← Product Code [LNH](/productcode/LNH) · K201389

# Resoundant Acoustic Driver System (K201389)

_Resoundant, Inc. · LNH · Jul 10, 2020 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K201389

## Device Facts

- **Applicant:** Resoundant, Inc.
- **Product Code:** [LNH](/productcode/LNH.md)
- **Decision Date:** Jul 10, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1000
- **Device Class:** Class 2
- **Review Panel:** Radiology

## Indications for Use

The Resoundant Acoustic Driver System is intended for use with magnetic resonance diagnostic devices (MRDD) that include legally marketed MR elastography capabilities. It is indicated for generating acoustic vibrations in the body during an MRI exam, in order to assess tissue elasticity for diagnostic purposes as part of magnetic resonance elastography (MRE). When interpreted by a trained physician, this information can be useful in determining a diagnosis.

## Device Story

Resoundant Acoustic Driver System generates low-audio frequency transverse acoustic waves (20-350 Hz) to induce small vibrations in patient body during MRI exams; facilitates assessment of soft tissue elasticity for MRE. System comprises Active Driver (function generator, power amplifier, linear motor, pump head, microprocessor controller) and Passive Driver (lightweight enclosure with diaphragm, no electrical components) connected via flexible tubing. Used in clinical MRI environment; operated by MRE-trained technician or radiologist. Active Driver generates vibrations; Passive Driver transmits them to patient. Physician interprets resulting MRE images to aid diagnosis. Device provides mechanical input for MRE imaging sequence; does not perform image processing itself.

## Clinical Evidence

Bench testing only. Verification and validation included mechanical testing (dimensional, operational performance), electrical safety testing per IEC ANSI/AAMI ES60601-1 and IEC 60601-1-2, firmware testing, and packaging/distribution testing per ASTM D4169-16.

## Technological Characteristics

Electro-mechanical system; Active Driver includes function generator, power amplifier, linear motor, pump head, and microprocessor controller. Passive Driver is a non-electrical enclosure with a diaphragm. Frequency range 20-350 Hz (nominal 60 Hz). Complies with IEC ANSI/AAMI ES60601-1 and IEC 60601-1-2. Packaging tested per ASTM D4169-16.

## Regulatory Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

## Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- GE MR-Touch Option ([K083421](/device/K083421.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

July 10, 2020

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Resoundant, Inc. % John Hartigan, DRSc. MBA VP QA/RA 421 First Avenue SW, Suite 204W ROCHESTER MN 55902

Re: K201389

Trade/Device Name: Resoundant Acoustic Driver System Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: Class II Product Code: LNH Dated: May 21, 2020 Received: May 27, 2020

Dear Mr. Hartigan:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see

{1}------------------------------------------------

https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

For

Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K201389

Device Name

Resoundant Acoustic Driver System

#### Indications for Use (Describe)

The Resoundant Acoustic Driver System is intended for use with magnetic resonance diagnostic devices (MRDD) that include legally marketed MR elastography capabilities. It is indicated for generating acoustic vibrations in the body during an MRI exam, in order to assess tissue elasticity for diagnostic purposes as part of magnetic resonance elastography (MRE). When interpreted by a trained physician, this information can be useful in determining a diagnosis.

| Type of Use (Select one or both, as applicable)                                                                                                                                    |                                                                                                                                                                                   |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| <span style="font-size: 10pt;"></span> <span style="font-size: 10pt;"><b></b></span> <span style="font-size: 10pt;"><b>    Prescription Use (Part 21 CFR 801 Subpart D)</b></span> | <span style="font-size: 10pt;"></span> <span style="font-size: 10pt;"><b></b></span> <span style="font-size: 10pt;"><b>    Over-The-Counter Use (21 CFR 801 Subpart C)</b></span> |

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## 510(k) Summary

#### Resoundant Acoustic Driver System

| Date Prepared:     | 7 July 2020                                                                                                                       |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------|
| Submitter:         | Resoundant Inc.<br>421 First Ave SW<br>Suite 204W<br>Rochester, MN 55902<br>Telephone: 507-322-0011                               |
| Contact:           | John Hartigan, DRSc., MBA<br>VP QA/RA<br>421 First Ave SW, Suite 204W<br>Rochester, MN 55902<br>Telephone: 507-322-0011           |
| Proprietary Name:  | Resoundant Acoustic Driver System                                                                                                 |
| Common/Usual Name: | Acoustic Driver System<br>Accessory for Magnetic Resonance Image System for MRE                                                   |
| Regulation Name:   | Magnetic Resonance Image System<br>21 CFR 892.1000, Class II<br>Product Code – LNH, Magnetic Resonance Image System, Radiological |

Description: The Resoundant Acoustic Driver System includes an Active Driver, connective tubing and a Passive Driver. The system generates transverse acoustic waves in the low-audio frequency range in the body during an MRI exam. This allows assessment of elastic properties of soft tissue to aid in medical diagnosis as part of magnetic resonance elastography (MRE).

