WallFlex Colonic Stent System with Anchor Lock Delivery System
K201159 · Boston Scientific Corporation · MQR · Aug 17, 2020 · General, Plastic Surgery
Device Facts
Record ID
K201159
Device Name
WallFlex Colonic Stent System with Anchor Lock Delivery System
Applicant
Boston Scientific Corporation
Product Code
MQR · General, Plastic Surgery
Decision Date
Aug 17, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.3610
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The WallFlex Colonic Stent System with Anchor Lock Delivery System is indicated for palliative treatment of colonic strictures caused by malignant neoplasms and to relieve large bowel obstructions prior to colectomy in patients with malignant strictures. The WallFlex Duodenal Stent System with Anchor Lock Delivery System is indicated for the palliative treatment of gastroduodenal obstructions produced by malignant neoplasms.
Device Story
Nitinol-based expandable metallic stents; delivered via coaxial Anchor Lock delivery system; used for palliative treatment of malignant colonic or gastroduodenal strictures/obstructions. Stents available in multiple diameters (22mm/25mm) and lengths (6cm, 9cm, 12cm). Delivery system uses retractable exterior tube to release stent. Operated by clinicians in clinical settings. Stent provides mechanical radial force to maintain patency of obstructed bowel segments; improves patient quality of life by relieving obstruction symptoms. Modification involves surface finish of nitinol wire; no change to fundamental technological characteristics.
Clinical Evidence
Bench testing only. Performance data includes deployment/reconstrainment force, unconstrained dimensions, hoop force, fatigue resistance, flexibility, corrosion resistance, and potentiodynamic testing. All tests met pre-defined specifications.
Indicated for patients with malignant neoplasms causing colonic strictures or large bowel obstructions (for palliative treatment or pre-colectomy relief) and patients with malignant neoplasms causing gastroduodenal obstructions.
Regulatory Classification
Identification
An esophageal prosthesis is a rigid, flexible, or expandable tubular device made of a plastic, metal, or polymeric material that is intended to be implanted to restore the structure and/or function of the esophagus. The metal esophageal prosthesis may be uncovered or covered with a polymeric material. This device may also include a device delivery system.
Special Controls
The special control for this device is FDA's “Guidance for the Content of Premarket Notification Submissions for Esophageal and Tracheal Prostheses.”
Predicate Devices
Wallflex Enteral Colonic Stent with Anchor Lock Delivery System (K061877)
Wallflex Enteral Duodenal Stent with Anchor Lock Delivery System (K062750)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). The logo consists of two parts: the Department of Health & Human Services logo on the left and the FDA text logo on the right. The FDA text logo is in blue and includes the letters "FDA" followed by "U.S. FOOD & DRUG ADMINISTRATION".
September 28, 2020
Boston Scientific Corporation Catherine Sanford Senior Regulatory Affairs Specialist 100 Boston Scientific Way Marlborough, MA 01752
Re: K201159
Trade/Device Name: WallFlex Colonic Stent System with Anchor Lock Delivery System WallFlex Duodenal Stent System with Anchor Lock Delivery System Regulation Number: 21 CFR 878.3610 Regulation Name: Esophageal prosthesis Regulatory Class: II Product Code: MQR, MUM
Dear Catherine Sanford:
The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated July 17, 2020. Specifically, FDA is updating this SE Letter, Indications for Use, and 510(k) trade names, which did match, as an administrative correction.
Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Glenn Bell, (301) 796-6531, Glenn.Bell(@fda.hhs.gov.
Sincerely,
## Joyce Whang
for Glenn B. Bell, Ph.D. Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health
#### Enclosure
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
August 17, 2020
Boston Scientific Corporation Catherine Sanford Senior Regulatory Affairs Specialist 100 Boston Scientific Way Marlborough, Massachusetts 01752
Re: K201159
Trade/Device Name: WallFlex Colonic Stent System with Anchor Lock Delivery System Regulation Number: 21 CFR 878.3610 Regulation Name: Esophageal Prosthesis Regulatory Class: Class II Product Code: MQR, MUM Dated: July 16, 2020 Received: July 17, 2020
#### Dear Catherine Sanford:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part
{2}------------------------------------------------
801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4. Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE(@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerelv.
# Daniel G. Walter Jr -S
Daniel G. Walter, Jr. Assistant Director DHT3A: Division of Renal, Gastrointestinal, Obesity and Transplant Devices OHT3: Office of Gastrorenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health
Enclosure
{3}------------------------------------------------
### Indications for Use
510(k) Number (if known) K201159
#### Device Name
WallFlex Colonic Stent System with Anchor Lock Delivery System WallFlex Duodenal Stent System with Anchor Lock Delivery System
#### Indications for Use (Describe)
The WallFlex Colonic Stent System with Anchor Lock Delivery System is indicated for palliative treatment of colonic strictures caused by malignant neoplasms and to relieve large bowel obstructions prior to colectomy in patients with malignant strictures.
