← Product Code [GEH](/productcode/GEH) · K201366

# Forwarts (K201366)

_Pronova Laboratories BV · GEH · Jan 21, 2021 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K201366

## Device Facts

- **Applicant:** Pronova Laboratories BV
- **Product Code:** [GEH](/productcode/GEH.md)
- **Decision Date:** Jan 21, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4350
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

Forwarts® is indicated for the treatment of warts on hands (common warts) and feet (plantar warts).

## Device Story

Forwarts is an over-the-counter cryosurgical device for wart removal. It consists of a pressurized aerosol canister containing 35 ml dimethyl ether (DME) housed in a polypropylene shell. The device features a distance holder and visor to guide and limit the spray pattern. Operation involves the user directing the cryogenic spray onto the wart; the rapid freezing causes cell injury and ischemic necrosis, leading the wart to fall off in 10-14 days. Protection patches are provided to shield healthy skin. The nozzle requires cleaning with 70% rubbing alcohol after each use. The device is intended for lay-person self-use at home. It provides a non-invasive method for wart treatment, with performance comparable to existing cryosurgical units.

## Clinical Evidence

No clinical trials were conducted. Evidence consists of bench testing, including comparative thermal performance (freeze temperature, duration, cycle length, spot size) using a simulated skin model, and human factors validation. A self-selection/label comprehension study demonstrated a >90% success rate for lay-person use. Biocompatibility was established per ISO 10993-1.

## Technological Characteristics

Pressurized aerosol canister containing 35 ml dimethyl ether (DME). Materials: polypropylene shell, PE foam protection patches. Principle: cryotherapy via direct cryogenic spray. Dimensions/form factor: handheld aerosol dispenser. Connectivity: none. Sterilization: non-sterile. Software: none.

## Regulatory Identification

(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.

## Predicate Devices

- Wart Freeze ([K130599](/device/K130599.md))
- Compound W® Nitro-Freeze ([K172373](/device/K172373.md))
- Wartie® Wart Remover ([K140314](/device/K140314.md))
- Wartner® Wart Removal System ([K032271](/device/K032271.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo features the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.

January 21, 2021

Pronova Laboratories BV % Stuart Goldman Regulatory Consultant Emergo Global Consulting, LLC 2500 Bee Cave Road, Building 1, Suite 300 Austin, Texas 78746

Re: K201366

Trade/Device Name: Forwarts Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit and Accessories Regulatory Class: Class II Product Code: GEH Dated: December 21, 2020 Received: December 22, 2020

Dear Stuart Goldman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrl/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

### Indications for Use

510(k) Number (if known) K201366

Device Name Forwarts®

Indications for Use (Describe)

Forwarts® is indicated for the treatment of warts on hands (common warts) and feet (plantar warts).

Type of Use (Select one or both, as applicable)

| Prescription Use (Part 21 CFR 801 Subpart D)

X Over-The-Counter Use (21 CFR 801 Subpart C)

### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

# 510(k) Summary

## Forwarts®

#### 1. Submission Sponsor

Pronova Laboratories BV Kruitpad 16 1398 CP Muiden The Netherlands Office Phone: +31 294 46 66 11 Email: diede.klever@pronovalab.com Contact: Diede Klever Title: D&D Manager

### 2. Submission Correspondent

Emergo Global Consulting, LLC 2500 Bee Cave Road Building 1, Suite 300 Austin, TX 78746 Office Phone: (512) 327-9997 Email: LST.AUS.ProjectManagement@ul.com Contact: Indraj Bamrah Title: Senior Regulatory Consultant

### 3. Date Prepared

January 19, 2021

### 4. Device Identification

| Trade/Proprietary Name: | Forwarts®                         |
|-------------------------|-----------------------------------|
| Common/Usual Name:      | Cryosurgical unit and accessories |
| Classification Name:    | Unit, Cryosurgical, Accessories   |
| Regulation Number:      | 878.4350                          |
| Product Code:           | GEH                               |
| Class:                  | Class 2                           |
| Classification Panel:   | General & Plastic Surgery         |

#### 5. Legally Marketed Predicate Device(s)

#### Primary Predicate

Device name: Wart Freeze

{4}------------------------------------------------

510(k) number: K130599 Manufacturer: Koninklijke Utermohlen NV

Wart Freeze is an over the counter cryosurgery product for the treatment of common and plantar warts. The device consists of a pressurized dispenser containing 38 ml dimethyl ether (DME) with a polypropylene applicator that is permanently attached (fixed) to the dispenser.

