K201497 · Pioneer Surgical Technology, Inc. (Dba Rti Surgical, Inc.) · JDQ · Oct 28, 2020 · Orthopedic
Device Facts
Record ID
K201497
Device Name
EVOS Cabling System
Applicant
Pioneer Surgical Technology, Inc. (Dba Rti Surgical, Inc.)
Product Code
JDQ · Orthopedic
Decision Date
Oct 28, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The EVOS Cabling System is intended to be used in general orthopaedic repair procedures including patellar fractures, general cerclage, trochanteric reattachment, femoral and tibial fractures, prophylactic banding, olecranon fractures, ankle fractures, fixation of spiral fractures in conjunction with intramedullary nail and screw fixation techniques.
Device Story
EVOS Cabling System is a bone fixation cerclage device used in orthopedic trauma and reconstructive surgeries to reduce and stabilize fractures or osteotomies. The system consists of a sterile, stainless-steel cable (ASTM F1314) and a stainless-steel crimp (ASTM F138), accompanied by non-sterile, reusable instruments. During surgery, the surgeon passes the cable around the bone, tensions it, and secures it using the crimp mechanism to provide supplementary fracture fixation. The device is intended for use in clinical settings by orthopedic surgeons. It is MR Conditional. The system provides mechanical stabilization to bone segments, facilitating healing and potentially reducing the need for more invasive fixation methods.
Clinical Evidence
No clinical data. Substantial equivalence is supported by non-clinical bench testing, including static and dynamic tensile testing of construct strength, MR safety evaluation (ASTM F2052-15, F2213-17, F2182-11a, F2119-07), sterilization validation, and biological safety evaluation (ISO 10993-1). Pyrogenicity was assessed via LAL assay with an endotoxin limit of 20 EU per device.
Technological Characteristics
Materials: Stainless steel cable (ASTM F1314) and crimp (ASTM F138). Design: Multistrand, bundled cable with crimp-based locking mechanism. Sterilization: Gamma radiation for implants; validated sterilization for reusable instruments. Packaging: Double sterile barrier. MR status: MR Conditional.
Indications for Use
Indicated for patients requiring orthopedic repair procedures including patellar, femoral, tibial, olecranon, and ankle fractures; general cerclage; trochanteric reattachment; prophylactic banding; and fixation of spiral fractures with intramedullary nails or screws.
Regulatory Classification
Identification
A bone fixation cerclage is a device intended to be implanted that is made of alloys, such as cobalt-chromium-molybdenum, and that consists of a metallic ribbon or flat sheet or a wire. The device is wrapped around the shaft of a long bone, anchored to the bone with wire or screws, and used in the fixation of fractures.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. Food & Drug Administration".
Pioneer Surgical Technology, Inc. (DBA RTI Surgical, Inc.) % Jaclyn Holli Regulatory Affairs Specialist RTI Surgical, Inc 375 River Park Circle Marquette, Michigan 49855
Re: K201497
Trade/Device Name: EVOS Cabling System Regulation Number: 21 CFR 888.3010 Regulation Name: Bone Fixation Cerclage Regulatory Class: Class II Product Code: JDQ Dated: October 9, 2020 Received: October 13, 2020
### Dear Jaclyn Holli:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
October 28, 2020
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For; Shumaya Ali, M.P.H. Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K201497
Device Name EVOS Cabling System
Indications for Use (Describe)
The EVOS Cabling System is intended to be used in general orthopaedic repair procedures including patellar fractures, general cerclage, trochanteric reattachment, femoral and tibial fractures, prophylactic banding, olecranon fractures, ankle fractures, fixation of spiral fractures in conjunction with intramedullary nail and screw fixation techniques.
