← Product Code [LAF](/productcode/LAF) · K201442

# Lin-Zhi International, Inc. (LZI) Methamphetamine Enzyme Immunoassay for Pictus Analyzers (K201442)

_Diatron Group · LAF · Jul 31, 2020 · Clinical Toxicology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K201442

## Device Facts

- **Applicant:** Diatron Group
- **Product Code:** [LAF](/productcode/LAF.md)
- **Decision Date:** Jul 31, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.3610
- **Device Class:** Class 2
- **Review Panel:** Clinical Toxicology

## Indications for Use

The Lin-Zhi International, Inc. (LZI) Methamphetamine Enzyme Immunoassay for Pictus Analyzers is intended for the qualitative determination of d-methamphetamine in human urine at a cutoff value of 500 ng/mL. The system was calibrated with d-methamphetamine. The assay provides a rapid screening procedure for determining the presence of d-methamphetamine in urine. The assay provides only a preliminary analytical result. A more specific alternative analytical chemistry method must be used in order to obtain a confirmed analytical result. Gas or Liquid Chromatography/Mass Spectrometry (GC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.

## Device Story

The device is a homogeneous enzyme immunoassay for the qualitative screening of d-methamphetamine in human urine. It is designed for use on Pictus automated clinical chemistry analyzers by trained laboratory professionals. The system measures absorbance at 340 nm using a photodiode detector; the analyzer automatically calculates the change in absorbance to determine the presence of the drug relative to a 500 ng/mL cutoff. Results are reported as positive or negative. Because the assay provides only preliminary results, positive findings require confirmation via more specific methods like GC/MS or LC/MS. Clinical decision-making should incorporate professional judgment alongside these results. The device benefits patients and clinicians by providing a rapid, automated screening tool for drug abuse testing in a clinical laboratory setting.

## Clinical Evidence

No clinical data. Bench testing only. Method comparison study performed on 98 human urine samples against LC/MS reference method, showing 100% agreement for positive samples and 98% for negative samples. Precision studies (within-run and total) conducted per CLSI EP05-A2. Cross-reactivity tested for structurally related compounds.

## Technological Characteristics

Homogeneous enzyme immunoassay. Reagents: mouse monoclonal anti-methamphetamine antibodies, G6PDH-labeled drug conjugate, NAD, G6P, sodium azide preservative. Detection: spectrophotometric (340 nm). Platform: Diatron Pictus 700 clinical chemistry analyzer. Qualitative output. Traceable to d-methamphetamine standard.

## Regulatory Identification

A methamphetamine test system is a device intended to measure methamphetamine, a central nervous system stimulating drug, in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of methamphetamine use or overdose.

## Special Controls

*Classification.* Class II (special controls). A methamphetamine test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

## Predicate Devices

- LZI Methamphetamine Immunoassay reagent ([K113661](/device/K113661.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

FDA

U.S. FOOD &amp; DRUG

ADMINISTRATION

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

ASSAY ONLY

## I Background Information:

A 510(k) Number

K201442

B Applicant

Diatron Group

C Proprietary and Established Names

Lin-Zhi International, Inc. (LZI) Methamphetamine Enzyme Immunoassay for Pictus Analyzers

D Regulatory Information

|  Product Code(s) | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  LAF | Class II | 21 CFR 862.3610 - Methamphetamine Test System | TX - Clinical Toxicology  |

## II Submission/Device Overview:

A Purpose for Submission:

Clearance of the LZI Methamphetamine assay (k113661) on the Diatron Pictus 700 clinical chemistry analyzer (k151487).

B Measurand:

d-Methamphetamine.

C Type of Test:

Qualitative, homogenous enzyme immunoassay.

Food and Drug Administration

10903 New Hampshire Avenue

Silver Spring, MD 20993-0002

www.fda.gov

{1}

III Intended Use/Indications for Use:

A Intended Use(s):
See Indications for Use below.

B Indication(s) for Use:
The Lin-Zhi International, Inc. (LZI) Methamphetamine Enzyme Immunoassay for Pictus Analyzers is intended for the qualitative determination of d-methamphetamine in human urine at a cutoff value of 500 ng/mL. The system was calibrated with d-methamphetamine. The assay provides a rapid screening procedure for determining the presence of d-methamphetamine in urine.

The assay provides only a preliminary analytical result. A more specific alternative analytical chemistry method must be used in order to obtain a confirmed analytical result. Gas or Liquid Chromatography/Mass Spectrometry (GC/MS or LC/MS) are the preferred confirmatory methods. Clinical consideration and professional judgment should be exercised with any drug of abuse test result, particularly when the preliminary test result is positive.

