K201490 · Med Systems, Inc. · BYI · Feb 12, 2021 · Anesthesiology
Device Facts
Record ID
K201490
Device Name
Electro Flo 6 Airway Clearance System
Applicant
Med Systems, Inc.
Product Code
BYI · Anesthesiology
Decision Date
Feb 12, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5665
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Electro Flo® 6 Airway Clearance System is intended to provide Airway Clearance Therapy and promote bronchial drainage where external manipulation of the thorax is the physician's choice of treatment. It is indicated for patients having difficulty with secretion clearance, or the presence of atelectasis caused by mucus plugging. The device is intended for home or institutional use for patients weighing at least 23 Kg.
Device Story
Electro Flo® 6 is a High Frequency Chest Wall Oscillation (HFCWO) system for home/institutional use. It consists of an electromechanical Control Box and a handheld Power Head. The device operates on a 'hammer and anvil' principle: the Control Box sends electrical pulses to separate the hammer (Power Head body) and anvil (contact surface). The therapist/patient applies manual pressure to the Power Head; when a pulse ends, the manual force drives the hammer into the anvil, delivering a mechanical impulse to the chest. This acts as a 'force multiplier,' creating pressure waves to loosen mucus. The user controls frequency (6 settings) and intensity (5 settings) via the Control Box. An optional strap allows for unassisted therapy. The output is a vibratory, shaking motion on the thorax, which helps the patient clear mucus via huffing, coughing, or suction. It benefits patients by providing consistent therapy without human fatigue and allowing for higher frequencies/force than manual percussion.
Clinical Evidence
Bench testing only. Testing characterized mechanical performance using a linear bearing and load cell assembly to measure vibration waveforms at a 1000 Hz sampling rate. Results confirmed the device functions as an HFCWO system with power output controllable by static load (hand pressure) and settings. Power spectra analysis showed output in the 16-127 Hz range. No clinical data provided.
Technological Characteristics
Electromechanical 'hammer and anvil' actuator. Frequency range: 5-25 Hz. Power: 120 V, 60 Hz AC. Dimensions: 1.5 Kg Power Head. Non-sterile. No software. Safety standards: UL 60601-1, CSA C22.2 #601.1, EN60601-1-2, EN55011 Class A, FCC Part 15 Subpart B.
Indications for Use
Indicated for patients ≥ 23 kg with difficulty clearing secretions or atelectasis due to mucus plugging. Contraindicated for patients with pacemakers/implants in/near chest, brittle bones, severe osteoporosis, collapsed lungs, pulmonary embolism, abscess, tuberculosis, angina, arrhythmia, aneurysms, severe asthma, or chest/spine/head/neck injuries.
Regulatory Classification
Identification
A powered percussor is a device that is intended to transmit vibration through a patient's chest wall to aid in freeing mucus deposits in the lung in order to improve bronchial drainage and that may be powered by electricity or compressed gas.
Med Systems Electro Flo Percussor Model 5000 (K031876)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, with the letters "FDA" in a blue square. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
February 12, 2021
Med Systems, Inc. % Frederick Cahn Principal BioMedical Strategies 133 Colonial Dr. White River Junction, Vermont 05001
Re: K201490
Trade/Device Name: Electro Flo® 6 Airway Clearance System Regulation Number: 21 CFR 868.5665 Regulation Name: Powered Percussor Regulatory Class: Class II Product Code: BYI Dated: August 26, 2020 Received: September 8, 2020
Dear Frederick Cahn:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal
{1}------------------------------------------------
statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Brandon L. Blakely, Ph.D. Acting Assistant Director DHT1C: Division of ENT, Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201490
Device Name
Electro Flo® 6 Airway Clearance
Indications for Use (Describe)
The Electro Flo® 6 Airway Clearance System is intended to provide Airway Clearance Therapy and promote bronchial drainage where external manipulation of the physician's choice of treatment. It is indicated for patients having difficulty with secretion clearance, or the presence of atelectasis caused by mucus plugging.
