← Product Code [GEH](/productcode/GEH) · K201183

# Vortex Radial Spray Catheter (K201183)

_CSA Medical, Inc. · GEH · Dec 7, 2020 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K201183

## Device Facts

- **Applicant:** CSA Medical, Inc.
- **Product Code:** [GEH](/productcode/GEH.md)
- **Decision Date:** Dec 7, 2020
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 878.4350
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic, Real-World Evidence

## Real-World Evidence

| Submission | Device | Sponsor | RWD Sources | RWE Use Summary | Key Tags |
| --- | --- | --- | --- | --- | --- |
| K201183 · Dec 7, 2020 | Vortex Radial Spray Catheter | CSA Medical, Inc. | Prospective, multi-center patient registry | The registry data was used to document the safety profile and clinical experience of circumferential liquid nitrogen spray ablation in patients with Barrett's Esophagus, supporting the substantial equivalence of the new radial spray catheter. | Barrett's Esophagus; Patient Registry; Circumferential Ablation; Safety Profile |

### Clinical Evidence

| Study Design | Population | Comparator | Key Endpoints |
| --- | --- | --- | --- |
| Ongoing prospective, multi-center patient registry; Prospective, multi-center patient registry | Patients with Barrett's Esophagus; Sample Size: 112 total patients (49 received circumferential ablation) | Not applicable for this study | Safety profile, incidence of adverse events, strictures, and serious adverse events |

## Indications for Use

The truFreeze System is intended for cryogenic destruction of tissue using Liquid Nitrogen spray that has a boiling point of -196°C, requiring either active or passive venting during surgical procedures. The Vortex™ Radial Spray catheter is indicated for use as a cryosurgical tool to ablate benign lesions (e.g., Barrett's Esophagus with high grade and/or low grade dysplasia) in the upper gastrointestinal tract using active venting.

## Device Story

The Vortex Radial Spray Catheter is a cryosurgical accessory for the truFreeze System. It delivers liquid nitrogen (-196°C) in a 3-cm long, 360° radial spray pattern to ablate circumferential esophageal tissue. The device includes a mesh that expands to 20 mm to dilate the lumen and center the spray head. It is used in clinical settings by physicians under endoscopic visualization. The system uses active suction via a Cryo-Decompression Tube (CDT) to evacuate nitrogen gas. The console controls spray duration via a foot pedal and provides audible cues. The device ensures safety through automated system tests, continuous monitoring, and pressure controls (valves, transducers, burst discs). The output allows physicians to perform controlled freeze-thaw cycles to destroy diseased tissue, benefiting patients with Barrett's Esophagus by providing a circumferential ablation tool.

## Clinical Evidence

Bench testing verified design specifications, including shelf-life (ISO 11607), sterilization (ISO 11135), EMC/safety (IEC 60601-1), and usability. Biocompatibility testing (ISO 10993) confirmed safety for limited mucosal contact. Three preclinical GLP swine studies demonstrated controlled 3-cm circumferential ablation depth without serosal injury. Clinical evidence includes peer-reviewed literature (73 patients) and real-world data from a prospective registry (112 patients, 49 treated circumferentially) demonstrating safety and efficacy for Barrett's Esophagus ablation, with adverse event/stricture rates below 4%.

## Technological Characteristics

Cryosurgical catheter for liquid nitrogen delivery. Features a 360° radial spray head with an expandable 20 mm mesh for lumen dilation. Materials comply with ISO 10993. Sterilized via Ethylene Oxide (SAL 10-6). Connectivity includes RFID for catheter validation. System uses pressure-propelled cryogen, active suction (CDT), and active defrost. Software performs pre-use system tests and continuous monitoring to abort freezing upon failure detection.

## Regulatory Identification

(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.

## Predicate Devices

- truFreeze System ([K171626](/device/K171626.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo of the U.S. Food & Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA logo on the right. The FDA logo is in blue and includes the letters "FDA" in a square, followed by the words "U.S. FOOD & DRUG ADMINISTRATION" in a sans-serif font.

