Arthrex MaxForce MTP Compression Plates and Screws
Applicant
Arthrex, Inc.
Product Code
HRS · Orthopedic
Decision Date
Jul 6, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Arthrex MaxForce MTP Compression Plates are intended to be used for internal bone fixation for bone fractures, fusions, or osteotomies in the Forefoot. The Arthrex Low Profile Screws (2.0-3.0mm solid) are intended to be used as stand-alone bone screws, or in a plate-screw system for internal bone fixation for bone fractures, fusions, osteotomies and non-unions in the ankle, foot, hand, and wrist. When used with a plate, the screws may be used with the Arthrex Low Profile Plates and MaxForce MTP Compression Plates. The Arthrex Low Profile Screws (3.5mm and larger, solid) are intended to be used as stand-alone bone screws, or in a plate-screw system for internal bone fractures, fusions, osteotomies and non-unions in the ankle, foot, hand, wrist, clavicle, scapula, olecranon, humerus, radius, ulna, tibia, calcaneous, femur and with a plate, the screws may be used with the Arthrex Low Profile Plates, Humeral Fracture Plates, Osteotomy Plates, and MaxForce MTP Compression Plates.
Device Story
System consists of titanium alloy plates and screws for internal bone fixation; used for fractures, fusions, osteotomies, and non-unions. Plates are contoured for metatarsal anatomy (Petite, Standard, Long, X-Long, Revision) with varying valgus/dorsiflexion angles. Screws (3.0-3.5mm diameter) provide fixation. Used in clinical settings by orthopedic surgeons. Device provides mechanical stability to bone segments to facilitate healing. Non-sterile, single-use.
Clinical Evidence
Bench testing only. Mechanical performance validated via Static Four-Point Bend (ASTM F382), Four-Point Bend Fatigue (ASTM F382), plate/instrument compression, pull-out strength (ASTM F543), torsional strength (ASTM F543), and insertion torque. MR safety validated per ASTM F2052, F2119, F2182, and F2213.
Indicated for internal bone fixation (fractures, fusions, osteotomies, non-unions) in the forefoot, ankle, foot, hand, wrist, clavicle, scapula, olecranon, humerus, radius, ulna, tibia, calcaneus, femur, and fibula. Patient population includes individuals requiring orthopedic bone fixation.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Arthrex Inc. Rebecca Homan Regulatory Affairs Specialist 1370 Creekside Boulevard Naples, Florida 34108-1945
## Re: K201235
Trade/Device Name: Arthrex MaxForce MTP Compression Plates and Screws Regulation Number: 21 CFR 888.3030 Regulation Name: Single/Multiple Component Metallic Bone Fixation Appliances And Accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: May 4, 2020 Received: May 7, 2020
Dear Rebecca Homan:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Shumava Ali. MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
## Indications for Use
510(k) Number (if known) K201235
Device Name
Arthrex MaxForce MTP Compression Plates and Screws
#### Indications for Use (Describe)
The Arthrex MaxForce MTP Compression Plates are intended to be used for internal bone fixation for bone fractures, fusions, or osteotomies in the Forefoot.
The Arthrex Low Profile Screws (2.0-3.0mm solid) are intended to be used as stand-alone bone screws, or in a platescrew system for internal bone fixation for bone fractures, fusions, osteotomies and non-unions in the ankle, foot, hand, and wrist. When used with a plate, the screws may be used with the Arthrex Low Profile Plates and MaxForce MTP Compression Plates.
The Arthrex Low Profile Screws (3.5mm and larger, solid) are intended to be used as stand-alone bone screws, or in a plate-screw system for internal bone fractures, fusions, osteotomies and non-unions in the ankle, foot, hand, wrist, clavicle, scapula, olecranon, humerus, radius, ulna, tibia, calcaneous, femur and with a plate, the screws may be used with the Arthrex Low Profile Plates, Humeral Fracture Plates, Osteotomy Plates, and MaxForce MTP Compression Plates.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| <div> <span> Prescription Use (Part 21 CFR 801 Subpart D) </span> </div> |
|---------------------------------------------------------------------------------------------|
| <div> <span> Over-The-Counter Use (21 CFR 801 Subpart C) </span> </div> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
# 510(k) Summary
| Date Prepared | July 6, 2020 |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Submitter | Arthrex Inc.<br>1370 Creekside Boulevard<br>Naples, FL 34108-1945 |
| Contact Person | Rebecca R. Homan<br>Regulatory Affairs Specialist<br>1-239-643-5553, ext. 73429<br>rebecca.homan@arthrex.com |
| Name of Device | Arthrex MaxForce MTP Compression Plates and Screws |
| Common Name | Plate, fixation, bone<br>Screw, fixation, bone |
| Product Code | HRS; HWC |
| Classification Name | 21 CFR 888.3030: Single/multiple component metallic bone fixation appliances<br>and accessories (Primary Classification);<br>21 CFR 888.3040: Smooth or Threaded metallic bone fixation fastener |
| Regulatory Class | II |
| Predicate Devices | K130510: Arthrex Compression Plates (Primary Predicate)<br>K111253: Arthrex Distal Extremity Plate System<br>K103705: Arthrex Low Profile Screws |
| Purpose of Submission | This Traditional 510(k) premarket notification is submitted to obtain clearance for<br>the Arthrex MaxForce MTP Compression Plates and Screws. |