#### Secondary Predicates

| Device name:   | Compound W® Nitro-Freeze     |
|----------------|------------------------------|
| 510(k) number: | K172373                      |
| Device name:   | Wartie® Wart Remover         |
| 510(k) number: | K140314                      |
| Device name:   | Wartner® Wart Removal System |
| 510(k) number: | K032271                      |

#### 6. Indication for Use Statement

Forwarts® is indicated for the treatment of warts on hands (common warts) and feet (plantar warts).

#### 7. Device Description

Forwarts® is a wart removal device that works by the principle of cryotherapy and is intended for the treatment of common and plantar warts. It delivers a precise cryogenic spray directly onto a wart or verruca which causes them to freeze and fall off the skin in approximately 10-14 days. The device consists of a pressurized aerosol can containing 35 ml dimethyl ether (DME). The aerosol can is held in a polypropylene shell (front and back). The front shell contains a distance holder for a safe and regulated application and a visor to point and limit the spray pattern. The translucent back shell contains a ridge for safe handling. The valve system will be extended by an actuator to point and minimize the spray pattern. Forwarts® is supplied with ten (10) protection patches per device to be placed around the wart prior to treatment to help prevent damage of healthy skin. Forwarts® is supplied non-sterile and not intended to be sterilized by the end user. The Forwarts® nozzle should be cleaned after each use with rubbing alcohol (70%).

### 8. Substantial Equivalence Discussion

The following table compares Forwarts® to the predicate device with respect to indications for use, principles of operation, technological characteristics, and performance, and forms the basis for the determination of substantial equivalence. The subject device does not raise any new questions of safety or effectiveness as compared to the predicate device.

{5}------------------------------------------------

|  |  | Table 5A - Comparison of Characteristics |
|--|--|------------------------------------------|
|--|--|------------------------------------------|