The device is intended for home or institutional use by patients weighing at least 23 Kg.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <span style="font-size: 16px;">☑</span> Prescription Use (Part 21 CFR 801 Subpart D) | ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
|--------------------------------------------------------------------------------------|-----------------------------------------------|
|--------------------------------------------------------------------------------------|-----------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
- I. SUBMITTER: Med Systems, Inc. 2631 Ariane Drive San Diego, CA 92117
Contact: Susan B. Davis, CEO (858) 483-9671
II. Electro Flo® 6 Airway Clearance System (K201490) DEVICE: Classification Code: BYI, "Percussor, Powered-Electric" Regulation 21 CFR 868.5665, "Powered Percussor" Class II
#### III. PREDICATE DEVICES
Primary Predicate: Monarch Airway Clearance System (K173603) Classification Code: BYI, "Percussor, Powered-Electric" Regulation 21 CFR 868.5665, "Powered Percussor" Class II
Reference Predicate: Med Systems Electro Flo Percussor Model 5000 (K031876) Classification Code: BYI, "Percussor, Powered-Electric" Regulation 21 CFR 868.5665, "Powered Percussor" Class II
- IV. DATE February 11, 2021
#### V. DESCRIPTION
### Device Identification
The Electro Flo® 6 Airway Clearance System is a High Frequency Chest Wall Oscillation System (HFCWO), including all accessories and supplies.
Electro Flo® 6 Airway Clearance System comprises:
- An electrically powered Control Box with a power switch on the back and two knobs on the front: one knob provides five intensity settings, and the other provides six frequency settings,
- The Power Head, connected by a cable to the Control Box, ●
- The Carrying Case,
- The Information Manual, ●
{4}------------------------------------------------
- The Optional Self-Administrator® Strap that can hold the Electro Flo® 6 system Power Head and enable a patient to apply therapy to the back, unassisted.
### Device Characteristics and Environment of Use
The Electro Flo® 6 Airway Clearance System is durable medical equipment for home use. It is non-sterile and has no software or patient-contacting materials. Treatment sessions can last up to 20 minutes.
## Device Description
The Electro Flo® 6 Airway Clearance System's operating mechanism is an electromechanical "hammer and anvil"; the hammer is the handheld body of the Power Head, and the anvil is the surface of the Power Head that is held in contact with the chest. Repeated electrical pulses from the Control Box separate the hammer and anvil. The Frequency knob of the Control Box controls the repetition rate, and the Power knob controls the pulse duration; there are 30 combinations.
When each electrical pulse ends, the force applied by the therapist's hand forces the hammer through the gap until it strikes the anvil and delivers a mechanical impulse to the chest. The resulting pressure wave radiates into the chest cavity to shake mucus loose. Because the energy transmitted to the chest vibrations is a multiple of the force applied by hand, the Electro Flo® 6 system acts as a "force multiplier" that provides the therapist with additional control.
#### VI. INTENDED USE
The Electro Flo® 6 System is designed for home or institutional use and provides HFCWO with controllable frequency and intensity similar to the hand percussion therapy provided by a therapist. The chest vibrations target and loosen thick, sticky mucus in the lungs so that the patient can clear it by huffing, coughing, or suction. The Electro Flo® 6 system has advantages over hand percussion because patients can administer therapy without assistance, because it does not suffer fatigue, and because it can operate at higher frequencies and force than a human therapist.
#### INDICATIONS FOR USE VII.
The Electro Flo® 6 Airway Clearance System is intended to provide Airway Clearance Therapy and promote bronchial drainage where external manipulation of the thorax is the physician's choice of treatment. It is indicated for patients having difficulty with secretion clearance, or the presence of atelectasis caused by mucus plugging. The device is intended for home or institutional use for patients weighing at least 23 Kg.
## VIII. CONTRAINDICATIONS
The Electro Flo® 6 system is not intended for the following classes of patients:
- Infants or children weighing less than 50 pounds (23 Kg). ●
- Patients with pacemakers or other implanted medical devices in or near the ● chest.
{5}------------------------------------------------
- . Patients with brittle bones, broken ribs or severe osteoporosis, severe calcium deficiencies, or severe osteoporosis.
