K201203 · Anest Iwata Sparmax Co., Ltd. · BTA · Sep 3, 2020 · General, Plastic Surgery
Device Facts
Record ID
K201203
Device Name
Cliq Aspirator
Applicant
Anest Iwata Sparmax Co., Ltd.
Product Code
BTA · General, Plastic Surgery
Decision Date
Sep 3, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The DV-XXX aspirator provides a portable, battery-powered medical vacuum source. The DV-XXX can also be powered from a standard wall outlet with the use of an AC power adapter. It is intended for use in the home care or hospital environment.
Device Story
Portable AC/DC high vacuum/high flow suction pump; creates negative pressure to draw fluids through tubing into a collection container. Components: maintenance-free pump unit, AC adapter, on/off switch, pressure relief valve, pressure adjustment knob, analog pressure gauge, inline filter, rechargeable battery pack. Used in home or hospital settings; operated by clinicians or patients under physician order. Provides vacuum source for fluid aspiration; output monitored via analog gauge and LED battery status indicators. Benefits include portable fluid management.
Clinical Evidence
Bench testing only. Device met all predefined criteria for performance, safety, and electromagnetic compatibility per IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11, and ISO 10079-1 standards.
Technological Characteristics
Portable suction pump; 12 VDC positive displacement reciprocating pump; AC/DC powered (100-240 VAC or NiMH battery). ABS plastic housing; IP22 ingress protection; Class II double-insulated; Type BF applied part. Max vacuum 620 mmHg; flow rate up to 20 l/min. Analog pressure gauge; LED battery status indicators. Sterilization not specified.
Indications for Use
Indicated for use as a portable, battery-powered or AC-powered medical vacuum source in home care or hospital environments. Must be used on the order of a physician.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
September 3, 2020
ANEST IWATA SPARMAX CO., LTD. % Robert Dean President Compliance Systems International, LLC 1083 Delaware Ave. Buffalo, New York 14223
Re: K201203
Trade/Device Name: Cliq Aspirator Regulation Number: 21 CFR 878.4780 Regulation Name: Powered Suction Pump Regulatory Class: Class II Product Code: BTA Dated: August 5, 2020 Received: August 21, 2020
### Dear Robert Dean:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K201203
### Device Name
Cliq Aspirator DV-XXX (model family)
Indications for Use (Describe)
The DV-XXX aspirator provides a portable, battery-powered medical vacuum source. The DV-XXX can also be powered from a standard wall outlet with the use of an AC power adapter. It is intended for use in the home care or hospital environment.
Type of Use (Select one or both, as applicable)
| <span style="font-family: Arial, sans-serif;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> | □ Over-The-Counter Use (21 CFR 801 Subpart C) |
|-----------------------------------------------------------------------------------------------------|-----------------------------------------------|
|-----------------------------------------------------------------------------------------------------|-----------------------------------------------|
### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
K201203
# 510(k) Summary
# Cliq Aspirator DV-XXX (Model Family)
# February 20, 2020
### 1. Applicant Identification
Anest Iwata Sparmax Co., Ltd. No. 121, Sec 3, Zhongshan Rd. Dacun, Taiwan 51542
Telephone: +886-2-2345-1868 +886-2-2345-3162 Fax: Establishment Registration: 3006789357
# 2. Contact Person
Robert O. Dean Compliance Systems International, LLC 1083 Delaware Ave. Buffalo, NY 14223
716-440-7362 Telephone: Email: compliancesystems@yahoo.com
{4}------------------------------------------------
#### Device Name for Which Clearance is Sought 3.
| Trade Name: | Cliq Aspirator DV-XXX (model family) |
|-------------------------|--------------------------------------|
| Common/Usual Name: | Suction Pump / Aspirator |
| Regulation Description: | Powered Suction Pump |
### 4. Device Classification
| Product Code: | BTA |
|--------------------|------------------------------------------------|
| Device: | Pump, Portable, Aspiration (Manual or Powered) |
| Regulation Number: | 878.4780 |
| Class: | II |
| Review Panel: | General and Plastic Surgery |
#### 5. Intended Use
The DV-XXX aspirator provides a portable, battery-powered medical vacuum source. The DV-XXX can also be powered from a standard wall outlet with the use of an AC power adapter. It is intended for use in the homecare or hospital environment.
