally, the total pass rating for R-wave amplitude assessment was 97.2% compared to the reference.
# Conclusion
The Apple ECG 2.0 app is substantially equivalent to the Apple ECG app as they are identical with respect to intended use and share very similar technological characteristics. The differences in technological characteristics were assessed and testing through extensive non-clinical testing. The testing results demonstrate that the differences between the subject and predicate device do not raise new questions of safety and effectiveness. The Apple ECG 2.0 app is substantially equivalent to the Apple ECG app.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.