VenSure Balloon Device, VenSure Nav Balloon Device
Applicant
Fiagon GmbH
Product Code
LRC · Ear, Nose, Throat
Decision Date
Aug 26, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.4420
Device Class
Class 1
Attributes
Therapeutic
Indications for Use
The VenSure™ Balloon Device and VenSure™ Nav Balloon Device are used to access and treat the frontal recesses, sphenoid sinus ostia and maxillary ostia/ethmoid infundibula in adults using a trans-nasal approach. The bony sinus outflow tracts are remodeled by balloon displacement of adjacent bone and paranasal sinus structures. The VenSure™ Nav Balloon Device is intended for use in conjunction with the Fiagon Navigation System during sinus procedures when surgical navigation or image-guided surgery may be necessary to locate and move tissue, bone or cartilaginous tissue surrounding the drainage pathways of frontal, maxillary, and sphenoid sinuses to facilitate dilation of the sinus ostia. The Fiagon Navigation System is intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. The Fiagon Navigation system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery can be identified relative to a CT or MR based model of the anatomy.
Device Story
VenSure™ and VenSure™ Nav are sterile, single-use sinus dilation devices. Distal end features re-shapeable tip for accessing frontal, maxillary, and sphenoid sinuses. VenSure™ Nav includes integrated sensor carrier for electromagnetic tracking with Fiagon Navigation System; enables real-time visualization of instrument tip relative to patient CT/MR images. Physician uses endoscopic visualization; Nav version adds image-guided navigation. Balloon inflated with saline via syringe/inflator to remodel bony sinus outflow tracts. Basic VenSure™ supports active suction or irrigation via proximal luer fitting. Device aids surgeons in precise anatomical localization during sinus procedures; facilitates ostia dilation; potentially improves surgical outcomes by providing navigation-assisted access to complex sinus anatomy.
Clinical Evidence
Bench testing only. Testing included balloon dimensional integrity, pressure stability, inflation/deflation time, mechanical integrity (torsion/tensile strength), and navigation accuracy. Biocompatibility testing per ISO 10993-1:2009 confirmed patient-contacting materials are safe. Sterilization validated per ISO 11135:2014. Shelf life supported by accelerated aging (ASTM F1980) and package integrity testing.
Technological Characteristics
Sterile, single-use sinus dilation instrument. Features re-shapeable distal tip and balloon. VenSure™ Nav includes electromagnetic sensor carrier for navigation. Materials biocompatible per ISO 10993-1. Sterilization per ISO 11135. Packaging validated per ISO 11607-1, ASTM F1886, F88, F1140, F1929. Connectivity: Electromagnetic tracking via Fiagon Navigation System (IEC 60601-1-2).
Indications for Use
Indicated for adults requiring access and treatment of frontal recesses, sphenoid sinus ostia, and maxillary ostia/ethmoid infundibula via trans-nasal approach for remodeling of bony sinus outflow tracts. VenSure™ Nav indicated for use with Fiagon Navigation System for image-guided surgery to locate and move tissue, bone, or cartilage surrounding sinus drainage pathways.
Regulatory Classification
Identification
An ear, nose, and throat manual surgical instrument is one of a variety of devices intended for use in surgical procedures to examine or treat the bronchus, esophagus, trachea, larynx, pharynx, nasal and paranasal sinus, or ear. This generic type of device includes the esophageal dilator; tracheal bistour (a long, narrow surgical knife); tracheal dilator; tracheal hook; laryngeal injection set; laryngeal knife; laryngeal saw; laryngeal trocar; laryngectomy tube; adenoid curette; adenotome; metal tongue depressor; mouth gag; oral screw; salpingeal curette; tonsillectome; tonsil guillotine; tonsil screw; tonsil snare; tonsil suction tube; tonsil suturing hook; antom reforator; ethmoid curette; frontal sinus-rasp; nasal curette; nasal rasp; nasal rongeur; nasal saw; nasal scissors; nasal snare; sinus irrigator; sinus trephine; ear curette; ear excavator; ear rasp; ear scissor, ear snare; ear spoon; ear suction tube; malleous ripper; mastoid gauge; microsurgical ear chisel; myringotomy tube inserter; ossici holding clamp; sacculotomy tack inserter; vein press; wire ear loop; microrule; mirror; mobilizer; ear, nose, and throat punch; ear, nose and throat knife; and ear, nose, and throat trocar.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logos of the Department of Health and Human Services and the Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
August 26, 2020
Fiagon GmbH Dirk Mucha Chief Technology Officer Neuendorfstrasse 23b 16761 Hennigsdorf Germany
Re: K201472
Trade/Device Name: VenSure Balloon Device, VenSure Nav Balloon Device Regulation Number: 21 CFR 874.4420 Regulation Name: Ear, Nose, And Throat Manual Surgical Instrument Regulatory Class: Class I, reserved Product Code: LRC Dated: June 2, 2020 Received: June 3, 2020
Dear Dirk Mucha:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Actinclude requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Malvina B. Eydelman, M.D. Director OHT1: Office of Ophthalmic. Anesthesia. Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201472
Device Name VenSure™ Balloon Device VenSure™ Nav Balloon Device
#### Indications for Use (Describe)
The VenSure™ Balloon Device and VenSure™ Nav Balloon Device are used to access and treat the frontal recesses, sphenoid sinus ostia and maxillary ostia/ethmoid infundibula in adults using a trans-nasal approach. The bony sinus outflow tracts are remodeled by balloon displacement of adjacent bone and paranasal sinus structures.