#### 6. Device Description
The DV-XXX is a portable AC/DC high vacuum / high flow suction pump. The DV-XXX can be powered from a standard wall outlet or from a rechargeable battery pack. The DV-XXX creates a negative pressure (vacuum) that draws fluids through tubing and into a collection container where the fluids are trapped for proper disposal. The device is comprised of a maintenance-free pump unit, an AC power
{5}------------------------------------------------
adapter with power cord, an on/off switch, a pressure relief valve and pressure adjustment knob, a pressure gauge, an inline filter, intermediate tubing, and a rechargeable battery pack. Optional accessories which may be included with the unit include a collection canister and suction tubing.
The DV-XXX produces a flow rate of up to 20 liters per minute, and has a maximum vacuum pressure of 620 mmHg. Housed in ABS plastic, the DV-XXX has an IP22 ingress protection rating; and, as a Class II (double-insulated) electrical appliance, the unit affords Type BF applied part protection against electric shock.
The DV-XXX must only be used on the order of a physician.
#### 7. Performance and Safety Testing
Non-clinical performance and safety tests conducted on the Cliq Aspirator DV-XXX (model family) include the following consensus standards:
- Basic safety and essential performance ● IEC 60601-1:
- IEC 60601-1-2: Electromagnetic compatibility ●
- IEC 60601-1-11: Requirements for medical electrical equipment ● used in the home healthcare environment
- ISO 10079: ● Electrically powered suction equipment safety requirements
Testing Conclusions:
The Cliq Aspirator DV-XXX (model family) met all predefined criteria, and passed all tests for performance, safety, and electromagnetic compatibility. Full test reports can be found in Section of Standard Conformance of this submission.
{6}------------------------------------------------
#### Predicate Device 8.
| Trade Name: | EasyGo Aspirator |
|----------------|-------------------|
| Manufacturer: | Precision Medical |
| 510(k) Number: | K971749 |
| Product Code: | BTA |
### Reference Device
| Trade Name: | EasyVac Aspirato |
|----------------|-------------------|
| Manufacturer: | Precision Medical |
| 510(k) Number: | K932494 |
| Product Code: | BTA |
#### 9. Substantial Equivalence
The Cliq Aspirator DV-XXX (model family) suction pump design is substantially equivalent to the legally marketed EasyGo Aspirator manufactured by Precision Medical (K971749).
- Intended Use: Both the DV-XXX and the predicate device are suction pumps that are intended for use as medical vacuum sources in the homecare or hospital environments.
- Technology: The pump units for both the DV-XXX and the predicate device are 12 VDC positive displacement reciprocating pumps. Both units use an AC adapter for connection to mains; and both units can be powered by rechargeable battery.
{7}------------------------------------------------
- Both the DV-XXX and the predicate device require connection Operation: to a hydrophobic filter, which, in turn, connects, via tubing, to a collection container. Both devices have a knob for adjustment of vacuum pressure; and both devices have an analog gauge that displays current pressure. Both devices use LED light(s) to indicate battery status.
- Performance: Both the DV-XXX and the predicate device meet substantially equivalent testing and acceptance criteria. Both devices conform to the requirements of the following consensus standards: IEC 60601-1, IEC 60601-1-2, and ISO 10079-1. Further, in accordance with ISO 10079-1.
