ALTAPORE MIS is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities, posterolateral spine and pelvis). ALTAPORE MIS can be used by itself, with autograft as a bone graft extender or with autogenous bone marrow aspirate. These osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. ALTAPORE MIS resorbs and is replaced with bone during the healing process.
Device Story
Altapore MIS is a bioactive, osteoconductive silicate-substituted calcium phosphate bone void filler; supplied as microgranules (1-2 mm) in an absorbable aqueous gel carrier within a 7.5ml cartridge. Used by surgeons in clinical settings to fill non-structural bony voids or gaps in extremities, posterolateral spine, or pelvis. Device acts as a scaffold; forms surface apatite layer in vivo to support natural bone growth; undergoes cell-mediated remodeling and resorption. Delivered via manual applicator; can be used standalone, with autograft, or with bone marrow aspirate. Benefits patient by facilitating bone healing and defect repair.
Clinical Evidence
No clinical data. Performance assessed via bench testing, including chemical analysis, biocompatibility (ISO 10993-1), pyrogenicity (ISO 10993-11, USP<85>), and sterility validation (ANSI/AAMI/ISO 11137-2). Shelf-life validated for two years.
Technological Characteristics
Silicate-substituted calcium phosphate (Ca, P, O, H, Si) microgranules (1-2 mm) with 80-85% porosity in aqueous gel carrier. Bioactive/osteoconductive. Sterilized via radiation per ANSI/AAMI/ISO 11137-2. Supplied in 7.5ml cartridge with applicator.
Indications for Use
Indicated for patients requiring filling of bony voids or gaps in the skeletal system, including extremities, posterolateral spine, and pelvis, resulting from surgical creation or traumatic injury. Not for use in defects intrinsic to the stability of the bony structure.
Regulatory Classification
Identification
A resorbable calcium salt bone void filler device is a resorbable implant intended to fill bony voids or gaps of the extremities, spine, and pelvis that are caused by trauma or surgery and are not intrinsic to the stability of the bony structure.
Special Controls
*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA.” See § 888.1(e) of this chapter for the availability of this guidance.
Actifuse™ MIS System Bone Graft Substitute (K082575)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: on the left, there is a symbol representing the Department of Health & Human Services - USA, and on the right, there is the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The word "FDA" is placed inside a blue square. The text is stacked in three lines, with "ADMINISTRATION" on the bottom line.
June 30, 2020
Baxter Healthcare Corporation Phillip Romei Specialist, Global Regulatory Affairs 32650 N. Wilson Road Round Lake, Illinois 60073
Re: K201464
Trade/Device Name: Altapore MIS Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable Calcium Salt Bone Void Filler Device Regulatory Class: Class II Product Code: MQV Dated: May 19, 2020 Received: June 4, 2020
Dear Mr. Romei:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Laura Rose, Ph.D. Acting Assistant Director DHT6C: Division of Restorative, Repair, and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) TBD
Device Name Altapore MIS
Indications for Use (Describe)
ALTAPORE MIS is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities, posterolateral spine and pelvis).
ALTAPORE MIS can be used by itself, with autograft as a bone graft extender or with autogenous bone marrow aspirate. These osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. ALTAPORE MIS resorbs and is replaced with bone during the healing process.
| Type of Use (Select one or both, as applicable) | | |
|--------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------|---------------------------------------------|
| <table><tr><td>Prescription Use (Part 21 CFR 201 Subpart D)</td></tr><tr><td>Over-The-Counter Use (21 CFR 201 Subpart C)</td></tr></table> | Prescription Use (Part 21 CFR 201 Subpart D) | Over-The-Counter Use (21 CFR 201 Subpart C) |
| Prescription Use (Part 21 CFR 201 Subpart D) | | |
| Over-The-Counter Use (21 CFR 201 Subpart C) | | |
|XJ Prescription Use (Part 21 CFR 801 Subpart D)
| | Over-The-Counter Use (21 CFR 801 Subpart C)
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
### Section 5. 510(k) Summary
June 30, 2020
#### OWNER:
Baxter Healthcare Corporation One Baxter Parkway Deerfield, Illinois 60015
#### CONTACT PERSON:
Phillip Romei Specialist, Global Regulatory Affairs 25212 W. Illinois Route 120 Round Lake, IL 60073 Telephone: (224) 948-2652 Fax: (224) 270-4119
#### IDENTIFICATION OF THE DEVICE:
Common Name: Filler, bone void, calcium compound Trade Name: ALTAPORE MIS Classification Panel: 87 Orthopedic Regulation Number: 21 CFR 888.3045 Regulation Name: Resorbable calcium salt bone void filler devices Regulatory Class: Class II Product Code: MQV
| Model Number | Name |
|--------------|---------------------------------------------------|
| 1508032 | Altapore MIS System Applicator + Cartridge, 7.5ml |
| 1508047 | Altapore MIS System Cartridge, 7.5ml |
#### Table 1. Model Numbers for Altapore MIS
{4}------------------------------------------------
# PREDICATE DEVICE:
| Device | Company | Predicate 510(k) | Clearance Date |
|-------------------------------------------------------------------------|----------------------------------|------------------|-------------------|
| Altapore Bone Graft<br>Substitute<br>(Predicate Device) | Baxter Healthcare<br>Corporation | K192363 | January 9, 2020 |
| Actifuse™ MIS<br>System Bone Graft<br>Substitute,<br>(Predicate Device) | ApaTech Limited1 | K082575 | November 25, 2008 |
#### Table 2. Predicate Device
1 ApaTech Limited was acquired by Baxter Healthcare Corporation in 2010
The proposed device is substantially equivalent to the predicate devices, Altapore and Actifuse MIS System. Actifuse MIS System is owned and manufactured by Baxter, and has identical packaging, chemical composition, and sterilization as the proposed device. The proposed and predicate Altapore devices have identical chemical composition, physical structure, sterilization, and manufacturing process.
