← Product Code [ONB](/productcode/ONB) · K201142

# TEVADAPTOR Bag Adaptor LL, TEVADAPTOR Bag Adaptor LL with ULTRASITE, TEVADAPTOR Bag Adaptor SP with ULTRASITE, TEVADAPTOR IV Secondary Safety Set with ULTRASITE, TEVADAPTOR IV Secondary Safety set (K201142)

_Simplivia Healthcare , Ltd. · ONB · Feb 25, 2021 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K201142

## Device Facts

- **Applicant:** Simplivia Healthcare , Ltd.
- **Product Code:** [ONB](/productcode/ONB.md)
- **Decision Date:** Feb 25, 2021
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5440
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

Tevadaptor® is a Closed System Transfer Device (CSTD) that mechanically prohibits the release of drugs, including antineoplastic and hazardous drugs, in vapor, aerosol or liquid form during preparation, reconstitution, compounding and administration, minimizing exposure of individuals, healthcare personnel, and the environment to hazardous drugs.

## Device Story

Tevadaptor IV Administration Sets are single-use, sterile, non-pyrogenic devices for delivering IV solutions/drugs to a patient's vascular system. As part of the Tevadaptor/Chemfort CSTD system, they feature a spike, tubing, clamp, ULTRASITE needless injection site, and Luer connections. The device functions by connecting to the Tevadaptor/Chemfort Syringe Adaptor (SA) or Luer Lock Adaptor (LLA) to create a closed fluid path. This mechanical interface prevents the escape of hazardous drug vapors, aerosols, or liquids during drug transfer and administration. Used in clinical settings by healthcare professionals, the device protects the user, patient, and environment from hazardous drug exposure. The healthcare provider manages the connection to the IV bag or primary set, ensuring a closed system throughout the infusion process.

## Clinical Evidence

Bench testing only. No clinical data. Testing included biocompatibility (ISO 10993 series), sterilization validation (ISO 11135), packaging integrity (ISO 11607-1), infusion set performance (ISO 8536-4), and connector compatibility (ISO 80369-7).

## Technological Characteristics

Single-use, sterile, non-pyrogenic IV administration set. Components: spike, drip chamber, tubing, flow regulator, clamp, ULTRASITE needless injection site, Luer connection. Materials are biocompatible. Sterilization: Ethylene Oxide (SAL 10^-6). Connectivity: Mechanical snap and Luer lock connections. Complies with ISO 80369-7 for small-bore connectors.

## Regulatory Identification

An intravascular administration set is a device used to administer fluids from a container to a patient's vascular system through a needle or catheter inserted into a vein. The device may include the needle or catheter, tubing, a flow regulator, a drip chamber, an infusion line filter, an I.V. set stopcock, fluid delivery tubing, connectors between parts of the set, a side tube with a cap to serve as an injection site, and a hollow spike to penetrate and connect the tubing to an I.V. bag or other infusion fluid container.

## Special Controls

*Classification.* Class II (special controls). The special control for pharmacy compounding systems within this classification is the FDA guidance document entitled “Class II Special Controls Guidance Document: Pharmacy Compounding Systems; Final Guidance for Industry and FDA Reviewers.” Pharmacy compounding systems classified within the intravascular administration set are exempt from the premarket notification procedures in subpart E of this part and subject to the limitations in § 880.9.

## Predicate Devices

- I.V. Administration Set ([K121269](/device/K121269.md))

## Reference Devices

- Chemfort™ Closed System Transfer Device (CSTD) ([K192866](/device/K192866.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logos of the U.S. Department of Health and Human Services and the U.S. Food and Drug Administration (FDA). The Department of Health and Human Services logo is on the left, and the FDA logo is on the right. The FDA logo includes the text "FDA U.S. FOOD & DRUG ADMINISTRATION" in blue. The logos are displayed side-by-side.

February 25, 2021

Simplivia Healthcare Ltd. Shay Shaham VP Qa/ra North Industrial Zone Kiryat Shmona, 1011801 Israel

## Re: K201142

Trade/Device Name: TEVADAPTOR Bag Adaptor LL, TEVADAPTOR Bag Adaptor LL with ULTRASITE, TEVADAPTOR Bag Adaptor SP with ULTRASITE, TEVADAPTOR IV Secondary Safety Set with ULTRASITE, TEVADAPTOR IV Secondary Safety set Regulation Number: 21 CFR 880.5440 Regulation Name: Intravascular Administration Set Regulatory Class: Class II Product Code: ONB, FPA Dated: January 19, 2021 Received: January 25, 2021

Dear Shay Shaham:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{1}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part

801); medical device reporting of medical device-related adverse events) (21 CFR 803) for devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.

For comprehensive regulatory information about medical devices and radiation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).

