TTBIO EVOCLEAN Ultrasonic Scaler is designed for use during dental cleaning and periodontal debridement of periodontal diseases.
Device Story
TTBIO EVOCLEAN Ultrasonic Scaler is a dental device used by trained professionals for cleaning and periodontal debridement. It utilizes an ultrasonic generator, handpiece, power supply, and foot pedal to convert electrical energy into mechanical micro-vibrations. The system automatically detects the resonant frequency of compatible 25 KHz or 30 KHz inserts, eliminating manual tuning. The device is a standalone unit operated via a foot pedal; the clinician applies the vibrating scaler tip to the patient's teeth to remove calculus and plaque. The output is mechanical vibration, which facilitates the removal of deposits during dental procedures. The device benefits patients by enabling effective periodontal therapy and dental prophylaxis. It is designed for clinical use in dental offices.
Clinical Evidence
No clinical data. Bench testing only. Verification performed per ISO 18397:2016. Biocompatibility evaluated per ISO 10993-1. Electrical safety and EMC testing conducted per IEC 60601-1-2:2014 and ANSI/AAMI ES60601-1:2005/(R)2012 & A1:2012. Cleaning/reprocessing validation performed per AAMI TIR30 and AAMI TIR12.
Technological Characteristics
Ultrasonic scaler; multi-frequency (25/30 KHz). Materials: ABS and PC (case), PA, TPU, PVC (components). Energy: 100-240 VAC, 50/60 Hz. Standalone form factor. Connectivity: Handpiece to generator, foot pedal control. Sterilization: Reprocessing per AAMI TIR30/TIR12. Software: Embedded firmware for frequency detection and operating modes (Intermittent, Boost, Purge, Cruise).
Indications for Use
Indicated for dental cleaning and periodontal debridement of periodontal diseases in patients requiring professional dental prophylaxis or periodontal therapy. Contraindicated for use in patients or operators with pacemakers and for amalgam restorative dental procedures.
Regulatory Classification
Identification
An ultrasonic scaler is a device intended for use during dental cleaning and periodontal (gum) therapy to remove calculus deposits from teeth by application of an ultrasonic vibrating scaler tip to the teeth.
Predicate Devices
BONART ART-M3II Ultrasonic Scaler UNITS WITH ACCESSORIES (K052028)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square and the words "U.S. FOOD & DRUG ADMINISTRATION".
December 4, 2020
Ttbio Corp. Shu-Ching Lee Section Chief 2F, NO.7, 6TH Road Industry Park Taichung, TAIWAN 40755 Taiwan
Re: K201317
Trade/Device Name: TTBIO EVOCLEAN Ultrasonic Scaler Regulation Number: 21 CFR 872.4850 Regulation Name: Ultrasonic Scaler Regulatory Class: Class II Product Code: ELC Dated: November 23, 2020 Received: November 27, 2020
Dear Shu-Ching Lee:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
For Srinivas "Nandu" Nandkumar, Ph. D. Director DHT1B: Division of Dental Devices OHT1: Office of Ophthalmic, Anesthesia, Respiratory, ENT and Dental Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known)
#### K201317
Device Name TTBIO EVOCLEAN Ultrasonic Scaler
Indications for Use (Describe)
TTBIO EVOCLEAN Ultrasonic Scaler is designed for use during dental cleaning and periodontal debridement of periodontal diseases.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
_ Over-The-Counter Use (21 CFR 801 Subpart C)
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for "TT BIO". The logo consists of two stylized "T" letters, one in blue and the other in green, connected to each other. To the right of the "T" letters, the word "BIO" is written in gray, with each letter in a bold, sans-serif font.
# 510(K) SUMMARY FOR K201317
# TTBIO EVOCLEAN Ultrasonic Scaler
# 1. Submitter Information
510(k) Owner: TTBIO CORP.
2F, No.7, 6th Road, Industry Park, Taichung, Taiwan, R.O.C. 40755
Contact Person: Shu-Ching Lee/Section Chief
Phone: 886-4-23595958
Email: jo@ttbio.com
- 2. Date Prepared: November 23, 2020
# 3. Device Name
Trade Name: TTBIO EVOCLEAN Ultrasonic Scaler Common Name: Ultrasonic Scaler Classification Name: Ultrasonic Scaler (21 CFR 872.4850) Regulatory Class: II Product Code: ELC
# 4. Predicate Device
Primary predicate device: BONART ART-M3II Ultrasonic Scaler UNITS WITH ACCESSORIES made by BONART CO.LTD. (K052028) No reference devices were used in this submission.