The VenSure™ Nav Balloon Device is intended for use in conjunction with the Fiagon Navigation System during sinus procedures when surgical navigation or image-guided surgery may be necessary to locate and move tissue, bone or cartilaginous tissue surrounding the drainage pathways of frontal, maxillary, and sphenoid sinuses to facilitate dilation of the sinus ostia.
The Fiagon Navigation System is intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. The Fiagon Navigation system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery can be identified relative to a CT or MR based model of the anatomy.
Type of Use (Select one or both, as applicable)
| Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-size: 20px;">☑</span> |
|----------------------------------------------|-----------------------------------------|
| Over-The-Counter Use (21 CFR 801 Subpart C) | <span>☐</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Fiagon. The logo consists of the word "fiagon" in a sans-serif font. The "f" is green and enclosed in a green circle. The rest of the letters are black, except for a green dot after the "n".
# 510(k) Summary
#### August 11, 2020
#### 1. Submitter Information
| Submitter: | Fiagon GmbH |
|------------|----------------------------------------------------|
| Address: | Neuendorfstrasse 23b<br>16761 Hennigsdorf, Germany |
| Telephone: | +49 3302 201 21 10 |
| Telefax: | +49 3302 201 21 15 |
Contact: Dirk Mucha Chief Technology Officer
#### 2. Device Information
| Trade Name: | VenSure™ Balloon Device<br>VenSure™ Nav Balloon Device |
|----------------------|--------------------------------------------------------|
| Common Name: | Sinus Dilation Device |
| Classification: | Class I per 21 CFR 874.4420 |
| Classification Name: | Ear, nose and throat manual surgical instrument |
| Product Code: | LRC |
### 3. Predicate Device Information
The VenSure™ Balloon Device and VenSure™ Nav Balloon Device are substantially equivalent to the following predicate devices:
- Primary predicate: XprESS Multi-Sinus Dilation Tool (K142252) ●
- Secondary predicate: Medtronic EM Sinus Dilation System (K132297) ●
### 4. Device Description
Fiagon's VenSure™ Balloon Device and VenSure™ Nav Balloon Device are sterile, singleuse devices designed to remodel the bony structures within the sinuses. The device comes in two versions a navigation ready version (VenSure™ Nav) that is compatible with the Fiagon electromagnetic navigation system, and a basic non-navigation ready version (VenSure™).
The VenSure™ and VenSure™ Nav devices, combine features of a malleable suction and a malleable probe with the tissue expansion effect of balloon dilation. The distal end of the device includes an atraumatic tip and can be shaped to fit the frontal, maxillary, and sphenoid sinuses using the Bending Tool provided with the device. Since the distal end of the device is re-shapeable, one balloon can be modified to work on multiple sinuses within the same patient.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for Fiagon. The logo consists of a green circle with a white "f" inside, followed by the word "fiagon" in black. There is a small green dot after the word "fiagon".
Both versions enable a physician to track the device into the sinuses using endoscopic visualization; while the VenSure™ Nav allows for image-guided visualization when connected to the Fiagon Navigation System. The VenSure™ Nav contains an integrated sensor carrier that enables the use of image guidance through "plug and play" tracking capability when used with the Fiagon Navigation System. The sensor carrier containing localizer elements detects a signal within a low-energy magnetic field delivered from the navigation unit. The navigation software then displays the location of the sinus dilation instrument's tip within multiple patient image planes and other anatomic renderings. After confirmation of placement, the balloon of the dilation device can be inflated with saline solution, using the inflator to expand the outflow track of the targeted sinus.