| Attribute                                                               | Forwarts®                                                                                                                                                                     | Wart Freeze /<br>Koninklijke<br>Utermohlen NV                                                                                                                           | Compound W®<br>Nitro-Freeze /<br>Medtech<br>Products Inc.                                                                                                                                           | Wartie® Wart<br>Remover /<br>YouMedical B.V.                                                                                                                                                              | Wartner® Wart<br>Removal System<br>/ Wartner USA<br>B.V.                                                                                                                                                                              | Comparison                                                                                                                                                                                                                                                      |
|-------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number                                                           | K201366 (pending)                                                                                                                                                             | K130599                                                                                                                                                                 | K172373                                                                                                                                                                                             | K140314                                                                                                                                                                                                   | K032271                                                                                                                                                                                                                               | N/A                                                                                                                                                                                                                                                             |
| Product Code                                                            | GEH, Unit,<br>Cryosurgical,<br>Accessories                                                                                                                                    | GEH, Unit,<br>Cryosurgical,<br>Accessories                                                                                                                              | GEH, Unit,<br>Cryosurgical,<br>Accessories                                                                                                                                                          | GEH, Unit,<br>Cryosurgical,<br>Accessories                                                                                                                                                                | GEH, Unit,<br>Cryosurgical,<br>Accessories                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                            |
| Regulation<br>Number                                                    | 21 CFR 878.4350                                                                                                                                                               | 21 CFR 878.4350                                                                                                                                                         | 21 CFR 878.4350                                                                                                                                                                                     | 21 CFR 878.4350                                                                                                                                                                                           | 21 CFR 878.4350                                                                                                                                                                                                                       | Same                                                                                                                                                                                                                                                            |
| Indications for<br>Use                                                  | Forwarts® is<br>indicated for the<br>treatment of warts<br>on hands (common<br>warts) and feet<br>(plantar warts).                                                            | Wart Freeze is<br>indicated for the<br>removal of common<br>and plantar warts                                                                                           | Compound W®<br>Nitro-Freeze is<br>intended for<br>over-the-counter<br>treatment of<br>common warts<br>and plantar warts<br>in adults and<br>children four<br>years of age or<br>older               | Wartie® Wart<br>Remover is<br>indicated for the<br>over-the-counter<br>treatment of<br>common warts<br>and plantar<br>warts, for<br>patients aged 4<br>years and older.                                   | WARTNER® Wart<br>Removal System<br>is indicated for<br>the over-the-<br>counter<br>treatment of<br>common warts<br>and plantar<br>warts.                                                                                              | Same                                                                                                                                                                                                                                                            |
| OTC or Rx                                                               | OTC                                                                                                                                                                           | OTC                                                                                                                                                                     | OTC                                                                                                                                                                                                 | OTC                                                                                                                                                                                                       | OTC                                                                                                                                                                                                                                   | Same                                                                                                                                                                                                                                                            |
| Technological<br>Characteristics                                        | Pressurized gas<br>canister that                                                                                                                                              | Pressurized gas<br>canister that applies                                                                                                                                | Pen with<br>pressurized<br>cartridge that                                                                                                                                                           | Pressurized gas<br>canister that                                                                                                                                                                          | Pressurized gas<br>canister that                                                                                                                                                                                                      | Similar;<br>Differences in<br>cryogen and cold                                                                                                                                                                                                                  |
| Attribute                                                               | Forwarts®                                                                                                                                                                     | Wart Freeze /<br>Koninklijke<br>Utermohlen NV                                                                                                                           | Compound W®<br>Nitro-Freeze /<br>Medtech<br>Products Inc.                                                                                                                                           | Wartie® Wart<br>Remover /<br>YouMedical B.V.                                                                                                                                                              | Wartner® Wart<br>Removal System<br>/ Wartner USA<br>B.V.                                                                                                                                                                              | Comparison                                                                                                                                                                                                                                                      |
|                                                                         | applies extreme<br>cold to target;<br>Cryogen (liquefied<br>dimethyl ether, BP<br>= -24°C) aerosol<br>spray is applied<br>directly to the wart,<br>causing rapid<br>freezing. | extreme cold to<br>target;<br>Cryogen (liquefied<br>dimethyl ether, BP =<br>-24°C) is applied<br>directly to the wart,<br>causing rapid<br>freezing.                    | applies extreme<br>cold to target;<br>Cryogen<br>(liquefied nitrous<br>oxide, BP = -89°C)<br>is loaded onto a<br>foam applicator,<br>which is applied<br>to the wart,<br>causing rapid<br>freezing. | applies extreme<br>cold to target;<br>Cryogen<br>(liquefied<br>dimethyl ether,<br>BP = -24°C) is<br>used to<br>refrigerate a steel<br>tip, which is<br>applied to the<br>wart, causing<br>rapid freezing. | applies extreme<br>cold to target;<br>Cryogen<br>(liquefied<br>dimethyl ether /<br>propane mixture,<br>BP = -24°C / -<br>48°C) is loaded<br>onto a foam<br>applicator, which<br>is applied to the<br>wart, causing<br>rapid freezing. | application<br>technique.<br>The added wind<br>chill effect of<br>Forwarts' DME<br>aerosol spray<br>results in lower<br>temperatures<br>compared to<br>Wart Freeze.                                                                                             |