December 7, 2020

CSA Medical, Inc. Heather Nigro SVP, Regulatory/Quality/Clinical Affairs 91 Hartwell Ave Lexington, Massachusetts 02421

Re: K201183

Trade/Device Name: Vortex Radial Spray Catheter Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical Unit And Accessories Regulatory Class: Class II Product Code: GEH Dated: November 4, 2020 Received: November 5, 2020

Dear Heather Nigro:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal

{1}------------------------------------------------

statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for

devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely,

Long Chen, Ph.D. Assistant Director DHT4A: Division of General Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

## Indications for Use

510(k) Number (if known) K201183

Device Name truFreeze® System, Vortex™ Radial Spray Catheter

Indications for Use (Describe)

Intended Use:

The truFreeze® System is intended for cryogenic destruction of tissue using Liquid Nitrogen spray that has a boiling point of -196°C, requiring either active or passive venting during surgical procedures.

Indications for Use. VortexTM Radial Spray catheter:

The Vortex™ Radial Spray catheter is indicated for use as a cryosurgical tool to ablate benign (e.g., Barrett's Esophagus with high grade and/or low grade dysplasia) in the upper gastrointestinal tract using active venting.

| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|

X Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

#### CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

## CSA Medical, Inc. truFreeze® System Vortex™ Radial Spray Catheter Traditional 510(k) Premarket Notification

## K201183

## 510(k) Summary

This 510(k) Summary is being submitted in accordance with the requirements of the Safe Medical Device Act (SMDA) of 1990. The content contained in this 510(k) summary has been provided in conformance with 21 CFR §807.92

Date Prepared:

December 7, 2020

Submitter's Information:

| Name:                                                  | CSA Medical, Inc.                                                     |
|--------------------------------------------------------|-----------------------------------------------------------------------|
| Address:                                               | 91 Hartwell Avenue<br>Lexington, MA 02421                             |
| FDA Establishment<br>Owner/Operator Number:            | 3010140265                                                            |
| Contact Person:                                        | Heather V. Nigro, MS, RAC<br>SVP, Regulatory/Quality/Clinical Affairs |
| Phone:                                                 | 781-538-4793 (office) or 781-640-8426 (mobile)                        |
| Fax:                                                   | 866-300-5183                                                          |
| e-mail:                                                | hnigro@csamedical.com                                                 |
| Legal Manufacturer:                                    | CSA Medical, Inc.<br>91 Hartwell Avenue<br>Lexington, MA 02421        |
| Facility Establishment<br>Identification (FEI) Number: | 3004534508                                                            |
| Trade/Proprietary Name:                                | truFreeze® System, Vortex™ Radial Spray Catheter                      |
| Common/Usual Name:                                     | Cryosurgical Catheter                                                 |
| Classification Name:                                   | Cryosurgical Unit, Cryogenic Surgical Device                          |
| Regulation Number:                                     | 21 CFR 878.4350                                                       |
| Product Code:                                          | GEH                                                                   |
| Device Classification:                                 | Class II                                                              |
| Predicate Device                                       | truFreeze System, K171626                                             |

{4}------------------------------------------------

## CSA Medical, Inc. truFreeze® System Vortex™ Radial Sprav Catheter Traditional 510(k) Premarket Notification

#### Device Description:

The CSA Medical (CSA) truFreeze® System consists of a truFreeze Console and a truFreeze Spray Catheter Kit. The truFreeze Console is to be used only with a truFreeze Spray Kit. This document specifically applies to the Vortex Radial Spray Kit (20-00360) for use with the truFreeze® System. Vortex Radial delivers liquid nitrogen in a 3-cm long, 360° spray pattern to ablate circumferential, epithelium such as Barrett's Esophagus.

The Vortex Radial Spray Kit (20-00360) consists of one (1) truFreeze Vortex Radial Spray Catheter (20-00187), one (1) 4-inch Suction Canister Connector, and one (1) Patient Tube.

