The Biocore9 Femoral Resurfacing Head Component, is indicated for the treatment of painful hip arthritis due to post-traumatic arthritis, or rheumatoid arthritis, as well as painful hip arthroplasty when sufficient femoral head and neck stock exist to stabilize a resurfacing femoral head. The Biocore9 Femoral Resurfacing Head Component is intended for cemented use only. This device is intended for hemi-arthroplasty only.
Device Story
Non-constrained orthopedic implant; replaces articular surface of femoral head. Axisymmetric truncated spherical shell with tapered central alignment stem; internal surface features porous titanium bead coating for cement fixation. Manufactured from Ti-6Al-4V alloy with Titanium Nitride (TiN) ceramic coating. Available in thirteen sizes. Used in surgical hip hemi-arthroplasty; implanted by orthopedic surgeons. Provides stable articular surface; restores joint function; reduces pain associated with arthritis or failed prior prostheses.
Clinical Evidence
Bench testing only. Testing included TiN coating characterization (adhesion, thickness, hardness, scratch testing), CP Ti porous coating characterization (tensile strength, shear strength, porosity), Mode 1 wear testing, abrasion resistance testing, and ISO 1143 rotating beam fatigue testing.
Technological Characteristics
Material: Ti-6Al-4V alloy (ASTM F136-13) with Titanium Nitride (TiN) ceramic coating. Design: Axisymmetric truncated spherical shell with tapered central alignment stem. Internal surface: Porous titanium bead coating. Energy source: None (mechanical implant). Sterilization: Gamma irradiation. Connectivity: None.
Indications for Use
Indicated for patients with painful hip arthritis (post-traumatic, osteoarthritis, rheumatoid) or painful hip arthroplasty requiring hemi-arthroplasty, provided sufficient femoral head and neck bone stock exists for stabilization. Intended for cemented use only.
Regulatory Classification
Identification
A hip joint femoral (hemi-hip) metallic resurfacing prosthesis is a device intended to be implanted to replace a portion of the hip joint. This generic type of device includes prostheses that have a femoral resurfacing component made of alloys, such as cobalt-chromium-molybdenum.
Predicate Devices
Endotec Inc. Modified New Jersey Femoral Hip Resurfacing Component (K904870)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA acronym and name on the right. The FDA part of the logo is in blue, with the acronym in a square and the full name, "U.S. Food & Drug Administration," written out next to it.
January 19, 2021
BioCore9 LLC Steve Peoples Consultant 9 Whippany Road, BLDG A1, Unit 12 A Whippany, New Jersey 07981
Re: K201219
Trade/Device Name: Biocore9 Femoral Head Resurfacing Component Regulation Number: 21 CFR 888.3400 Regulation Name: Hip Joint Femoral (Hemi-Hip) Metallic Resurfacing Prosthesis Regulatory Class: Class II Product Code: KXA Dated: December 21, 2020 Received: December 22, 2020
Dear Steve Peoples:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{1}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely,
Vesa Vuniqi Assistant Director DHT6A: Division of Joint Arthroplasty Devices OHT6: Office of Orthopedic Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201219
Device Name Biocore9 Femoral Resurfacing Head Component
### Indications for Use (Describe)
The Biocore9 Femoral Resurfacing Head Component, is indicated for the treatment of painful hip arthritis due to post-traumatic arthritis, or rheumatoid arthritis, as well as painful hip arthroplasty when sufficient femoral head and neck stock exist to stabilize a resurfacing femoral head. The Biocore9 Femoral Resurfacing Head Component is intended for cemented use only. This device is intended for hemi-arthroplasty only.
| Type of Use (Select one or both, as applicable) | |
|-------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|
| <label><input checked="" type="checkbox"/> Prescription Use (Part 21 CFR 801 Subpart D)</label> | <label><input type="checkbox"/> Over-The-Counter Use (21 CFR 801 Subpart C)</label> |
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for Biocore9 LLC. The logo is blue and features the company name in a stylized font. The 'O' in Biocore is replaced with a circular design that has a star in the middle.
### Biocore9, LLC. Femoral Resurfacing Head Component Traditional 510(k)
## 510(k) SUMMARY
In accordance with the Food and Drug Administration Rule to implement provisions of the Safe Medical Devices Act of 1990 and in conformance with 21 CFR 807.92, this information serves as a Summary of Safety and Effectiveness for the use of the Femoral Resurfacing Head Component.
| Submitted by: | Biocore9, LLC.<br>9 Whippany Road, Bldg A1, Unit 12.<br>Whippany NJ 0798<br>(973)-585-4281 |
|------------------------------------|----------------------------------------------------------------------------------------------|
| Contact Person: | Stephen J. Peoples |
| | Email: speoplesVMD@gmail.com |
| Proprietary Name: | Biocore9 Femoral Resurfacing Head<br>Component |
| Common Name: | Femoral Resurfacing Head |
| Classification Name and Reference: | 21 CFR 888.3400, Class II<br>Hip joint femoral (hemi-hip) metallic<br>resurfacing prosthesis |
| Device Product Code, Device Panel: | KXA |
| Predicate Device: | Endotec Inc. Modified New Jersey<br>Femoral Hip Resurfacing Component<br>510(k) #K904870 |
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for BIOCORE9. The logo is in blue and features the word "BIOCORE9" in a stylized font. The "O" in "BIOCORE" is replaced by a circular symbol with a gear-like design in the center. The overall design is modern and professional.
