← Product Code [HRS](/productcode/HRS) · K201229

# Intai Anatomic Locking Plate and Screw System (K201229)

_Intai Technology Corporation · HRS · Mar 11, 2022 · Orthopedic · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K201229

## Device Facts

- **Applicant:** Intai Technology Corporation
- **Product Code:** [HRS](/productcode/HRS.md)
- **Decision Date:** Mar 11, 2022
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Intai Anatomy Locking Plate and Screw System is provided non-sterile. The system is intended for use in fixation of fractures to the various bones, including; clavicle, radius, ulna, humerus, femur, tibia, fibula, and acetabulum for adults. This system is not indicated for use in the spine.

## Device Story

System consists of anatomically contoured metallic plates and screws for bone fracture fixation; provides both locking and non-locking screw options; includes holes for cortex screws to enhance compression pressure. Used by surgeons in clinical settings to stabilize fractures; provides mechanical support to bone segments during healing. Benefits include improved fracture stability through combined locking and compression technology.

## Clinical Evidence

Bench testing only. Performance demonstrated via mechanical testing including single cycle bend and bending fatigue (ASTM F382), and torsional properties, driving torque, and axial pullout testing (ASTM F543).

## Technological Characteristics

Materials: Titanium alloy (ASTM F136) or Stainless Steel (ASTM F138/F139). Design: Anatomically contoured locking compression plates with locking and non-locking screw holes. Mechanical fixation principle. Non-sterile delivery.

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Synthes 3.5mm/4.5mm Curved Narrow and Broad Locking Compression Plates (LCP) ([K092609](/device/K092609.md))
- Synthes 3.5mm LCP Clavicle Plate System ([K073186](/device/K073186.md))
- Synthes Clavicle Hook Plate ([K061753](/device/K061753.md))
- Stryker Leibinger Universal Distal Radius System ([K040022](/device/K040022.md))
- Zimmer Periarticular Locking Plate System: Distal Humeral and Proximal Ulna Plate ([K082078](/device/K082078.md))
- Synthes LCP Proximal Humerus Plates, Long ([K041860](/device/K041860.md))
- Zimmer Periarticular Locking Plate System: Distal Humeral and Proximal Ulna Plate ([K082078](/device/K082078.md))
- Synthes 3.5mm Titanium LCP Proximal Tibia Plate ([K030597](/device/K030597.md))
- Synthes LCP Distal Tibia Plate ([K013248](/device/K013248.md))
- Synthes 2.7 mm/ 3.5 mm LCP Distal Fibula Plates ([K073460](/device/K073460.md))
- Synthes LCP Distal Femur Plates ([K062564](/device/K062564.md))
- Synthes 3.5mm/4.5mm LCP Metaphyseal Plate ([K033805](/device/K033805.md))
- Synthes 3.5 mm Quadrilateral Surface Plates ([K093928](/device/K093928.md))
- Synthes Low Profile Reconstruction Plates ([K042377](/device/K042377.md))
- Synthes 3.5mm Cortex Screw ([K043185](/device/K043185.md))
- Zimmer Universal Locking System, 3.5mm Plates And Screws-4936 Plates, 4835 Screws ([K060710](/device/K060710.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

March 11, 2022

Image /page/0/Picture/1 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

Intai Technology Corporation Dale Chang Head of Implant Business Unit No. 9 Jingke Rd., Nantun Dist. Taichung, Taichung 40852 Taiwan

Re: K201229

Trade/Device Name: Intai Anatomic Locking Plate and Screw System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: April 14, 2020 Received: May 7, 2020

Dear Dale Chang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's

{1}------------------------------------------------

requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Shumaya Ali, MPH Assistant Director DHT6C: Division of Restorative, Repair and Trauma Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K201229

Device Name

Intai Anatomical Locking Plate and Screw System

Indications for Use (Describe)

The Intai Anatomy Locking Plate and Screw System is provided non-sterile.

The system is intended for use in fixation of fractures to the various bones, including; clavicle, radius, ulna, humerus, femur, tibia, fibula, and acetabulum for adults.

This system is not indicated for use in the spine.

| Type of Use (Select one or both, as applicable)                                    |                                                                                   |
|------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| <span style="font-size:12px">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | <span style="font-size:12px">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for INTAI. The logo features a stylized "i" with a yellow circle above it and a blue curved shape to the left. The word "INTAI" is written in blue, bold letters to the right of the "i". The logo has a reflection effect below it.

## K201229 - 510(k) SUMMARY

This summary regarding 510(k) information is being submitted in accordance with the requirements of 21 CFR 807.92.