A suction tube may be connected directly to the proximal luer fitting of the basic VenSure™ balloon dilation device to provide active suction. Alternately, an Extension Line connected to a syringe may be connected directly to the proximal luer fitting to provide irrigation. Suction and irrigation are not possible on the VenSure™ Nav.
### 5. Intended Use
The VenSure™ and VenSure™ Nav and the predicate devices have the same Intended Use, which is to access and treat the frontal, maxillary and sphenoid sinuses using a transnasal approach.
The Indications for Use for the VenSure™ and VenSure™ Nav Balloon Devices are the same as the XprESS Multi-Sinus Dilation Tube. Both, the VenSure™ Nav and the Medtronic EM Sinus Dilation System have an additional indication for use with a compatible image-guided navigation system (Fiagon Navigation System) as an aid for locating anatomical structures in either open or percutaneous procedures.
| Device | Indications for Use |
|-----------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| VenSure™ Balloon Device | The VenSure™ Balloon Device and VenSure™ Nav Balloon Device are used to access and treat the frontal recesses, sphenoid sinus ostia and maxillary ostia/ethmoid infundibula in adults using a trans-nasal approach. The bony sinus outflow tracts are remodeled by balloon displacement of adjacent bone and paranasal sinus structures. |
| VenSure™ Nav Balloon Device | The VenSure™ Nav Balloon Device is intended for use in conjunction with the Fiagon Navigation System during sinus procedures when surgical navigation or image-guided surgery may be necessary to locate and move tissue, bone or cartilaginous tissue surrounding the drainage pathways of frontal, maxillary, and sphenoid sinuses to facilitate dilation of the sinus ostia. |
| | The Fiagon Navigation System is intended as an aid for precisely locating anatomical structures in either open or percutaneous procedures. The Fiagon Navigation system is indicated for any medical condition in which the use of stereotactic surgery may be appropriate, and where reference to a rigid anatomical structure in the field of ENT surgery can be identified relative to a CT or MR based model of the anatomy. |
| Xpress Multi-Sinus Dilation Tool<br>(K142252) | To access and treat the frontal recesses, sphenoid sinus ostia and maxillary ostia/ethmoid infundibula in adults using a trans-nasal approach. The bony sinus outflow tracts are remodeled by balloon displacement of adjacent bone and paranasal sinus structures. |
| Indications for Use for VenSure™ / VenSure™ Nav and Predicate Devices | | | |
|-----------------------------------------------------------------------|--|--|--|
|-----------------------------------------------------------------------|--|--|--|
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for Fiagon. The logo consists of a green circle with a white "f" inside, followed by the word "fiagon" in black sans-serif font. A small green circle is located to the right of the word.
| Device | Indications for Use |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| EM Sinus Dilation<br>System /<br>(K132297) | The EM Sinus Dilation System is intended for use in sinus procedures when surgical<br>navigation or image-guided surgery may be necessary to locate and remove tissue,<br>bone or cartilaginous tissue surrounding the drainage pathways of the frontal,<br>maxillary, and sphenoid sinuses.<br>The EM Sinus Dilation system is used in conjunction with the Medtronic computer-<br>assisted surgery system. |
| The Medtronic computer-assisted surgery system and its associated applications are<br>intended as an aid for precisely locating anatomical structures in either open or<br>percutaneous procedures. Their use is indicated for any medical condition in which the<br>use of stereotactic surgery may be appropriate, and where reference to a rigid<br>anatomical structure, such as the skull, can be identified relative to a CT- or MR-based<br>model, or digitized landmarks of the anatomy. | |
| The system and its associated applications should be used only as an adjunct for<br>surgical guidance. They do not replace the surgeon's knowledge, expertise or<br>judgment. | |
# 6. Comparison of Technological Characteristics
The VenSure™ and VenSure™ Nav Balloon Devices have the same fundamental scientific technology as the predicate devices [K142252 and K132297]. The subject devices have the same technological characteristics; in particular, basic design, performance, and principle of operation.
