K201239 · Bios S.R.L. · NGX · Dec 31, 2020 · Physical Medicine
Device Facts
Record ID
K201239
Device Name
NuEra Tight Family, EMS Model
Applicant
Bios S.R.L.
Product Code
NGX · Physical Medicine
Decision Date
Dec 31, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NuEra Tight Family, EMS Model, is intended for: - improvement of abdominal tone, strengthening of the abdominal muscles, development of firmer abdomen; - strengthening, toning and firming of buttocks, thighs and calves; - improvement of muscle tone and firmness, for strengthening muscles in arms.
Device Story
NuEra Tight Family, EMS Model, is a powered muscle stimulator using electromagnetic fields to induce electrical currents in muscles. Device comprises a main unit with a touch screen interface and two EMS handpieces (small and large) for treating various body parts. Operation involves generating intense magnetic fields to initiate nerve action potentials, resulting in muscle contraction. Used in clinical settings by healthcare providers to strengthen, tone, and firm muscles in the abdomen, buttocks, thighs, calves, and arms. The device is prescription-only. It features patient override controls and firmware-based control of stimulation parameters (1-150 Hz frequency, biphasic sinusoidal waveform). Benefits include non-invasive muscle conditioning and improved muscle tone.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified through electrical safety (IEC 60601-1, IEC 60601-1-2, IEC 60601-2-10), software verification (IEC 62304), and functional testing against technical specifications. Biocompatibility assessed per ISO 10993-1; device has no direct skin contact.
Technological Characteristics
Powered muscle stimulator; generates electromagnetic fields via magnetic coils in handpieces. Materials: steel and injection-molded plastics. Energy: magnetic field (0.5–2.0 T). Waveform: biphasic sinusoidal. Connectivity: touch screen interface, firmware-controlled. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-6, IEC 60601-2-10, IEC 62304, ISO 14971, ISO 10993-1.
Indications for Use
Indicated for improvement of abdominal tone, strengthening of abdominal muscles, development of firmer abdomen; strengthening, toning, and firming of buttocks, thighs, and calves; and improvement of muscle tone and firmness for strengthening arm muscles.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logos of the Department of Health & Human Services and the Food and Drug Administration (FDA). The Department of Health & Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the letters "FDA" in a blue square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in blue text.
December 31, 2020
Bios s.r.l. Maurizio Bianchi Head of QA/RA Via Guido Rossa 10/12 Vimodrone, MI 20090 Italy
Re: K201239
Trade/Device Name: NuEra Tight Family, EMS Model Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: NGX Dated: October 5, 2020 Received: October 8, 2020
Dear Maurizio Bianchi:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's
{1}------------------------------------------------
requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-devices/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance)and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Heather Dean, PhD Assistant Director, Acute Injury Devices Team DHT5B: Division of Neuromodulation and Physical Medicine Devices OHT5: Office of Neurological and Physical Medicine Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201239
Device Name NuEra Tight Family, EMS Model
***Indications for Use (Describe)***
| The NuEra Tight Family, EMS Model, is intended for: | |
|-----------------------------------------------------|--|
|-----------------------------------------------------|--|
- improvement of abdominal tone, strengthening of the abdominal muscles, development of firmer abdomen;
- strengthening, toning and firming of buttocks, thighs and calves;
- improvement of muscle tone and firmness, for strengthening muscles in arms.
| Type of Use (Select one or both, as applicable) | |
|-----------------------------------------------------------------------------------------------|----------------------------------------------------------|
| <span style="text-decoration: underline;">Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span>Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
# CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Contact Details | |
|-----------------------|-----------------------------------------------------------------------|
| 510(k) Number | K201239 |
| 510(k) Type | Traditional |
| Applicant Information | Bios s.r.l.<br>Via Guido Rossa, 10/12<br>20090 Vimodrone (MI) – Italy |
| Contact | Dr Eliana Russo |
| Date Prepared | 24 June 2020 |
| Device Name(s): | NuEra Tight Family, EMS Model |
| Model Refs | APMD151 |
| Common Name | Powered Muscle Stimulator |
| Regulatory Class | Class II |
| Product Codes | NGX |
| Regulation Names | Stimulator, muscle, powered, for muscle conditioning |
| Predicate Device for EMS Treatments | | | |
|-------------------------------------|---------------|----------------------------------------|----------------------|
| K190456 | NGX, 890.5850 | Powered Muscle Stimulator (BTL 799-2L) | BTL Industries, Inc. |
# Device Description
NuEra Tight Family, EMS Model, is a family of devices designed to:
- produce an electromagnetic field that induces electrical current in the muscles. By muscle stimulation, the system helps to strengthen, tone and firm the abdomen, buttocks, thighs and calves.