C Special Conditions for Use Statement(s):
Rx - For Prescription Use Only

D Special Instrument Requirements:
Diatron Pictus 700 clinical chemistry analyzer

IV Device/System Characteristics:

A Device Description:
The LZI Methamphetamine Enzyme Immunoassay for Pictus Analyzers is a kit comprised of two reagents, an R1 and R2, which are bottled separately but sold together within the kit. Antibody/Substrate Reagent (R1): Contains mouse monoclonal anti-methamphetamine antibodies, glucose-6-phosphate (G6P), nicotinamide adenine dinucleotide (NAD), stabilizers, and sodium azide (0.09%) as a preservative. Enzyme-drug Conjugate Reagent (R2): Contains methamphetamine-labeled glucose-6-phosphate dehydrogenase (G6PDH) in buffer with sodium azide (0.09%) as a preservative.

B Principle of Operation:
The Lin-Zhi International, Inc. (LZI) Methamphetamine Enzyme Immunoassay for Pictus Analyzers is a homogeneous enzyme immunoassay with ready-to-use liquid reagent. The assay is based on competition between drug in the sample and drug labeled with the enzyme glucose-6-phosphate dehydrogenase (G6PDH) for a fixed amount of antibody in the reagent. Enzyme activity decreases upon binding to the antibody, and the drug concentration in the sample is measured in terms of enzyme activity. In the absence of drug in the sample, methamphetamine-labeled G6PDH conjugate is bound to antibody, and the enzyme activity is inhibited. On the other hand, when drug is present in the sample, antibody binds to the free drug; and the unbound methamphetamine-labeled G6PDH then exhibits its maximal enzyme activity. Active enzyme converts nicotinamide adenine dinucleotide (NAD) to NADH, resulting in an absorbance change that can be measured spectrophotometrically at a 340 nm primary wavelength.

K201442 - Page 2 of 7

{2}

V Substantial Equivalence Information:

A Predicate Device Name(s):
LZI Methamphetamine Enzyme Immunoassay

B Predicate 510(k) Number(s):
K113661

C Comparison with Predicate(s):

|  Device & Predicate Device(s): | K201442 | K113661  |
| --- | --- | --- |
|  Device Trade Name | Lin-Zhi International, Inc. (LZI) Methamphetamine Enzyme Immunoassay for Pictus Analyzers | LZI Methamphetamine Enzyme Immunoassay  |
|  General Device Characteristic Similarities |  |   |
|  Intended Use/Indications For Use | For the qualitative determination of methamphetamine in human urine, at a cutoff value of 500 ng/mL. | Same  |
|  Mode of Detection | Photometric/photodiode | Same  |
|  Wavelength to measure Methamphetamine reagent reactions | 340 nm | Same  |
|  Cut off level | 500 ng/mL | Same  |
|  Sample type | Human urine | Same  |
|  General Device Characteristic Differences |  |   |
|  Results output | Qualitative | Qualitative and semi-quantitative  |
|  Instrument Platform | Diatron Pictus 700 analyzer | Cleared for clinical chemistry analyzers (open system)  |

VI Standards/Guidance Documents Referenced:

CLSI EP05-A2: User protocol for evaluation of Qualitative Test Performance; Approved Guideline, 2004.
CLSI EP06-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline, 2003.
CLSI EP07-A2: Interference Testing in Clinical Chemistry, Approved Guideline, 2005.

K201442 - Page 3 of 7

{3}

VII Performance Characteristics (if/when applicable):

A Analytical Performance:

1. Precision/Reproducibility:

Two studies were conducted to estimate the within-run and total imprecision of the test system following the CLSI EP05-A2 guidelines. Nine samples at 0, 125, 250, 375, 500, 625, 750, 875 and 1000 ng/mL of methamphetamine (cutoff, ±25 %, ±50 %, ±75%, and ±100 % of the cut-off) were evaluated in these studies. The within-run imprecision study was performed by testing each of the nine samples in 20 consecutive testing. The total imprecision study was performed by testing each of the nine samples in duplicate twice per day over 21 days. The data were assessed for “positive” and “negative” interpretation at either side of the 500 ng/mL cutoff. The results are shown in the two summary tables.