| Device Description | The Arthrex MaxForce MTP Compression Plates and Screws consists of a series of<br>plates and screws of varying lengths and diameters. The plates are contoured to<br>fit the various aspects of the metatarsals including Petite, Standard, Long, X-Long<br>and Revision with 0° and 5° Valgus and Dorsiflex in left and right configurations.<br>The plates are manufactured from titanium alloy conforming to ASTM F136. The<br>plates are attached to bone with 3.0 mm cortical, 3.0 mm cortical, hybrid and 3.0<br>mm variable locking screws. The screws range from 3.0 mm to 3.5 mm in<br>diameter and from 10 mm to 26 mm in length. The screws are manufactured<br>from titanium alloy conforming to ASTM F136. The plates and screws are sold<br>non-sterile and single-use. |
| Indications for Use | The Arthrex MaxForce MTP Compression Plates are intended to be used for<br>internal bone fixation for bone fractures, fusions, or osteotomies in the Forefoot.<br><br>The Arthrex Low Profile Screws (2.0-3.0mm solid) are intended to be used as<br>stand-alone bone screws, or in a plate-screw system for internal bone fixation for<br>bone fractures, fusions, osteotomies and non-unions in the ankle, foot, hand, and<br>wrist. When used with a plate, the screws may be used with the Arthrex Low<br>Profile Plates, Small Fragment Plates and MaxForce MTP Compression Plates.<br><br>The Arthrex Low Profile Screws (3.5mm and larger, solid) are intended to be used<br>as stand-alone bone screws, or in a plate-screw system for internal bone fixation<br>for bone fractures, fusions, osteotomies and non-unions in the ankle, foot, hand,<br>wrist, clavicle, scapula, olecranon, humerus, radius, ulna, tibia, calcaneous, femur<br>and fibula. When used with a plate, the screws may be used with the Arthrex<br>Low Profile Plates, Small Fragment Plates, Humeral Fracture Plates, Osteotomy<br>Plates, and MaxForce MTP Compression Plates. |
| | |
| <b>Performance Data</b> | Engineering analysis, Static Four-Point Bend (ASTM F382), Four-Point Bend<br>Fatigue (ASTM F382), and plate/instrument compression testing were conducted<br>on the subject plates. Pull-out strength testing (ASTM F543), torsional strength<br>testing (ASTM F543) and Insertion Torque testing were conducted on the subject<br>screws. The testing demonstrates that the Arthrex MaxForce MTP Compression<br>Plates and Screws performs statistically equivalent to the predicate devices<br>cleared under K130510, K111253 and K103705.<br><br>MRI force, torque, and image artifact testing were conducted in accordance with<br>FDA guidance Testing and Labeling Medical Devices for Safety in the Magnetic<br>Resonance (MR) Environment, ASTM F2052 Standard Test Method for<br>Measurement of Magnetically Induced Displacement Force on Medical Devices in<br>the Magnetic Resonance Environment, ASTM F2119 Standard Test Method for<br>Evaluation of MR Image Artifacts from Passive Implants, ASTM F2182 Standard<br>Test Method for Measurement of Measurement of Radio Frequency Induced<br>Heating Near Passive Implants During Magnetic Resonance Imaging and ASTM<br>F2213 Standard Test Method for Measurement of Magnetically Induced Torque<br>on Medical Devices in the Magnetic Resonance Environment. |
| <b>Technological<br/>Comparison</b> | The Arthrex MaxForce MTP Compression Plates and Screws are substantially<br>equivalent to the predicate devices in which the basic design features, intended<br>use, fundamental scientific technology, indications for use, sterility, packaging,<br>and shelf-life are identical.<br><br>The Arthrex MaxForce MTP Compression Plates are contoured to fit the various<br>aspects of the metatarsals including Petite, Standard, Long, X-Long and Revision<br>with 0° and 5° Valgus and Dorsiflex in left and right configurations. The<br>predicate Arthrex Compression Plates cleared under K130510 are<br>contoured to fit the various aspects of the foot, ankle, hand and wrist<br>including Square, Linear, T- configurations. The predicate Arthrex<br>Contoured Plates cleared under K111253 are contoured to fit the metatarsals<br>including Petite, Standard, and Long in left and right configurations.<br><br>The Arthrex MaxForce MTP Compression Plates and Screws are manufactured<br>from titanium; whereas the predicate Arthrex Compression Plates cleared under<br>K130510 are manufactured from stainless steel. However, the titanium alloy<br>used to manufacture the Arthrex MaxForce MTP Compression Plates and Screws<br>is identical is identical to the titanium alloy cleared for the Arthrex Contoured<br>Plates cleared under K111253.<br><br>The Arthrex MaxForce MTP Compression Plates and Screws were evaluated for<br>MR Conditional labeling; whereas the predicate devices were not evaluated for<br>MR Conditional labeling. |
| <b>Conclusion</b> | The Arthrex MaxForce MTP Compression Plates and Screws are substantially<br>equivalent to the predicate devices in which the basic design features and<br>intended uses are the same. Any differences between the proposed devices and<br>the predicate devices are considered minor and do not raise different questions<br>concerning safety or effectiveness. |
| The Arthrex MaxForce MTP Compression Plates and Screws MR compatibility<br>testing supports the devices MR Conditional labeling. There is no increased risk<br>from this difference in technology. | |
| The submitted mechanical testing data demonstrates that the Static Four-Point<br>Bend, Four-Point Bend Fatigue, Compression Force, Compression Distance, Pull-<br>out and Torque strength of the Arthrex MaxForce MTP Compression Plates and<br>Screws is substantially equivalent to that of the predicate devices for the desired<br>indications. | |
| Based on the indications for use, technological characteristics, and the summary<br>of data submitted, Arthrex Inc. has determined that the proposed devices are<br>substantially equivalent to the currently marketed predicate devices. | |
{4}------------------------------------------------
{5}------------------------------------------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.