The device belonging to the NuEra Tight Family, EMS Model, generates an intense magnetic field which is used to stimulate muscles trough two EMS handpieces, one small and one large, which can treat body parts of different sizes.
# Indications for Use
The NuEra Tight Family, EMS Model, is intended for:
- . Improvement of abdominal tone, strengthening of the abdominal muscles, development of firmer abdomen;
- ı Strengthening, toning and firming of buttocks, thighs and calves;
- . Improvement of muscle tone and firmness, for strengthening muscles in arms.
# Predicate Device Comparison
The product specification, functionality, indications for use, and treatment parameters of the NuEra Tight Family, EMS Model, are the same or very similar to the legally marketed predicate and reference devices.
{4}------------------------------------------------
For the electromagnetic stimulation function, the subject device and predicate device have many identical, similar or substantially equivalent properties or features. No differences that could affect safety or effectiveness have been identified, as indicated in Table 1.
| Table 1: Predicate device comparison table – NuEra Tight Family, EMS Model – Electromagnetic Stimulation | | | |
|------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| Feature | Subject device | Predicate device | Similarity |
| Device name | NuEra Tight Family (EMS Model) | BTL 799-2L | N/A |
| Device Manufacturer | Bios S.r.l. | BTL Industries, Inc. | N/A |
| 510(K) Number | K201239 | K190456 | N/A |
| Product Code | NGX - Stimulator, Muscle, Powered | NGX - Stimulator, Muscle, Powered | Same |
| Regulation | 21 CFR 890.5850 | 21 CFR 890.5850 | Same |
| Indications for Use | The NuEra Tight Family (EMS Model) is intended:<br>• for improvement of abdominal tone, strengthening of the abdominal muscles, development of firmer abdomen;<br>• for strengthening, toning and firming of buttocks, thighs and calves;<br>• for improvement of muscle tone and firmness, for strengthening muscles in arms. | BTL 799-2L is indicated to be used for:<br>• Improvement of abdominal tone, strengthening of the abdominal muscles, development of firmer abdomen.<br>• Strengthening, toning and firming of buttocks, thighs and calves.<br>• Improvement of muscle tone and firmness, for strengthening muscles in arms. | Same |
| Principle of action | Initiating action potential of nerves results in muscle contraction. | Initiating action potential of nerves results in muscle contraction. | Same |
| Clinical use | Prescription Use | Prescription Use | Same |
| Electrical Protection | Class I type BF | Class I type BF | Same |
| User Interface | Touch Screen | Touch Screen | Same |
| Firmware Controlled | Yes | Yes | Same |
| Type of energy | Magnetic Field | Magnetic Field | Same |
| Number of outputs | 2 | 2 | Same |
| Number of magnetic<br>coils in the applicator | 1 | 1 | Same |