- Patients with collapsed lungs, pulmonary embolism, lung abscess, or ● tuberculosis.
- Patients with angina, arrhythmia, aneurism in the head, neck, chest, or abdomen. ●
- Patients with severe asthma. ●
- Patients with injuries to the chest wall, spine, head, or neck that could be ● exacerbated by strong vibrations.
#### WARNINGS AND PRECAUTIONS IX.
WARNING - To avoid the risk of electrocution, fire, or serious injuries:
- Keep the Electro Flo® 6 system out of the reach of children. ●
- Do not use the Electro Flo® 6 system with, near, or in water. ●
- Do not use the Electro Flo® 6 system near flammable vapors or fire. ●
- Do not insert any object into any opening on the Electro Flo® 6 system.
- Inspect the Control Box and Power Head before use. Do not use the Electro ● Flo® 6 system if there is damage to the power cord or any other part of the device.
WARNING - Do not use the Electro Flo 6 while you are experiencing conditions such as:
- Recent heart attack.
- Recent surgery to the chest, abdomen, head, or neck. ●
- Bleeding from the lungs or coughing up blood. ●
- Open wounds or burns.
- Vomiting.
- . Pressure in the skull.
- Intense pain. ●
- Active hemorrhage.
- Acute asthma.
### CAUTION
- The Electro Flo® 6 system is available by prescription only and is intended . solely for the person for whom it was prescribed and only for the prescribed use.
- . Users should always observe all contraindications, warnings, and cautions listed in the Information Manual.
- Do not use the Electro Flo® 6 system directly after eating. ●
- Do not percuss over bare skin. ●
- Do not percuss directly on the shoulder blade, collar bone, spine, breastbone (sternum), or breasts.
- Percussion should not hurt. If it does, STOP. Reduce the hand pressure on the ● Power Head's rear cover or reduce the power setting on the Control Box.
- Do not place Self-Administrator® strap over the head or around the neck.
{6}------------------------------------------------
## NOTICES
To avoid damage or excessive wear:
- Do not unplug or switch off the unit until the cool-down is complete. ●
- Do not block the fan vent with hand or fingers.
- . Limit percussion to 20 minutes maximum per session to avoid overheating the system.
- Avoid using the unit in a dusty environment or laying the device on a blanket or ● carpet as lint may be drawn in by the fan and damage the unit.
- . Do not spray disinfectant on the Control Box and device; fluid could enter the device and damage the system.
- If any interferences from a television, computer, cell phone, or other electronic ● device are experienced when using the Electro Flo® 6 system, move the Electro Flo® 6 system to another room or location or turn off the interfering device.
#### COMPARISONS WITH THE PREDICATE DEVICES X.
## Comparison with Primary Predicate
We compare the Electro Flo® 6 Airway Clearance System to a legally marketed predicate device: The Monarch Airway Clearance System (K173603). The predicate Monarch Airway Clearance System is classified in product code BYI, "Percussor, Powered-Electric", classification 868.5665, "Powered Percussor." The devices have identical indication statements.