# Device Description
The Biocore9 Femoral Resurfacing Head component is a non-constrained orthopedic implant intended to replace the articular surface of the existing femoral head in the patient's hip joint. The Biocore9 Femoral Resurfacing Head Component is intended for in the reconstruction of painful and/or severely disabled hip joints resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis, avascular necrosis, or previously failed prosthesis. Sufficient bone stock on the femoral head and neck is necessary to stabilize a resurfacing femoral component. This device is intended for cemented use only. Components are available in thirteen sizes with available outside diameters 39.5, 41, 42.5, 47, 49, 51, and 52.5 mm.
The Biocore9 Femoral Resurfacing Head Component is manufactured from Ti-6Al-4V alloy (ASTM F136-13) with a Titanium Nitride (TiN) thin film ceramic coating. The titanium nitride coating is an inert, highly adherent, near diamond-hard surface coating.
The Femoral Resurfacing component is axisymmetric with a truncated spherical shell and a tapered central alignment stem. The component consists of a thin-walled spherical shell. The internal surface of the spherical shell is covered with a porous coating consisting of titanium beads. The stem is not porous coated.
## Indications for Use
The Biocore9 Femoral Resurfacing Head Component, used with cement, is indicated for the treatment of painful hip arthritis due to post-traumatic arthritis, osteoarthritis, or rheumatoid arthritis, as well as painful hip arthroplasty when sufficient femoral head and neck stock exist to stabilize a resurfacing femoral head. The Biocore9 Femoral Resurfacing Head Component is intended for cemented use only. This device is intended for hemi-arthroplasty only.
## Substantial Equivalence
Biocore9 Femoral Resurfacing Head Component is substantially equivalent to the Endotec - Modified New Jersey Femoral Hip Resurfacing Component - K904870, since there are no significant differences in materials, form, fit, or functionality. See Table 1 for reference:
## Performance Testing
Performance testing was conducted to provide support for establishing substantial equivalence. This testing included the following:
- TiN coating characterization; adhesion testing, thickness, hardness, scratch testing -
- -CP Ti porous coating characterization; Tensile strength, shear strength, porosity
- Mode 1 wear testing -
- -Abrasion resistance testing
- ISO 1143 Rotating beam fatigue testing -
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the Biocores logo in blue. Below the logo is the text "Table 1: Device Comparison Summary". The logo is a stylized representation of the company name.
Table 1: Device Comparison Summary.
| Device | Predicate:<br>Endotec - Modified New Jersey Femoral<br>Hip Resurfacing Component - K904870 | Biocore9 Femoral Resurfacing Head<br>Component |
|-------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Design | Component is Axisymmetric with a<br>truncated spherical shell with central<br>tapered stem and porous coated inner<br>surface. Available in several sizes. | Component is Axisymmetric with a<br>truncated spherical shell with central<br>tapered stem and porous coated inner<br>surface. Available in several sizes. |
| Materials | Titanium alloy (ASTM F-136) with<br>Ultracoat TiN Ceramic coating,<br>Biocoat® porous coating on the non-<br>articulating, inner surface. | Titanium alloy (ASTM F-136) with TiN<br>Ceramic coating, components are porous<br>coated on the non-articulating, inner<br>surface. |
| Indications | The Endotec Modified New Jersey<br>Femoral Hip Resurfacing Component is<br>intended to treat painful hip arthritis due<br>to post-traumatic arthritis, osteoarthritis,<br>or rheumatoid arthritis, as well as hip cup<br>arthroplasty when sufficient femoral head<br>and neck stock exist to stabilize a<br>resurfacing femoral head. | The Biocore9 Femoral Resurfacing Head<br>Component, used with cement, is indicated<br>for the treatment of painful hip arthritis due<br>to post-traumatic arthritis, osteoarthritis, or<br>rheumatoid arthritis, as well as painful hip<br>arthroplasty when sufficient femoral head<br>and neck stock exist to stabilize a<br>resurfacing femoral head. The Biocore9<br>Femoral Resurfacing Head Component is<br>intended for cemented use only. This device<br>is intended for hemi-arthroplasty only. |
| Size availability | 36.5, 38, 39.5, 41,42.5, 44, 45.5, 47, 49,<br>51, 52.5 (mm) | 39.5, 41,42.5, 44, 45.5, 47, 49, 51, 52.5<br>(mm) |
| Anatomical Site | Femoral Head | Femoral Head |
| Sterilization | Gamma Irradiation | Gamma Irradiation |
| 510(K) Number | K904870 | K201219 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.