K201387 · Dentlight, Inc. · NVK · Feb 26, 2021 · General, Plastic Surgery
Device Facts
Record ID
K201387
Device Name
Ultrafast, Ultrafast Plus, Ultrafast Lite
Applicant
Dentlight, Inc.
Product Code
NVK · General, Plastic Surgery
Decision Date
Feb 26, 2021
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Dental soft tissue indications: incision, vaporization, ablation, and coagulation of oral soft tissues including the following: · Gingival troughing for crown impression · Gingivectomy · Gingivoplasty · Gingival incision and excision · Hemostasis and coagulation · Excisional and incisional biopsies · Exposure of unerupted teeth · Fibroma removal · Frenectomy and frenotomy · Implant recovery · Incision and drainage of abscess · Leukoplakia · Operculectomy · Oral papillectomies · Pulpotomy · Pulpotomy as an adjunct to root canal therapy · Reduction of gingival hypertrophy · Soft tissue crown lengthening · Treatment of canker sores, herpetic and ulcers · Vestibuloplasty · Tissue retraction for impression Laser periodontal indications including: · Sulcular debridernent (curettage, removal of diseased, inflamed and necrosed soft tissue in the periodontal pocket to improve clinical indices including: gingival bleeding index, probe depth, attachment loss, and tooth mobility). · Laser removal of diseased, inflamed and necrosed soft tissue within the periodontal pocket.
Device Story
Cordless diode laser system for dental soft tissue procedures; utilizes solid-state laser diodes (808nm or 980nm) to generate infrared radiation. Energy delivered via single-use fiber optic tip to treatment site. Handpiece contains laser diode, control circuitry, microprocessor, and rechargeable battery. Includes green aiming beam for procedure guidance. Operated by dentists in clinical settings. Provides visual/audible notifications via LCD interface. Enables precise tissue management (incision, coagulation, ablation); benefits include improved clinical indices in periodontal pockets and efficient soft tissue surgery.
Clinical Evidence
Bench testing only. Device performance verified through optical, thermal, electrical, and mechanical safety testing (IEC 60601-1, IEC 60601-1-2, IEC 60825-1, IEC 60601-2-22) and biocompatibility testing per ISO 10993-1.
Technological Characteristics
Diode laser system; wavelengths 808nm or 980nm; max CW power 1.5W-3W. Handpiece constructed of anodized aluminum. Single-use fiber optic delivery (400μm core). Powered by rechargeable Lithium-ion battery. Includes LCD interface, green aiming beam (525nm). Sterilization/Biocompatibility per ISO 10993-1. Compliance with IEC 60601-1, 60601-1-2, 60825-1, 60601-2-22.
Indications for Use
Indicated for dental soft tissue procedures including incision, excision, vaporization, ablation, and coagulation in patients requiring oral soft tissue management, periodontal pocket debridement, or treatment of oral lesions.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue box, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
February 26, 2021
DentLight Inc. Richard Liu President 1825 Summit Ave. Ste 210 Plano, Texas 75074
Re: K201387
Trade/Device Name: Ultrafast, Ultrafast Plus, Ultrafast Lite Regulation Number: 21 CFR 878.4810 Regulation Name: Laser Surgical Instrument For Use In General And Plastic Surgery And In Dermatology Regulatory Class: Class II Product Code: NVK, GEX Dated: February 9, 2021 Received: February 12, 2021
Dear Richard Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Michael Adjodha Assistant Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### III. Indications for Use
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K201387
#### Device Name
Ultrafast/Ultrafast Plus/Ultrafast Lite
#### Indications for Use (Describe)
Dental soft tissue indications: incision, vaporization, ablation, and coagulation of oral soft tissues including the following:
- · Gingival troughing for crown impression
- · Gingivectomy
- · Gingivoplasty
- · Gingival incision and excision
- · Hemostasis and coagulation
- · Excisional and incisional biopsies
- · Exposure of unerupted teeth
- · Fibroma removal
- · Frenectomy and frenotomy
- · Implant recovery
- · Incision and drainage of abscess
- · Leukoplakia
- · Operculectomy
- · Oral papillectomies
- · Pulpotomy
- · Pulpotomy as an adjunct to root canal therapy
- · Reduction of gingival hypertrophy
- · Soft tissue crown lengthening
- · Treatment of canker sores, herpetic and ulcers
- · Vestibuloplasty
- · Tissue retraction for impression
Laser periodontal indications including:
· Sulcular debridernent (curettage, removal of diseased, inflamed and necrosed soft tissue in the periodontal pocket to improve clinical indices including: gingival bleeding index, probe depth, attachment loss, and tooth mobility).
