K201468 · Shenzhen Creative Industry Co., Ltd. · DQA · Jan 25, 2021 · Cardiovascular
Device Facts
Record ID
K201468
Device Name
Pulse Oximeter
Applicant
Shenzhen Creative Industry Co., Ltd.
Product Code
DQA · Cardiovascular
Decision Date
Jan 25, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2700
Device Class
Class 2
Attributes
Pediatric
Indications for Use
This Pulse Oximeter is intended for measuring the functional oxygen saturation (SpO2) and pulse rate (PR). It is intended for spot check and continuous recording of SpO2, PR of adult or pediativ patients in hospitals, clinics, or home. This device is not intended for continuous monitoring.
Device Story
Wrist-worn pulse oximeter; measures SpO2 and pulse rate via transmittance optical sensor (red 660nm, infrared 905nm) applied to finger. Device displays results on 1.44" color TFT LCD; stores up to 500 hours of data. Data transfer to PC/smartphone via USB or Bluetooth. Used in hospitals, clinics, or home by patients or clinicians. Provides spot-check and continuous recording; output used by healthcare providers for clinical assessment of oxygenation status. Benefits include portable, non-invasive monitoring of physiological parameters.
Clinical Evidence
Clinical testing conducted per Annex EE of ISO 80601-2-61:2011 to evaluate SpO2 accuracy in human subjects. Results demonstrate compliance with accuracy requirements (±3% for 70-100% SpO2 range).
Technological Characteristics
Wrist-worn device; transmittance optical sensor (660nm/905nm); 1.44" color TFT LCD; rechargeable 3.7V 500mAh Li-ion polymer battery. Connectivity: USB, Bluetooth. Biocompatibility per ISO 10993-1. Standards: ANSI AAMI ES60601-1, IEC 60601-1-2, IEC 60601-1-11, ISO 80601-2-61.
Indications for Use
Indicated for adult and pediatric patients in hospitals, clinics, or home settings for spot check and continuous recording of functional oxygen saturation (SpO2) and pulse rate (PR). Not intended for continuous monitoring.
Regulatory Classification
Identification
An oximeter is a device used to transmit radiation at a known wavelength(s) through blood and to measure the blood oxygen saturation based on the amount of reflected or scattered radiation. It may be used alone or in conjunction with a fiberoptic oximeter catheter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. Food & Drug Administration".
Shenzhen Creative Industry Co., Ltd. % Charles Mack Principal Engineer International Regulatory Consultants (IRC) 2950 E Lindrick Drive Chandler, Arizona 85249
Re: K201468
Trade/Device Name: Pulse Oximeter, AP-10 Regulation Number: 21 CFR 870.2700 Regulation Name: Oximeter Regulatory Class: Class II Product Code: DQA Dated: December 11, 2020 Received: December 28, 2020
Dear Charles Mack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Actinclude requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Todd Courtney Assistant Director DHT1C: Division of ENT. Sleep Disordered Breathing, Respiratory and Anesthesia Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known)
Device Name Pulse Oximeter, AP-10
Indications for Use (Describe)
This Pulse Oximeter is intended for measuring the functional oxygen saturation (SpO2) and pulse rate (PR). It is intended for spot check and continuous recording of SpO2, PR of adult or pediativ patients in hospitals, clinics, or home. This device is not intended for continuous monitoring.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
#### 510(k) Summarv
This 510(k) summary is submitted in accordance with the requirements of 21 CFR Part 807.92.
Date: December 11, 2020
- 1. Company and Correspondent submitting:
Shenzhen Creative Industry Co., Ltd. Floor 5, BLD 9, Baiwangxin High-Tech Industrial Park, Songbai Road, Xili Street, Nanshan District, Shenzhen, P.R.China 518110 Tel: +86-755-26433325 Fax: +86-755-26435433 Contact - Mrs. Jia Wang General Manager US Agent and Correspondent Mr. Charles Mack Principal Engineer IRC USA 2950 E Lindrick Drive, Chandler, Arizona 85249 USA Tel: 931-6254938 Email: charliemack@irc-us.com
- 2. Device:
Trade/proprietary name: Pulse Oximeter, AP-10 Common Name : Pulse Oximeter Classification Name : Oximeter Product Code : DQA Regulation Number: 21CFR870.2700 Device Class : 2
{4}------------------------------------------------
- 3. Predicate Devices:
| Manufacturer | Device | 510(k)<br>Number |
|----------------------------------------|--------------------------------|------------------|
| Shenzhen Creative<br>Industry Co., Ltd | Handheld Pulse Oximeter, SP-20 | K172792 |
#### Reason for submission: New device 4.
#### Device Description: 5.
This AP-10 Pulse Oximeter is wrist-worn. It measures the SPO2 physiological parameter, and the measurement results are displayed and stored. The data can be transferred to a phone or PC via a USB connection or via Bluetooth. The AP-10 Pulse Oximeter measures SPO2 and Pulse Rate simultaneously. The device can store up to 500 hours of SPO2 and Pulse rate data.