Sincerely.

for Payal Patel Acting Assistant Director DHT3C: Division of Drug Delivery and General Hospital Devices, and Human Factors OHT3: Office of GastroRenal, ObGyn, General Hospital and Urology Devices Office of Product Evaluation and Ouality Center for Devices and Radiological Health

Enclosure

{2}------------------------------------------------

# Indications for Use

510(k) Number (if known) K201142

#### Device Name

I.V. Administration Set: TEVADAPTOR Bag Adaptor LL; TEVADAPTOR Bag Adaptor LL with ULTRASITE; TEVADAPTOR Bag Adaptor SP with ULTRASITE; TEVADAPTOR IV Secondary Safety Set with ULTRASITE; TEVADAPTOR IV Secondary Safety

### Indications for Use (Describe)

Tevadaptor® is a Closed System Transfer Device (CSTD) that mechanically prohibits the release of drugs, including antineoplastic and hazardous drugs, in vapor, aerosol or liquid form during preparation, compounding and administration, minimizing exposure of individuals, healthcare personnel, and the environment to hazardous drugs.

Type of Use (Select one or both, as applicable)

| <span style="font-family: DejaVu Sans, sans-serif">✘</span> Prescription Use (Part 21 CFR 801 Subpart D) | <span style="font-family: DejaVu Sans, sans-serif">□</span> Over-The-Counter Use (21 CFR 801 Subpart C) |
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|
|----------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------|

## CONTINUE ON A SEPARATE PAGE IF NEEDED.

This section applies only to requirements of the Paperwork Reduction Act of 1995.

## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image contains the logo for Simplivia. The logo consists of a green circular symbol on the left and the word "SIMPLIVIA" in gray, sans-serif font on the right. The circular symbol appears to be an abstract design, possibly representing a stylized letter S.

# K201142 510(k) Summary

| Name:             | Simplivia Healthcare Ltd.<br>North Industrial Zone<br>Kiryat Shmona<br>Israel 1101801<br>Tel. (972)4-690-8826<br>Fax (972)4-8921665                                                                                                           |
|-------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:   | Shay Shaham<br>VP QA & RA<br>Simplivia Healthcare Ltd.<br>Tel. (972)52-831-3378<br>Fax (972)4-8921665                                                                                                                                         |
| Date Prepared:    | February 21, 2021                                                                                                                                                                                                                             |
| Name of Device:   | Intravascular Administration Set:<br>TEVADAPTOR Bag Adaptor LL<br>TEVADAPTOR Bag Adaptor LL with ULTRASITE<br>TEVADAPTOR Bag Adaptor SP with ULTRASITE<br>TEVADAPTOR IV Secondary Safety Set with ULTRASITE<br>TEVADAPTOR IV Secondary Safety |
| Common Name:      | Closed Drug Reconstitution and Transfer System                                                                                                                                                                                                |
| Classification:   | Product Code: ONB (primary), FPA (secondary)<br>Regulation No: 21 C.F.R. §880. 5440<br>Class: II<br>Regulation name: Intravascular administration set<br>Classification Panel: General Hospital                                               |
| Predicate Device: | I.V. Administration Set (K121269)                                                                                                                                                                                                             |
| Reference Device: | Chemfort™ Closed System Transfer Device (CSTD) (K192866)                                                                                                                                                                                      |

## Intended Use / Indications for Use

Tevadaptor® is a Closed System Transfer Device (CSTD) that mechanically prohibits the release

{4}------------------------------------------------

of drugs, including antineoplastic and hazardous drugs, in vapor, aerosol or liquid form during preparation, reconstitution, compounding and administration, minimizing exposure of individuals, healthcare personnel, and the environment to hazardous drugs.

## Device Description

The I.V. Administration Set is single use, sterile, non-pyrogenic device used to administer intravenous solutions and/or drugs solutions from a container to a patient's vascular system.

The I.V. Administration set is comprised of various components such as: Spike, tubing, clamp, ULTRASITE® needless injection site (Y-site), 'twist-off' and Luer connection.

The I.V. Administration set is part of the Tevadaptor® closed system transfer device (CSTD) and is intended to interact with Tevadaptor® and/or Chemfort™ system's Syringe Adaptor (SA), Syringe Adaptor Lock (SAL) and Luer Lock Adaptor (LLA) components that were cleared under K192866. The Tevadaptor® and Chemfort™ CSTD are intended to protect the healthcare professional, the patient and the environment during preparation, reconstitution, compounding and administration of hazardous drugs in the form of liquid, vapor or aerosol.