### Comparison of Indication Statements
| Electro Flo® 6 Airway | Monarch Airway | Commentary |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------|
| Clearance System | Clearance System | |
| The Electro Flo® 6 Airway Clearance<br>System is intended to provide Airway<br>Clearance Therapy and promote bronchial<br>drainage where external manipulation of the<br>thorax is the physician's choice of<br>treatment. It is indicated for patients having<br>difficulty with secretion clearance, or the<br>presence of atelectasis caused by mucus<br>plugging. | The Monarch product is intended to<br>provide Airway Clearance Therapy<br>and promote bronchial drainage where<br>external manipulation of the thorax is<br>the physician's choice of treatment. It<br>is indicated for patients having<br>difficulty with secretion clearance, or<br>the presence of atelectasis caused by<br>mucus plugging. | The indications<br>statements are<br>identical. |
The devices are both electromechanical devices intended for airway clearance and have substantially equivalent technology, as shown in the following table:
{7}------------------------------------------------
| Characteristic | Electro Flo® 6 Airway<br>Clearance System | Monarch Airway<br>Clearance System | Commentary |
|--------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Mechanism | Variable frequency<br>electromechanical<br>actuators. | Variable frequency<br>electromechanical<br>actuators. | Substantially equivalent:<br>both vibrate the chest. |
| Anatomic<br>positions of the<br>actuator(s) on the<br>thorax | Determined by the<br>therapist or patient<br>manually or aided by a<br>repositionable fabric self-<br>administrator strap. | Determined by the eight<br>pockets at fixed anatomic<br>locations in a fabric vest. | Substantially equivalent:<br>The Electro Flo® 6 system<br>can be positioned at the<br>same anatomic locations as<br>the Monarch System's<br>actuators, but it is not<br>limited to those positions. |
| Actuator motion | The electromechanical<br>actuator of the Electro<br>Flo® 6 system provides a<br>vibratory, shaking motion<br>to the body. | The electromechanical<br>actuator of the Monarch<br>system provides a<br>vibratory, shaking motion<br>to the body. | Substantially equivalent: |
| Type of device | Durable medical<br>equipment for home use. | Durable medical<br>equipment for home use. | Substantially equivalent. |
| Controller<br>Frequency range | 5 — 25 Hz | 5-20 Hz | Substantially equivalent. |
| Power supply | 120 V, 60 Hz AC | Rechargeable battery. | Substantially equivalent.<br>Both devices are electrically<br>powered |
| Weight | 1.5 Kg (Power Head) | 6 Kg (vest) | The lighter weight of the<br>Electro Flo® 6 system<br>enables it to be easily<br>handheld. The Monarch vest<br>system is worn on the body,<br>making the higher weight<br>easily tolerated. |
Comparison of Technical Characteristics
## Comparison with Reference Predicate
We compare the Electro Flo® 6 Airway Clearance System our previously cleared device: The Electro Flo Percussor Model 5000 (K021876). The Electro Flo Percussor Model 5000 is also classified in product code BYI, "Percussor, Powered-Electric", classification 868.5665, "Powered Percussor."
{8}------------------------------------------------
| Electro Flo® 6 Airway<br>Clearance System | Electro Flo Percussor Model 5000 | Commentary |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| The Electro Flo® 6 Airway<br>Clearance System is<br>intended to provide Airway<br>Clearance Therapy and<br>promote bronchial drainage<br>where external manipulation<br>of the thorax is the<br>physician's choice of<br>treatment. It is indicated for<br>patients having difficulty<br>with secretion clearance, or<br>the presence of atelectasis<br>caused by mucus plugging. | The intended use of the Med Systems Electro Flo<br>Percussor Model 5000 is the same as the predicate device,<br>which is to provide airway clearance therapy when<br>external manipulation of the thorax is the physician's<br>choice of treatment. Indications for this form of therapy<br>are described by the American Association for<br>Respiratory Care (AARC) in the Clinical Practices<br>Guidelines for Postural Drainage Therapy (1991).<br>According to AARC guidelines, specific indications for<br>external manipulation of the thorax include evidence or a<br>suggestion of retained secretions, evidence that the patient<br>is having difficulty with the secretion clearance, or<br>presence of atelectasis caused by mucus plugging. In<br>addition, the Med Systems Electro Flo Percussor Model<br>5000 is also indicated for external manipulation of the<br>thorax to promote airway clearance or improve bronchial<br>drainage for purposes of collecting mucus for diagnostic<br>evaluation. | Substantially<br>equivalent. |
The devices have the same intended use and are the identical technologically, as shown in the following table:
| Characteristic | Electro Flo® 6 Airway<br>Clearance System | Electro Flo Percussor<br>Model 5000 | Commentary |