The following page exhibits a brief tabular display of the device functions and operation as compared to the predicate device.
# Indication for use:
This Pulse Oximeter is intended for measuring and recording the functional oxygen saturation (SpO2) and pulse rate (PR). It is intended for spot check and continuous recording of SpO2, PR of adult or pediatric patients in hospitals, clinics, or home. This device is not intended for continuous monitoring.
# Comparison with the predicate device:
Shenzhen Creative Industry Co., Ltd. believes that the Pulse Oximeter, AP-10 is substantially equivalent to the Shenzhen Creative Industry Co., Ltd; Handheld Pulse Oximeter, Model SP-20 (K172972).
{5}------------------------------------------------
#### Comparison to Predicate Devices
| Characteristics | Submitted Device | Predicate Device | Differences Discussion |
|---------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | Shenzhen Creative Industry Co., LTD. | Shenzhen Creative Industry Co., LTD. | N/A |
| Name and model | Pulse Oximeter, AP-10 | Handheld Pulse Oximeter, SP-20 | N/A |
| 510(K) Number | Pending | K172792 | N/A |
| Product Code | DQA | DQA | Same |
| Indication for Use | This Pulse Oximeter is intended for measuring and<br>recording the functional oxygen saturation (SpO2)<br>and pulse rate (PR). It is intended for spot check and<br>continuous recording of SpO2, PR of adult or<br>pediatric patients in hospitals, clinics, or home. This<br>device is not intended for continuous monitoring. | This Handheld Pulse Oximeter is intended for<br>measuring and recording the pulse rate,<br>functional oxygen saturation (SpO2), and<br>temperature (optional). It is intended for spot<br>check and recording of SpO2, pulse rate, and<br>temperature of adult and pediatric patients in<br>clinical institutions and homes. The Handheld<br>Pulse Oximeter is not intended for active<br>continuous monitoring. | Similar. |
| Type of SpO2<br>Sensor | Transmittance Optical Sensor | Transmittance Optical Sensor | The difference of SpO2 Sensor<br>does not raise any new questions |
| SpO2 Module | KM-SPO-04 SpO2 Module with KS-AR01 Probe | KM-SPO-01 SpO2 Module with KS-C01, KS-<br>YW02 probe | of safety and effectiveness and<br>still complies with the ISO80601-<br>2-61 |
| Application Site | Finger | Finger | |
| Output | Red (660 nm) @ 2 mW maximum | Red (660 nm) @ 2 mW maximum | |
| Wavelength and<br>Radiant Power of<br>SpO2 Sensor | Infrared (905 nm) @ 2 mW maximum | Infrared (905 nm) @ 2 mW maximum | |
| Display | 1.44" color TFT LCD | 3.5" color TFT LCD | The display size difference does<br>not raise any new questions of<br>safety and effectiveness. |
| Measuring Mode | Spot-check and Continuous recording | Spot-check and Continuous recording | Spot-check and Continuous<br>recording |
| Display Range | SpO2: 0-100 %<br>Pulse Rate: 30-250 bpm | SpO2: 0-100 %<br>Pulse Rate: 30-240 bpm | The Pulse Rate display range of<br>AP-10 is wider than SP-20, but<br>still complies with ISO80601-2-61 |
| Characteristics | Submitted Device | Predicate Device | Differences Discussion |
| Display | SpO2: 1% | SpO2: 1% | SpO2: 1% |
| Resolution | Pulse Rate: 1 bpm | Pulse Rate: 1 bpm | Pulse Rate: 1 bpm |
| SpO2 | SpO2: | SpO2: | SpO2: |
| Measurement | Adult and Pediatric: | Adult and Pediatric: | Adult and Pediatric: |
| Accuracy | ±3% (during 70%-100%) | ±3% (during 70%-100%) | ±3% (during 70%-100%) |
| | Undefined (during 0-70%) | Undefined (during 0-70%) | Undefined (during 0-70%) |
| Pulse Rate | Pulse Rate: ±2bpm or ±2% (whichever is greater) | Pulse Rate: ±2bpm or ±2% (whichever is | Pulse Rate: ±2bpm or ±2% |
| Accuracy | | greater) | (whichever is greater) |
| Data Memory | Up to 24 hours data storage (500 pieces of record at | Up to 500 hours of data storage for SpO2 and | The little difference does not raise |
| | most) | PR and can be recalled | any new questions of safety and |
| | | | effectiveness. |
| Power Supply | Rechargeable Lithium-lon Polymer Battery | Rechargeable Lithium-Ion Polymer Battery | Both devices are powered by a |
| | (3.7V,500mAh) | (3.7V d.c.,1000mAh) | lithium battery, which matches the |
| | | | design requirement and complies |
| | | | with the applicable standards, |
| | | | including IEC60601-1, IEC60601- |
| | | | 1-2, and IEC62133, etc. |
| AC Power for | 100-240VAC, 50-60Hz,15VA max | 100-240VAC, 50-60Hz,0.5A | Both devices are powered by a |
| Battery Charger | | | lithium battery that full match the |
| | | | design requirement and comply |
| | | | with the applicable standards, |
| | | | including IEC60601-1 and |
| | | | IEC60601-1-2. |
| Type of | Internal Powered | Internal Powered | Internal Powered |
| Protection | | | |