| Magnetic Field Intensity | Large HP: 0.5 – 1.8 T ± 20% | BTL 299-6 applicator: 0.5 - 1.8 T ± 20% | Same |
| | Small HP: 0.7 – 2.0 T ± 20% | BTL 299-7 applicator: 0.7 - 2.0 T ± 20% | Same |
| Pulse Repetition Rate | 1-150 Hz | 1-150 Hz | Same |
| Pulse Duration | Large HP: 280 ± 20% µs | BTL 299-6 applicator: 280 ± 20% µs | Same |
| | Small HP: 190 ± 20% µs | BTL 299-7 applicator: 190 ± 20% µs | Same |
| Therapy Time | Up to 60 min | Up to 60 min | Same |
| Characteristics | Bios Device | Predicate Device | Notes |
| Device name | NuEra Tight Family (EMS Model) | BTL 799-2L | N/A |
| Device Manufacturer | Bios s.r.l. | BTL Industries, Inc. | N/A |
| 510(K) Number | K201239 | K190456 | N/A |
| Power Source(s)<br>- Method of Line<br>Current Isolation<br>Patient Leakage<br>Current<br>- Normal condition<br>- Single fault condition | IEC 60601-1 compliant<br>< 0,1 μΑ<br>< 5 μΑ | IEC 60601-1 compliant<br>< 0,1 μΑ<br>< 5 μΑ | N/A |
| Average DC current<br>through electrodes<br>when device is on but<br>no pulses are being<br>applied (µA) | N/A | N/A | No electrodes.<br>Applicators are<br>not connected<br>to patient. |
| Number of Output<br>Modes | 1 | 1 | N/A |
| Number of Output<br>Channels<br>- Synchronous or<br>Alternating? | 2<br>Synchronous | 2<br>Synchronous | No electrodes.<br>Applicators are<br>not connected<br>to patient. |
| - Method of Channel<br>Isolation | N/A | N/A | |
| Regulated Current or<br>Regulated Voltage? | Regulated voltage | Regulated Voltage | N/A |
| Software /Firmware<br>/Microprocessor<br>Control? | Yes | Yes | N/A |
| Automatic Overload<br>Trip? | N/A | N/A | No electrodes.<br>Applicators are<br>not connected<br>to patient. |
| Automatic No-Load<br>Trip? | N/A | N/A | No electrodes.<br>Applicators are<br>not connected<br>to patient. |
| Automatic Shut Off? | No | Not available | N/A |
| Patient Override<br>Control? | Yes | Not available | N/A |
| Indicator Display:<br>- On/Off Status?<br>- Low Battery?<br>- Voltage/Current<br>Level? | Yes<br>N/A<br>No | Yes<br>N/A<br>No | N/A |
| Timer Range (minutes) | Up to 30 minutes | Up to 30 minutes | N/A |
| Compliance with<br>Voluntary Standards?<br>(If yes, specify) (If yes,<br>specify) | N/A | N/A | N/A |
| Compliance<br>with 21 CFR 898? | N/A | N/A | N/A |
| Weight | 100 kg | Not available | N/A |
| Dimensions (in.) [W x<br>H x D] | 1350x560x710 mm<br>(53x22x28 in) | 580x1380x580 mm (23x55x23 in) | N/A |
| Housing Materials and<br>Construction | Steel and Injection Molded<br>Plastics | Steel and Injection Molded Plastics | N/A |
{5}------------------------------------------------
#### Table 2: Basic Unit Characteristics
This section is intended to describe basic unit characteristics. The parameters listed are assumed to be independent of the selected output mode. If this is not the case, or if the information is not applicable to the device, an explanation should be provided. If a specific parameter is not applicable (N/A), this should be noted.