The Active Driver component is an electro-mechanical device that consists of a function generator, power-amplifier, linear motor and pump head, along with a microprocessor-based controller and power supply electronics, in an enclosure. The Passive Driver component is connected to the Active Driver through flexible tubing and is used to induce small vibrations in the scan subject. The Passive Driver is a lightweight enclosure containing no electrical components. It features a connection for flexible tubing and a diaphragm that is placed securely over patient clothing.

Indications for Use: The Resoundant Acoustic Driver System is intended for use with magnetic resonance diagnostic devices (MRDD) that include legally marketed MR elastography capabilities. It is indicated for generating acoustic vibrations in the body during an MRI exam, in order to assess tissue elasticity for diagnostic purposes as part of magnetic resonance

{4}------------------------------------------------

elastography (MRE). When interpreted by a trained physician, this information can be useful in determining a diagnosis.

### Substantial Equivalence:

The Resoundant Acoustic Driver System is substantially equivalent to the following predicate device:

Primary Predicate: GE MR-Touch Option 510(k): K083421 GE Medical Systems This predicate has not been subject to a design related recall.

### Comparison with Predicate:

The predicate, MR-Touch from GE Medical Systems, consists of the Resoundant Active Acoustic Driver System and acquisition/processing software in its MRI system. The hardware portion of the MR-Touch option is the Resoundant Acoustic Driver System, which is identical in intended use and technology/configuration.

All electro-mechanical functional specifications of the proposed Resoundant Acoustic Driver System remain the same as the original Acoustic Driver System utilized in the predicate MR-Touch.

A comparison of the devices is provided in the following table:

| Device<br>Parameter    | Resoundant Acoustic Driver<br>System                                                                                                                                                                                                                                                                                                                             | GE Mr-Touch Option                                                                                                                                                                                                                                                                                                                              | Difference if<br>any                                                                                                                                                   |
|------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k)                 |                                                                                                                                                                                                                                                                                                                                                                  | K083421                                                                                                                                                                                                                                                                                                                                         | NA                                                                                                                                                                     |
| Classification         | 21 CFR 892.1000, Class II                                                                                                                                                                                                                                                                                                                                        | 21 CFR 892.1000, Class II                                                                                                                                                                                                                                                                                                                       | Same                                                                                                                                                                   |
| Product Code           | Product Code – LNH<br>Magnetic Resonance<br>Diagnostic Device,<br>Radiological                                                                                                                                                                                                                                                                                   | Product Code – LNH Magnetic<br>Resonance Diagnostic Device,<br>Radiological                                                                                                                                                                                                                                                                     | Same                                                                                                                                                                   |
| Indications<br>for Use | The Resoundant Acoustic<br>Driver System is intended for<br>use with magnetic resonance<br>diagnostic devices (MRDD)<br>that include legally marketed<br>MR elastography capabilities.<br>It is indicated for generating<br>acoustic vibrations in the body<br>during an MRI exam, in order<br>to assess tissue elasticity for<br>diagnostic purposes as part of | MR-TouchTM is a software and<br>hardware option intended for<br>use with GE Signa® MR<br>systems. it is indicated for<br>magnetic resonance imaging of<br>the human body.<br>MR-TouchTM generates<br>transverse sectional information<br>related to the relative stiffness<br>of soft tissue. It consists of<br>hardware as well as acquisition | Overall<br>Intended Use<br>is substantially<br>equivalent; the<br>acoustic<br>vibration<br>generation is<br>identical. The<br>Acoustic<br>Driver System<br>is the same |
|                        | magnetic resonance                                                                                                                                                                                                                                                                                                                                               | and reconstruction software.                                                                                                                                                                                                                                                                                                                    | hardware                                                                                                                                                               |
|                        | elastography (MRE).                                                                                                                                                                                                                                                                                                                                              | The hardware components                                                                                                                                                                                                                                                                                                                         | device that                                                                                                                                                            |
|                        |                                                                                                                                                                                                                                                                                                                                                                  | induce vibrations into the scan                                                                                                                                                                                                                                                                                                                 | generates the                                                                                                                                                          |
|                        | When interpreted by a trained                                                                                                                                                                                                                                                                                                                                    | subject. The MR-Touch™                                                                                                                                                                                                                                                                                                                          | acoustic                                                                                                                                                               |
|                        | physician, this information can                                                                                                                                                                                                                                                                                                                                  | acquisition software is an                                                                                                                                                                                                                                                                                                                      | vibrations,                                                                                                                                                            |
|                        | be useful in determining a                                                                                                                                                                                                                                                                                                                                       | evolutionary improvement to                                                                                                                                                                                                                                                                                                                     | enabling MR                                                                                                                                                            |
|                        | diagnosis.                                                                                                                                                                                                                                                                                                                                                       | the gradient echo sequence. The                                                                                                                                                                                                                                                                                                                 | Elastography.                                                                                                                                                          |
|                        |                                                                                                                                                                                                                                                                                                                                                                  | sequence synchronizes the                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                        |
|                        |                                                                                                                                                                                                                                                                                                                                                                  | induced vibrations to acquire a                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                        |
|                        |                                                                                                                                                                                                                                                                                                                                                                  | series of phase-contrast images                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                        |
|                        |                                                                                                                                                                                                                                                                                                                                                                  | over time. The phase-contrast                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                        |
|                        |                                                                                                                                                                                                                                                                                                                                                                  | imaging technique measures                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                        |
|                        |                                                                                                                                                                                                                                                                                                                                                                  | motion or displacement. The                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                        |
|                        |                                                                                                                                                                                                                                                                                                                                                                  | displacement from the induced                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                        |
|                        |                                                                                                                                                                                                                                                                                                                                                                  | vibrations is detected using the                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                        |
|                        |                                                                                                                                                                                                                                                                                                                                                                  | time-series of phase-contrast                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                        |
|                        |                                                                                                                                                                                                                                                                                                                                                                  | images. The displacement                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                        |
|                        |                                                                                                                                                                                                                                                                                                                                                                  | information is                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                        |
|                        |                                                                                                                                                                                                                                                                                                                                                                  | reconstructed and presented as                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                        |
|                        |                                                                                                                                                                                                                                                                                                                                                                  | strain wave and relative                                                                                                                                                                                                                                                                                                                        |                                                                                                                                                                        |
|                        |                                                                                                                                                                                                                                                                                                                                                                  | stiffness images.                                                                                                                                                                                                                                                                                                                               |                                                                                                                                                                        |
|                        |                                                                                                                                                                                                                                                                                                                                                                  | When used with a GE Signa®                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                        |
|                        |                                                                                                                                                                                                                                                                                                                                                                  | MR system, MR-Touch™ is                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                        |
|                        |                                                                                                                                                                                                                                                                                                                                                                  | capable of producing transverse                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                        |
|                        |                                                                                                                                                                                                                                                                                                                                                                  | images of internal body                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                        |
|                        |                                                                                                                                                                                                                                                                                                                                                                  | structures such as muscle and                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                        |
|                        |                                                                                                                                                                                                                                                                                                                                                                  | liver.                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                        |
|                        |                                                                                                                                                                                                                                                                                                                                                                  | When interpreted by a trained                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                        |
|                        |                                                                                                                                                                                                                                                                                                                                                                  | physician, these images may                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                        |
|                        |                                                                                                                                                                                                                                                                                                                                                                  | provide information that can be                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                        |
|                        |                                                                                                                                                                                                                                                                                                                                                                  | useful in determining a                                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                        |
|                        |                                                                                                                                                                                                                                                                                                                                                                  | diagnosis.                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                        |
| Vibration              | The Resoundant Acoustic                                                                                                                                                                                                                                                                                                                                          | The Resoundant Acoustic                                                                                                                                                                                                                                                                                                                         | Same                                                                                                                                                                   |
| Product                | Driver System includes the                                                                                                                                                                                                                                                                                                                                       | Driver System includes the                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                        |
| Configuration          | Active Driver, tubing and                                                                                                                                                                                                                                                                                                                                        | Active Driver, tubing and                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                        |
|                        | Passive Driver.                                                                                                                                                                                                                                                                                                                                                  | Passive Driver.                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                        |
| Output                 | Nominally 60 Hz with a range                                                                                                                                                                                                                                                                                                                                     | Nominally 60 Hz with a range                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                   |
|                        | of 20-350 Hz                                                                                                                                                                                                                                                                                                                                                     | of 20-350 Hz                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                        |
| Intended               | MRE-trained technician or                                                                                                                                                                                                                                                                                                                                        | MRE-trained technician or                                                                                                                                                                                                                                                                                                                       | Same                                                                                                                                                                   |
| User                   | radiologist                                                                                                                                                                                                                                                                                                                                                      | radiologist                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                        |