| Feature | VenSure™ and<br>VenSure™ Nav<br>Balloon Device | Xpress Multi-Sinus<br>Dilation Tool<br>[K142252] | EM Sinus Dilation<br>System [K132297] | Equivalence |
|--------------------------------------------------------------------|--------------------------------------------------------|------------------------------------------------------------|----------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------|
| Class | Class I<br>21 CFR 874.4420<br>Product code: LRC | Class I<br>21 CFR 874.4420<br>Product code: LRC | Class I<br>21 CFR 874.4420<br>Product code: LRC | Same |
| Indications for<br>Use | (see above) | (see above) | (see above) | Equivalent<br>For Dilation same as<br>Predicate 1<br>For image guidance<br>same as Predicate 2 |
| Balloon design | Fixed balloon | Sliding balloon<br>mechanism | Fixed balloon | Same as Predicate 1 |
| Balloon<br>Dimensions<br>[D x L (mm)] | 6 x 18 | 5 x 8, 5 x 18, 5 x 20, 6<br>x 8, 6 x 18, 6 x 20, 7 x<br>18 | 5 x 7, 6 x 7, 7 x 7<br>5 x 17, 6 x 17, 7 x<br>17 | Equivalent<br>Same size available<br>in Predicate 1<br>Less additional sizes<br>are available |
| Ability to access<br>multiple<br>locations with<br>single balloon? | Yes - tip angle is<br>reshapable using<br>bending tool | Yes - tip angle is<br>reshapable using<br>bending tool | N/A - Rigid tip.<br>Multiple versions of<br>device with fixed<br>balloon shapes<br>based on sinus. | Same as Predicate 1 |
| Inflation Device | Syringe barrel and<br>plunger | Syringe barrel and<br>plunger | Syringe barrel and<br>plunger | Same |
| Visualization | Endoscopy & Image<br>guided option<br>(VenSure™ Nav) | Endoscopy | Endoscopy & Image<br>guided option ( | Same as Predicate 2 |
| Image-guided<br>Tracking<br>Method | Electromagnetic<br>(VenSure™ Nav<br>only) | N/A (Compatible) | Electromagnetic | Same |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for Fiagon. The logo consists of the word "fiagon" in black letters, with a green circle containing a white "f" to the left of the word. There is also a green dot to the right of the word.
# 7. Performance Data
#### Bench Testing
Bench testing was conducted to ensure that the VenSure™ and VenSure™ Nav met the predefined acceptance criteria to demonstrate safety and performance. Testing included the following:
- Balloon dimensional integrity ●
- Balloon pressure stability /Inflation/Deflation ●
- Inflation and Deflation Time ●
- Mechanical integrity ●
- o Torsion strength
- Tensile force o
- o Dimensions
- Navigation compatibility (VenSure™ Nav only) .
All tests met the predefined acceptance criteria. The test results demonstrated that the minor differences in device characteristics between the subject device and predicate devices do not raise any new questions of safety or effectiveness.
#### Biocompatibility
The biocompatibility evaluation for the VenSure™ and VenSure™ Nav balloon dilation devices was conducted in accordance with FDA recognized consensus standard ISO 10993-1:2009. Biocompatibility testing included cytotoxicity, irritation, sensitization and acute systemic toxicity testing. All tests successfully met the required acceptance criteria, demonstrating that the patient contacting materials used in the devices are biocompatible.
### Sterilization & Shelf Life
Validation of the sterilization cycle has been conducted in accordance with ISO 11135:2014, ISO10993-7:2018, ISO 11737-1:2018.
Shelf life and packaging testing were performed to support the labeled shelf life:
- Accelerated aging study has been performed (ASTM F 1980) and in parallel a real-● time aging study was started
- package integrity was determined by:(referenced standards: ISO11607-1, ASTM F1886, ASTM F88, ASTM F1140, ASTM F1929)
- packaging system performance testing was performed according to reference standard ● ISTA-2A.
All tests were successfully completed. Real time aging study has been started.
Navigation Compatibility (VenSure™ Nav)
Bench testing was conducted to determine the imaging accuracy of the device (VenSure™ Nav). Test results demonstrate functionality and compatibility with the Fiagon Navigation
{7}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for Fiagon. The logo consists of a green circle with a white "f" inside, followed by the word "fiagon" in black lowercase letters. There is a green dot after the word "fiagon".
system and support the claim of substantial equivalence to the predicate, Medtronic EM Sinus Dilation System.
# Electromagnetic Compatibility
The device VenSure™ Nav is used in connection with the Fiagon Navigation system. Compliance to IEC 60601-1-2, 4th edition of the new device could be demonstrated by leveraging existing data of cleared navigation probes used with the Fiagon Navigation System.
### 8. Conclusion
Based on the indications for use, technological characteristics, performance testing, and comparison to the predicates, Fiagon GmbH demonstrated that the VenSure™ and VenSure™ Nav are substantially equivalent to the predicate devices identified in this submission, and they are as safe and perform in an equivalent manner to the stated predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.