# DESCRIPTION OF THE DEVICE:
Altapore MIS System is a is a bioactive and osteoconductive silicate-substituted calcium phosphate bone void filler. The interconnected and open porous structure of the silicatesubstituted calcium phosphate phase of Altapore is similar to human cancellous bone and is intended to support bone growth with macro and micro- porosity. Altapore is composed solely of elements that exist naturally in normal bone (Ca. P. O, H. Si).
Altapore MIS System is supplied in a sterile 7.5ml cartridge and contains Altapore microgranules, sized 1-2 mm, 80-85% total porosity, suspended in an absorbable aqueous gel carrier. Altapore does not set in-situ following implantation.
Altapore MIS System (hereafter referred to as "Altapore MIS) is designed for use as a standalone bone graft substitute or as an autograft extender. While not necessary, the product can be mixed with Bone Marrow Aspirate (BMA) or autologous bone at the discretion of the surgeon.
Altapore MIS is bioactive based on in vitro studies that show it forms a surface apatite layer when submerged in simulated body fluid that contains the same ion concentrations as human blood plasma. This apatite layer provides scaffolding onto which the patient's new bone will grow, allowing complete repair of the defect.
{5}------------------------------------------------
Altapore MIS is osteoconductive based on in vivo animal studies that show it achieves bone healing in a critical defect model as confirmed with radiographic. histolopathological, histomorphometric, and mechanical analyses. Altapore MIS undergoes cell-mediated remodeling and is replaced by natural bone.
### INDICATIONS FOR USE:
Altapore MIS is an implant intended to fill bony voids or gaps of the skeletal system (i.e., extremities, posterolateral spine and pelvis).
Altapore MIS can be used by itself, with autograft as a bone graft extender or with autogenous bone marrow aspirate. These osseous defects may be surgically created or the result of traumatic injury to the bone and are not intrinsic to the stability of the bony structure. Altapore MIS resorbs and is replaced with bone during the healing process.
### PURPOSE OF SUBMISSION:
The basis for this premarket notification is the addition of the Altapore MIS System configurations to the Altapore product line. The additional configurations consist of a 7.5ml cartridge and delivery applicator and a separate 7.5ml refill cartridge. The proposed new cartridge and applicator materials are currently used in Baxter's Altapore and Actifuse™ MIS System product lines and have been previously cleared under 510(k) premarket notifications K192363 (cleared date 9th Jan, 2020) and K082575 (cleared date 25th Nov, 2008) respectively. The hydroxyapatite granules used in the proposed Altapore MIS System configurations and the predicate devices Actifuse MIS and Altapore are chemically identical. These modifications do not impact the intended use or the fundamental technology of the devices.
# TECHNOLOGICAL CHARACTERISTICS AND SUBSTANTIAL EQUIVALENCE:
The proposed Altapore MIS device has equivalent technological characteristics relative to Baxter's currently cleared Altapore product line, cleared under 510(k) premarket notification K192363 (cleared January 9th, 2020).
### DISCUSSION OF NON-CLINICAL DATA:
Baxter Healthcare Corporation conducts risk analyses and design verification tests based on the result of these analyses. All test results meet the acceptance criteria, and support that the proposed devices are appropriately designed for their intended use. All testing was performed on the configuration of the devices presented in this premarket notification.
{6}------------------------------------------------
### Performance Data
The performance of the Altapore MIS System device configurations was assessed by analyzing the Altapore MIS System final product for chemical changes relative to the Altapore predicate product's specification. All testing confirmed that the new configurations had no impact to the chemistry or structure of the Altapore MIS System product.
# Biocompatibility
All materials found in these devices that are the subject of this submission have been previously cleared under Baxter's 510(k) premarket notifications K192363 (clearance date January 9th, 2020), and K082575 (clearance date November 25th, 2008).
Biocompatibility assessments were conducted based on ISO-10993-1, Biological Evaluation of Medical Devices for permanent duration, implant device, tissue/bone contact, and FDA Guidance, Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process", as recommended in the "Class II Special Controls Guidance Document: Resorbable Calcium Salt Bone Void Filler Device; Guidance for Industry and FDA".
ALTAPORE is non-pyrogenic based on material-mediated pyrogenicity testing conducted per ISO 10993-11 and bacterial endotoxin testing performed per USP<85>.
# Sterility
The Altapore MIS System product line is sterilized with radiation. The minimum sterilizing dose (MSD) required to provide a 106 Sterility Assurance Level (SAL) for this (sub) category was established and validated at the manufacturing facility as described in ANSI/AAMI/ISO 11137-2, "Sterilization of health care products- Radiation-Part 2: Establishing the Sterilization Dose."
These products are labeled "Sterile". Package Verification testing is based on Visual Inspection, Seal Strength, and Bubble Leak testing.
# Shelf Life
Testing performed on representative samples supports a shelf-life claim of two (2) years for Altapore MIS.
# CONCLUSION:
Altapore MIS System is as safe and effective as the predicate Altapore Bone Graft Substitute. Both devices share indications for use, technological characteristics, and
{7}------------------------------------------------
principles of operation. The only differences between the two devices are the changes to the dimensions and materials of the containers housing the devices. The non-clinical data demonstrate that the subject device is substantially equivalent and performs comparably to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.