K201141 · Nous Imaging, Inc. · LNH · Aug 26, 2020 · Radiology
Device Facts
Record ID
K201141
Device Name
FIRMM
Applicant
Nous Imaging, Inc.
Product Code
LNH · Radiology
Decision Date
Aug 26, 2020
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Indications for Use
The Nous FIRMM system is an accessory to an MRI scanner to calculate and display patient motion during a head scan. The motion results are derived from the MR image data as it is being acquired. MR images can be imported during the scan, and analyzed to detect patient motion in real-time. FIRMM enables the operator to become aware of patient motion during the scanning session and can be used to support scanning efficiency. The device is intended for prescription use only.
Device Story
FIRMM (Framewise Integrated Real-time MRI Monitoring) is a software-based accessory for MRI scanners; monitors patient head motion during brain MRI acquisitions. Input: MR image data as it is being generated/acquired. Processing: continuously analyzes image data to detect and quantify small rotational or translational displacements of anatomy; calculates frame displacement (FD) values. Output: real-time 'seismograph' plot of motion and overall motion quality figure displayed on a dedicated tablet; optional biofeedback display provided to patient via in-bore entertainment system. Used in clinical MRI settings by scan operators (technologists/radiologists). Benefits: enables operators to manage studies in real-time, potentially reducing motion artifacts and the need for repeat scans. FIRMM has read-only access to scanner data; does not interact with the acquisition process or adjust scan parameters.
Clinical Evidence
Bench testing only. Accuracy validated using synthetic datasets with known frame displacement (FD) values. Compatibility verified with MRI scanners, projection systems, and screen resolutions. Performance compared against a reference off-line method (4dfp) using diverse adult and infant datasets, demonstrating strong correlation.
Technological Characteristics
Software-based system running on off-the-shelf tablet (kiosk mode). Tablet compliant with IEC 60601-1 (electrical safety) and IEC 60601-1-2 (EMC). Connectivity: read-only access to MRI scanner data; HDMI output for patient biofeedback. Operates as a standalone appliance. No components inside the imaging volume.
Indications for Use
Indicated for use as an accessory to MRI scanners to monitor, calculate, and display patient head motion during MRI acquisitions in adult and infant patients. Used to support scanning efficiency by providing real-time motion feedback to the operator and optional biofeedback to the patient.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
Predicate Devices
KinetiCor Motion Correction System (KinetiCor)
Submission Summary (Full Text)
{0}------------------------------------------------
August 26, 2020
Image /page/0/Picture/1 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.
Nous Imaging, Inc. % Mr. Rob Newman Regulatory Consultant Scion Opportunity Consulting 680 Webb Road CHADDS FORD PA 19317
Re: K201141
Trade/Device Name: FIRMM Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: Class II Product Code: LLZ Dated: August 3, 2020 Received: August 4, 2020
Dear Mr. Newman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. Although this letter refers to your product as a device, please be aware that some cleared products may instead be combination products. The 510(k) Premarket Notification Database located at https://www.accessdata.fda.gov/scripts/cdrh/cfdocs/cfpmn/pmn.cfm identifies combination product submissions. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803) for
{1}------------------------------------------------
devices or postmarketing safety reporting (21 CFR 4, Subpart B) for combination products (see https://www.fda.gov/combination-products/guidance-regulatory-information/postmarketing-safety-reportingcombination-products); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820) for devices or current good manufacturing practices (21 CFR 4, Subpart A) for combination products; and, if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to https://www.fda.gov/medical-device-safety/medical-device-reportingmdr-how-report-medical-device-problems.
For comprehensive regulatory information about mediation-emitting products, including information about labeling regulations, please see Device Advice (https://www.fda.gov/medicaldevices/device-advice-comprehensive-regulatory-assistance) and CDRH Learn (https://www.fda.gov/training-and-continuing-education/cdrh-learn). Additionally, you may contact the Division of Industry and Consumer Education (DICE) to ask a question about a specific regulatory topic. See the DICE website (https://www.fda.gov/medical-device-advice-comprehensive-regulatoryassistance/contact-us-division-industry-and-consumer-education-dice) for more information or contact DICE by email (DICE@fda.hhs.gov) or phone (1-800-638-2041 or 301-796-7100).
Sincerely.