## 5. Device Description
TTBIO EVOCLEAN Ultrasonic Scaler consists of an ultrasonic generator, a handpiece, a power supply, a foot pedal and other minor accessories. There are two models of TTBIO EVOCLEAN Ultrasonic Scaler, one is EVOCLEAN and the other is EVOCLEAN+. The specifications of these two models are almost the same; the major difference is the connection type, fix and plug in, between the handpiece and main unit. TTBIO EVOCLEAN Ultrasonic Scaler uses ultrasonic energy to generate mechanical micro-vibration of the available inserts to perform the dental procedures defined in its intended use. TTBIO EVOCLEAN Ultrasonic Scaler is a multi-frequency available device which is compatible with 25 KHz or 30 KHz inserts. The system will automatically detect the insert oscillation frequency, need not to operate it manually. The handpiece is connected directly to the device's ultrasonic generator, from which it receives the functional drive signals, search and locate the resonant frequency of the insert, which varies according to the insert in use.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for TT BIO. The logo consists of two stylized "T" shapes, one in blue and one in green, followed by the word "BIO" in gray. The "T" shapes are interconnected and have rounded edges, giving the logo a modern and sleek appearance. The overall design is simple yet eye-catching, with the use of contrasting colors adding visual interest.
# 6. Indications for Use
TTBIO EVOCLEAN Ultrasonic Scaler is designed for use during dental cleaning and periodontal debridement of periodontal diseases.
# 7. Substantial Equivalence
| Substantial Equivalence Comparison Table | | | | |
|------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------|--------------|
| | Proposed Device | Primary Predicate Device | Variations | |
| 510 (k) Number | K201317 | K052028 | N/A | |
| Device Trade<br>Name | TTBIO EVOCLEAN Ultrasonic Scaler | BONART ART-M3II Ultrasonic<br>scaler Unit with accessories | Analysis (1) | |
| Regulation<br>Intended Use | "for use during dental cleaning and<br>periodontal (gum) therapy to remove<br>calculus deposits from teeth by<br>application of an ultrasonic vibrating<br>scaler tip to the teeth." | "for use during dental cleaning<br>and periodontal (gum) therapy to<br>remove calculus deposits from<br>teeth by application of an<br>ultrasonic vibrating scaler tip to<br>the teeth." | Identical | |
| Indications for use | The device is designed for use during<br>dental cleaning and periodontal<br>debridement of periodontal diseases. | Ultrasonic procedures:<br>-Removal of calculus and plaque<br>during dental prophylaxis.<br>-General supra and sub-gingival<br>scaling applications.<br>-Periodontal debridement for all<br>types of periodontal diseases.<br>-Endodontic procedures. | Similar | |
| contraindications | -Do not use TTBIO EVOCLEAN<br>Ultrasonic Scaler for amalgam<br>restorative dental procedures.<br>-Do not use this device if the patient or<br>operator is wearing a pacemaker. | -Do not use the BONART<br>ART-M3II for amalgam restorative<br>dental procedures.<br>-Do not use the BONART<br>ART-M3II if the patient or operator<br>is wearing a pacemaker. | Identical | |
| | Prescription/<br>over-the-counter<br>use | Prescription use | Prescription use | Identical |
| Components | -Main Unit<br>-Handpiece<br>-Foot pedal<br>-Power supply, AC power cord<br>-Water supply line with quick<br>disconnect<br>-Directions For Use | -Main Unit<br>-Handpiece<br>-Foot control / Foot switch<br>-AC power cord set<br>-PU water tubing, quick<br>disconnect<br>-User Manual<br>-Inserts | Analysis (1) | |
| | Direct<br>patient-contacting<br>Material | No | Insert: Plastic, Metal | Analysis (1) |
| | | Proposed Device | Primary Predicate Device | Variations |
| | User Contact<br>Material | Case : ABS and PC | Case : Ferrous cover | Analysis (2) |
| | | Others : PA, TPU and PVC | Others : Unknown | |
| | Treatment | Ultrasonic Scaling | Ultrasonic Scaling | Identical |
| Mechanism of<br>treatment | application of an ultrasonic<br>vibrating scaler tip to the teeth | application of an ultrasonic<br>vibrating scaler tip to the teeth | Identical | |