### 10. Substantial Equivalence Comparison Chart
See next page
{8}------------------------------------------------
| | Cliq Aspirator<br>DV-XXX | Precision Medical<br>EasyGo Aspirator | Substantially<br>Equivalent? |
|-----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------|
| 510(k) Number | Unknown | K971749 | N/A |
| Manufacturer | Anest Iwata Sparmax<br>Co., Ltd. | Precision Medical | N/A |
| Trade Name | Aspirator | Aspirator | N/A |
| Model Number | DV-XXX (model family) | EasyGo | N/A |
| Device Classification | 878.4780<br>Powered Suction Pump<br>Class II<br>Product Code: BTA | 878.4780<br>Powered Suction Pump<br>Class II<br>Product Code: BTA | Equivalent |
| Intended Use | The DV-XXX aspirator<br>provides a portable,<br>battery-powered<br>medical vacuum<br>source. The DV-XXX<br>can also be powered<br>from a standard wall<br>outlet with the use of<br>an AC power adapter.<br>It is intended for use in<br>the homecare or<br>hospital environment. | The EasyGo Aspirator<br>provides a portable,<br>battery power medical<br>vacuum source. The<br>EasyGo Aspirator can<br>also be powered from a<br>standard wall outlet<br>with the use of an AC<br>power adapter. It is<br>intended for use in the<br>homecare or hospital<br>environment. | Equivalent |
{9}------------------------------------------------
K201203
| Consensus<br>Standards | IEC 60601-1:2005<br>IEC 60601-1-2:2015<br>EN ISO 10079-1:2015+<br>A: 2018<br>IEC 60601-1-11:2012 | IEC 60601-1:2006<br>IEC 60601-1-2:2007<br>EN ISO 10079-1:2009 | Equivalent |
|------------------------|--------------------------------------------------------------------------------------------------|---------------------------------------------------------------|------------|
|------------------------|--------------------------------------------------------------------------------------------------|---------------------------------------------------------------|------------|
| | Technological and System Specifications | Cliq Aspirator<br>DV-XXX | Precision Medical<br>EasyGo Aspirator | Substantially<br>Equivalent? |
|--------------------------------------|-----------------------------------------|-------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------|------------------------------|
| Electrical<br>requirements | | AC Adaptor:<br>100-240 VAC, 50-60Hz<br>Pump Unit: 12 VDC | AC Adaptor:<br>100-240 VAC, 50-60Hz<br>Pump Unit: 12 VDC | Equivalent |
| Protection against<br>electric shock | | Class II with Type BF<br>applied part | Class I with Type BF<br>applied part | Equivalent |
| Battery type | | NiMH rechargeable<br>battery pack | Lead acid battery | Equivalent |
| Battery status<br>indicator | | Colored LED lights used<br>to indicate:<br>- battery fully charged<br>- battery low | Colored LED lights used<br>to indicate:<br>- battery fully charged<br>- battery low | Equivalent |
| Vacuum Pressure | | Max: ~620 mmHg<br>Vacuum adjustable<br>Vacuum gauge | Max: 533 mmHg<br>Vacuum adjustable<br>Vacuum gauge | Equivalent |
| Pump Type | | Positive displacement<br>reciprocating pump | Positive displacement<br>reciprocating pump | Equivalent |
| Flow | | Up to ~20 l/min | Up to ~14 l/min | Equivalent |
{10}------------------------------------------------
| Sound level | < 53 dBa | Not known |
|-----------------------|-----------------------|----------------------------------|
| Weight | 3.5 kg / 7.7 lbs | 4.4 kg / 9.6 lbs |
| Dimensions | L30 x W16.5 x H19 cm | L30 x W24 x H25 cm |
| Operating Environment | Temperature: | Temperature: 10 - 40°C |
| | Humidity: | Humidity: not known |
| | Atmospheric Pressure: | Atmospheric Pressure: not known |
| Storage Environment | Temperature: | Temperature: -20 - 50°C |
| | Humidity: | Humidity: 15 - 93% |
| | Atmospheric Pressure: | Atmospheric Pressure: 50-106 kPa |
## 11. Conclusion
The Anest Iwata Sparmax Co., Ltd. Cliq Aspirator DV-XXX (model family) described in this 510(k) submission is substantially equivalent in design, technology, specifications, intended use, operation, and performance to the predicate device (K971749). Further the Cliq Aspirator DV-XXX (model family) does not raise any new safety or effectiveness issues when compared to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.