| Mechanism of<br>Action                                                  | Cryotherapy<br>Rapid freezing<br>followed by slow<br>thawing, causing<br>cell injury and<br>ischemic necrosis.                                                                | Cryotherapy;<br>Rapid freezing<br>followed by slow<br>thawing, causing cell<br>injury and ischemic<br>necrosis.                                                         | Cryotherapy;<br>Rapid freezing<br>followed by slow<br>thawing, causing<br>cell injury and<br>ischemic necrosis.                                                                                     | Cryotherapy;<br>Rapid freezing<br>followed by slow<br>thawing, causing<br>cell injury and<br>ischemic necrosis.                                                                                           | Cryotherapy;<br>Rapid freezing<br>followed by slow<br>thawing, causing<br>cell injury and<br>ischemic necrosis.                                                                                                                       | Same                                                                                                                                                                                                                                                            |
| Materials with<br>direct patient<br>contact                             | DME,<br>polypropylene<br>Optional protection<br>patch: PE foam                                                                                                                | DME and<br>polypropylene                                                                                                                                                | Nitrous oxide and<br>PU foam                                                                                                                                                                        | Nickel                                                                                                                                                                                                    | DME, propane<br>and PU foam                                                                                                                                                                                                           | Similar to primary<br>predicate device;<br>Different from<br>supplemental<br>predicate devices.                                                                                                                                                                 |
| Attribute                                                               | Forwarts®                                                                                                                                                                     | Wart Freeze /<br>Koninklijke<br>Utermohlen NV                                                                                                                           | Compound W®<br>Nitro-Freeze /<br>Medtech<br>Products Inc.                                                                                                                                           | Wartie® Wart<br>Remover /<br>YouMedical B.V.                                                                                                                                                              | Wartner® Wart<br>Removal System<br>/ Wartner USA<br>B.V.                                                                                                                                                                              | Comparison                                                                                                                                                                                                                                                      |
| Biocompatibility                                                        | Established<br>conform ISO 10993-<br>1                                                                                                                                        | Established conform<br>ISO 10993-1                                                                                                                                      | Established<br>conform ISO<br>10993-1                                                                                                                                                               | Established<br>conform ISO<br>10993-1                                                                                                                                                                     | Established<br>conform ISO<br>10993-1                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                            |
| Maximum<br>allowed<br>consecutive<br>treatments<br>(2-week<br>interval) | 4                                                                                                                                                                             | 4                                                                                                                                                                       | 3                                                                                                                                                                                                   | 3                                                                                                                                                                                                         | 3                                                                                                                                                                                                                                     | Similar                                                                                                                                                                                                                                                         |
| Sterility                                                               | Non-sterile                                                                                                                                                                   | Non-sterile                                                                                                                                                             | Non-sterile                                                                                                                                                                                         | Non-sterile                                                                                                                                                                                               | Non-sterile                                                                                                                                                                                                                           | Same                                                                                                                                                                                                                                                            |
| Re-Usable                                                               | Yes;<br>Except for optional<br>protection patch                                                                                                                               | Yes;<br>includes applicator                                                                                                                                             | Yes;<br>except for<br>disposable<br>applicator foam<br>tips                                                                                                                                         | Yes;                                                                                                                                                                                                      | Yes;<br>except for<br>disposable<br>applicator foam<br>tips                                                                                                                                                                           | Similar;<br>Skin-contacting<br>tips of Forwarts<br>and Wart Freeze<br>are re-used and<br>to be cleaned.<br>Those of<br>Compound W®<br>Nitro-Freeze and<br>Wartner® Wart<br>Removal System<br>are disposable.                                                    |
| Attribute                                                               | Forwarts®                                                                                                                                                                     | Wart Freeze /<br>Koninklijke<br>Utermohlen NV                                                                                                                           | Compound W®<br>Nitro-Freeze /<br>Medtech<br>Products Inc.                                                                                                                                           | Wartie® Wart<br>Remover /<br>YouMedical B.V.                                                                                                                                                              | Wartner® Wart<br>Removal System<br>/ Wartner USA<br>B.V.                                                                                                                                                                              | Comparison                                                                                                                                                                                                                                                      |
| Thermal<br>Performance<br>Testing                                       | Measured<br>attributes included:<br>freeze temperature<br>extreme, freeze<br>temperature<br>plateau, freeze<br>duration, freeze<br>cycle length and<br>freeze spot size       | Measured attributes<br>included: Freeze<br>temperature<br>extreme, freeze<br>temperature<br>plateau, freeze<br>duration, freeze<br>cycle length and<br>freeze spot size | Measured<br>attributes<br>included: freeze<br>temperature<br>extreme, freeze<br>temperature<br>plateau, freeze<br>duration, freeze<br>cycle length and<br>freeze spot size                          | Measured<br>attributes<br>included: freeze<br>temperature<br>extreme, freeze<br>temperature<br>plateau, freeze<br>duration, freeze<br>cycle length and<br>freeze spot size                                | Measured<br>attributes<br>included: freeze<br>temperature<br>extreme, freeze<br>temperature<br>plateau, freeze<br>duration, freeze<br>cycle length and<br>freeze spot size                                                            | Similar; Forwarts®<br>achieves thermal<br>performance<br>characteristics<br>within the ranges<br>of the predicates<br>demonstrating<br>comparable<br>efficacy.<br>Temperatures<br>achieved<br>demonstrating<br>that there are no<br>new questions of<br>safety. |

