K201352 · Philips Health Care · LLZ · Aug 17, 2020 · Radiology
Device Facts
Record ID
K201352
Device Name
3D Auto LAA
Applicant
Philips Health Care
Product Code
LLZ · Radiology
Decision Date
Aug 17, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Indications for Use
The 3D Auto LAA is a software application designed to view and quantify 3D image data acquired by Philips Ultrasound Systems for use in measuring the area, circumference, and diameter of a Left Atrial Appendage (LAA) orifice.
Device Story
3D Auto LAA is a semi-automated software application for Philips EPIQ and Affiniti ultrasound systems; processes 3D TEE images to quantify LAA orifice dimensions. Workflow: clinician aligns multi-planar reconstruction (MPR) views; algorithm performs greyscale intensity differentiation with shape regularization to generate preliminary LAA border; clinician reviews and manually edits contour via trackball or rotary knob. Output: area, circumference, and diameter measurements of LAA orifice. Used on-cart by clinicians to assist in cardiac assessment; provides semi-automated alternative to manual tracing. Benefits: standardized, efficient LAA quantification for clinical decision-making.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including software verification and validation, and a performance validation study comparing the semi-automated 3D Auto LAA application to manual tracing methods.
Technological Characteristics
Software-only application; runs on-cart on Philips EPIQ and Affiniti ultrasound systems. Utilizes 3D TEE image data. Employs semi-automated border detection algorithm based on greyscale intensity differentiation and shape regularization. Provides editable measurements (area, circumference, diameter).
Indications for Use
Indicated for clinicians using Philips Ultrasound systems to view and quantify 3D image data for measuring the area, circumference, and diameter of the Left Atrial Appendage (LAA) orifice.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
Predicate Devices
Philips QLAB Advanced Quantification Software System (K191647)
Submission Summary (Full Text)
{0}------------------------------------------------
August 17, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Philips Healthcare % Travis Catania Senior Regulatory Affairs Specialist 22100 Bothell Everett Highway BOTHELL WA 98021
Re: K201352
Trade/Device Name: 3D Auto LAA Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: July 21, 2020 Received: July 22, 2020
Dear Travis Catania:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmp/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see
{1}------------------------------------------------
https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201352
Device Name 3D Auto LAA
Indications for Use (Describe)
The 3D Auto LAA is a software application designed to view and quantify 3D image data acquired by Philips Ultrasound Systems for use in measuring the area, circumference, and diameter of a Left Atrial Appendage (LAA) orifice.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
| Philips Ultrasound, Inc. | Traditional 510(k) | Page 1 of 7 |
|--------------------------|--------------------|-------------|
| | 3D Auto LAA | |
## Section 8: 510(k) Summary K201352
{4}------------------------------------------------
# Philips 3D Auto LAA Cardiac Quantification Application
This summary of safety and effectiveness information is submitted in accordance with 21 CFR §807.92
### 1. Submitter's name, address, telephone number, contact person
| Sponsor: | Philips Ultrasound, Inc.<br>22100 Bothell Everett Hwy<br>Bothell, WA 98021-8431 | | |
|------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|
| Contact Person: | Travis Catania<br>Senior Regulatory Affairs Specialist<br>22100 Bothell Everett Hwy<br>Bothell, WA 98021-8431<br>Phone: (908) 227-9423<br>Fax: 425-402-3481 | | |
| Secondary Contact: | Hebe Sun<br>Senior Manager, Regulatory Affairs | | |
| Date Prepared | May 20, 2020 | | |
| 2.<br>Name of the device, including the trade of proprietary name if applicable, the common or usual name, and<br>the classification name, if known: | | | |
| Proprietary Name: | 3D Auto LAA |
|-------------------|-------------------------------------------------------------------|
| Common Name: | 3D Auto LAA<br>Picture Archiving and Communications System (PACS) |
## Regulation Description:
| Classification Description | 21 CFR Section | Product Code |
|---------------------------------------------|----------------|--------------|
| Picture Archiving and Communications System | 892.2050 | LLZ |
As stated in 21 CFR, part 892.2050, each of the generic types of devices that meet this classification description have been classified as Class II.
Device Class:
Class II
{5}------------------------------------------------
#### 3. Indications for Use
The 3D Auto LAA is a software application designed to view and quantify 3D image data acquired by Philips Ultrasound Systems for use in measuring the area, circumference, and diameter of a Left Atrial Appendage (LAA) orifice.