The WallFlex Duodenal Stent System with Anchor Lock Delivery System is indicated for the palliative treatment of gastroduodenal obstructions produced by malignant neoplasms.
| Type of Use (Select one or both, as applicable) |
|---------------------------------------------------------------------------------------------------------------|
| <span style="text-decoration: underline;"><b> ☑ </b></span> Prescription Use (Part 21 CFR 801 Subpart D) |
| ☐ Over-The-Counter Use (21 CFR 801 Subpart C) |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
#### 510(k) SUMMARY
#### 1. Submitter:
Boston Scientific Corporation 100 Boston Scientific Way Marlborough, MA 01752
Contact: Catherine Sanford Senior Regulatory Affairs Specialist Telephone: 508-683-4498 E-mail: Catherine.sanford@bsci.com
Date Prepared: April 29th, 2020
#### 2. Proposed Device:
Trade Name: Wallflex Colonic Stent System with Anchor Lock Delivery System Classification Name: Expandable, metallic colonic stent Regulation Number: 21 CFR 878.3610 Product Code: MQR Classification and Panel: Class II, Gastroenterology/Urology
Trade Name: Wallflex Duodenal Stent System with Anchor Lock Delivery System Classification Name: Expandable, metallic duodenal stent Regulation Number: 21 CFR 878.3610 Product Code: MUM Classification and Panel: Class II, Gastroenterology/Urology
#### 3. Predicate Device:
Trade Name: Wallflex Enteral Colonic Stent with Anchor Lock Delivery System Submitter / 510(k) Holder: Boston Scientific Clearance Number: K061877 Classification Name: Expandable, metallic colonic stent Regulation Number: 21 CFR 878.3610 Product Code: MQR Classification and Panel: Class II, Gastroenterology/Urology
Trade Name: Wallflex Enteral Duodenal Stent with Anchor Lock Delivery System Submitter / 510(k) Holder: Boston Scientific Clearance Number: K062750
{5}------------------------------------------------
Classification Name: Expandable, metallic duodenal stent
Regulation Number: 21 CFR 878.3610 Product Code: MUM Classification and Panel: Class II, Gastroenterology/Urology
#### 4. Device Description:
The WallFlex Colonic Stent System with Anchor Lock Delivery System and WallFlex Duodenal Stent System with Anchor Lock Delivery System each consist of two components: the implantable metal stent and the anchor lock delivery system.
The proposed stent is manufactured of Nitinol. The stent will be offered in two diameters, 22mm with a 27mm flare and 25mm with a 30mm flare. Each diameter will be available in three lengths, 6cm, 9cm, and 12cm. The WallFlex Duodenal Stent will only be offered in the 22mm/27mm stent diameter (all lengths) preloaded on the 230cm Anchor Lock Delivery System. The WallFlex Colonic Stent will be offered in the 22mm/27mm and 25mm/30mm diameters (all lengths) preloaded on the 230cm or 135cm Anchor Lock Delivery System. The Anchor Lock delivery system consists of a coaxial tubing assembly that constrains the stent on the delivery catheter shaft until the stent is released by retracting the exterior tube.
#### 5. Indications for Use:
The WallFlex Colonic Stent System with Anchor Lock Delivery System is indicated for palliative treatment of colonic strictures caused by malignant neoplasms and to relieve large bowel obstructions prior to colectomy in patients with malignant strictures.
The WallFlex Duodenal Stent System with Anchor Lock Delivery System is indicated for the palliative treatment of gastroduodenal obstructions produced by malignant neoplasms.
The proposed Indications for Use are identical to the predicate devices Indications for Use.
#### Technological Characteristics 6.
The only change introduced in this submission is to the surface finish of the nitinol wire. There is no impact to the technological characteristics as a result of this change as demonstrated by the biocompatibility testing (Section 15) and performance testing (Section 18).
#### 7. Performance Data
#### Bench Testing:
The proposed WallFlex Colonic Stent System with Anchor Lock Delivery System and WallFlex Duodenal Stent System with Anchor Lock Delivery System successfully passed all pre-defined product specifications for the tests performed. Below is a summary of the tests performed to show the proposed device satisfied all design verification requirements. In addition,
{6}------------------------------------------------
| Section | Test | Results (Pass/ Fail) |
|---------|----------------------------------------------|----------------------|
| 18.1 | Deployment Force | Pass |
| 18.2 | Reconstrainment Force | Pass |
| 18.3 | Unconstrained Stent Length | Pass |
| 18.4 | Unconstrained Stent Diameter | Pass |
| 18.5 | Flare Diameter | Pass |
| 18.6 | Stent Hoop Force (Compression and Expansion) | Pass |
| 18.7 | Stent Fatigue Resistance | Pass |
| 18.8 | Stent Flexibility | Pass |
| 18.9 | Stent Corrosion Resistance | Pass |
potentiodynamic testing and foreshortening was performed to address recent FDA guidance and to further support substantial equivalence. The results of this testing are passing.
#### 8. Conclusion
The information provided by Boston Scientific Corporation in this submission demonstrates that the proposed WallFlex Colonic Stent System with Anchor Lock Delivery System and WallFlex Duodenal Stent System with Anchor Lock Delivery System are substantially equivalent to the currently cleared Wallflex Enteral Colonic Stent with Anchor Lock Delivery System and Wallflex Enteral Duodenal Stent with Anchor Lock Delivery System (K061877, K062750).
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.