For
Thalia T. Mills, Ph.D. Director Division of Radiological Health OHT7: Office of In Vitro Diagnostics and Radiological Health Office of Product Evaluation and Quality Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K201141
Device Name
FIRMM
Indications for Use (Describe)
The Nous FIRMM system is an accessory to an MRI scanner to calculate and display patient motion during a head scan. The motion results are derived from the MR image data as it is being acquired.
MR images can be imported during the scan, and analyzed to detect patient motion in real-time. FIRMM enables the operator to become aware of patient motion during the scanning session and can be used to support scanning efficiency.
The device is intended for prescription use only.
Type of Use (Select one or both, as applicable)
|X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW! *
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: 06/30/2020 See PRA Statement below.
{3}------------------------------------------------
### 510(k) Summary 5
# K201141
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of the Safe Medical Devices Act 1990 and 21 CFR Section 807.92.
| Establishment: | Nous Imaging Inc.<br>4220 Duncan Ave<br>St Louis MO 63110<br>Tel: +1 (314) 776-4516<br>Email: INFO@firmm.io<br>Registration Number: TBD |
|----------------|-----------------------------------------------------------------------------------------------------------------------------------------|
| Date Prepared: | 10 Aug 2020 |
| Manufacturer: | Nous Imaging Inc.<br>4220 Duncan Ave<br>St Louis MO 63110<br>Phone: (314) 296-5648 |
#### 5.1 General Information
#### 5.2 Contact Information
Rob Newman Scion Consulting, LLC 680 Webb Road Chadds Ford, PA 19317 Phone: 484-574-9633 Email: rob.newman@firmm.io
| Device Name: | FIRMM |
|-----------------------|---------------------------------------------|
| | |
| Trade Name: | FIRMM |
| Classification Name: | Magnetic Resonance Diagnostic Device (MRDD) |
| Classification Panel: | Radiology |
| CFR Code: | 21 CFR Sec 892.1000 |
| Classification: | Class II |
#### 5.3 Device Name and Classification
{4}------------------------------------------------
| Product Code: | LNH |
|---------------|-----|
|---------------|-----|
#### 5.4 Legally Marketed Predicate Device
| Trade Name: | KinetiCor Motion Correction System |
|-----------------------|---------------------------------------------|
| Classification Name: | Magnetic Resonance Diagnostic Device (MRDD) |
| Classification Panel: | Radiology |
| CFR Code: | 21 CFR Sec 892.1000 |
| Classification: | Class II |
| Product Code: | Primary: LNH |
#### ર રંડ Indications For Use
The Nous FIRMM system is an accessory to an MRI scanner to calculate and display patient motion during a head scan. The motion results are derived from the MR image data as it is being acquired.
MR images can be imported during the scan, and analyzed to detect patient motion in realtime. FIRMM enables the operator to become aware of patient motion during the scanning session and can be used to support scanning efficiency.
The device is intended for prescription use only.
#### 5.6 Device Description
Magnetic resonance imaging (MRI) has unrivaled clinical and research utility as a diagnostic imaging modality in use worldwide. However, the utility of MRI has one significant limitation: patient motion during an MRI scan can greatly diminish the diagnostic quality of the images. The result is a decision to either to repeat the scan series, or to use the images regardless of their compromised diagnostic information.
The company has developed a software-based system that performs Framewise Integrated Real-time MRI Monitoring (FIRMM®). The FIRMM system continuously monitors the MR image data as it is being generated during the exam to detect small displacements (rotation or translation) of the anatomy along any axis during MRI acquisitions of time series datasets. It quantifies these movements and provides the MRI scanner operator with a plot of patient motion displayed as type of 'seismograph', as well as providing a figure of overall motion quality. This system allows for non-invasive, non-contact monitoring of patient motion during brain MRIs, and displays motion metrics in real-time to the scanner operator, with an optional ability to provide biofeedback to the patient.
{5}------------------------------------------------
Real-time feedback on patient motion during the scan can be very helpful to the scan operator who, as a result may be able to better manage the study, allowing the patient and operator to complete the exam yielding diagnostic quality images with fewer motion artifacts.
The FIRMM system itself has no interaction with the data acquisition process, and there is no component of the FIRMM system inside the imaging volume. Because FIRMM only has read access to the scanner, FIRMM has no direct impact on scan time or diagnostic image generation of the MRI scanner.