| Submitted By                                    | Intai Technology Corp.<br>Registration No.3011187779<br>No.9 Jingke Rd., Nantun District, Taichung City, Taiwan, R.O.C.<br>Tel. +886 4 23595336<br>Fax.+886 4 36013076                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |                                     |                      |                                                                                  |  |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------|----------------------|----------------------------------------------------------------------------------|--|
| Contact Person                                  | Kevin Wang                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                     |                      |                                                                                  |  |
| Date Prepared                                   | March 11th, 2022                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                     |                      |                                                                                  |  |
| Device Name                                     | Intai Anatomical Locking Plate and Screw System                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                     |                      |                                                                                  |  |
| Classifications<br>Product Codes                | Class II, 21 CFR 888.3030 - Single/multiple component metallic bone<br>fixation appliances and accessories. (primary)<br>Class II, 21 CFR 888.3040 - Smooth or threaded metallic bone fixation<br>fastener.<br>HRS (primary), HWC                                                                                                                                                                                                                                                                                                                                                                                                                                         |                                     |                      |                                                                                  |  |
| Predicate Devices<br>Information                | The device is substantially equivalent to the marketed:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                     |                      |                                                                                  |  |
|                                                 | No.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Premarket<br>Notification<br>Number | Name                 | Device Name                                                                      |  |
|                                                 | 013                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K092609                             | Synthes              | 3.5mm/4.5mm Curved Narrow and<br>Broad Locking Compression Plates<br>(LCP)       |  |
|                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | Other Predicate Devices             |                      |                                                                                  |  |
|                                                 | 001                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K073186                             | Synthes              | 3.5mm LCP Clavicle Plate System                                                  |  |
|                                                 | 002                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K061753                             | Synthes              | Clavicle Hook Plate                                                              |  |
|                                                 | 003                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K040022                             | Stryker<br>Leibinger | Universal Distal Radius System                                                   |  |
|                                                 | 004                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K082078                             | Zimmer               | Periarticular Locking Plate System:<br>Distal Humeral and Proximal Ulna<br>Plate |  |
|                                                 | 005                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K041860                             | Synthes              | LCP® Proximal Humerus Plates,<br>Long                                            |  |
|                                                 | 006                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K082078                             | Zimmer               | Periarticular Locking Plate System:<br>Distal Humeral and Proximal Ulna<br>Plate |  |
|                                                 | 007                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K030597                             | Synthes              | 3.5mm Titanium LCP Proximal Tibia<br>Plate                                       |  |
|                                                 | 008                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K013248                             | Synthes              | LCP Distal Tibia Plate                                                           |  |
|                                                 | 009                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K073460                             | Synthes              | 2.7 mm/ 3.5 mm LCP Distal Fibula<br>Plates                                       |  |
|                                                 | 010                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K062564                             | Synthes              | Synthes LCP Distal Femur Plates                                                  |  |
|                                                 | 011                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K033805                             | Synthes              | 3.5mm/4.5mm LCP Metaphyseal<br>Plate                                             |  |
|                                                 | 012                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K093928                             | Synthes              | 3.5 mm Quadrilateral Surface Plates                                              |  |
|                                                 | 014                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K042377                             | Synthes              | Low Profile Reconstruction Plates                                                |  |
|                                                 | 015                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K043185                             | Synthes              | 3.5mm Cortex Screw                                                               |  |
|                                                 | 016                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | K060710                             | Zimmer               | Universal Locking System, 3.5mm<br>Plates And Screws-4936<br>Plates, 4835 Screws |  |
| Material                                        | Titanium-alloy per ASTM F136 or Stainless Steel per ASTM F138/F139                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        |                                     |                      |                                                                                  |  |
| Device Description                              | The Intai Anatomical Locking Plate and Screw System offers anatomically<br>contoured plates for use with non-locking and locking screws. The plates are<br>provided with locking holes as well as holes for cortex screws to enhance<br>compression pressure. With the combination of locking and compression<br>technology, the system further ensures the stability of the implant devices.                                                                                                                                                                                                                                                                             |                                     |                      |                                                                                  |  |
| Indication for use                              | The Intai Anatomy Locking Plate and Screw System is provided non-sterile.<br>The system is intended for use in fixation of fractures to the various bones,<br>including: clavicle, radius, ulna, humerus, femur, tibia, fibula, and<br>acetabulum for adults. This system is not indicated for use in the spine.                                                                                                                                                                                                                                                                                                                                                          |                                     |                      |                                                                                  |  |
| Summary of<br>Technological<br>Characteristics: | The subject device and the predicate devices have the same intended use and<br>have the same or similar technological characteristics. The subject and<br>predicate devices are all fabricated from the same or similar materials and<br>share similar design characteristics. Any differences in the technological<br>characteristics do not raise new issues of safety or efficacy.  Intended Use: Both the subject and predicate devices are intended for<br>use in fixation fracture of bones. Materials: Use similar material.<br>-Stainless Steel<br>-Titanium Alloy Design Features: Both the subject and predicate devices include<br>Locking Compression Plates. |                                     |                      |                                                                                  |  |
| Performance Data<br>(Non-clinical) discussion   | Performance of the Intai Anatomical Locking Plate and Screw System has<br>been demonstrated substantial equivalence through the test parameters<br>outlined in ASTM F382 to conduct Single Cycle Bend and Bending fatigue<br>testing. In addition, Torsional Properties, Driving Torque, and Axial Pullout<br>testing was conducted on the screws per ASTM F543.                                                                                                                                                                                                                                                                                                          |                                     |                      |                                                                                  |  |
| Conclusion                                      | The subject device has been compared to the predicate devices with respect<br>to intended use, materials, design features, and non-clinical tests. The<br>technological characteristics of the subject device do not raise new type of<br>questions regarding safety and efficacy. These comparison demonstrate that<br>the Intai Anatomical Locking Plate and Screw System is substantially<br>equivalence to the predicate devices.                                                                                                                                                                                                                                     |                                     |                      |                                                                                  |  |

Intai Anatomical Locking Plate and Screw System for 510(k)

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for INTAI. The logo consists of a blue stylized "i" with three curved lines inside the left side of the "i". There is a yellow circle above the "i" and a yellow vertical line on the right side of the "i". The word "INTAI" is written in blue next to the "i".

---

**Source:** [https://fda-staging.innolitics.com/device/K201229](https://fda-staging.innolitics.com/device/K201229)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com