The catheter is connected to the console and transports cryogen from the console to the targeted ablation area. The Suction Cannector and Patient Tube provide the connection from the suction source in the console to the Cryo-Decompression Tube (CDT) at the end of the catheter, to provide active evacuation (suction) of nitrogen gas at the ablation site. The spray head is surrounded by the Mesh that expands to dilate the lumen to a diameter of 20 mm and center the device during the delivery of cryogen spray.

#### Indications for use:

The Indications for Use statement for the truFreeze® System is:

The truFreeze System is intended for cryogenic destruction of tissue using Liquid Nitrogen spray that has a boiling point of -196°C requiring either active or passive venting during surgical procedures. The Vortex™ Radial Spray catheter is indicated for use as a cryosurgical tool to ablate benign lesions (e.g., Barrett's Esophagus with high grade and/or low grade dysplasia) in the upper gastrointestinal tract using active venting.

#### Technological Characteristics:

The proposed device and predicate device are based on the same technological characteristics including principle of operation via pressure-propelled cryogen, application of liquid nitrogen to ablate unwanted tissue, and delivery of equivalent cooling power density per unit area. The differentiator for the Vortex™ catheter is that the cryogen is sprayed radially as compared to the previously cleared catheters which spray linearly.

| Characteristic                                                                          | Predicate truFreeze® System, (171626)                                                                                                                                                                                                                                 | Proposed truFreeze® System, Vortex™ Radial Spray Kit                                                                                                                                                                                                                                                                                                                                                                                                             |
|-----------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use                                                                            | Intended for cryogenic destruction of tissue using liquid nitrogen spray that has a boiling point of -196°C requiring either active or passive venting during surgical procedures.                                                                                    | Intended for cryogenic destruction of tissue using liquid nitrogen spray that has a boiling point of - 196°C requiring either active or passive venting during surgical procedures.                                                                                                                                                                                                                                                                              |
| Indication for Use                                                                      | The truFreeze® System is<br>indicated for use as a<br>cryosurgical tool in the fields of<br>dermatology, gynecology, and<br>general surgery, to ablate<br>benign (eg. Barrett's<br>Esophagus with high grade<br>and/or low grade dysplasia)<br>and malignant lesions. | The truFreeze System is intended<br>for cryogenic destruction of tissue<br>using Liquid Nitrogen spray that<br>has a boiling point of -196°C<br>requiring either active or passive<br>venting during surgical<br>procedures. The Vortex™ Radial<br>Spray catheter is indicated for use<br>as a cryosurgical tool to ablate<br>benign lesions (e.g., Barrett's<br>Esophagus with high grade<br>and/or low grade dysplasia) in the<br>upper gastrointestinal tract |
| Cryogen                                                                                 | Liquid nitrogen                                                                                                                                                                                                                                                       | Liquid nitrogen                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Principle of Operation                                                                  | Pressure Propelled Cryogen                                                                                                                                                                                                                                            | Pressure Propelled Cryogen                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Mode of Ablation                                                                        | ≥1 freeze-thaw cycles,<br>Quantity determined by<br>physician                                                                                                                                                                                                         | ≥1 freeze-thaw cycles,<br>Quantity determined by physician                                                                                                                                                                                                                                                                                                                                                                                                       |
| Delivery/Cryoprobe                                                                      | Spray Tip (Linear)                                                                                                                                                                                                                                                    | Spray Tip (Circumferential)                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Output Temperature                                                                      | -196°C                                                                                                                                                                                                                                                                | -196°C                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Cooling Power Density,<br>Low Flow (12.5 W, N/A<br>W)                                   | 2.7 to 4.0 W/cm²                                                                                                                                                                                                                                                      | N/A                                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Cooling Power Density,<br>Normal Flow (25 W,<br>100 W)                                  | 5.3 to 8.0 W/cm²                                                                                                                                                                                                                                                      | 5.3 W/cm²                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
| 1Depth of Freeze in<br>Hydrogel when used<br>per IFU at nominal<br>settings (mean ± SD) | 1.0 mm ± 0.4 mm                                                                                                                                                                                                                                                       | 1.0 mm ± 0.2 mm                                                                                                                                                                                                                                                                                                                                                                                                                                                  |
| Delivery of Cryogen<br>(spray dosimetry)                                                | Physician-controlled duration<br>(GUI activated timer with<br>audible cues)                                                                                                                                                                                           | Physician-controlled duration<br>(pedal activated timer with audible<br>cues)                                                                                                                                                                                                                                                                                                                                                                                    |
|                                                                                         | 0 s to 60 s, 5-s increment                                                                                                                                                                                                                                            | 10 s to 20 s, 1-s increment                                                                                                                                                                                                                                                                                                                                                                                                                                      |
| Procedure<br>Visualization                                                              | Direct visualization via<br>endoscope                                                                                                                                                                                                                                 | Direct visualization via endoscope                                                                                                                                                                                                                                                                                                                                                                                                                               |