| Input Power | 100-240 VAC , 50/60 Hz | 115V±5% ~50/60Hz | Analysis (3) | |
| | | 230V±5% ~50/60Hz | | |
| Power (Max) | 60 W | 92 VA | Analysis (3) | |
| Scaler tip<br>frequency | 25 / 30 KHz | 25 / 30 KHz | Identical | |
| Water supply<br>pressure | 25 to 40 psi (172 to 275 kPa) | 25 to 60 psi (172~414KPa) | Similar | |
| Weight | 4.4 lbs (2kg) | 3.5kg (include handpiece) | Analysis (4) | |
| Dimensions | 15cm(D)x12.5cm(W)x 8.5cm(H) | 26cm(L)x 18.5cm(W)/x 7cm(H) | Analysis (4) | |
| | Handpiece Cable: 250cm | Handpiece Cable: 250cm | | |
| | Foot Pedal Cable: 250cm | Footswitch Cable: 250cm | | |
| Operating Mode | Intermittent, Boost, Purge and Cruise | Continuous Operation | Analysis (5) | |
| Flow rate<br>adjustment | Mechanical regulator | Mechanical regulator | Identical | |
| Housing | Stand alone device | Stand alone device | Identical | |
## Substantial Equivalence Comparison Table
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for "TT BIO". The logo consists of two stylized "T" letters, one in blue and the other in green, followed by the word "BIO" in gray. The "T" letters are connected and have a modern, curved design.
#### TTBIO CORP. 2F, NO.7, 6th ROAD INDUSTRY PARK, TAICHUNG, TAIWAN, R.O.C. 40755 TEL : 886-4-23595958 FAX : 886-4-23594196
TTBIO EVOCLEAN ultrasonic scaler is same with primary predicate device which intended for cleaning and periodontal therapy, and the device is only for operation by a trained professional in the field of general dentistry. The proposed device is essentially identical or similar to the primary predicate device in indications for use, principles of operation, energy source and technological characteristics. The minor differences are:
Analysis (1) The insert is not included in proposed device as its accessories.
Analysis (2) The material of proposed device is different from primary predicate device. However, the biocompatibility for proposed device has been evaluated, so no new issue of biocompatibility is raised with regard to the proposed devices.
Analysis (3) The electrical safety for proposed device has been tested and the test result is acceptable. This difference does not affect substantially equivalence.
Analysis (4) The proposed device has lighter weight and smaller size, these differences are negligible.
Analysis (5) The software validation report was provided to ensure these operating modes fulfills its intended purpose.
Based on performance testing methods and results, there is no significant difference in accordance with international standard ISO 18397:2016 (Dentistry - Powered scaler).
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for TT BIO. The logo consists of two stylized "T" letters, one in blue and the other in green, followed by the word "BIO" in gray. The two "T" letters are connected at the top, forming a single shape. The logo is simple and modern, and the colors are bright and eye-catching.
# 8. Non-clinical Testing:
The functions of TTBIO EVOCLEAN Ultrasonic Scaler were verified according to ISO18397:2016.
Cleaning validation testing is performed in accordance with recommended evaluations as listed in AAMI TIR30, AAMI TIR12, and Guidance for Industry and FDA Staff -
Processing/Reprocessing Medical Devices in Health Care Settings.
The biocompatibility evaluation for TTBIO EVOCLEAN Ultrasonic Scaler was conducted in accordance with Guidance for Industry and FDA Staff - Use of International Standard ISO 10993-1, "Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process".
Software verification and validation testing were conducted, and documentation was provided as recommended by FDA's Guidance for Industry and FDA Staff, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices". EMC and Electrical safety testing comply with IEC 60601-1-2:2014 and ANSI/AAMI ES60601-1:2005/(R)2012 & A1:2012.
# 9. Clinical Testing:
No clinical testing was conducted for this submission.
## 10. Conclusion:
TTBIO EVOCLEAN Ultrasonic Scaler, model EVOCLEAN and model EVOCLEAN+, is substantially equivalent to the primary predicate device. They are same as the primary predicate device in terms of its intended use, operating principles and functions.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.