{6}------------------------------------------------

{7}------------------------------------------------

{8}------------------------------------------------

{9}------------------------------------------------

#### 9. Non-Clinical Performance Data

To demonstrate safety and effectiveness of Forwarts® and to show substantial equivalence to the predicate device, Pronova Laboratories BV completed the following non-clinical tests. Results confirm that the design inputs and performance specifications for the device are met. Forwarts® passed the testing in accordance with internal requirements, national standards, and international standards shown below, supporting its safety and effectiveness, and its substantial equivalence to the predicate device:

- . Biocompatibility Testing per ISO 10993-1 - Supports biocompatibility of the device.
	- O ISO 10993-5 (cytotoxicity)
	- ISO 10993-10 (irritation) o
	- ISO 10993-10 (sensitization) o
	- ISO 10993-17 (toxicological risk assessment (TRA)) O
	- O ISO 10993-18 (chemical characterization)

Based on the chemical characterization testing and TRA that was performed; therefore, no carcinogenicity testing was required.

- Shelf Life Testing per ASTM D3090-72 – Supports shelf life of twenty months.
Shelf life testing was based on device stability and functionality testing when stored under worst case simulated conditions. The results of the test demonstrate device stability for twenty months.

- Transportation Testing per ASTM D4169 – Demonstrates package integrity maintained.
Forwarts® in its final shipping configuration was subjected to impact, drop, vibration and low-pressure testing under simulated transport conditions. The results of the testing demonstrated that the packaging configuration is suitable to maintain device integrity.

- Bench Testing - Supports device performance.
Forwarts® was subjected to comparative thermal performance testing that included thermoelectric testing and a simulated skin model to determine freeze temperature extreme, freeze temperature plateau, freeze duration, freeze cycle length and freeze spot size. The test provided data on both the temperature profile at the target surface and on the tissue freezing capacity of the results of the tests demonstrated comparable performance to the predicate devices.

- Self-selection / Label Comprehension Study – Supports device self-selection and labeling comprehension.
A self-selection study was performed on Forwarts® to demonstrate a success rate of at least 90% when used as indicated. The study subjects were provided with the Forwarts® labeling (package and instructions for use) for this study.

{10}------------------------------------------------

- . Human Factors Study – Supports safe usability of the device for lay-person use.
A Formative Study was conducted on Forwarts® that included Simulated Use Testing focusing on Critical Tasks, combined with Contextual Enquiry and a semi-guided interview with test subjects. A Human Factor Validation Test was then performed to validate that Forwarts® is safe for its intended use.

#### 10. Statement of Substantial Equivalence

Forwarts® has the same intended use as the Wart Freeze predicate device, and the same or similar technological characteristics. The differences in technological characteristics do not raise new or different questions of safety and effectiveness. Performance testing has demonstrated that Forwarts® is as safe and effective as the predicate device. Therefore, Forwarts® is substantially equivalent to the predicate device.

---

**Source:** [https://fda-staging.innolitics.com/device/K201366](https://fda-staging.innolitics.com/device/K201366)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