| Degree of | Type BF - applied part | Type BF - applied part | Type BF - applied part |
| Protection - | | | |
| sensor | | | |
| Enclosure | IP22 | IP22 | The enclosure Degree of Ingress |
| Degree of | | | Protection complies with the |
| Ingress | | | applicable standards, including |
| Protection | | | IEC60601-1 and ISO80601-2-61. |
| Characteristics | Submitted Device | Predicate Device | Differences Discussion |
| Dimension<br>(LxWxH) | Watch Case: D 56mmx W 44mmxH 16mm | 158 mm (L) x 73 mm (W) x 25 mm (H) | The physical dimension difference<br>does not raise any new questions<br>of safety and effectiveness. |
| Weight | Net Weight: about 45g | 230g (including battery) | The weight difference does not<br>raise any new questions of safety<br>and effectiveness. |
| Patient Contact<br>Materials | Complies with ISO10993-1 | Complies with ISO10993-1 | Complies with ISO10993-1 |
| Data update | USB, Bluetooth | N/A | The AP-10 can upload data to the<br>PC through USB or to a<br>smartphone through Bluetooth. It<br>complies with the applicable<br>standard FCC part 15B. |
| Operating | Operating temperature: 5~40°C<br>Operating humidity: 15%~93% (non-condensing)<br>Atmospheric pressure: 70kPa~106kPa | Operating Temperature: 5°C ~40°C<br>Operating Humidity: 15%~93%, non-<br>condensing<br>Atmospheric pressure: 70kPa~106kPa | Operating Temperature: 5°C<br>~40°C<br>Operating Humidity: 15%~93%,<br>non-condensing<br>Atmospheric pressure:<br>70kPa~106kPa |
| Storage | Ambient temperature: -20°C ~60°C,<br>Relative humidity 10%~95%,<br>Atmospheric pressure: 50kPa~107.4kPa. | Ambient temperature: -20°C~60°C<br>Relative humidity: 10%~95%<br>Atmospheric pressure: 50kPa~107.4kPa | Ambient temperature: -20°C<br>~60°C<br>Relative humidity: 10%~95%<br>Atmospheric pressure:<br>50kPa~107.4kPa |
{6}------------------------------------------------
{7}------------------------------------------------
{8}------------------------------------------------
### Testing Summary:
The following performance data were provided in support of the substantial equivalence determination.
## Non-Clinical:
## Safety and EMC
To verify the basic safety and essential performance of the AP-10, we performed the tests as noted below:
ANSI AAMI ES60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012
IEC 60601-1-2:2007 Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests
IEC 60601-1-11 Edition 2.0 2015-01Medical electrical equipment - Part 1-11: General requirements for basic safety and essential performance - Collateral Standard: Requirements for medical electrical equipment and medical electrical systems used in the home healthcare environment
The test results demonstrate the subject device complies with the standard's requirements.
### Biocompatibility
Patient contact classification is surface contact <24 hours. The biocompatibility evaluation for the subject device was conducted in accordance with ISO 10993-1. and the testing included the following test results of which demonstrate the biocompatibility of the subject device:
- Cytotoxicity
- Skin Sensitization
- Skin Irritation
### Performance
For the subject device, AP-10, it meets the performance standards: ISO 80601-2-61 First edition 2011-04-01 Medical electrical equipment - Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment
{9}------------------------------------------------
## Software Verification and Validation
Software documentation, including verification & validation, was provided following FDA Guidance: Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices for software with a moderate level of concern.
# Cleaning Validation
The AP-10 was cleaned following the procedure defined in the User's Manual. The device was checked for performance following the cleaning and met all performance requirements following the cleaning.
# Shelf Life:
The Pulse Oximeter, AP-10, is not subject to the shelf life, as the device doesn't contain any sterile or degradable components.
# Clinical:
Clinical testing was conducted per Annex EE Guideline for evaluating and documenting SpO2 accuracy in human subjects of ISO 80601-2-61:2011 Medical electrical equipment – Part 2-61: Particular requirements for basic safety and essential performance of pulse oximeter equipment.
# Conclusions:
In accordance with the Federal Food, Drug and Cosmetic Act, 21 CFR Part 807, and based on the information provided in this premarket notification Shenzhen Creative Industry Co., Ltd. concludes that the Pulse Oximeter, AP-10 is substantially equivalent to predicate device as described herein. The subject device has the same intended use as the predicate device and the technological differences do not raise different questions of safety and effectiveness
END
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.