## Hardware
The FIRMM software is run on a dedicated off-the-shelf computer tablet with a touch screen display operating in kiosk mode. The tablet is IEC 60601-1 and IEC 60601-1-2 compliant and sits next to the MR scan operator. The FIRMM application is the only function provided by the system. The computer is strictly an appliance and the user cannot alter its function outside of the provided administration and configuration functions accessed from the touch screen.
The system includes a video output with a HDMI signal that can be connected to an in-bore entertainment system (customer supplied item) and can also provide a biofeedback display to the patient of any detected motion. This system may be helpful to make patients aware of small voluntary motions.
#### 5.7 Risk Management, General Safety and Effectiveness
The device labeling contains instructions for use and any necessary cautions and warnings to provide for safe and effective use of the device.
Risk management is ensured via a risk analysis. which is used to identify potential hazards and mitigations. These potential hazards are controlled by design means, protection measures and user instructions. To confirm that the measures are effective and that the product meets its intended uses, verification of requirements and standards, and validation of the clinical workflow was performed. Nous Imaging adheres to recognized and established industry practice and relevant international standards where indicated.
#### 5.8 Technological Characteristics and Substantial Equivalence
Nous FIRMM System is substantially equivalent to the predicate device; both systems have the fundamental purpose of detecting and reducing the effects of patient motion on MR exams. FIRMM has the same intended use, and similar indications, technological characteristics and principles of operation as its predicate device.
While the predicate device tracks patient motion via a fiducial/tracking camera, the FIRMM system uses the MR images as they are being produced. FIRMM provides feedback on patient motion to the scan operator during the MRI scans, whereas the predicate provides direct viewing of the patient's face was well a view of the raw motion
{6}------------------------------------------------
traces from the tracking system on an optional monitor. Additionally, the predicate provides the patient motion correction coordinates to the MRI scan, to actively adjust the MR image acquisitions minimize the effects of motion in the images. The technological differences between the FIRMM and its predicate devices raise no new issues of safety or effectiveness. Therefore, the technological characteristics, and principles of operation are substantially equivalent to the predicate device.
#### 5.9 Nonclinical Tests
- Software Verification and Validation testing was performed in accordance with the FDA guidance document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices", dated May 11, 2005 and with Design Control Regulations (21 CFR 820).
- . Electrical Safety: The tablet hardware was certified by the OEM manufacturer to the general standard IEC 60601-1 "Medical electrical equipment - Part 1: General requirements for basic safety and essential performance". Certification is included which shows that the tablet hardware is in compliance with IEC 60601-1.
- Electromagnetic Compatibility: The tablet hardware was tested to IEC 60601-1-2 "Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral Standard: Electromagnetic disturbances -Requirements and tests". Certification is included which shows that the tablet hardware is in compliance with IEC 60601-1-2.
- Accuracy: An accuracy study was conducted using a synthetic data set that ● included known frame displacement (FD) values. The data set was processed by the FIRMM system and the resulting FD values were compared to the known FD values. The conclusion is that the accuracy exceeds the specification.
- . Compatibility: Evidence of compatibility with the scanner, projection systems, and screen resolutions was based upon a combination of studies:
- Scanner and fMRI Compatibility Study O
- Biofeedback Compatibility Study O
- Screen Resolution Compatibility Study o
- . 4dfp comparison study: A wide variety of data sets from adult and infant subjects were processed by FIRMM as well as a reference off-line method. The results showed the that FIRMM correlates very strongly with the reference method.
The nonclinical data support the safety of the device and the hardware and software verification and validation demonstrate that the FIRMM system should perform as intended in the specified use conditions.
{7}------------------------------------------------
#### 5.10 Conclusions
The Nous FIRMM System is substantially equivalent to the predicate device; both systems have the fundamental purpose of detecting and reducing the negative effects of patient motion on MR exams. FIRMM has the same intended uses and similar indications, and principles of operation as its predicate device. While the predicate uses a camera/fiducial system in the bore of the magnet and direct video observation of the patient, and FIRMM uses the MR images as they are being produced, the technological differences between the FIRMM and its predicate devices raise no new issues of safety or effectiveness.
Nous believes that FIRMM as an accessory to a medical device is substantially equivalent to the currently marketed device KinetiCor's Motion Correction System. The subject device does not raise any new questions with respect to safety or effectiveness. The conclusions drawn from the nonclinical tests demonstrate that the device is safe, and effective.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.