| Early spray termination                                                                 | User control via foot pedal<br>release or emergency stop<br>button                                                                                                                                                                                                    | User control via foot pedal<br>release or emergency stop button                                                                                                                                                                                                                                                                                                                                                                                                  |
| Prevent use/reuse of<br>expired or invalid<br>catheter                                  | Confirms use of valid catheter<br>using RFID                                                                                                                                                                                                                          | Confirms use of valid catheter<br>using RFID                                                                                                                                                                                                                                                                                                                                                                                                                     |
| Notifies physician to<br>stop spraying                                                  | Audible beeper to coincide with<br>visual display of timer                                                                                                                                                                                                            | Audible beeper to coincide with<br>visual display of timer. Console<br>terminates spray at physician-<br>selected duration                                                                                                                                                                                                                                                                                                                                       |
| Computerized test of<br>system prior to use                                             | Software test confirms system<br>is properly operating before<br>exposure of cryotherapy                                                                                                                                                                              | Software test confirms system is<br>properly operating before<br>exposure of cryotherapy                                                                                                                                                                                                                                                                                                                                                                         |
| Computerized<br>continuous monitoring<br>of system during<br>procedures                 | Uses computer program to<br>abort freezing if a system failure<br>is detected                                                                                                                                                                                         | Uses computer program to abort<br>freezing if a system failure is<br>detected                                                                                                                                                                                                                                                                                                                                                                                    |
| Ensure patient is not<br>exposed to high<br>pressure gases                              | Uses active suction pump and<br>CDT (active), or a natural orifice<br>(passive), as per instructions for<br>use. IFU provides venting<br>guidance for proper gas egress.                                                                                              | Uses active suction pump and<br>integral CDT (active) and passive<br>egress channels as per<br>instructions for use. Console<br>pump activation during spray<br>delivery. IFU provides venting<br>guidance for proper gas egress.                                                                                                                                                                                                                                |
| Protect healthy tissue<br>from excessive<br>temperatures                                | Provides sufficient insulation to<br>maintain safe temperature on<br>patient and user exposure<br>surfaces                                                                                                                                                            | Provides sufficient insulation to<br>maintain safe temperature on<br>patient and user exposure<br>surfaces                                                                                                                                                                                                                                                                                                                                                       |
| Pressure Controls                                                                       | Valves and pressure transducer<br>to control pressure of liquid<br>nitrogen; redundant pressure<br>switch; mechanical relief valve;<br>redundant burst disc.                                                                                                          | Valves and pressure transducer<br>to control pressure of liquid<br>nitrogen; redundant pressure<br>switch; mechanical relief valve;<br>redundant burst disc.                                                                                                                                                                                                                                                                                                     |
| Thermal/Defrost                                                                         | Active defrost capability to thaw<br>catheter using warm nitrogen<br>gas                                                                                                                                                                                              | Active defrost capability to thaw<br>catheter using warm nitrogen gas                                                                                                                                                                                                                                                                                                                                                                                            |
| Safe Storage of<br>Cryogen                                                              | Internal Pressure-rated vacuum<br>insulated Dewar                                                                                                                                                                                                                     | Internal Pressure-rated vacuum<br>insulated Dewar                                                                                                                                                                                                                                                                                                                                                                                                                |
| Product Label                                                                           | Uniquely identifies catheter as<br>linear spray for active or<br>passive venting procedures.                                                                                                                                                                          | Uniquely identifies catheter as<br>radial spray for active venting<br>procedures.                                                                                                                                                                                                                                                                                                                                                                                |
| Sterility                                                                               | Sterile catheter provided using<br>Ethylene Oxide with SAL 10-6                                                                                                                                                                                                       | Sterile catheter provided using<br>Ethylene Oxide with SAL 10-6                                                                                                                                                                                                                                                                                                                                                                                                  |
| Biocompatibility                                                                        | Patient (direct and indirect)<br>contacting materials comply<br>with ISO 10993                                                                                                                                                                                        | Patient (direct and indirect)<br>contacting materials comply with<br>ISO 10993                                                                                                                                                                                                                                                                                                                                                                                   |