The Tevadaptor® and/or Chemfort™ SA connects to the Tevadaptor® port located on the I.V. Administration set's Spike component. After the connection is made, a fluid path is opened which allows to inject of the drug to the patient's IV bag. The Tevadaptor® and/or Chemfort™ LLA connects to any female luer connection located on the I.V. Administration set to create a Tevadaptor® port connection that enables the Tevadaptor® and/or Chemfort™ SA to administer the drug directly to the patient.

The purpose of this submission is to add five new designs of I.V. Administration Sets to the Tevadaptor® and/or Chemfort™ Closed System Transfer Device (CSTD) that was cleared for sale in the US under K192866, the descriptions of the added sets, subject of this submission:

- Tevadaptor® IV Secondary Safety Set with ULTRASITE®
- Tevadaptor® IV Secondary Safety Set .
- Tevadaptor® Bag Adaptor LL ●
- Tevadaptor® Bag Adaptor LL with ULTRASITE® .
- . Tevadaptor® Bag Adaptor SP with ULTRASITE®

All of the components used in the five new I.V. Administration Sets were previously used in the predicate device cleared under K121269, except for the Spike component which has been redesigned in comparison to the predicate's Spike component that was cleared under K121269. The redesigned Spike component is substantially equivalent to the reference device, Chemfort™ Bag Adaptor SP's Spike component, which was cleared under K192866. The Spike component of both the IV Administration Set and of the reference device are identical in terms of design and raw material.

{5}------------------------------------------------

# Summary of Technological Characteristics Including Modifications to the Device:

The following table compares the I.V. Administration sets to the predicate device with respect to intended use, technological characteristics and principles of operation, providing detailed information regarding the basis for the determination of substantial equivalence.

| Item                                                    | Predicate Device:<br>Simplivia Healthcare Ltd.<br>IV Administration Set<br>(K121269)                                                                                                                                                                                                                                                                                                                                                                                          | Proposed Device:<br>Simplivia Healthcare Ltd.<br>IV Administration Set<br>(K201142)                                                                                                                                                                                                                                                                                                                                                                                           | Equivalence to predicate                                                                                                                                                                                                                                                                                                                                                                                                                   |
|---------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code                                            | FPA                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | ONB (primary)<br>FPA (secondary)                                                                                                                                                                                                                                                                                                                                                                                                                                              | K201142 sets were tested<br>according to NIOSH and<br>proved to be a CSTD                                                                                                                                                                                                                                                                                                                                                                  |
| Regulation Number                                       | 880.5440                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 880.5440                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | Same                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Indications for use                                     | The I.V. Administration Set<br>is a single use, sterile I.V.<br>set for administration of<br>fluids from a container to a<br>patient's vascular system.                                                                                                                                                                                                                                                                                                                       | Tevadaptor® is a Closed<br>System Transfer Device<br>(CSTD) that mechanically<br>prohibits the release of<br>drugs, including<br>antineoplastic and<br>hazardous drugs, in vapor,<br>aerosol or liquid form<br>during preparation,<br>reconstitution,<br>compounding and<br>administration, minimizing<br>exposure of individuals,<br>healthcare personnel, and<br>the environment to<br>hazardous drugs.                                                                     | Since K201142 sets are<br>under the ONB code, the<br>intended use includes the<br>following: The sets<br>mechanically prohibit the<br>release of drugs, including<br>antineoplastic and hazardous<br>drugs, in vapor, aerosol or<br>liquid form during<br>preparation, reconstitution,<br>compounding and<br>administration, minimizing<br>exposure of individuals,<br>healthcare personnel, and the<br>environment to hazardous<br>drugs. |
| Components                                              | Spike, drip chamber, tubing,<br>flow regulator, clamp,<br>ULTRASITE® Y injection<br>site, Luer connection                                                                                                                                                                                                                                                                                                                                                                     | Spike, drip chamber, tubing,<br>flow regulator, clamp,<br>ULTRASITE® Y injection<br>site, Luer connection, twist-<br>off                                                                                                                                                                                                                                                                                                                                                      | Similar                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Sterilization                                           | Ethylene Oxide<br>validated cycle SAL 10-6                                                                                                                                                                                                                                                                                                                                                                                                                                    | Ethylene Oxide<br>validated cycle SAL 10-6                                                                                                                                                                                                                                                                                                                                                                                                                                    | Same                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Biocompatibility                                        | Biocompatible                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Biocompatible                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Same                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Interconnecting<br>features                             | Mechanical snap<br>connections, Luer lock<br>connection, spike                                                                                                                                                                                                                                                                                                                                                                                                                | Mechanical snap<br>connections, Luer lock<br>connection, spike                                                                                                                                                                                                                                                                                                                                                                                                                | Same                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Interaction with<br>patient                             | No direct interaction - this is<br>a secondary set                                                                                                                                                                                                                                                                                                                                                                                                                            | No direct interaction - this is<br>a secondary set                                                                                                                                                                                                                                                                                                                                                                                                                            | Same                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Indirect interaction<br>with patient                    | Indirect interaction with the<br>patient is achieved through<br>the passage of IV fluids<br>through the central tubing of<br>the administration set                                                                                                                                                                                                                                                                                                                           | Indirect interaction with the<br>patient is achieved through<br>the passage of IV fluids<br>through the central tubing<br>of the administration set                                                                                                                                                                                                                                                                                                                           | Same                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Item                                                    | Predicate Device:<br>Simplivia Healthcare Ltd.<br>IV Administration Set<br>(K121269)                                                                                                                                                                                                                                                                                                                                                                                          | Proposed Device:<br>Simplivia Healthcare Ltd.<br>IV Administration Set<br>(K201142)                                                                                                                                                                                                                                                                                                                                                                                           | Equivalence to predicate                                                                                                                                                                                                                                                                                                                                                                                                                   |
| Interaction with<br>other devices                       | For most uses, this<br>secondary administration set<br>will connect to an IV<br>solution container and a<br>primary administration set.<br>Other connections may be<br>made through specific<br>components of the set, such<br>as the Spike Port and the<br>ULTRASITE® Y injection<br>site.<br>while using the following<br>Chemfort™ or Tevadaptor™<br>components:<br>- Syringe Adaptor (K192866)<br>- Syringe Adaptor Lock<br>(K192866)<br>- Luer Lock Adaptor<br>(K192866) | For most uses, this<br>secondary administration<br>set will connect to an IV<br>solution container and a<br>primary administration set.<br>Other connections may be<br>made through specific<br>components of the set, such<br>as the Spike Port and the<br>ULTRASITE® Y injection<br>site,<br>while using the following<br>Chemfort™ or Tevadaptor™<br>components:<br>- Syringe Adaptor (K192866)<br>- Syringe Adaptor Lock<br>(K192866)<br>- Luer Lock Adaptor<br>(K192866) | Same                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Intended Shelf life                                     | 3 years                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | 3 years                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Same                                                                                                                                                                                                                                                                                                                                                                                                                                       |
| Meets the NIOSH<br>and ISOPP<br>definition of a<br>CSTD | N/A (the first draft protocol<br>was issued in 2015)                                                                                                                                                                                                                                                                                                                                                                                                                          | Yes                                                                                                                                                                                                                                                                                                                                                                                                                                                                           | The first draft protocol was<br>issued after predicate 510(k)                                                                                                                                                                                                                                                                                                                                                                              |