|--------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|------------|
| Mechanism | Variable frequency<br>electromechanical<br>actuators. | Variable frequency<br>electromechanical<br>actuators. | Identical |
| Anatomic<br>positions of the<br>actuator(s) on the<br>thorax | Determined by the<br>therapist or patient<br>manually or aided by a<br>repositionable fabric self-<br>administrator strap. | Determined by the<br>therapist or patient<br>manually or aided by a<br>repositionable fabric self-<br>administrator strap. | Identical |
| Actuator motion | The electromechanical<br>actuator provides a<br>vibratory, shaking motion<br>to the body. | The electromechanical<br>actuator provides a<br>vibratory, shaking motion<br>to the body. | Identical |
| Type of device | Durable medical<br>equipment for home use. | Durable medical<br>equipment for home use. | Identical |
| Controller<br>Frequency range | 5 - 25 Hz | 5 - 25 Hz | Identical |
| Power supply | 120 V, 60 Hz AC | 120 V, 60 Hz AC | Identical |
| Weight | 1.5 Kg (Power Head) | 1.5 Kg (Power Head) | Identical |
## Comparison of Technical Characteristics
#### XI. PERFORMANCE DATA
## Safety and Electromagnetic Radiation
The Electro Flo® 6 system meets the following safety and electromagnetic radiation standards:
{9}------------------------------------------------
| Standard Reference | Titles |
|--------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------|
| UL 60601-1:2003<br>Ed.1 +R:26Apr2006 | Medical Electrical Equipment - Part 1: General Requirements for<br>Safety |
| CSA C22.2 #601.1:1990<br>Ed.1+G1;S1;A2;G2] | Medical Electrical Equipment - General Requirements<br>For Safety (R2005) |
| EN60601-1-2 | Medical electrical equipment - Part 1: General requirements for<br>basic safety and essential performance |
| EN55011 Class A | Industrial, scientific and medical equipment- Radio-frequency<br>disturbance characteristics, limits, and methods of measurement |
| FCC Part 15 Subpart B | Unintentional radiators |
## Bench Data
Med Systems. Inc. contracted with Aeromethod Precision Engineering and Manufacturing, San Diego, CA, to conduct testing to characterize the Electro Flo® 6 system's mechanical performance. The test objectives were to verify and document that the Electro Flo® 6 system can be described as a high frequency chest wall oscillator (HFCWO) and to demonstrate that the output can be controlled by the force applied by the user to the device, as expected from its force multiplier design principle.
The test assembly was designed to collect one-dimensional vibration data along the axis of the Power Head. The Power Head body was rigidly attached to a linear bearing that could slide along a linear guide track. An elastic material pressed on the Power Head created a Static Load to simulate hand pressure. The Power Head's anvil end was attached to a plastic cvlinder in contact with a load cell that measured the waveforms at a sampling rate of 1000 Hz over a one-second interval. Waveforms were collected at a full factorial sampling with three frequency settings (1,3,6) on the Control Box, three power settings (1,3,5) on the Control Box, and three static loads of one, three, and five pounds of force applied to the Power Head. BioMedical Strategies, White River Junction, VT computed power spectra of the waveforms in arbitrary power units.
### Results
At all settings, most of the power generated by the Electro Flo® 6 system was in the acoustic frequency range of 16 - 127 Hz. Varying the static load could control the power output, as predicted by the force multiplier design principle.
### Comparison with Predicates
The Electro Flo and the Monarch primary predicate have variable frequency and power output. The Electro Flo has additional control over power output via hand pressure. They operate in overlapping frequency ranges.
The Electro Flo 6 and the reference predicate Electro Flo 5000 are the identical hardware, with the same frequency range, power, and controllability.
#### SUBSTANTIAL EQUIVALENCE CONCLUSION XII.
Substantial Equivalence between the Electro Flo® 6 Airway Clearance System and the primary predicate Monarch Airway Clearance System is demonstrated by:
- Identical indications statements. ●
- Identical intended use. ●
{10}------------------------------------------------
- Substantially equivalent technological characteristics: .
- o Both employ electromechanical actuators;
- Both treat the same anatomic locations of the thorax; O
- Both provide a vibratory, shaking motion to the body; and o
- o Both operate over the same fundamental frequency range.
Substantial Equivalence between the Electro Flo® 6 Airway Clearance System and the reference predicate Electro Flo 5000 system is demonstrated by:
- Substantially equivalent indications statements. ●
- . Identical intended use.
- Identical technological characteristics. ●
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.