{5}------------------------------------------------

# CSA Medical, Inc. truFreeze® System
Vortex™ Radial Spray Catheter Traditional 510(k) Premarket Notification

{6}------------------------------------------------

## CSA Medical, Inc. truFreeze® System Vortex™ Radial Spray Catheter Traditional 510(k) Premarket Notification

## Summary of Performance Testing - Bench, Non-clinical, Clinical:

A number of bench tests were conducted to verify the product design met the predetermined product performance specifications. Testing was inclusive of the following: Product Design and Software verification, Shelf-life in accordance with ISO 11607, Sterilization in accordance with ISO 11135:2014, EMC/Safety per IEC 60601-12:2014, Distribution Testing per ISTA 2A:2011 and Usability. The Vortex™ Radial Spray catheter is classified as surfacting, mucosal membrane device having limited contact duration (<24 hours) therefore requiring cytotoxicity,

{7}------------------------------------------------

## CSA Medical, Inc. truFreeze® System Vortex™ Radial Sprav Catheter Traditional 510(k) Premarket Notification

sensitization and irritation testing. This testing was conducted in accordance with ISO 10993-1:2018, including 10993:5 and 10993:10. All testing confirmed that the product specifications had been met, demonstrated that the device is adequately designed for the labeled indications for use and the risk management report concluded that the benefits of the device outweigh the associated risks.

Three preclinical GLP studies were completed utilizing the Vortex™ Radial Spray catheter in swine for the purposes of dose ranging and gathering evidence of procedural safety consistent with other cleared truFreeze® accessories. Histopathology data confirmed a 3-cm-long circumferential treatment zone and showed controlled depth of injury without reaching the serosa. All animals survived and no serious adverse events were reported.

Three scientific, peer-reviewed literature articles have been published demonstrating the ability of the predicate device (i.e. linear spray) to achieve circumferential freezing of diseased tissue in the esophagus. The scientific, peer-reviewed literature provided evidence that 73 patients have been treated with spray LN2 circumferentially since 2012 to ablate BE lesions with a safety and efficacy profile. Additionally, Real-World Data (RWD) has been extracted from an ongoing prospective, multi-center patient registry which further documents evidence of circumferential spray of LN2 with a safety profile. Of the 112 patients identified in the registry as having Barrett's Esophagus, 49 of those patients received circumferential ablation. The mean BE segment length from the first procedure was 4.8 cm (range 1.0 - 13.0 cm) and the mean number of circumferentially ablated segments per procedure was 2.64 (range 2.0 - 7.0). The dosimetry used was consistent with the IFU for the linear spray catheter as most patients were treated with two sprays for 20 seconds. Procedures associated with an adverse event, stricture or serious adverse event are all below 4%.

#### Comparison to Predicate:

The truFreeze® Vortex™ catheter has the same intended use and utilizes the same fundamental technology as the predicate truFreeze® System and catheters (K171626). The truFreeze® Vortex™ catheter has been compared to the legally marketed predicate device as cleared through K171626 (August 30, 2017) and there are no new safety or effectiveness questions raised with this proposed modification. Therefore, it is believed that the proposed Vortex™ Radial Spray catheter is substantially equivalent to the truFreeze® System described in K171626.

---

**Source:** [https://fda-staging.innolitics.com/device/K201183](https://fda-staging.innolitics.com/device/K201183)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).