### Device Description 4.
The purpose of this Traditional 510(k) Pre-market Notification is to introduce the new 3D Auto LAA cardiac quantification application for use on the Philips EPIQ and Affiniti Diagnostic Ultrasound systems, which were most recently cleared under K201012. The 3D Auto LAA application is compatible with 3D images generated via Philips Transesophageal Echocardiogram (TEE) transducers such as the X8-2t and X7-2t. The Philips 3D Auto LAA cardiac quantification application is a semi-automated application intended to provide measurements of the Left Atrial Appendage. The 3D Auto LAA application is designed to provide automatic and editable area, circumference, and diameter measurements of the Left Atrial Appendage (LAA) orifice. When the 3D Auto LAA application is launched, going through the workflow provides the clinician with a semi-automated preliminary border of the LAA in three planes based on greyscale intensity differentiation with some shape regularization.
#### ഗ് Substantially Equivalent Devices
## Primary Predicate Device
Philips QLAB Advanced Quantification Software System K191647 December 20, 2019
#### 6. Technological Comparison to Predicate Devices
The introduction of the new 3D Auto LAA cardiac quantification has an equivalence intended use and similar technological characteristics as the legally marketed primary QLAB System predicate device with the distinction being that the 3D Auto LAA application is specific to LAA viewing and quantification. A comparison of the subject 3D Auto LAA application to the currently marketed predicate QLAB System device is provided in the table below:
{6}------------------------------------------------
| Table 8-1: Comparison of The Subject Philips 3D Auto LAA Cardiac Quantification Application to the predicate Philips QLAB System<br>(the 3DQ Q-App) | | | | |
|-----------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Subject Device | Predicate Device | Explanation of Differences | |
| Manufacturer | Philips Ultrasound, Inc. | Philips Ultrasound, Inc. | None | |
| Trade Name | 3D Auto LAA | QLAB System | None | |
| Feature | 3D Auto LAA | 3DQ | None | |
| 510(k) Number | Pending | K191647 | None | |
| Regulation Number | 21 CFR 892.2050 | 21 CFR 892.2050 | | |
| Regulation Name | System, Image processing, Radiological -<br>Picture Archiving and Communications<br>System (PACS) | System, Image processing, Radiological -<br>Picture Archiving and Communications System<br>(PACS) | The Regulation Number, Regulation Name,<br>Classification, and Product Code are identical<br>between the subject device and the primary<br>predicate device. | |
| Classification | Class II | Class II | | |
| Product Code(s) | LLZ | LLZ | | |
| Indications for Use | The 3D Auto LAA is a software application<br>designed to view and quantify 3D image data<br>acquired by Philips Ultrasound Systems for<br>use in measuring the area, circumference,<br>and diameter of a Left Atrial Appendage<br>(LAA) orifice. | QLAB Quantification software is a software<br>application package. It is designed to view and<br>quantify image data acquired on Philips<br>ultrasound systems. | The Indications for Use of the subject 3D Auto<br>LAA application and the primary predicate<br>QLAB System are equivalent with the<br>underlying difference being that the 3D Auto<br>LAA application specifically identifies the<br>anatomical structure / region (the LAA) and<br>the measurements that are reported to the<br>end user (area, circumference, etc.). | |
| System Components | Software only system | Software only system | The System Components of the subject 3D<br>Auto LAA application and the QLAB System<br>are identical as they both are software only<br>systems. | |
| Availability | 3D Auto LAA application is available only<br>on-cart for the Philips Ultrasound EPIQ and<br>Affiniti Ultrasound Systems | The QLAB System is available either as a stand-<br>alone product that can function on a standard<br>PC, a dedicated workstation, and on-board<br>Philips Ultrasound Systems | The availability of the 3D Auto LAA and the<br>QLAB System are similar in that both<br>software systems are available on the Philips<br>Diagnostic Ultrasound Systems as dedicated<br>on-cart applications. However, the QLAB<br>System offers the added freedom to be used | |
| | | Subject Device | Predicate Device | Explanation of Differences |
| Manufacturer | | Philips Ultrasound, Inc. | Philips Ultrasound, Inc. | None |
| Trade Name | | 3D Auto LAA | QLAB System | None |
| Feature | | 3D Auto LAA | 3DQ | None<br>off-cart in variable capacities whereas the 3D Auto LAA application does not allow for this. |
| Software<br>Design | Application<br>Description | The 3D Auto LAA cardiac quantification<br>application is a semi-automatic tool that is<br>intended to assist the end user with LAA<br>border detection and provide specific<br>measurements for the orifice. This<br>application is provided as an on-cart option<br>only. | The 3DQ Q-App provides a manual way to view,<br>slice, and display 3D volumes and measure<br>distance and areas from MPR views to get<br>biplane volume and ejection fraction in addition<br>to mass calculations. QLAB Q-Apps are available<br>both on and off-cart options. | The subject 3D Auto LAA application<br>introduces a semi-automated workflow to<br>provide the user with a dedicated tool for<br>LAA quantification for on-cart use only.. The<br>primary predicate QLAB 3DQ Q-App does also<br>allow the user to measure and quantify<br>several various cardiac structures but this is<br>an all manual process |