## Feature and Technical Characteristic Comparison

{5}------------------------------------------------

{6}------------------------------------------------

## Verification and Validation:

The Resoundant Acoustic Driver System has been extensively tested through the design verification and system validation phases of its development. The following provides an overview of the testing successfully performed:

- Mechanical Testing: Dimensional, operational performance, and material testing was . performed to assure compliance to specifications.
- Electrical Testing: Testing was conducted by an authorized third party laboratory to . verify compliance of the Acoustic Driver System to product specifications as well as the following international standards for Safety for Medical Electrical Equipment: IEC ANSI/AAMI ES60601-1: A1:2012, C1:2009/(R)2012 and A2:2010/(R)2012 and IEC 60601-1-2:2014 (Fourth Edition).
- Firmware: The firmware utilized in the Active Driver microprocessor was developed and ● tested to assure compliance to specifications.
- Packaging: The packaged Acoustic Driver System was tested to verify that the packaging ● and product was not damaged during processing, shipping, or storage when subjected to simulated distribution testing per ASTM D4169-16 Distribution Cycle 13.

## Conclusion:

Through the data and information presented. Resoundant Inc. considers the Acoustic Driver System substantially equivalent to the predicate device already on the market (cleared by the 510(k) process) in terms of indications for use, scientific technology, design, and functional performance and presents no new concerns about safety and effectiveness. The verification and validation tests conducted demonstrate that the device is as safe, as effective, and performs as well as the legally marketed predicate device for its intended use.

---

**Source:** [https://fda-staging.innolitics.com/device/K201389](https://fda-staging.innolitics.com/device/K201389)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