· Laser removal of diseased, inflamed and necrosed soft tissue within the periodontal pocket.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: 06/30/2020
See PRA Statement below.
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
{3}------------------------------------------------
#### IV. 510(k) Summary
### K201387
This summary of 510(k) substantial equivalence is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
### 1. APPLICANT
DentLight Inc. 1825 Summit Ave., Suite 210 Plano, TX 75074
Phone: 972-889-8857 Fax: 972-346-6550 Contact Person: Richard Liu Date Prepared: Feb. 8, 2021
## 2. DEVICE NAME
Proprietary Name: Ultrafast/Ultrafast Plus/Ultrafast Lite
Common/Usual Name: Dental Diode Laser
Classification Name: Laser Surgical Instrument for Use in General and Plastic Surgery and in Dermatology (878.4810)
Product Code: GEX
## 3. PRIMARY PREDICATE DEVICE
Styla Microlaser (K081214), manufactured by Zap Lasers, LLC.
### REFERENCE DEVICE
K2 Mobile (K200693), manufactured by Hu Laser.
### 4. INTENDED USE / INDICATIONS FOR USE
Dental soft tissue indications: incision, excision, vaporization, and coagulation of oral soft tissues including the following:
- Gingival troughing for crown impression
- · Gingivectomy
- Gingivoplasty
- · Gingival incision and excision
- Hemostasis and coagulation
- · Excisional and incisional biopsies
- Exposure of unerupted teeth
- Fibroma removal
- Frenectomy and frenotomy
- Implant recovery
- · Incision and drainage of abscess
- · Leukoplakia
{4}------------------------------------------------
- Operculectomy
- · Oral papillectomies
- · Pulpotomy
- Pulpotomy as an adjunct to root canal therapy
- · Reduction of gingival hypertrophy
- · Soft tissue crown lengthening
- · Treatment of canker sores, herpetic and ulcers
- · Vestibuloplasty
- · Tissue retraction for impression
Laser periodontal indications including:
· Sulcular debridement (curettage, removal of diseased, inflamed and necrosed soft tissue in the periodontal pocket to improve clinical indices including gingival bleeding index, probe depth, attachment loss, and tooth mobility).
· Laser removal of diseased, inflamed and necrosed soft tissue within the periodontal pocket.
## 5. DEVICE DESCRIPTION
The Ultrafast series (Ultrafast, Ultrafast Plus, and Ultrafast Lite) are cordless diode laser systems designed for a wide variety of dental soft tissue procedures. The products have three separate models under the model names: Ultrafast, Ultrafast Plus and Ultrafast Lite. They use solid-state laser diodes as sources of infrared radiation. The energy from the infrared radiation is delivered to the treatment site via a single-use fiber optic tip.
The product kit consists of a laser handpiece, laser goggles, charging stand, power adapter, laser shield, barrier sleeves, product warning labels and instructions for use. The laser handpiece is constructed from anodized aluminum and contains a single-use, disposable fiber optic tip and main body containing laser diode, main control and a rechargeable battery assembly. The main control houses the control circuitry with microprocessor and user interface to control and deliver the laser and a green aiming beam to guide the laser procedures. The charging stand allows for the placement of the handpiece in a holder with electrical connection to the power adapter as the source for charging the rechargeable battery.
Based on the wavelength and power of the laser diodes used, the kit can be preconfigured into three separate models:
| Trade Name | Wavelength (nm) | Max CW Power (Watt) |
|----------------|-----------------|---------------------|
| Ultrafast | 808 | 3 |
| Ultrafast Plus | 980 | 3 |
| Ultrafast Lite | 808 | 1.5 |
## 6. DEVICE CHARACTERISTICS AND SUBSTANDTIAL EQUIVALENCE
Ultrafast/Ultrafast Plus/Ultrafast Lite are substantially equivalent to the primary predicate device Styla Microlaser (K081214), manufactured by Zap Lasers, LLC (now Denmat) and the reference device K2 Mobile (K200693), manufactured by Hu Laser. A comparison table on performance and technology with the predicate and reference devices lists key features and characteristics of the subject and predicate devices. It shows the subject products have technology characteristics as the predicate devices.