| | Quantification<br>Technology for<br>Cardiac | Manual MPR alignment that results in semi-<br>automated border generation of the LAA<br>orifice and provides measurements based on<br>preliminary border generated by algorithm.<br>The border and the measurements can be<br>further refined by the end user. | Manual border drawing / tracing ability in<br>addition to manual MPR alignment and<br>provides preliminary measurement estimates<br>based on user defined border. | The subject 3D Auto LAA application<br>introduces a semi-automatic border detection<br>functionality for the LAA orifice while<br>specifying the measurements specific to the<br>LAA. The ability to align the MPRs based on<br>the anatomical structure is a shared feature<br>across both the primary and reference<br>predicate devices. |
| | Contour<br>Generation | LAA border is semi-autonomously generated<br>following the MPR alignment by the end user<br>by utilizing greyscale intensity<br>differentiation with some shape<br>regularization. The user can then modify the<br>entire contour (utilizing the rotary knob) or<br>an individual point(s) utilizing the trackball. | Contour is generated manually by the end user<br>utilizing the trackball and if the contour needs<br>modification, the end user must retrace the<br>contour. | The 3D Auto LAA introduces a semi-<br>automated workflow for the border<br>generation as compared to both predicate<br>devices. Similarly to the 3D Auto LAA<br>application, the predicate devices both allow<br>for the user to modify and adjust the MPRs to<br>better fit the anatomy. However, the<br>predicate devices both allow<br>for the user to modify and adjust the MPRs to<br>better fit the anatomy. However, the |
| Table 8-1: Comparison of The Subject Philips 3D Auto LAA Cardiac Quantification Application to the predicate Philips QLAB System<br>(the 3DQ Q-App) | | | | |
| | Subject Device | Predicate Device | Explanation of Differences | |
| Manufacturer | Philips Ultrasound, Inc. | Philips Ultrasound, Inc. | None | |
| Trade Name | 3D Auto LAA | QLAB System | None | |
| Feature | 3D Auto LAA | 3DQ | None | |
| | | | draw / trace the contour utilizing the<br>trackball. | |
| | Measurements<br>Parameters | *LAA Area (mm²)<br>*LAA Circumference (mm)<br>*LAA Max Diameter (mm)<br>*LAA Minimum Diameter (mm) | *Area (cm²)<br>*Circumference (cm)<br>*Distance (cm) (can be utilized to draw a line<br>and quantify both the max and min diameters) | The measurements provided by the 3D Auto<br>LAA application are presented to the user as<br>specific to the LAA orifice, however; the<br>primary predicate and reference predicate<br>devices are also capable of quantifying those<br>measurement parameters as the 3D Auto LAA<br>application. |
{7}------------------------------------------------
Table 8-1: Comparison of The Subject Philips 3D Auto LAA Cardiac Quantification to the predicate Philips QLAB System (the 3DQ Q-App)
{8}------------------------------------------------
{9}------------------------------------------------
#### Non-Clinical Testing 7.
The proposed introduction of the subject Philips 3D Auto LAA cardiac quantification was tested in accordance with Philips internal processes. Verification and software validation test data are provided to support the newest cardiac quantification application, the 3D Auto LAA application, relative to the currently marketed manual LAA tracing and measuring options.
Design Control activities to assure the safe and effective performance of the 3D Auto LAA application include but are not limited to the following:
- Requirements Review O
- Risk Analysis and Management O
- Product Specifications O
- Design Reviews O
- Software Verification and Validation O
Non-clinical V&V testing also included the Performance Validation Study for the proposed 3D Auto LAA clinical application.
Software Verification and Validation testing were used to support substantial equivalence of the new 3D Auto LAA cardiac quantification application (as part of the EPIQ / Affiniti System software release version 7.0) to the currently marketed manual LAA tracing and measuring options.
#### 8. Clinical Testing
The subject Philips 3D Auto LAA cardiac quantification did not require clinical data in order to make a determination for substantial equivalence when compared to the predicate device(s).
#### 9. Conclusion
Based on the conformance to standards, development under Philips Ultrasound's Quality Management System, the successful verification and validation testing, Philips Ultrasound believes that the proposed Philips 3D Auto LAA cardiac quantification application is substantially equivalent to the predicate device Philips QLAB System (K191647). Testing performed demonstrated that the proposed 3D Auto LAA application meets the defined requirements and performance claims.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.