{6}------------------------------------------------

{7}------------------------------------------------

## Performance Data

Simplivia conducted several performance tests to demonstrate that the modified I.V. Administration sets comply with the following standards and that they function as intended.

- ISO 10993-1:2009, Biological Evaluation of Medical Devices. Part 1: Evaluation and . testing within a risk management process.
- ISO 10993-4:2017, Biological Evaluation of Medical Devices. Part 4: Selection of tests for interactions with blood.
- . ISO 10993-5:2009, Biological Evaluation of Medical Devices. Part 5: Tests for in vitro cytotoxicity.
- . ISO 10993-7:2008, Biological Evaluation of Medical Devices. Part 7: Ethylene oxide sterilization residuals.
- . ISO 10993-10:2010, Biological Evaluation of Medical Devices. Part 10: Tests for irritation and skin sensitization.
- ISO 10993-11:2017, Biological Evaluation of Medical Devices. Part 11: Tests for . systemic toxicity.
- ISO 10993-18:2005. Biological Evaluation of Medical Devices. Part 18: Chemical . characterization of medical device materials within a risk management process.
- ISO 11135:2014, Sterilization of health-care products Ethylene oxide Requirements . for the development, validation and routine control of a sterilization process for medical devices.
- ISO 11607-1:2019, Packaging for terminally sterilized medical devices Part 1: ● Requirements for materials, sterile barrier systems and packaging systems.
- . ISO 8536-4:2010, Infusion equipment for medical use — Part 4: Infusion sets for single use, gravity feed.
- ISO 80369-7:2016. Small-bore connectors for liquids and gases in healthcare . applications — Part 7: Connectors for intravascular or hypodermic applications.
- ISO 14971:2007 Medical devices Medical devices Application of risk management . to medical devices

## Substantial Equivalence

Simplivia Healthcare's modified I.V. Administration sets have the same intended use and principles of operation as its predicate device, K121269. Performance data demonstrate that the modified I.V. Administration sets are as safe and effective as their predicate device and do not raise any new safety and effectiveness issues. Thus, Simplivia Healthcare's modified I.V. Administration Sets are substantially equivalent to their predicate device, K121269.

---

**Source:** [https://fda-staging.innolitics.com/device/K201142](https://fda-staging.innolitics.com/device/K201142)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).