{6}------------------------------------------------
| Table 3: Output Specifications | | | |
|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------|----------------------------------------------------------------------------------------------|-------------------------------------------------------------------|
| An output mode is defined (for reporting purposes) as a<br>version of a waveform produced by the unit. For example,<br>biphasic symmetrical, biphasic asymmetrical, and<br>monophasic would all be considered separate output<br>modes. A copy of the following information should be<br>completed for each output mode. If a specific parameter is<br>not applicable (N/A), this should be noted. | | | |
| | | Unique output mode.<br>No electrical output delivered to the body for this EMS<br>technology | |
| Characteristics | Bios Device | Predicate Device | Notes |
| Waveform (e.g.,<br>pulsed monophasic,<br>biphasic) | Biphasic | Biphasic | N/A |
| Shape (e.g.,<br>rectangular, spike,<br>rectified sinusoidal) | Sinusoidal | Sinusoidal | N/A |
| Maximum Output<br>Voltage (specify units) | N/A | N/A | No electrodes.<br>Applicators are<br>not connected<br>to patient. |
| Maximum Output<br>Current (specify units) | N/A | N/A | No electrodes.<br>Applicators are<br>not connected<br>to patient. |
| Pulse Width (specify<br>units) | Large HP: 280 ± 20% µs | BTL 299-6 applicator: 280 ± 20% µs | N/A |
| | Small HP: 190 ± 20% µs | BTL 299-7 applicator: 190 ± 20% µs | N/A |
| Frequency (Hz) | 1-150 Hz | 1-150 Hz | N/A |
| For interferential<br>modes only:<br>- Beat Frequency (Hz) | N/A | N/A | N/A |
| For multiphasic<br>waveforms only:<br>- Symmetrical phases?<br>- Phase Duration<br>(include units)<br>(state range, if<br>applicable)<br>(both phases, if<br>asymmetrical) | Yes | Yes | N/A |
| Net Charge (mC per pulse) | | | |
| Net Charge (mC per pulse) | N/A | N/A | No electrodes.<br>Applicators are not connected to patient. |
| Maximum Phase Charge, (mC) | N/A | N/A | No electrodes.<br>Applicators are not connected to patient. |
| Maximum Current Density (mA/cm²) | N/A | N/A | No electrodes.<br>Applicators are not connected to patient. |
| Maximum Average Current | N/A | N/A | No electrodes.<br>Applicators are not connected to patient. |
| Maximum Power Density (W/cm²) (using smallest electrode conductive surface area) | N/A | N/A | No electrodes.<br>Applicators are not connected to patient. |
| Burst Mode (i.e., pulse trains)<br>a. Pulses per burst<br>b. Bursts per second<br>c. Burst duration (seconds)<br>d. Duty Cycle [Line (b) x Line (c)] | N/A | N/A | N/A |
| ON Time (seconds) | N/A | N/A | N/A |
| OFF Time (seconds) | N/A | N/A | N/A |
| Additional Features (if applicable) | N/A | N/A | N/A |
{7}------------------------------------------------
{8}------------------------------------------------
# Performance Data
#### Electrical safety and electro-magnetic compatibility
The NuEra Tight Family, EMS Model, has been tested and is in compliance with:
- . IEC 60601-1; Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance;
- . IEC 60601-1-2; Medical Electrical Equipment - Part 1-2: General Requirements For Basic Safety And Essential Performance - Collateral Standard: Electromagnetic Disturbances - Requirements And Tests;
- IEC 60601-1-6; Medical Electrical Equipment Part 1-6: General Requirements For Basic Safety And ● Essential Performance - Collateral Standard: Usability;
- IEC 60601-2-10/A1; Medical Electrical Equipment - Part 2-10: Particular Requirements For The Basic Safety And Essential Performance Of Nerve And Muscle Stimulators;
- IEC 62304; Medical Device - Software Life Cycle Processes;
- ISO 14971; Medical Devices Application Of Risk Management To Medical Devices .
# Software Verification and Validation
In addition to the electrical safety testing performed, software verification was conducted to FDA regulations, standards and guidance document requirements. The results of this testing conclude the software has met these requirements. Design verification was also performed on the NuEra Tight Family, EMS Model, in compliance with internal design control procedures. The results of this testing conclude the NuEra Tight Family, EMS Model, is determined to be safe and effective.
# Functional Testing
Functional tests have been performed to confirm that the performance of the NuEra Tight Family, EMS Model, devices are aligned with the device technical specifications.
#### Biocompatibility
Based on the type and duration of body contact, existing biocompatibility data has been assessed in accordance with ISO 10993-1:2018 (FDA standards recognition # 2-258) and FDA guidance document 'Use of International Standard ISO 10993-1, "Biological evaluation of medical devices - Part 1: Evaluation and testing within a risk management process": Guidance for Industry and Food and Drug Administration Staff', 16 June 2016.
The NuEra Tight Family, EMS Model, has no contact with the skin of the treated patients.
# Conclusion
Based upon the indications for use and known technical information provided in this pre-market notification, the NuEra Tight Family, EMS Model, devices has been shown to be substantially equivalent to currently marketed predicate and reference devices. Any differences are considered minor and do not raise new issues of the safety and effectiveness of the NuEra Tight Family, EMS Model, devices when compared to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.