{5}------------------------------------------------
## Subject Device Performance and Technological Comparison
| Parameter\ Name | | Ultrafast / Ultrafast Plus /Ultrafast Lite | Styla Microlaser (Primary Predicate Device) | K2 MOBILE (Reference Device) |
|-------------------------------|-------------------------------|--------------------------------------------------------------------|---------------------------------------------|---------------------------------------------|
| | Manufacturer | DentLight | Zap Laser/Denmat | Hu Laser |
| 510(k) Number | | K201387 | K081214 | K200693 |
| | Light Source | Diode Laser | Diode Laser | Diode Laser |
| | Regulation Number | 878.4810 | 878.4810 | 878.4810 |
| | Device Class | II | II | II |
| | Product Code | GEX | GEX | GEX |
| Peak Wavelength | | 808±10 nm (Ultrafast/Ultrafast Lite)<br>980±10 nm (Ultrafast Plus) | 808±5 nm | 980±10 nm |
| Maximum CW Power<br>(Watt) | | 3 (Ultrafast/Ultrafast Plus)<br>1.5 (Ultrafast Lite) | 1.5 | 3.5 |
| | | | | |
| Maximum Pulse Power<br>(Watt) | | 5 (Ultrafast/Ultrafast Plus)<br>2 (Ultrafast Lite) | 2 | 6 |
| | | | | |
| | Power Accuracy (W) | ±20% | ±20% | ±20% |
| | Fiber Core Diameter<br>(μm) | 400 | 400 | 400 |
| | Aiming Beam | 525 (±10nm, 0.6mW) | 650 (±10nm, 0.5mW) | 635 (5 mW) |
| | OPERATION<br>MODE | Continuous / Pulse | Continuous / Pulse | Continuous / Pulse |
| | Display | LCD | OLED | OLED |
| | Audible Notification | Yes | Yes | Yes |
| | Visual Notification | Yes | Yes | Yes |
| | Length<br>(without fiber tip) | 163 | 175 | 205 |
| Weight | Handpiece | 100 | 55 | 135 |
| (grams) | Charging<br>Stand | 88 | 340 | 150 |
| | Battery | Lithium ion | Lithium ion | Lithium ion |
| | Power Supply Input | 100-240V, 50/60Hz | Input: 100-240V, 50-60Hz | Input: 100-240V, 50-60Hz |
| | Structure | Ergonomic aluminum wand | Ergonomic aluminum wand | Ergonomic aluminum/plastic wand |
| Compliance Standards | | IEC 60601-1; IEC 60601-1-2; IEC<br>60825-1; IEC 60601-2-22 | IEC 60601-1; IEC 60825-1;<br>IEC 60601-2-22 | IEC 60601-1; IEC 60825-1; IEC<br>60601-2-22 |
## Subject Device Indications for Use Comparisons
| Parameter Name | Ultrafast / Ultrafast Plus<br>/Ultrafast Lite | Styla Microlaser<br>(Primary Predicate Device) | K2 MOBILE<br>(Reference Device) |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer | DentLight | Zap Laser/Denmat | Hu Laser |
| 510(k) Number | K201387 | K081214 | K200693 |
| Indications for Use | Dental soft tissue indications:<br>incision, excision, vaporization,<br>ablation, and coagulation of oral<br>soft tissues including the<br>following:<br>• Gingival troughing for crown<br>impression<br>• Gingivectomy<br>• Gingivoplasty<br>• Gingival incision and excision<br>• Hemostasis and coagulation<br>• Excisional and incisional | Intended for ablating,<br>incising, excising,<br>vaporizing, and coagulation<br>of oral soft tissues using a<br>contact fiber optic delivery<br>system. The following are the<br>oralpharngeal indications for<br>use for which the device will<br>be marketed:<br>• Excisional and incisional<br>biopsies<br>• Hemostasis assistance | Intended for use by dentists<br>for excision, incision,<br>vaporization, ablation and<br>coagulation of oral soft tissue<br>procedures, including Tooth<br>Whitening and the temporary<br>relief of pain. The Specific<br>indications are as follows:<br>• Biopsies<br>• Crown lengthening<br>• Exposure of unerupted teeth<br>• Fibroma removal |
DentLight Inc. 510(k) - K201387 Feb. 8, 2021
{6}------------------------------------------------
| biopsies | • Treatment of aphthous<br>ulcers | • Frenectomy |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| • Exposure of unerupted teeth | • Frenectomy and frenotomy | • Gingival troughing |
| • Fibroma removal | • Gingival Incision and<br>Excision | • Gingivectomy |
| • Frenectomy and frenotomy | • Gingivectomy | • Gingivoplasty |
| • Implant recovery | • Gingivoplasty | • Pulpotomy |