Total Precision

|  Level (ng/mL) | N | % Negative | # of Negative | % Positive | # of Positive  |
| --- | --- | --- | --- | --- | --- |
|  0 | 84 | 100% | 84 | 0% | 0  |
|  125 | 84 | 100% | 84 | 0% | 0  |
|  250 | 84 | 100% | 84 | 0% | 0  |
|  375 | 84 | 100% | 84 | 0% | 0  |
|  500 | 84 | 49% | 41 | 51% | 43  |
|  625 | 84 | 0% | 0 | 100% | 84  |
|  750 | 84 | 0% | 0 | 100% | 84  |
|  875 | 84 | 0% | 0 | 100% | 84  |
|  1000 | 84 | 0% | 0 | 100% | 84  |

Within Run Precision

|  Level (ng/mL) | N | % Negative | # of Negative | % Positive | # of Positive  |
| --- | --- | --- | --- | --- | --- |
|  0 | 20 | 100% | 20 | 0% | 0  |
|  125 | 20 | 100% | 20 | 0% | 0  |
|  250 | 20 | 100% | 20 | 0% | 0  |
|  375 | 20 | 100% | 20 | 0% | 0  |
|  500 | 20 | 60% | 12 | 40% | 8  |
|  625 | 19 | 0% | 0 | 100% | 19  |
|  750 | 20 | 0% | 0 | 100% | 20  |
|  875 | 20 | 0% | 0 | 100% | 20  |
|  1000 | 20 | 0% | 0 | 100% | 20  |

2. Linearity:

Not applicable as this is a qualitative assay.

K201442 - Page 4 of 7

{4}

K201442 - Page 5 of 7

3. Analytical Specificity/Interference:

Four structurally related compounds were tested in duplicate at the concentrations described in the table below. The reagents used in this assay are the same as those used in k113661, and the information describing analytical specificity of the device reviewed k113661 was also considered in this review.

|  Compound | Target Concentration (ng/mL)  |
| --- | --- |
|  d-Methamphetamine | 500  |
|  d-Amphetamine | 50,000  |
|  Methylenedioxyamphetamine (MDA) | 72,500  |
|  Methylenedioxymethylamphetamine (MDMA) | 1,500  |

Results:

|  Compound | Target Concentration (ng/mL) | P700 Qualitative Result | % Cross Reactivity  |
| --- | --- | --- | --- |
|  d-Methamhetamine | 500 | Positive | 103.8  |
|  d-Amphetamine | 50,000 | Positive | 1.1  |
|  Methylenedioxyamphetamine (MDA) | 72,500 | Positive | 0.8  |
|  Methylenedioxymethylamphetamine (MDMA) | 1,500 | Positive | 34.5  |

The sponsor claims that the results from the testing of these four compounds are similar to the results from the labeling of the predicate device. These results are adequate to demonstrate that the interference and cross reactivity testing performed in support of k113661 are representative of the expected cross reactivity and interference of these compounds with the candidate test system.

4. Assay Reportable Range:

Not applicable. This is a qualitative test.

5. Traceability, Stability, Expected Values (Controls, Calibrators, or Methods):

The Lin-Zhi International, Inc. (LZI) Methamphetamine Enzyme Immunoassay for Pictus Analyzers is traceable to a commercially available d-methamphetamine standard.

6. Detection Limit:

Refer to the precision/reproducibility study section, above.

7. Assay Cut-Off:

Refer to the precision/reproducibility study section, above.

{5}

K201442 - Page 6 of 7

# B Comparison Studies:

1. Method Comparison with Predicate Device:

A method comparison study was conducted using 98 human urine samples spanning in methamphetamine concentration from zero to 2000 ng/mL as measured by LC/MS method. One discordant result was obtained. The results are summarized in the below tables.

|  Candidate device results | Negative by LC/MS analysis | < 50 % of the cutoff concentration by LC/MS analysis | Near Cutoff Negative (Between 50 % below the cutoff and the cutoff concentration by LC/MS analysis) | Near Cutoff Positive (Between the cutoff and 50 % above the cutoff concentration by LC/MS analysis) | High Positive (Greater than 50 % above the cutoff concentration by LC/MS analysis)  |
| --- | --- | --- | --- | --- | --- |
|  Positive | 0 | 0 | 1 | 20 | 28  |
|  Negative | 15 | 23 | 11 | 0 | 0  |

Discordant Sample:

|  Sample # | LC/MS (ng/mL) | Candidate device result  |
| --- | --- | --- |
|  61634620 | 488 | Positive  |

2. Matrix Comparison:

Not applicable. The test is only for urine specimens.

# C Clinical Studies:

1. Clinical Sensitivity:

Not applicable

2. Clinical Specificity:

Not applicable

3. Other Clinical Supportive Data (When 1. and 2. Are Not Applicable):

Not applicable

# D Clinical Cut-Off:

Not applicable; the device is for determining positive or negative. See Section above on Precision for analytical cutoff information.

{6}

E Expected Values/Reference Range:

Not applicable

VIII Proposed Labeling:

The labeling supports the finding of substantial equivalence for this device.

IX Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

K201442 - Page 7 of 7

---

**Source:** [https://fda-staging.innolitics.com/device/K201442](https://fda-staging.innolitics.com/device/K201442)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