| Type of Use (Select one or both, as applicable) | |
|---------------------------------------------------------------------------------------|--|
| <span style="font-size: large;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | |
| <span style="font-size: large;">☐</span> Over-The-Counter Use (21 CFR 801 Subpart C) | |
X Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary per 21 CFR 807.92 EVOS Cabling System
| Date Prepared | October 27, 2020 |
|----------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Owner/<br>Manufacturer | Pioneer Surgical Technology, Inc.<br>(DBA RTI Surgical, Inc.)<br>375 River Park Circle<br>Marquette, MI 49855, USA<br>Registration #1833824 |
| Submitter/Contact<br>Person | Jaclyn Holli – Regulatory Affairs Specialist<br>RTI Surgical, Inc.<br>Phone: 1.906.226.9909<br>Email: jholli@rtix.com |
| Trade Name of<br>Device | EVOS Cabling System |
| Common Name | Cerclage, Fixation |
| Classification<br>Name | Bone Fixation Cerclage |
| Classification | Class II per 21 CFR 888.3010<br>Product code JDQ |
| Panel | Orthopedic Device Panel |
| Predicate | Primary Predicate: Orthopedic Cabling System (K031162)<br>Additional Predicate: SDB Cerclage System (K992616) |
| Description of<br>Device | The EVOS Cabling System is used in orthopedic trauma and<br>reconstructive surgeries to reduce and stabilize fractures and<br>osteotomies. The EVOS Cabling System may be used for<br>supplementary fracture fixation when used with bone plates or<br>screws.<br>The EVOS Cabling System includes a sterile, stainless-steel cable<br>implant (ASTM F1314) packaged together with a stainless-steel<br>crimp (ASTM F138). Non-sterile, reusable instruments are also<br>provided to facilitate proper implantation of the cable-crimp implant. |
| Purpose of<br>Submission | To obtain FDA Clearance of the EVOS Cabling System. There are<br>no prior submissions for the subject system. |
| Indications for<br>Use | The EVOS Cabling System is intended to be used in general<br>orthopaedic repair procedures including patellar fractures, general<br>cerclage, trochanteric reattachment, femoral and tibial fractures,<br>prophylactic banding, olecranon fractures, ankle fractures, fixation<br>of spiral fractures in conjunction with intramedullary nail and screw<br>fixation techniques. |
| Summary of<br>Technological<br>Characteristics | The subject EVOS Cabling System has the same intended use and<br>technological characteristics as the predicates and the following<br>similarities: |
| | Fundamental technology and intended uses: orthopedic<br>cable cerclage to reduce and stabilize fractures for bone<br>fixation Overall design: multistrand, bundled cable with a<br>mechanism to resist tension and lock the construct (crimp<br>or clamp cerclage) Principles of operation (cable passage around bone,<br>cerclage tensioning and fixation) Materials: metallic, stainless steel alloys with well-<br>established manufacturing processes and biological safety Sterility: sterile via gamma radiation Single Use, Rx only Packaging: double sterile barrier packaging with a 5-year<br>shelf life Mechanical performance sufficient to perform its intended<br>use Same instrumentation types The minor differences in the design characteristics do not raise<br>different issues of safety or effectiveness as evidenced by the non-<br>clinical testing. The subject EVOS Cabling System will be labeled<br>as magnetic resonance (MR) conditional for which performance<br>testing has been submitted for support of this claim. There are no<br>other differences in technological characteristics between the<br>subject and predicate devices. |
| Discussion of<br>Supporting<br>Clinical Evidence<br>and Non-Clinical<br>Evidence | The following non-clinical evidence was submitted and relied upon<br>for a determination of substantial equivalence:<br>Mechanical performance: static and dynamic tensile testing<br>of the construct strength. MR Safety evaluation determined the devices are MR<br>Conditional via the following standard methods: ASTM F2052-15, Standard Test Method for<br>Measurement of Magnetically Induced Displacement<br>Force on Medical Devices in the Magnetic<br>Resonance Environment ASTM F2213-17, Standard Test Method for |
| | Medical Devices in the Magnetic Resonance<br>Environment ASTM F2182-11a, Standard Test Method for<br>Measurement of Radio Frequency Induced Heating<br>Near Passive Implants During Magnetic Resonance<br>Imaging ASTM F2119-07 (Reapproved 2013), Standard Test<br>Method for Evaluation of MR Image Artifacts from<br>Passive Implants Implant packaging and sterilization testing and evaluations,<br>relying on equivalency to predicate device design and<br>package configuration Instrument (and case components that keep the instruments<br>in place during transit and sterilization) sterilization testing<br>and evaluations, relying on subject validation testing ISO 10993-1 Biological safety evaluation, relying on<br>material and manufacturing process equivalency to<br>predicate devices. Pyrogenicity evaluation of the devices based on current<br>industry practice and FDA guidance for batch release<br>testing: endotoxin limit of 20 EU per device using the<br>Limulus amebocyte lysate (LAL) assay. There are no clinical tests relied on in this premarket notification<br>submission for a determination of substantial equivalence. |
| Conclusion | The evidence provided in this premarket notification submission<br>supports substantial equivalence of the subject EVOS Cabling<br>System to the predicate devices. |
{4}------------------------------------------------
{5}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.