| • Incision and drainage of abscess | • Incision and drainage of<br>abscess | • Root canal therapy |
| • Leukoplakia | • Operculectomy | • Hemostasis and coagulation |
| • Operculectomy | • Oral papillectomy | • Leukoplakia |
| • Oral papillectomies | • Removal of fibromas | • Implant recovery |
| • Pulpotomy | • Soft tissue crown<br>lengthening | • Incision and drainage of<br>abscess |
| • Pulpotomy as an adjunct to root<br>canal therapy | • Sulcular debridement<br>(removal of diseased or<br>inflamed soft tissue in the<br>periodontal pocket) | • Operculectomy |
| • Reduction of gingival<br>hypertrophy | • Vestibuloplasty | • Papillectomies |
| • Soft tissue crown lengthening | | • Reduction of gingival<br>hypertrophy |
| • Treatment of canker sores,<br>herpetic and aphthous ulcers | | • Treatment of Aphthous-<br>ulcer canker sores and<br>herpetic |
| • Vestibuloplasty | | •Vestibuloplasty |
| • Tissue retraction for impression | | |
| Laser periodontal indications<br>including: | | Periodontal procedures: |
| • Sulcular debridement (curettage,<br>removal of diseased, infected,<br>inflamed and necrosed soft tissue<br>in the periodontal pocket to<br>improve clinical indices including<br>gingival index, gingival bleeding<br>index, probe depth, attachment<br>loss, and tooth mobility). | | • Sulcular debridement |
| • Laser removal of diseased,<br>infected, inflamed and necrosed<br>soft tissue within the periodontal<br>pocket. | | • Topical heating for the<br>purpose of elevating tissue<br>temperature for temporary<br>relief of minor muscle<br>and joint pain and stiffness<br>minor arthritis pain, or<br>muscle spasm, minor sprains<br>and minor muscular back<br>pain; the temporary increase<br>in local blood circulation; the<br>temporary relation of muscle |
| | | • Laser Assisted whitening |
Similarities between Ultrafast Plus/Ultrafast Lite and Predicate/Reference Devices:
- · Ultrafast/Ultrafast Plus/Ultrafast Lite have the same indications for use as the predicate/reference devices.
- · Ultrafast/Ultrafast Plus/Ultrafast Lite have substantially equivalent operating modes as the predicate/reference devices.
- · Ultrafast/Ultrafast Plus/Ultrafast Lite have substantially equivalent design, construction materials, weight, size, performance characteristics and key features as the predicate/reference devices.
- · Ultrafast/Ultrafast Plus/Ultrafast Lite are substantially equivalent to the predicate/reference devices in standards compliance.
## Differences between the Ultrafast Plus/Ultrafast Lite and Predicate/Reference Devices:
Ultrafast/Ultrafast Plus/Ultrafast Lite have an LCD display instead of an OLED display by the predicate and reference devices. They use green aiming beams instead of red aiming beams by the predicate and reference devices. The difference in display and aiming beam improves the operator's ease of use and does not raise any substantial equivalence concerns as it has similar performance of the predicate and reference devices. The device modifications do not potentially alter the fundamental scientific technology of the device.
# 7. NON-CLINICAL PERFORMANCE TESTING
The subject devices have been tested for optical, thermal, electrical and mechanical safety and comply with the electrical and optical safety requirements as the predicate devices: IEC 6060-1, IEC 60601-1-2, IEC 60825-1 and IEC 60601-2-22.
{7}------------------------------------------------
The subject devices have been tested for biocompatibility and comply with ISO 10993-1 Biological Evaluation of Medical Devices- Part 1: Evaluation and Testing within a Risk Management Process. The subject devices have tested and verified to meet all design specifications that are substantially equivalent to the predicate devices.
#### 8. CONCLUSION
Based on similarities in indications for use and technology together with results from performance testing, we believe that Ultrafast Plus/Ultrafast Lite Diode Lasers are substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.